CardiAid CT0207RS Application guide

User Manual and
Introduction to the Device
For Models
CT0207RS Semi-Automatic Defibrillator
&
CT0207RF Fully Automatic Defibrillator


User Manual
and
Introduction to the Device
For Models
CT0207RS Semi-Automatic Defibrillator
&
CT0207RF Fully Automatic Defibrillator
ENGLISH

2
ENGLISH
Contents
Overview of CardiAid CT0207RS & CT0207RF............................................... 3
1. Introduction to the Device ................................................... 6
1.1 Intended Use......................................................................................... 6
1.2 User Qualification................................................................................. 6
1.3 Description of the Functions................................................................. 6
1.4 Indications for Use................................................................................ 8
1.5 Contraindications for Use..................................................................... 8
1.6 Important Points in an Emergency........................................................ 8
2. User Manual .................................................................................... 9
2.1 Description of the User Manual........................................................... 9
2.2 Labels on the Device and its Accessories.......................................... 10
2.3 Safety Rules........................................................................................ 11
General Rules............................................................................................ 11
Defibrillation / Use ................................................................................... 12
Electrodes ................................................................................................. 13
2.4 Side Effects......................................................................................... 14
3. Preparing CardiAid for Use ............................................... 15
4. Operation ........................................................................................ 16
4.1 Before Using CardiAid ........................................................................ 16
4.2 Providing Reanimation........................................................................ 16
4.3 After Using CardiAid........................................................................... 20
4.4 Operation Documentation .................................................................. 20
5. Hygiene ............................................................................................. 21
6. Function Test................................................................................. 22
6.1 Periods of User Tests.......................................................................... 22
6.2 Function Check.................................................................................... 22
7. Troubleshooting.......................................................................... 23
8. Disposal............................................................................................ 25
9. Storage.............................................................................................. 26
10. Maintenance................................................................................. 27
10.1 After-Use Maintenance.................................................................... 27
10.2 Periodical Maintenance.................................................................... 27
11. Package Content........................................................................ 28
12. Technical Information ............................................................ 29
12.1 Technical Specifications................................................................... 29
12.2 Pulse Form ........................................................................................ 35
13. Address Information .............................................................. 36

3
ENGLISH
Overview of CardiAid CT0207RS & CT0207RF
1. Device Cover
2. Socket for Electrode Plug
3. Loudspeaker
4. Emergency Kit
5. Defibrillation Electrodes
6. Protection Bag
7. Indicator Window
8. Indicator: “Remove clothing from chest and stick on electrodes”
9. Indicator: “Do not touch the patient from now on”
10. Indicator: “Patient may be touched”
11.
Indicator: “Ready for Shock” (only for CT0207RS)
12. Shock Button (only for CT0207RS)
13. Info-Button
14. Repair Symbol
15. Battery Symbol
16. OK Symbol
CT0207RS

ENGLISH
4
Overview of CardiAid CT0207RS & CT0207RF
CardiAid Emergency Kit
a. CPR Mask
b. Razor
c. Scissors
d. Gloves
e. Alcohol Pad
CT0207RF

5
ENGLISH
Overview of CardiAid CT0207RS & CT0207RF
1. Device Cover
CardiAid is switched on by opening the cover and switched off by closing the cover.
Closing the cover will switch off the device only after the electrodes are disconnected.
2. Socket for Electrode Plug
Electrodes are connected to CardiAid through this socket. Electrodes supplied with the device are
already connected to this socket.
3. Loudspeaker
Audio warnings of the CardiAid are heard through this loudspeaker.
4. Emergency Kit
Emergency kit includes scissors, shaver, respiration mask, gloves, and alcohol pad.
Emergency kit should be replaced after use.
5. Defibrillation Electrodes
Electroshock is delivered to the patient through these electrodes. Electrodes should be replaced
after each use.
6. Protection Bag
Protection bag is used to store, carry, and protect the device.
7. Indicator Window
The status of the device can be observed through this window on the protection bag.
8. Indicator: “Remove clothing from chest and stick on electrodes”
When this indicator lights, you should stick the electrodes on patient’s bare chest.
9. Indicator: “Do not touch the patient from now on”
Patient should not be touched when this indicator is flashing. For example: During heart rhythm
analysis and shock application.
10. Indicator: “Patient may be touched”
While this indicator is lighting, patient may be touched. For example: During cardiopulmonary
resuscitation.
11. Indicator: Ready for Shock (only for CT0207RS)
This indicator flashes when CardiAid is ready to deliver shock.
12. Shock Button (only for CT0207RS)
Shock button starts flashing after shock is prepared. This button is pressed to deliver electroshock.
13. Info-button
When pressed, an audible notification indicating the duration of use and number of shocks ap-
plied is heard.
14. Repair Symbol
The device should not be used, if the repair symbol is flashing or lighting. In this case, it should
be repaired by Cardia International or an authorized service provider of Cardia International.
15. Battery Symbol
The device is not ready for use, if the battery symbol is flashing or lighting. In this case, contact
Cardia International or an authorized service provider of Cardia International immediately.
16. OK Symbol
CardiAid is ready for use, if the OK symbol is flashing while the device is switched off.

