Capnomed HIMA Plus User manual

HIMA Plus Handheld Vital Sign Monito
User Manual
Company name: Qinhuangdao Kapunuomaite Medical Equipment S&T Co.,Ltd.
Address: 2 Area, 5th Floor, No.10 Standard Workshop, Yanghe Road,
Economic-Technological Development Area,Qinhuangdao City, Hebi Province,
China.
ZIP code: 066000 Tel: +86-335-8595611

HIMA Plus Handheld Vital Sign Monito User Manual(1.0)

HIMA Plus Handheld Vital Sign Monito User Manual(1.0)
-3-
Contents
1. Product Overview....................................................................................................................................... - 4 -
1.1.Application background.........................................................................................................................- 4 -
1.2.Intended use........................................................................................................................................... - 6 -
1.3.Product structure....................................................................................................................................- 6 -
1.4.Working, transportation and storage conditions............................................................................... - 6 -
1.5.Technical specifications........................................................................................................................ - 6 -
1.6.Product Safety Classification............................................................................................................... - 8 -
1.7.Contraindications................................................................................................................................... - 8 -
2. Precautions, warnings and tips..............................................................................................................- 8 -
2.1. Warning...................................................................................................................................................- 8 -
2.2. Matters needing attention.................................................................................................................. - 11 -
2.3. Icon description................................................................................................................................... - 11 -
3. Appearance of the instrument..............................................................................................................- 12 -
4. Instructions for use................................................................................................................................. - 13 -
4.1. Unpacking............................................................................................................................................ - 13 -
4.2. Install SpO2 Probe and EtCO2 Module.......................................................................................... - 13 -
4.3. Factors affecting performance..........................................................................................................- 15 -
4.4. Boot....................................................................................................................................................... - 16 -
4.5. Monitoring process............................................................................................................................. - 16 -
4.6. Shutdown............................................................................................................................................. - 17 -
5. Settings........................................................................................................................................................- 17 -
5.1. Volume setting.....................................................................................................................................- 18 -
5.2. Alarm settings......................................................................................................................................- 19 -
5.2.1 Physiological alarm.......................................................................................................................... - 19 -
5.3. Time setting......................................................................................................................................... - 21 -
5.4. Record settings................................................................................................................................... - 21 -
5.5. System Settings.................................................................................................................................. - 22 -
5.6. File System.......................................................................................................................................... - 23 -
6. Maintenance and maintenance of the instrument.............................................................................. - 24 -
6.1. Troubleshooting.................................................................................................................................. - 24 -
6.2. Battery use and maintenance...........................................................................................................- 26 -
6.3. Instrument cleaning............................................................................................................................ - 26 -
6.4. Waste treatment..................................................................................................................................- 27 -
6.5. Packing list...........................................................................................................................................- 27 -
7. Alarm and indication................................................................................................................................... - 28 -
7.1. Alarm signal......................................................................................................................................... - 28 -
7.2. Default alarm settings........................................................................................................................ - 28 -
7.3. Mute...................................................................................................................................................... - 29 -
7.4. Battery indicator.................................................................................................................................. - 29 -
7.5. Alarm threshold setting...................................................................................................................... - 29 -
7.6. Sound characteristics.........................................................................................................................- 30 -
7.7. Alarm function verification................................................................................................................. - 30 -
8. EMC statement..............................................................................................................................................- 30 -
8.1.Description............................................................................................................................................ - 31 -
8.2.precautions in use................................................................................................................................- 31 -
8.3. precautions regarding........................................................................................................................ - 31 -
9. quality promise.............................................................................................................................................- 34 -

