Fingertip Pulse Oximeter
User’s Manual (Capnomed)
Ver: 2.0
General description
The Fingertip pulse oximeter, based on all digital technology, is intended
for noninvasive spot-check measurement of functional oxygen
saturation of arterial hemoglobin (SpO2). Advanced DSP algorithm can
reduce the influence of motion artifact and improve the measurement
accuracy during low perfusion.
Technical Parameters
Display Method: Color OLED display, 4 display directions.
Power Requirements: Two AAA size 1.5V Alkaline Bаttеrу
Power Consumption: 30mA(Normal)
Measurement Performance in Low perfusion Condition: 0.3%
Measurement specifications
SpO2
Measuring Range 0-100%
Resolution 1%
Accuracy 70~100%,2%
0-69%,unspecified
PR
Measuring Range 25-250bpm
Resolution 1bpm
Accuracy 2bpm
Product Properties
Operation of the product is simple and convenient.
The product is small in size, light in weight (total weight is about 50g
including batteries) and convenient in carrying.
Power consumption of the product is low and the two
originally-equipped two AAA batteries can be used continuously for 24
hours normally.
Battery voltage will be indicated in visual window when the voltage is so
low that normal operation of the Oximeter might be influenced.
The product will automatically be powered off when no finger is in the
device for longer than 16 seconds.
Product Operation Scope
The fingertip Oximeter can be used to measure human Hemoglobin
Saturation and heart rate through finger. The product is suitable for use
in family, hospital (including clinical use in internist/surgery, Anesthesia,
pediatrics, intensive care and etc.) Oxygen Bar, social medical
organizations, physical care in sports and etc.
The product is not suitable to monitor patient continuously.
Operation Instructions
Press the function button on the front panel to turn on the oximeter.
Nip the oximeter, then insert one of fingers into the rubber hole of the
Oximeter thoroughly before releasing the oximeter.
Change the display direction by pressing the function button again.
Read correspondent datum from display screen.
After turn on the oximeter, each time you press the function button, the
oximeter will switch to another display mode, there are 4 display modes
totally.
Long press the function button to enter the settings screen.
Short press to switch among the settings, long press to change the
current settings of sounds and alarm limits.
When your finger is plugged into the Oximeter, your nail surface
must be upward.
The results may be wrong if you did not plug the finger thoroughly
in the Oximeter.
Declaration: Please use the medical alcohol to clean the rubber touching
the finger inside of Oximeter, and clean the test finger using alcohol
before and after each test. (The rubber inside of the Oximeter belongs to
medical rubber, which has no toxin and no harmful to the skin of human
being).
Brief Description of Front Panel
The length of Bargraph indicates the intensity of the pulse.
Product Accessories
One hang lace.
Two batteries.
One instruction manual.