6
ENGLISH
1.1 Intended Use
CardiAid is a public access defibrillator (PAD), i.e. an automated external defibrillator (AED) which is
available for public use. CardiAid can be used for the resuscitation of patients older than 8 years (>25
kg) with standard electrodes and patients 1 to 8 years old (<25 kg) with special paediatric electrodes.
If a patient displays symptoms of a cardiac arrest due to ventricular fibrillation or ventricular
tachycardia, CardiAid can be used to deliver the required defibrillation therapy directly on the site of
the emergency. The user is guided through the resuscitation process with clear and comprehensible
instructions. The device automatically records and analyses the ECG signal and, if required, prepares
itself to deliver a shock to the patient.
The process by which the shock is delivered varies according to the version of CardiAid being used:
• In the semi-automatic version (CT0207RS), the user is asked to press a button to release the
shock.
• In the full-automatic version (CT0207RF), the device warns the user not to touch the patient and
then proceeds to releasing the shock automatically.
Important! CardiAid should be used only for the purposes described above.
1.2 User Qualification
In most countries, public access defibrillators (PAD’s) like CardiAid can be used by any rescuer who
is present when a person has sudden cardiac arrest. In some countries, CardiAid can be used only by
rescuers who are qualified with training in basic life support, use of automated external defibrillators
and use of CardiAid.
1.3 Description of the Functions
CardiAid is used to deliver defibrillation to a person having sudden cardiac arrest due to ventricular
fibrillation or ventricular tachycardia. It analyses the heart rhythm of the patient and decides whether
an electroshock is necessary or not. If one is necessary, it prepares the shock automatically. The shock
delivery method depends on the model used (semi-automatic or fully-automatic). After the shock (or
when no shock is advised), CardiAid directs the rescuer to basic life support (CPR) and guides the user
with verbal instructions and metronome. During the incident, duration of use and number of shocks
delivered can be heard by pressing “info-button”. CardiAid also records the ECG and event data in its
internal memory and this data can be obtained from the device as a report. Below, the functions of the
devices are explained briefly, and also will be explained in further detail through this user manual.
Visual and Acoustic Instructions for the User
CardiAid is designed to guide the user with verbal instructions together with
pictures, flashing lights of different colours at the same time, thus maximizing
the performance. CardiAid starts verbal instructions as soon as its cover is
opened and guides the user step by step through the resuscitation process.
Simultaneously, clear images support the verbal instructions. The pictures are simple and explanatory;
the flashing lights are designed to emphasize the pictures and buttons, with white flashing light
showing the stage of the process and red flashing light only indicating the shock button. In this
way, all steps are ensured to be implemented accurately even if the user has limited knowledge or
experience in resuscitation.
1. Introduction to the Device