HIMA Plus Handheld Vital Sign Monito User Manual(1.0)
-4-
1. Product Overview
1.1 Application background
The handheld vital signs monitor integrates blood oxygen measurement, end-expiratory
carbon dioxide monitoring, body temperature monitoring and other functions. It uses a color
display to display blood oxygen saturation, pulse rate, EtCO2, respiration rate, body
temperature, and system information. , And error messages (prompts such as falling off of
each parameter measuring probe) and alarm messages.
a Principle of pulse oximetry
Firstly, the difference between the absorption of red light and infrared light by
oxyhemoglobin and reduced hemoglobin (as measured by spectroscopic measurement);
secondly, the flow of arterial blood in the tissue is determined by the pulse (as recorded by
hematogram). ) Change when pulsating. The pulse oximeter adds red light and infrared
light to the arteries and measures the changes in light absorbed during the pulse cycle to
determine SpO2. Low-pressure red and infrared light-emitting diodes (LEDs) in the blood
oxygen probe are used as light sources, and photodiodes are used as light detectors.
Oxyhemoglobin and reduced hemoglobin absorb different light. The amount of red light and
infrared light absorbed by blood is related to the oxygen saturation of hemoglobin. In order
to identify the oxygen saturation of arterial hemoglobin, the oximeter uses the pulsating
characteristics of arterial blood flow. During systole, new arterial blood enters the ventricle,
blood volume and light absorption increase; during diastole, blood volume and light
absorption reach their lowest point. The SpO2 measurement of the oximeter depends on
the difference between the maximum and minimum absorption (eg, measured during
systole and diastole). In this way, the light absorbed by the arterial blood pulsation is the
focus of the problem. It is necessary to eliminate non-pulsatile absorption such as the
absorption effect of tissues, bones and venous blood. The vital signs monitor uses a
photodetector to measure SpO2 and pulse rate through two wavelengths of red light and
infrared light through human tissue. During the measurement, the color and thickness of
the human tissue, the position of the probe, the brightness of the light source, and the
absorption of blood in the arteries and veins of the human tissue affect the result of the
signal strength.

HIMA Plus Handheld Vital Sign Monito User Manual(1.0)
-5-
b End-breathing carbon dioxide concentration (EtCO2)
Human tissue cells produce CO2 through metabolism, which is then transported to the
lungs through capillaries and veins through the body, and is excreted from the body during
exhalation. In this process, the inhaled oxygen is transported by hemoglobin in the blood,
and the CO2 produced in the human body is dissolved in the blood and brought out by
ventilation. According to the law of gas diffusion, CO2 diffuses from high-concentration
areas to low-concentration areas. Therefore, CO2 diffuses from pulmonary blood vessels
(high-concentration areas) to alveoli (low-concentration areas), and then is discharged from
the lungs.
The carbon dioxide monitor adopts infrared monitoring technology. This method is based
on the principle that the absorption rate of infrared light is related to the concentration of
carbon dioxide when the infrared light passes through the gas sample to be tested: CO2
mainly absorbs infrared light with a wavelength of 426nm, which attenuates the infrared
beam, and its attenuation degree It is proportional to the concentration of CO2. When
infrared rays pass through the gas, the CO2 in the gas will absorb part of the infrared rays
and weaken the remaining infrared signal strength. By measuring the initial infrared
intensity and the infrared intensity after the exhalation passes, the CO2 component exhaled
by the patient is calculated, that is, the end-tidal carbon dioxide (EtCO2). details as follows:
The invisible infrared light emitted by the passed infrared light source passes through the
airway adapter, and the respiratory airflow flows in the airway adapter. When the light beam
passes through the airway adapter, some of the light is absorbed by the gas mixture. The
absorbed light beam reaches the infrared detector fixed at the other end. The infrared
detector converts the light beam into an electrical signal. The electrical signal is converted
into a digital value by the ADC and sent to the microprocessor. The microprocessor will

HIMA Plus Handheld Vital Sign Monito User Manual(1.0)
-6-
calculate the carbon dioxide concentration in the breathing gas mixture based on the ratio
of the light of the two measurement channels.
1.2 Intended use
Hand-held vital signs monitors are suitable for medical institutions to perform spot
measurement or continuous dynamic monitoring of patient (or animal) blood oxygen
saturation, pulse rate, carbon dioxide (EtCO2) and respiration rate, and can send out
alarms.
1.3 Product structure
It consists of a host (composed of circuit board, housing, display screen and buttons),
mainstream or sidestream carbon dioxide module (optional), blood oxygen sensor, body
temperature sensor, data transfer cable (optional) and power adapter.
1.4 Working, transportation and storage conditions
Working conditions
Ambient temperature: 5℃~40℃;
Relative humidity: ≤95%;
Atmospheric pressure: 70kPa~106kPa;
Power supply mode: There are two modes to choose from,
which are AC 220V powered by network power and 3.7V
powered by internal lithium battery.
Transportation and
storage conditions
Ambient temperature: -20℃~55℃;
Relative humidity: ≤95%;
Atmospheric pressure: 70kPa~106kPa
1.5 Technical specifications
1.5.1 The technical specifications of the host
SPO2
Measuring range 0~100%
Precision 1%
Accuracy 70~100%:± 2%
sensor The pulsed oxygen sensor contains light-emitting diodes, which emit red
light with a wavelength of about 660 nanometers and infrared light with a
wavelength of about 905 nanometers.

