Canon CXDI-80C User manual

C
CX
XD
DI
I-
-8
80
0C
C
W
Wi
ir
re
el
le
es
ss
s
Canon Inc. Japan
Copyright(C) Canon Inc. Medical Technical Service Dept. All rights reserved.
S
Se
er
rv
vi
ic
ce
e
M
Ma
an
nu
ua
al
l
Printed by Canon Inc.
October.2011 Rev.01
Confidential

Manual Control No. : BY8-2317-0E0
Name of Product : CXDI-80C Wireless
Distribution Control No.
Issued on
Service Manual Introduction
This service manual belongs to a series of after-service guides Canon Inc. publishes as part of
its comprehensive product quality guarantee program.
This service manual consists of nine chapters; “General”, “Installation Manual”, “Functions”,
“Repair Guide”, “Parts Catalog”, “Troubleshooting”, “Service Manual Report”, “Tools” and
“Appendix”. It describes an overview of the product, its functions, product configuration,
installation procedures, dimensions, specifications, and notes.
If the product undergoes a large modification, a revised edition of the service manual will be
sent to you. In other cases, a service manual report will be sent to you to update the manual.
Note 1:
This service manual is published by Canon Inc. in accordance with Article 6 (Furnishing
the Referring Materials) of the Service Assignment Contract it has concluded with your
company.
Note 2:
This service manual is the property of Canon Inc. and the company may seek to have it
returned, depending on the circumstances. You are expected to keep it until then.
Note 3:
Your inquiries, suggestions, etc. about the contents of this service manual should be
addressed to:
Medical Equipment Technical Service Dept.
Canon Inc. Headquarters
30-2, Shimomaruko 3-chome, Ohta-ku, Tokyo 146-8501, Japan

Caution Regarding Service
This product was precisely assembled under strict manufacturing process control. There
are several hazardous locations inside of this product. Careless work while the cover is
removed can result in the pinching of fingers or electrical shock. Please perform the work
with the following important points in mind:
1. Setup, Repair, and Maintenance
In order to ensure safety, the best performance, setup, repair, and maintenance work can only
be performed by technicians who have received service training specified by Canon Inc. If
there are order required certificates or restrictions specified by the law or ordinances, those
regulations of the country must be observed.
2. Removing the external cover
When removing the cover during maintenance, repair, etc., perform the work after switching
the power off. Never touch the device with wet hands, as there is a risk of electric shock.
3. Connecting the grounding wire
The provided ground wire must be connected to the ground terminal indoors. Make sure that
the device is properly grounded.
4. Alternation prohibition
Never modify the medical device in any way.
5. Waste control
The service provider is responsible for the disposal of used service parts, packing material,
etc. resulting from the setup, repair, or maintenance of the medical device. However, the
customer is responsible for the disposal of the medical device. Disposal activities must
follow the regulations (especially controlled industrial waste) of the country where the
device is used.

VORSICHT
Befolgen Sie die unten angegebenen Sicherheitsanweisungen.
Mißachtung kann zu erletzungenoder Unfällen führen.
1.Zerlegung, Zusammenbau, Einstellung und Wartung
Zerlegung, Zusammenbau, Einstellung und Wartung dürfen nur von einem Wartungstechniker
durchgeführt werden, der an einem von Canon vorgeschriebenen Wartungslehrgang teilgenommen hat.
2.Entfernen von Abdeckungen
Schalten Sie unbedingt die Stromversorgung des Instruments aus, bevor Sie die Abdeckungen zwecks
Wartung und Reparatur entfernen.
Vermeiden Sie auch eine Berührung des Instruments mit nassen Händen.
Anderenfalls können Sie einen elektrischen Schlag erleiden, der zum Tod oder schwerer Verletzung
führen kann.
3.Sicherung
Wenn die Sicherung ausgewechselt werden muß, schalten Sie unbedingt die Stromversorgung des
Instruments aus, und beheben Sie die Ursache für das Durchbrennen der Sicherung.
Ersetzen Sie die Sicherung nur durch den vorgeschriebenen Typ.
Anderenfalls kann es zu einem Brand oder elektrischen Schlag kommen.
4.Erdleiter
Erden Sie das Instrument unbedingt an einer Schukosteckdose.
Anderenfalls kann es zu einem Brand oder elektrischen Schlag durch Leckstrom kommen.
5.Umbau
Jeder Umbau des Produktes ist strengstens untersagt, da dies zu einem Brand oder elektrischen Schlag
führen kann.