7
ENGLISH
1. Introduction to the Device
ECG Analysis and Recording
When electrodes are placed on the patient correctly, CardiAid immediately starts analysing and
recording the ECG. According to the result of this analysis, CardiAid decides whether defibrillation
is necessary or not; and informs the user accordingly. CardiAid continues analysing the heart rhythm
until the device is turned off. The ECG analysis is continued also during the charging of the device.
The electroshock is aborted if the device detects a change in the rhythm. A change from “Shock
necessary” to “Shock not advised” is the result of a change in patient’s condition and is not a
malfunction.
Attention!
CardiAid is turned off by unplugging the electrodes and closing the cover of the device.
Closing the cover of the device when electrodes are still plugged in and connected to the
patient does not stop the operation.
Defibrillation
If CardiAid detects a rhythm which requires defibrillation (Venticular Fibrillation (VF) or Venticular
Tachycardia (VT)); it informs the user and prepares the electroshock. In semi-automatic model
(CT0207RS), CardiAid instructs the user to press the shock button to deliver the electroshock. In
the full-automatic model (CT0207RF), the device warns the user and delivers the electroshock
automatically.
The user cannot deliver an electroshock unless the device detects a shockable rhythm and prepares
the electroshock.
CPR Guidance
In the basic life support phase, CardiAid guides the user according to the latest resuscitation
guidelines. It provides metronome signals so that the user can perform chest compressions with the
correct rhythm and number. After 30 signal tones, verbal instruction “Now give 2 mouth to mouth
breaths” is heard, followed by a short silence for rescue breaths. Then, the user is directed to chest
compressions with the verbal instruction: “Now make 30 chest compressions”. This cycle is repeated
for 2 minutes according to the latest resuscitation guidelines.
Info-Button
By pressing the Info-button during basic life support stage, duration of use and number of shocks
delivered can be heard. During this time, the timer for basic life support continues in the background.
Application Documentation
CardiAid records ECG and incident data in its internal memory. These data may be obtained from the
device as a report to be analysed by specialists to define subsequent treatment.
Self-testing
CardiAid performs an automatic self-test daily, monthly and each time the cover is opened (i.e. when
the device is switched on). The status of the device is indicated with flashing status symbols on the
front side of the device.

8
ENGLISH
1. Introduction to the Device
1.4 Indications for Use
CardiAid is indicated to be used on victims of sudden cardiac arrest when;
• The patient is unconscious an unresponsive,
• The breathing is absent or not normal.
CardiAid CA-10ES Adult Defibrillation Electrodes should be used for patients older than 8 years or
weighing more than 25 kg. CardiAid CR-13P Paediatric Electrodes should be used if the patient is 1 to
8 years old or weighs less than 25 kg. Therapy should not be delayed to determine the exact age or
weight of the patient.
1.5 Contraindications for Use
CardiAid should not be used if any of the following signs are present:
• Consciousness and/or responsiveness
• Breathing
1.6 Important Points in an Emergency
If you suspect that a person is having sudden cardiac arrest, keep in mind the following points:
1. Keep calm and proceed rapidly.
2. Check consciousness and breathing.
• Check the victim for a response. Gently shake the shoulders and ask loudly: “Are you all right?”
• Check for normal breathing
Caution!
CardiAid should be used and defibrillation can be delivered only when the person is
unconscious and the breathing is absent or not normal.
3. Phone emergency services and provide the following information:
• Your name
• Your current location
• Number of patients
• Type of emergency (suspicion of sudden cardiac arrest)
• Presence of a defibrillator (PAD/AED)
Attention!
While starting resuscitation, make sure the emergency number is called without delay
(preferably by other people around you).
4. Open the cover of CardiAid. The device will switch on automatically.
5. Follow the instructions exactly. See Section 4 “Operation” for detailed information regarding the
verbal instructions.
Caution!
Note that the information in this user manual does not substitute a basic life support
training.

9
ENGLISH
2. User Manual
2.1 Description of the User Manual
Please read this user manual carefully to ensure safe and effective use of CardiAid and to be prepared
in case of an emergency. If you have additional questions about information in the user manual, you
may contact the local distributor or Cardia International directly. Keep this manual where it can be
reached easily.
The following safety warning icons are used throughout the manual:
Danger!
The icon defines a danger which can result in serious injury or death.
Caution!
The icon defines a possible danger which can result in serious injury or death.
Warning!
The icon defines a possible danger which can result in simple - mild injury. This symbol is
also used to indicate user errors which can result in damage to the device.
Attention!
This icon provides necessary additional information.

