
A-1
APPENDIX A
A.1 LI200S PYRANOMETER
LI200S pyranometers have a 100 ohm shunt
resistor built into the cable. They can be directly
measured by Campbell Scientific dataloggers.
The input range and multipliers vary from one
pyranometer to another. See Sections A.3 and
A.4 for calculating the proper input range and
multiplier.
A.1.1 WIRING
The red lead is connected to the high side (H)
of a differential input channel and the black lead
to the corresponding low side (L). On the CR10
a jumper wire is installed between the low side
and analog ground (AG). The clear lead is
connected to ground (G). On the 21X the
jumper wire is installed between the low side
and ground (G) and the clear lead is also
connected to ground (G). The measurement is
then made with Instruction 2 (see Section 4).
A.2. UNMODIFIED PYRANOMETERS
Pyranometers that do not have variable or fixed
shunt resistors built into the cable can still be
measured by Campbell Scientific dataloggers.
This is done by wiring in a 100 Ωshunt resistor
directly onto the datalogger wiring panel. The
input range and multipliers vary from one
pyranometer to another. See Sections A.3 and
A.4 for calculating the proper input range and
multiplier.
A.2.1 WIRING
Signal positive is connected into the high
side(H) of a differential input channel and signal
negative to the corresponding low side (L). A
jumper wire is installed between the low side (L)
and analog ground (AG) on the CR10 wiring
panel or ground on the 21X. A 100 Ω1%
resistor (P/N 191) is installed on the wiring
panel between the high and low sides the
measurement channel. The measurement is
then made with Instruction 2 (see Section 5).
FIGURE A.2-1. Unmodified Pyranometer
Wiring Schematic
A.3. INPUT RANGE
The following is an example of how to
determine the optimum input range for a given
sensor calibration and maximum expected
irradiance. This is an example only. Your
values will be different.
This example uses the calibration provided by
LI-COR, Inc. Assume that the sensor calibration is
87 µA kW-1 m2. The pyranometer outputs current
which is converted to voltage by the 100 Ωshunt
resistor in the cable or on the wiring panel. To
convert the calibration from current to voltage,
multiply the LI-COR calibration by 0.1 KΩ(shunt
resistor). The example calibration changes to
8.7 mV kW-1 m2.
A reasonable estimate of maximum of
irradiance at the earth's surface is 1 kW m-2.
Thus, an estimate of the maximum input
voltage is obtained by multiplying the calibration
by the maximum expected irradiance. In this
example that product is 8.7 mV. Now, select
the smallest input range which is greater than
the maximum expected input voltage. In this
case the 25 mV slow range for the CR10 and
15 mV slow range for the 21X are selected.