C
Ca
au
ut
ti
io
on
n
R
Re
eg
ga
ar
rd
di
in
ng
g
t
th
he
e
S
Se
et
tu
up
p
According to the “IEC 60601-1-1:2000”, equipments installed in the patient environment are
restricted to “medical electric equipment conforming to IEC 60601-1”.
Note that the entire radiography system, including this product, is required to comply with IEC
60601-1-1. The details are as follows: Medical equipment such as an X-ray generator
complying with IEC 60601-1, an image capture computer and a display monitor complying with
IEC 60601-1 or IEC 60950-1.
The image capture computer, optional devices such as network equipment or magnetic card
readers, and display monitors that are classified as other than medical electrical equipment (IEC
60601), for example as information technology equipment (IEC 60950), cannot be installed in
the patient environment.
The appended documentation is to be provided to the user, operator, and installer of the
equipment configuring the system, such as the image capture computer and display monitor,
upon installation of such equipment.
Also, for the installation, the above items must be connected to a common protective earth.
Equipment connected to the image capture computer must satisfy state No. 3 in table BBB.201
of IEC 60601-1-1.
Either connect the image capture computer to a common protective ground, add a dedicated
protective ground, or install an isolation device.
The patient environment described below is an example cited from the “IEC 60601-1-1:2000” –
the measurements are only guidelines. However, the “IEC 60601-1-1:2000” example must be
treated as the standard.
Therefore, the image capture computer, optional devices such as network equipment or
magnetic card readers, and display monitors must be installed in a location separated by at least
the dimensions indicated below (outside the patient environment). In addition, the X-ray
interface box should be installed outside the patient environment.
* The areas through which patients move at times other than X-ray exposure (when
entering and leaving, etc.) are also included in the patient environment. Therefore,
always confirm (discuss) with the customer that the installation location is outside the
patient environment.
Example of patient environment

Note: These measurements are only guidelines.

C
CX
XD
DI
I-
-8
80
0C
C
W
Wi
ir
re
el
le
es
ss
s
Canon Inc. Japan
Copyright(C) Canon Inc. Medical Technical Service Dept. All rights reserved.
1
1.
.
O
Ov
ve
er
rv
vi
ie
ew
w
Printed by Canon Inc.
October.2011 Rev.01

1
1.
.
O
Ov
ve
er
rv
vi
ie
ew
w
CONTENTS
1. Overview ............................................................. 1
2. System diagram ........................................................ 2
3. Description of system ................................................... 3
3.1 CXDI-80C Wireless System configuration ............................... 3
3.2 System compatibility with CXDI-70C Wireless ........................... 4
3.3 Restrictions on connections ........................................... 4
4. CXDI image processing ................................................. 5
4.1 Processing flow ..................................................... 5
4.2 Concerning each type of image ........................................ 5
5. Specifications .......................................................... 6
5.1 Detector specifications ............................................... 6
5.2 Power supply specifications ........................................... 9
5.3 Recommended LAN cable specifications ................................ 9
5.4 Environmental resistance specifications ................................. 9
5.5 Required PC specifications ........................................... 10
5.6 Wireless communication specifications ................................. 10
5.7 Infrared communication specifications ................................. 16
5.8 Battery and battery charger specifications ............................... 17
5.9 Grid specifications .................................................. 20
5.10 AC/DC power supply specifications of wiring unit (option) ............... 21
5.11 [Supplementary notes] Long-term storage of batteries .................... 22

