Cal Test Electronics
CT4422 Fused Probe DMM Electrical Kit, User Guide
© 2021 Cal Test Electronics. All rights reserved.
Cal Test Electronics products are covered by US and foreign patents, issued and
pending. Information on this publication supersedes all earlier versions. Speci cations
are subject to change without notice.
Made in Taiwan
Safety Notice and Symbols
This product is designed to IEC 61010 safety requirements
WARNING A warning statement calls attention to an operating procedure, practice, or
condition, which, if not followed correctly, could result in injury or death to
personnel.
CAUTION A caution statement calls attention to an operating procedure, practice,
or condition, which, if not followed correctly, could result in damage to or
destruction of parts or the entire product.
The following symbols may appear on the product or in its documentation:
Caution, see documentation for details
Fuse inside
Safety is compromised if instructions are not read, understood, and followed.
WARNING Never exceed the maximum interrupt current of the fuse. The fuse protects
the equipment against short-circuit but not against over-currents.
WARNING Before use check fuse for continuity and good condition.
WARNING Install only speci ed fuse, or equivalent, sizing fuse to the required short
circuit current.
Never open fused probe while connected or energized.
Disconnect from all devices and leads before replacing fuse.
Clean with cloth lightly moistened with water only. Dry thoroughly before use.
Do not use if any part is damaged. All maintenance should be referred to quali ed
service personnel only.
Do not use in wet or explosive atmosphere.
Always be aware of CAT environment
Description
The CT4422 Fused Probe DMM Electrical Kit includes a 1,000 V 500 mA fast acting
fused probe to protect both equipment and personnel. Used for both handheld and
bench DMMs, this kit features a unique locking tip cover, fully insulated medium
alligator clips tting conductors up to 7 mm, and silicone jacketed straight to right angle
4 mm sheathed banana plug test leads.
Fuse
• 6.3 x 32 mm, 0.5A, 1,000 V, 50 kA, p/n CT4049-0.5A
Speci cations
• Voltage Rating: 1,000 V CAT
III/600 V CAT IV, reinforced
insulation, pollution degree 2
• Current Rating: See rating on the
fuse (CT4300), 10 A (CT3754,
CT4439)
• Operating temperature: 0°C to
60°C (32°F to 140°F) and max
RH 85%; altitude up to 2000 m.
• Storage temperature: -20°C to
70°C (-4°F to 158°F)
CE Compliance
This product meets the essential
requirements of the applicable European
Directives as follows:
• Low-Voltage Directive (safety):
2014/35/EU
Standard EN 61010-031:2015
2nd Ed.
• Restriction of Hazardous
Substances Directive (RoHS):
2011/65/EU+AMD 2015/863
Fuse Replacement
• Disconnect from all devices and
test leads
• Unscrew fused probe body
by turning probe tip half
counterclockwise.
• Remove and replace fuse
• Screw probe pieces together
Tip Cover Removal
• Use small blade screwdriver to pry
up tab and slide cover off. Lifting
too much on tab can damage
and reduce the lifetime of the
cover.
Alligator Clip Attachment
• Remove tip cover and push
tip.
alligator clip directly onto 2 mm
2