HIMA Plus Handheld Vital Sign Monito User Manual(1.0)
-7-
The total light output power of the sensor led is less than 15mw.
Anti-interference Automatic anti-interference
Anti-interference
ability
Strong anti-hypoperfusion ability, strong anti-hypoperfusion ability, PR
amplitude can reach 0.2% under the displayed value of blood oxygen.
alarm Select high and low alarm limits for blood oxygen
Alarm adjustment
value
stand by
Pulse rate
Measuring range 25 bpm ~250 bpm
Resolution 1 bpm
Precision ±1% or ± 1 bpm
alarm Select the high and low alarm limits of PR
Alarm adjustment
value
stand by
Temperature
Measuring range 25.0℃~45.0℃
Resolution 0.1℃
Precision ±0.3℃
alarm Select high and low alarm limits for body temperature
Alarm adjustment
value
stand by
1.5.2 Technical specifications of the end-tidal carbon dioxide module
If the end-tidal carbon dioxide module is selected, please refer to the module manual for
the technical specifications of the end-tidal carbon dioxide module.
1.5.3 Main functions
Storage: The instrument can store and query the patient's test results and test trend
graphs.
It can monitor whether the blood oxygen sensor falls off.
It can detect real-time waveforms, end-tidal carbon dioxide concentration, inhaled
carbon dioxide concentration and respiratory rate.
The measurement results of end-tidal carbon dioxide are available in mmHg, %, and
kPa.
You can select the display speed, waveform amplitude and other parameters of the
end-tidal carbon dioxide waveform;
The instrument has two operation interfaces in English and Chinese for users to
choose.

HIMA Plus Handheld Vital Sign Monito User Manual(1.0)
-8-
1.6 Product Safety Classification
Classified by type of protection against electric shock: internal power supply equipment;
According to the degree of protection against electric shock: BF.
Classified according to the degree of protection against ingress of liquid: IPX3;
According to the safety level of flammable anesthetic gas mixed in the air or flammable
anesthetic mixed with oxygen or nitrous oxide: this instrument cannot be used in flammable
anesthetic gas or mixed gas with oxygen or nitrous oxide The equipment used in the case.
Classified by operating mode: continuous operation;
The rated voltage of the equipment: power supply AC 220V, 50Hz; internal lithium
battery power supply 3.7V;
Input power of the equipment: 30VA;
Does the equipment have an applied part to protect against the effects of defibrillation
discharge: none;
Whether the equipment has a signal input or output part: there is a signal input part;
Permanent installation equipment or non-permanent installation equipment:
non-permanent installation equipment;
1.7 Contraindications
Not found yet.
2. Precautions, warnings and tips
2.1. Warning
Explosion hazard: Do not use in an environment where flammable anesthetics or
flammable gases are present. There is a risk of explosion when using a carbon dioxide
monitor in this environment.
Risk of electric shock: Always disconnect before cleaning, and do not use damaged
instruments.
Danger of electric shock: There are no user-serviceable parts inside, so
non-professionals should not disassemble it.
This product needs to be used by professional medical staff after training.
Follow measures to prevent electrostatic discharge (ESD) and electromagnetic
interference (EMI) between equipment.
If the product does not achieve the expected function, please do not use it unless

