GasAlert Extreme
Quick Reference Guide
18
(1 ppm increments)
GasAlert Extreme PH3: 0-5.0 ppm (0.1 ppm increments)
GasAlert Extreme SO2: 0-100.0 ppm (0.1 ppm increments)
GasAlert Extreme Cl2: 0-50.0 ppm (0.1 ppm increments)
GasAlert Extreme NH3: 0-100 ppm (1 ppm increments)
GasAlert Extreme NH3 (high range): 0-400 ppm
(1 ppm increments)
GasAlert Extreme NO2: 0-100.0 ppm (0.1 ppm increments)
GasAlert Extreme HCN: 0-30.0 ppm (0.1 ppm increments)
GasAlert Extreme ETO: 0-100.0 ppm (0.1 ppm increments)
GasAlert Extreme ClO2: 0-1.00 ppm (0.01 ppm increments)
GasAlert Extreme O3: 0-1.00 ppm (0.01 ppm increments)
GasAlert Extreme NO: 0-250 ppm (1 ppm increments)
Sensor type: Plug-in electrochemical cells
Calibration: Auto zero, set span, and span sensor
Oxygen sensor: Automatic span upon activation (enable/dis-
able)
Alarm conditions: TWA alarm, STEL alarm, low alarm, high
alarm, sensor alarm, low battery alarm, confidence beep, and
automatic shutdown alarm.
Audible alarm: 95 dB at 1 ft. (0.3 m) typical
Visual alarm: Red light-emitting diode (LED)
Display: Alpha-numeric liquid crystal display (LCD)
Backlight: Automatically activates for 3 seconds whenever
there is insufficient light to view the LCD and during alarm con-
ditions (unless the backlight is disabled in user options). Any
pushbutton reactivates the backlight for 6 seconds.
Self-test: Initiated at activation
User field options: automatic backlight, passcode protection,
latching alarm, calibration due lock, language selection, enable/
disable automatic oxygen calibration, datalog interval, data
transfer option, confidence beep, calibration interval, stealth
mode, set alarm setpoints, calibration span concentration
Year of manufacture: The detector’s year of manufacture is
determined from the serial number. The second and third num-
ber after the first letter determines the year of manufacture.
E.g. J410-H001000 = 2010 year of manufacture
Battery test: Every 0.5-second
Battery: 3 V lithium Panasonic CR-2PE/BN
Event logs: A maximum of 10 events are stored. When 10
events have been recorded, the next event overwrites the old-
est event log.
Warranty: 2 years including sensors (1 year Cl2, NH3, O3,
ETO, and ClO2 sensor)
Intrinsic Safety:
Classified by UL to both U.S. and Canadian Standards as intrin-
sically safe for Class I, Division 1, Group A, B, C, D and Class
II, Group E, F, G
European Explosives Protection
CE 0539 g II 1G
Ex ia IIC T4 Ga DEMKO 04 ATEX 0336363X
IECEx Ex ia IIC T4 Ga IECEx UL 08.0001X
ABS Type Approved: VA-348-169-X
KTL 12-KB4BO-0199X1