1
1.
.
O
Ov
ve
er
rv
vi
ie
ew
w
- 1 -
1. Overview
CXDI-80C Wireless
The CXDI-80C Wireless is a slim-line lightweight flat panel detector (FPD) which has no
cables for communication or power supply and which has other features as well that are
designed for optimum ease of use. It uses the IEEE 802.11n (2.4 GHz/5 GHz) standard and can
display images at higher speeds than existing products.
The size of this FPD is the same as a film cassette complying with the ISO4090 standard, and it
enables digital radiography using existing analog radiography equipment. It has a light weight
of about 2.3 kg including the battery, and it delivers excellent operability.
The FPD realizes high resolution with a pixel pitch of 125 µm and employs cesium iodide (CsI)
for the scintillator to achieve enhanced sensitivity.
The CXDI-80C Wireless maintains system compatibility with the CXDI-70C Wireless,
allowing both to be used side-by-side in the same environment. In addition, an existing
installation can be used as is with the CXDI-80C Wireless, thereby minimizing the costs
associated with adding the CXDI-80C Wireless.
CXDI system software CXDI Control Software NE supported:
With Ver.1.40 or later
CXDI Controller RF: With Ver.1.40 or later
80C detector body No. 100001 or later
X-ray interface box body No. 100001 or later
AC/DC box body No. 100001 or later
* Intended uses and applications
Diagnosis, prevention and monitoring of the diseases of patients
Exposing patients to X-rays, detecting the X-rays passed through the bodies of the patients and
converting them into digital images which are then provided to the physicians for diagnosis
Anticipated usage environments: X-ray exposure rooms, wards in general, health screening
vans
Anticipated users: Radiologists, physicians
Parties expected to install unit: Sales companies or agents which have received specialized
training
Anticipated examinees: Patients
* Precautions for use
No guarantees are given for the system in terms of risk management control in regard to
providing protection from blood and chemicals in operating theaters (water-proof performance)
or assuring safety and hygiene or when it is used together with cardiac defibrillators in ICUs and
other locations, used outdoors or used in applications oriented toward animals. Operation of the
product in such circumstances must be performed under the responsibility and assurances of SI
(System Integrator) after the operating ranges allowed by the product’s specifications have been
fully understood and acknowledged.

1
1.
.
O
Ov
ve
er
rv
vi
ie
ew
w
- 2 -
2. System diagram

1
1.
.
O
Ov
ve
er
rv
vi
ie
ew
w
- 3 -
3. Description of system
3.1 CXDI-80C Wireless System configuration
To use this system, users can select between a wireless connection and wired connection as shown in the
figures below. The wired connection option is useful when the detector is to be permanently installed in
an upright stand for use. Communication and the supply of power are made possible through the
detector cables.
X-ray I/F Box
IR Data
Communication Unit
HUB
LAN (20m)
LAN
LAN
USB
IF Cable (20m)
Operation Unit
Sensor Unit
Battery Pack X-ray Generator
(< 5m)
(2m)
(1.8m)
(1.8m)
(2.3m)
Access Point
Handle Unit/ Grid Unit
Battery Charger
Figure showing system configuration with wireless connection
Figure showing system configuration with wired connection
Operation Unit
HUB
LAN
LAN
Status Indicato
r
X
-
ray Generato
r
AC/DC Box
Senso
r
Uni
t
Buil
t
-
in IF Cable (20m)
(10m)
(4.5m)
LAN (20m)
(1.8m)
(3m)
(1.8m)
X
-
ra
y
I/F Bo
x
80C Detector Unit
80C Detector Unit