HIMA Plus Handheld Vital Sign Monito User Manual(1.0)
-9-
approved by authorized personnel
It can only be used in the occasions specified in the manual, and can only be used for
the purposes described in the manual.
During use, do not apply external force to the well-connected gas path.
Read the instruction manual carefully before use.
After the instrument and its accessories have reached the end of their service life, you
should dispose of electronic waste in accordance with local regulations.
The measurement results of this device will be interfered by mobile phones or other
radio frequency communication devices. The user must ensure that it is used for
measurement in the specified electromagnetic environment.
This product is only an auxiliary device for clinical evaluation of patients, and must be
used with clinical signs and clinical diagnosis equipment.
Nursing staff may not hear audible alarms in a noisy environment, such as a siren
calling or the nursing staff is far away from the equipment. Therefore, the nursing staff
should do their own monitoring before use to ensure that any alarm sounds can be heard
and confirmed by the nursing staff in time.
During use and maintenance, using parts or materials beyond those specified in this
manual will reduce the minimum safety and product measurement accuracy.
This instrument cannot be used in a magnetic resonance imaging scanning system
(MRI). The induced current generated can be a potential cause of combustion. The
instrument may affect MRI images, and MRI may also affect the accuracy of blood oxygen
and carbon dioxide measurements.
Only use blood oxygen probes and end-breathing carbon dioxide modules that match
this instrument.
This instrument is mainly used in some health care centers that have been specially
trained. The operator must carefully understand the detailed information in the manual
before using it.
Before use, please set an appropriate alarm limit according to the actual situation of the
patient.
For long-term use or according to the needs of the patient, the position of the blood
oxygen probe should be changed regularly. In order to allow the patient's skin to breathe,
the probe should be repositioned at least every 4 hours.
The chemical substance after the display is broken is poisonous and needs to be
handled with care.

HIMA Plus Handheld Vital Sign Monito User Manual(1.0)
- 10 -
Before use, the operator should check whether the host is damaged to ensure the
normal operation of the instrument.
If there is a problem with the measurement accuracy, check the patient's physical signs
and whether the instrument is normal.
This instrument may be interfered by some strong electromagnetic equipment, such as
electrosurgical equipment.
High-intensity light can interfere with the blood oxygen measurement. If necessary, the
blood oxygen probe can be covered with a surgical towel.
Severe cardiopulmonary disease, clogged or twisted sampling tube, and fast breathing
rate all affect EtCO2 measurement. Prevent the patient's sputum and secretions from
contaminating the tracheal tube and affecting the measurement of EtCO2.
The use of blood pressure cuffs or the strong resistance of blood vessels throughout
the body causes obstruction of blood flow, which will result in the inability to accurately read
the value of blood oxygen saturation and pulse rate.
The dysfunction of carboxyhemoglobin or methemoglobin will affect the measurement
accuracy of blood oxygen saturation.
This instrument cannot be used for apnea monitoring.
When two or more blood oxygen probes are placed in close positions, the
measurement accuracy will be affected.
Change the position of the blood oxygen probe at least every 4 hours, and always
observe whether there are signs of tissue damage. The frequency of changing/checking the
position of the blood oxygen probe depends on the ambient light, patient's skin condition,
total exposure time or other factors. The probe position can be changed multiple times.
The blood oxygen cable and CO2 module cable entangle the patient, which may cause
suffocation. Please prevent the cable from getting entangled in the patient. If necessary,
bundle the cables.
In some clinical situations, the monitor may display invalid values ("---"), or even
fail to measure blood oxygen saturation and pulse rate values. In these situations such as
patient movement, hypoperfusion, arrhythmia, rapid pulse, or a combination of the above
conditions, the monitor may also display incorrect values.
In some clinical situations, the CO2 module may display the wrong value of EtCO2 due
to air leakage, twisted catheters, blocked trachea, or the patient's malignant hyperthermia,
elevated body temperature, and increased CO2 production due to injection of a large
amount of NaHCO3.