1
1.
.
O
Ov
ve
er
rv
vi
ie
ew
w
- 4 -
3.2 System compatibility with CXDI-70C Wireless
The CXDI-80C Wireless uses the same system components as the CXDI-70C Wireless, as shown
below.
3.3 Restrictions on connections
The control software that can be connected to the product is Ver.1.40 and up. The number of units
which can be connected is restricted as below by the software requirements.
Registered units: 10, simultaneously connected units: 4
The hardware requirements do not impose restrictions, but communication performance may be
degraded by interference when wireless channels are shared.
No such restrictions apply to wired connections.
◆Handling of ferrite cores
(1) [70C Wireless (2.4 GHz compatible)]
The system conforms to IEC 60601-1-2 when ferrite cores are attached to the detector cable and to the
LAN cable used to connect to the X-ray interface box.
It is therefore necessary to attach ferrite cores to these two cables in order to use the system.
(2) [70C Wireless (2.4 GHz/5 GHz compatible)]
The system conforms to IEC 60601-1-2 when no ferrite cores are attached to the sensor cable and to the
LAN cable used to connect to the X-ray interface box.
It is therefore not necessary to attach ferrite cores to these two cables in order to use the system.
Note: Caution on mixed usage of different detector models
If any of the detectors used is the 70C Wireless (2.4 GHz compatible) model, ferrite cores must be
attached to the two cables.
X
-ra
y
I/F Bo
x
IR Data
Communication Uni
t
HU
B
LAN (20m)
LAN
LAN
US
B
IF Cable (20m)
Operation Uni
t
Senso
r
Battery Pac
k
X
-ray Generato
r
(< 5m)
(1.8m)
(1.8m)
(2.3m)
Access Poin
t
80C Detector Unit
70C Detector Unit

1
1.
.
O
Ov
ve
er
rv
vi
ie
ew
w
- 5 -
4. CXDI image processing
4.1 Processing flow
4.2 Concerning each type of image
(1) Raw images
These are the images obtained by LANMIT. They have not been compensated or corrected
in any way.
(2) Original images
These images are produced from the raw images which have been subjected to offset
processing and gain compensation.
The characteristics intrinsic to the LANMIT detector have been corrected in these images.
They must never be taken outside.
(3) Diagnosis images
These images are produced from the original images which have been subjected to
automatic analysis and diagnosis image processing.
Raw images
Original images
Pre-processing
• Offset compensation
• Gain compensation
• Defect correction
Diagnosis images
Laser imager
DICOM output
Automatic analysis
• Exposure field
• ROI analysis
• Dynamic range analysis
• EI (exposure index)
Image processing for diagnoses
• Grid stripe suppression processing
• Gray scale conversion processing
• Dynamic range adjustment processing
• Enhancement processing
• Noise reduction processing
• Line noise reduction processing
Dtstore
Image display
unit Storage device
Processing from this point
onward is handled by the
image capture computer.

1
1.
.
O
Ov
ve
er
rv
vi
ie
ew
w
- 6 -
5. Specifications
5.1 Detector specifications
Item CXDI-80C Remarks
Usage formats
General imaging by wireless
connection or wired connection,
film cassette compatible
Wiring unit required for
wired connection.
Effective imaging range 35.0 cm x 27.4 cm
(14” x 11”)
Number of effective pixels 2800 x 2192
Pixel pitch 125μm x 125μm
DQE@1mR, RQA5 60% or more
MTF@2lp/mm 30% or more
Scintillator CsI
Distance between patient touch
screen (CFRP surface) and
detector surface (glass surface)
3.0 mm
Imaging cycle Approx. 15 sec.
Image display
time*1
Full-screen
image display 5 sec. or less
Preview
display 3 sec. or less
Battery
performance*2
Max mode Approx. 800 images
(about 3 hours)
15-second cycle, 1-second
sleep
Ave mode Approx. 140 images
(about 4 hours)
100-second cycle, 1-second
sleep
Save mode Approx. 6.5 hours In sleep mode
Full charge
time Approx. 3 hours In 25°C environment
Wireless standard IEEE 802.11n (2.4 GHz/5 GHz)
Dimensions 384 (W) x 307 (D) x 15 (H) mm
(15” x 12” x 0.6”)
Colors of detector exterior
Top cover: Light sky blue
Side covers: Metallic satin silver
Function switches: Medical blue
Rear cover: Titanium white
Detector weight 2.3 kg (including battery)
*1: Display time values reflect performance when using an image capture computer with the
recommended specifications.
*2: This is the performance for a new battery which has been 100% charged.

1
1.
.
O
Ov
ve
er
rv
vi
ie
ew
w
- 7 -
Item CXDI-80C Remarks
Service life
7 years
Operation time: 21,000 hours (10 hours a day,
300 days a year, 7 years)
Standby time: 61,500 hours (24 hours a day,
365 days a year, 7 years)
Ambient temperature:
+23°C ±2°C
Relative humidity:
60% ±15% RH
Atmospheric pressure:
860 to 1060 hPa
When used under the
ambient conditions
above
Power
consumption
/heat
generation*1
Momentary
max. load Approx. 70 VA or less
Momentary max.
value, excluding inrush
currents
Assumed
continuous
max. load
mode
Approx. 20 W or less (17 kcal/h, 72 kJ/h or
less)*2
Mean value during
continuous operation
with exposure at
15-second intervals
Standby Approx. 15 W or less Sleep mode
Detector withstand load
operation
Load applied over entire surface: 1.47 kN
(150 kgf) for 1 minute
Load applied in local areas: 0.98 kN (100 kgf),
40 mm diameter for 1 minute
There must be no problems in operation after
the loads have been applied under the above
conditions and then removed.
Test conditions:
The detector shall be
placed on a rigid flat
surface with its X-ray
incidence surface face
up, and images shall be
taken with the loads on
the left brought to bear
on the effective area.
Control software CXDI Control Software NE
Ver.1.40 and up supported
Photo timer Cannot be incorporated
Allowable temperature of
detector panel
Max. +45°C (with upper limit of usable
environment estimated at +35°C, a relative
temperature of 10°C or less must be satisfied
at room temperature)
When an operating
temperature of +49°C
or more has been
detected for the
detector, a restriction
applies to the detector
where it is forcefully
set to the sleep mode
and imaging is
prohibited, and the
imaging prohibition is
released when the
temperature drops to
under +48°C.
*1: The target unit is a complete CXDI-80C Wireless set (including detector, X-ray interface box,
wiring unit, and status indicator). The values indicated incorporate a large margin from the
actual values or indicate attainable values.
1) Power consumption [VA]: Information for the purpose of designing the power supply

1
1.
.
O
Ov
ve
er
rv
vi
ie
ew
w
- 8 -
system for the installation site
2) Power consumption [W]: Information for the purpose of designing the power supply
system for the installation site (Note that this is valid only
when the power factor is high [ex. 90% or more].)
3) Heat generation [cal or J]: Information for the purpose of designing the cooling system
for the installation site
*2: The heat generation is calculated from the power consumption (W) as follows:
Heat generation (kcal/h) = power consumption (W) * 0.23889 * 60 * 60 / 1000
Heat generation (kJ/h) = power consumption (W) * 60 * 60 / 1000

1
1.
.
O
Ov
ve
er
rv
vi
ie
ew
w
- 9 -
5.2 Power supply specifications
Item Details Remarks
Battery charger
Battery charger AC 100 to 240 V,
50/60 Hz, 70 to 90 VA
Power cord set
X-ray interface box
X-ray interface box DC 12V, 0.25 A
AC adapter
AC 100 to 240 V,
47/63 Hz, 0.33 to
0.18 A
Power cord set
Wiring unit (option)
AC/DC box
AC 100 to 240 V,
50/60 Hz, 0.29 to
0.18 A, 29 to 41 VA
Power cord set
5.3 Recommended LAN cable specifications
The LAN used by the CXDI system shall be a local network, and it shall be isolated from the
clinic/hospital network. etc. The specifications of the LAN cables are given below.
CAT-5 or above
Recommended cable length: Max. 30 meters
Straight type (type may differ depending on the hub functions)
Item Details Remarks
For connecting PC and hub 1000Base-T
For connecting X-ray interface box and hub 100Base-Tx
For connecting wireless LAN access point and hub 1000Base-T
For connecting wiring unit and hub 100Base-Tx
5.4 Environmental resistance specifications
Item Operation and
installation conditions Transportation and
storage conditions* Remarks
Temperature
range +5 to +35 [°C] -30 to +50 [°C]
Use in an ambient temperature
exceeding 35°C is not
considered as constituting
normal use.
Humidity
range 30 to 80 [%RH] 10 to 95 [%RH] No condensation
Vibration 0.05 [G], 10 to 100 [Hz] 1.0 [G], 10 to 100 [Hz]
Atmospheric
pressure 700 to 1060 [hPa] 700 to 1060 [hPa]
* The storage environment specifications apply in the formally packaged status. The
specifications in the non-formally packaged status (including no packaging) are +5 to
+40 [°C] for the temperature range and 30 to 85 [%RH] for the humidity range.