HIMA Plus Handheld Vital Sign Monito User Manual(1.0)
- 11 -
After the instrument is turned on, please check whether all the indicators on the
instrument are normal.
Since SpO2 measurement passes through arteries, obstruction of blood flow under any
conditions may result in inaccurate measurement of blood oxygen saturation and pulse
rate.
2.2. Matters needing attention
Disposable consumables used up during the use of this product shall be treated as
medical waste, and specific treatment methods shall be implemented in accordance with
hospital regulations.
Do not autoclave the instrument or probe, disinfect with ethylene oxide, or immerse the
instrument in liquid.
If the instrument is damp, thoroughly wipe off the moisture on the surface of the
machine before use to make it dry enough.
Do not use any sharp instruments to operate the keys, as this will damage the keys,
please use your fingers to operate.
Do not splash water or other liquids into the instrument
Some cleaning agents containing chemical substances can cause embrittlement of
plastics. Please clean the instrument according to the instructions on the manual.
Do not use cleaning items made of abrasive materials, such as wire brushes or
polishes.
All materials used by users and patients are non-toxic.
2.3 Icon description
icon meaning icon meaning
Production Date Not recyclable
Product number caveat
Product number Refer to the instruction manual
The degree of protection
against electric shock is
classified as BF
Manufacturer Information
IPX1 waterproof level

HIMA Plus Handheld Vital Sign Monito User Manual(1.0)
- 12 -
3. Appearance of the instrument
The appearance of the instrument is shown in the figure
1Keys: The instrument has six keys, which are as follows
button
Features menu
Enter
Alarm mute
button
Cursor up Cursor
down
return key Power
switch key
②Display screen: display monitoring results and information;
③Power indicator light: when it emits blue light, it means that the instrument is in a
power-on working state;
④Alarm indicator light: understand the alarm status through the power indicator light.
Green means normal operation, red means an alarm has occurred;
⑤Probe socket: connect the monitor and adapter cable or blood oxygen sensor;
⑥Body temperature sensor interface: connect the monitor and body temperature probe;
⑦Power charging line socket;
⑧Loudspeaker: Provides audible alarm and pulse sound. Please make sure that the
loudspeaker is not blocked during work.

HIMA Plus Handheld Vital Sign Monito User Manual(1.0)
- 13 -
4. Instructions for use
4.1 Unpacking
Take out the host and accessories from the packing box, and save the packing box for
use during transportation or storage.
Check whether the instrument is damaged, if there is any problem, please contact the
supplier immediately.
Install the packing list and check whether the accessories are complete.
4.2. Install SpO2 Probe and EtCO2 Module
Attention! 1 Before use, please read carefully the instructions for use of the blood oxygen
probe and ETCO2 module, including all warnings, precautions and instructions.
2 Long-term use and according to the patient's condition, the position of the
blood oxygen probe needs to be changed regularly. Change the position of the probe at
least every 4 hours, check the patient's skin condition, the state of tissue circulation,
and reposition it.
3 Reused probes need to be cleaned and disinfected.
4 Do not use improper blood oxygen probe and EtCO2 module, otherwise
the probe and EtCO2 module may be damaged, and the data may be inaccurate.
5 Before use, please read carefully the instructions for use of the blood
oxygen probe and ETCO2 module, including all warnings, precautions and instructions
4.2.1. First, insert the Y-type adapter cable into the probe socket of the instrument as
shown in the figure. If only monitoring the blood oxygen saturation and pulse rate, insert the
blood oxygen sensor directly into the adapter port
SpO2
sensor interface
EtCO
2interface
Y-type adapter cable
4.2.2. Insert the body temperature sensor into the body temperature sensor interface as
shown below:

HIMA Plus Handheld Vital Sign Monito User Manual(1.0)
- 14 -
4.2.3. Connect the blood oxygen probe to the Y-shaped blood oxygen probe interface, or
you can directly insert the blood oxygen sensor into the probe socket, and then insert the
carbon dioxide sensor module into the Y-shaped carbon dioxide sensor interface, and
buckle it to prevent Falling off during the test
4.2.4. Connect the blood oxygen probe to the patient, please put the wire on the back of
your hand as shown in the figure
below:
4.2.5. Connect the blood oxygen probe to the patient, please put the wire on the back of
your hand as shown in the figure below:

HIMA Plus Handheld Vital Sign Monito User Manual(1.0)
- 15 -
4.2.6 Inspection of the blood oxygen probe and EtCO2 module
Before use, carefully read the instructions for use of the probe and EtCO2, including all
warnings, cautions and instructions.
Check the probe and EtCO2 module carefully before use to ensure that there is no
damage.
Before connecting the blood oxygen probe and EtCO2 module to the patient, check the
probe and EtCO2 module to ensure that the red indicator light inside the blood oxygen
probe is on and the external indicator light of the EtCO2 module is on. Make sure that the
probe and EtCO2 module are clean and free of obstacles, otherwise it may affect
measurement accuracy.
Before measuring the patient, clean or disinfect the blood oxygen probe and EtCO2.
For specific methods, see the instructions for use of the probe and EtCO2.
Disconnect the blood oxygen probe and EtCO2 module from the host before cleaning
and disinfection.
4.3. Factors affecting performance
note! 1. The readings of the blood oxygen module and EtCO2 module may be affected by
the environment, the probe and the patient's own conditions.
2. Please place the probe and EtCO2 module according to the method and position
on the manual!
3. Exercise may cause inaccurate measurement。
4.3.1. Inaccurate blood oxygen measurement may be caused by the following reasons:
Unsuitable probes are used;
Wrongly place the blood oxygen probe on the blood pressure cuff or the arterial vessel
that is measuring the invasive blood pressure;
High-intensity ambient lighting;
The patient moves for a long time;
4.3.2. The pulse signal may be lost in the following situations
The blood oxygen probe is too tight;
Inflate the blood pressure cuff placed on the same limb as the blood oxygen probe;
Arterial occlusion occurs close to the probe;
4.3.3. Inaccurate EtCO2 measurement may be caused by the following reasons:

HIMA Plus Handheld Vital Sign Monito User Manual(1.0)
- 16 -
In patients with severe cardiopulmonary diseases, the V/Q ratio is imbalanced. EtCO2
measured by nasal oxygen tube sampling cannot be used as an indicator of ventilatory
function. PaCO2 must be measured at the same time as a reference.
The sampling tube may be blocked or twisted by secretions, which may affect the
EtCO2 test results.
If the breathing rate is too fast, the exhaled air cannot be completely discharged during
the expiration period.
4.3.4. Under the following circumstances, the value of EtCO2 may increase or decrease:
Reasons for the
increase in EtCO2
The depth of anesthesia is not enough, and the patient
has pain and tremor;
Insufficient ventilation
Upper respiratory tract obstruction;
Repeat breathing
The mechanical void increases.
Reasons for the
decrease of EtCO2
value
Pulmonary embolism;
Hyperventilation
The ventilation system is leaking.
4.3.5. If the patient exercises, the following measures can be used to remedy:
Check whether the probe position is correct;
Place the probe in a position with little movement;
When the patient needs to move, stick the probe together with the patient;
4.4. Boot
Press and hold the power button for about 5 seconds, and the instrument will turn on and
enter the monitoring interface.
4.5. Monitoring process
Warning! 1. In order to ensure the safety of the patient, please do not place the host in any
place where the patient may be injured.
2. Make sure that the speaker is normal to avoid inaudible or low sound when
alarming.
note! The use of the module is carried out strictly in accordance with the operation manual
of the module.

HIMA Plus Handheld Vital Sign Monito User Manual(1.0)
- 17 -
The specific operations are as follows:
The power indicator light is green. Enter the main menu, the instrument emits a beep
meter.
After connecting the patient to the instrument with the blood oxygen probe, the pulse
search icon will be displayed in the status bar. The "---" number will keep flashing until
the blood oxygen saturation and pulse rate readings are stable. The monitor will search
for pulse for approximately 8 seconds.
Start measuring and monitoring the patient's blood oxygen saturation, pulse, EtCO2
and breathing rate.
When the maximum or minimum time limit is exceeded, the upper warning bar will flash
and alarm.
4.6. Shutdown
After the measurement is completed, remove the blood oxygen probe and EtCO2 module
from the patient and save it or dispose of it in accordance with local regulations. When not
in use, press the power button for 5 seconds and then shut down.
5. Settings
Press the menu key to enter the menu interface, the menu
interface is as follows:

HIMA Plus Handheld Vital Sign Monito User Manual(1.0)
- 18 -
In the main menu interface, press the OK key to enter the menu interface, the cursor
stays at the "Volume Setting", press the up and down keys to move the cursor, when the
cursor stays at the content that needs to be set, press the OK key to set the content where
the cursor stays.
After setting, press the return key to return to the previous interface.
5.1. Volume setting
From the main interface, press the enter key to enter the menu interface, and then press
the enter key to enter the volume setting interface, the volume setting interface is as
follows:
The volume of key tone, alarm tone and pulse tone can be adjusted. The volume can
be adjusted up to 5 levels. Press the up and down keys to increase or decrease the
volume.
After setting, press the return key to return to the previous interface.