1
1.
.
O
Ov
ve
er
rv
vi
ie
ew
w
- 10 -
5.5 Required PC specifications
Use of the CXDI-80C Wireless requires a PC (hereinafter image capture computer) running
Microsoft Windows 7 and on which CXDI Control Software NE (Ver. 1.40 or later) or CXDI
Controller RF (Ver. 1.40 or later) made by Canon has been installed, and hardware that satisfies
the required specifications must be procured from a Canon sales company or agent. For details
of the required specifications of the image capture computer, refer to the CXDI Control
Software NE (Ver. 1.40 or later) or CXDI Controller RF (Ver. 1.40 or later) service manual.
5.6 Wireless communication specifications
The relay unit used for wireless communication between the CXDI-80C Wireless and Image
Capture Computer is called a wireless LAN access point. The table below lists the wireless
communication specifications.
Item Details Remarks
Communication
system
IEEE standard IEEE 802.11n
Frequency band 5 GHz/2.4 GHz
Security
(encryption
system)
WPA2-PSK
(AES)
AES is an encryption
system which has been
certified by the
National Institute of
Standards and
Technology (NIST) of
the United States.
Channel
bonding
ON (40 MHz)
(Recommended)
Channel settings are
used to avoid
interference in
adjacent imaging
rooms.
Communication
environment
Communication
range Within 8 meters Recommended in
order to meet detector
specifications.
Ambient
conditions
• Do not place any structures that
will block the wireless signals
between components of the system.

1
1.
.
O
Ov
ve
er
rv
vi
ie
ew
w
- 11 -
5.6.1 Communication system
2451 2462 2473
1611
2401 2412 2423 2426 2437 2448
2456 2467 2478
2712
2406 2417 2428 2431 2442 2453
2461 2472 2483
3813
2411 2422 2433 2436 2447 2458
2441 2452
49
2421 2432 2443 2446 2457
24632416 2427 2438
2468
510
2400 2410 2420 2430 2440 2450 2460 2470
Japan (Radio Law), Europe (R&TTE), Others (13 channel)
2480 2490
(MHz)
North America (FCC), Taiwan (11 channel)
Figure Channels in 2.4 GHz band
*1: HT20 (20 MHz band) allocates a frequency band of 22 MHz to each channel, so only three channels can operate at the same time.
*2: The channels in the HT40 (40 MHz band) that can be used with channel bonding are listed below.

1
1.
.
O
Ov
ve
er
rv
vi
ie
ew
w
- 12 -
As indicated above, in the 2.4 GHz band only 1 channel in HT20 can be used without interference when HT40 is used.
Primary channel Secondary channel offset Secondary channel Channels that can be used simultaneously without
interference
1CH above 5CH
HT20: Any one of 10CH, 11CH, 12CH, and 13CH
HT40: None
5CH below 1CH
2CH above 6CH
HT20: Any one of 11CH,12CH, and 13CH
HT40: None
6CH below 2CH
3CH above 7CH
HT20: Any one of 12CH, and 13CH
HT40: None
7CH below 3CH
4CH above 8CH
HT20:13CH
HT40: None
8CH below 4CH
5CH above 9CH
HT20: None
HT40: None
9CH below 5CH
6CH above 10CH
HT20:1CH
HT40: None
10CH below 6CH
7CH above 11CH
HT20: Any one of 1CH, and 2CH
HT40: None
11CH below 7CH
Table of contents
Other Canon Security Sensor manuals