HIMA Plus Handheld Vital Sign Monito User Manual(1.0)
- 19 -
5.2. Alarm settings
From the main interface, press the confirm key to enter the menu interface, press the
down key, when the cursor moves to the alarm setting, press the confirm key to enter the
alarm setting interface,
the
alarm setting interface is as follows:
The volume of key tone, alarm tone and pulse tone can be adjusted. The volume can be
adjusted up to 5 levels. Press the up and down keys to increase or decrease the volume.
After setting, press the return key to return to the previous interface.
The default alarm threshold is set when the product leaves the factory. According to the
needs of use, each operator enters the alarm setting interface after entering the password,
and enters it through the up and down keys. The specific settings are as follows:
5.2.1 Physiological alarm
Alarm event Alarm
level Whether adjustable Visual alarm Audible alarm
Hypoxemia
(SPO2)
high
Adjustable in the range
of 88~100
The blood oxygen
value flashes;
The red bar
flashes in the
alarm bar;
The warning light
flashes red;
The alarm sounded
quickly, Dididi-Dididi
Hypoxemia
(SPO2)
Adjustable in the range
of 70~85
Pulse rate in
Adjustable in the range
of 50~120
Pulse rate value
flashes
Flashing with red
bars in the alarm
bar;
The alarm tone is high
Didi Didi Didi

HIMA Plus Handheld Vital Sign Monito User Manual(1.0)
- 20 -
Yellow warning
light flashes
temperature in
Adjustable in the range
of 35~38.5
Pulse rate value
flashes
Flashing with red
bars in the alarm
bar;
Yellow warning
light flashes
The alarm tone is high
Didi Didi Didi
End tidal
carbon
dioxide
in
Adjustable in the range
of 25~50
Flashing with red
bars in the alarm
bar;
Yellow warning
light flashes
The alarm tone is high
Didi Didi Didi
Respiration
rate in
Adjustable in the range
of 5~30
Flashing with red
bars in the alarm
bar; Yellow
warning light
flashes
The alarm tone is high
Didi Didi Didi
Inhale
carbon
dioxide
in
Adjustable in the range
of 0~150
Flashing with red
bars in the alarm
bar; Yellow
warning light
flashes
The alarm tone is high
Didi Didi Didi
asphyxia high The limit is adjustable
from 20 to 60s;
Flashing with red
bars in the alarm
bar;
Yellow warning
light flashes
The alarm sounded
quickly, Dididi-Dididi
5.2.2 Technical alarm
Alarm event Alarm
level
Visual alarm Audible alarm
Finger falling off low In the alarm bar, it shows that the
finger is off and flashes;
The alarm tone is low.
Module falls off low In the alarm bar, display the module
falling off and flashing;
The alarm tone is low.
SpO2 sensor
falls off low The alarm bar shows that the
sensor is off and flashing;
The alarm tone is low.
Table of contents
Other Capnomed Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Accriva Diagnostic
Accriva Diagnostic AVOXimeter 4000 Operator's manual

bort medical
bort medical VarioPlus quick guide

NYMAS
NYMAS Nyma Pro GR-12/35 Series Installation and operating instructions

Physio Control
Physio Control LIFEPAK 15 operating instructions

ResMed
ResMed Mirage Swift II user guide

Gima
Gima Jetpull tourniquet Use and maintenance book

Otto Bock
Otto Bock 1C63 Triton Low Profile Instructions for use

Vendlet
Vendlet STANDARD quick guide

DAKEWE
DAKEWE 6250 user manual

Welch Allyn
Welch Allyn PROPAQ CS user manual

Vispero
Vispero Enhanced Vision Merlin Elite Pro HD OCR user manual

Bio Compression Systems
Bio Compression Systems SC-3004-DL operating instructions