10
ENGLISH
2. User Manual
2.2 Labels on the Device and its Accessories
Type Plate Label
Battery Life Label Battery Product Label (on battery) Electrode Product Label
Battery Pack for CT0207/CT0207RS/CT0207RF
Model No: CA-4BP 12V-15Ah-Alkaline
Cardia International A/S, Hersegade 20, 4000 Roskilde, Denmark
www.cardiaid.com
Do not re-use
Keep dry
Avoid physical impact
Avoid physical damage
Keep away from fire
Do not dispose in household waste
Keep away from sunlight
Do not use if package is damaged
Use-by period after package opened
Part number
Use-by date
Batch code
Humidity limits
Atmospheric pressure limits
Serial number of the device
Date of manufacture
Do not dispose the device in common
household waste
Defibrillation protected,
type BF patient connection
Bluetooth
Consult instructions for use
Dust protected,
protected against water jet
Manufacturer
High voltage
Replace battery before this date.
Caution:
Further information in user manual
Temperature limits
IP55
Fragile, handle with care
Transport and store this side up
IMQ S.p.A
2018
Low High
Adult 170J 270J
Pediatric 50J 75J 12V, 15Ah

11
ENGLISH
2. User Manual
2.3 Safety Rules
To ensure safety of the user, the patient and the bystanders, please pay attention to the following
safety instructions which are the requirements of Directive 93/42/EEC:
General Rules
Danger!
To prevent danger of explosion, keep CardiAid away from oxygen sources, flammable
anaesthesia gases and other flammable substance or gas mixtures.
Caution!
CardiAid can be used for the resuscitation of patients older than 8 years (>25 kg) with
standard adult electrodes and patients 1 to 8 years old (<25 kg) with special paediatric
electrodes.
Caution!
If you suspect that a person is having sudden cardiac arrest, check for signs of life, i.e.
consciousness and breathing, before using the device. CardiAid should be used and
defibrillation can be delivered only when the person is unconscious and the breathing is absent or not
normal.
Caution!
Check CardiAid and its accessories for visual damage before using the device. If you
observe damage on the device or its accessories, do not use it. Otherwise there may be
functional errors resulting in the injury of both the patient and the user.
Caution!
Do not use the device if you observe differences in the procedures from the ones described
in the user manual. In this case, immediately contact Cardia International or a Cardia
International authorized service provider.
Caution!
CardiAid can be used only after self-test is completed successfully and any damage or
misuse is not detected.
Caution!
Check periodically whether the device and its accessories are ready for use (See Section
6.2. Function Check for further details).
Caution!
CardiAid can be used to defibrillate a patient on either a wet or metal surface as long as
the appropriate safety precautions are taken. During use of the CardiAid, please be sure
that no one is touching the patient when the shock is being delivered.

12
ENGLISH
2. User Manual
Caution!
Do not attempt to deliver electroshock if defibrillation electrodes are in contact with each
other or are not connected to the patient.
Danger!
Be sure that the electrode cable is not wedged when closing the cover of CardiAid. It may
damage the electrode cables.
Caution!
Charging and delivering electroshock may affect the electronic devices nearby. Check the
function of these devices before using CardiAid.
Warning!
Operation of CardiAid can be affected from electrical and magnetic fields. Keep CardiAid at
least 2 meters away from electrical devices such as cellular phones, walkie-talkies, X-ray
machines etc.
Warning!
Do not immerse CardiAid or its accessories in any liquid. Liquid ingression may cause
serious damage and the device can be unusable.
Warning!
Use only original accessories and spare parts. Using incompatible accessories or spare
parts can cause irreversible damage to the device and serious injuries. Use of non-
approved accessories and spare parts invalidates the warranty of your device and the manufacturer
will not be responsible for any damages caused.
Warning!
Do not open and modify CardiAid. Opening and modifying CardiAid can cause irreversible
damage to the device. This invalidates the warranty of your device and the manufacturer
will not be responsible for any damages caused.
Caution!
Liquid ingressing in the holes for loudspeaker output can significantly reducethe audibility
of voice prompts. Avoid ingress of liquids and position the device vertically in case liquids
have already been ingressed.
Defibrillation / Use
Caution!
Always observe national / regional laws and regulations in effect regarding the use of an
automated external defibrillator.

13
ENGLISH
2. User Manual
Caution!
To prevent injuries of the user, the patient and any by-stander, make sure the patient is
not being touched or moved during defibrillation. Do not touch metal objects or equipment
which are in contact with the patient during defibrillation.
Caution!
Apply the electrodes on patient’s bare chest as shown on the electrode pads. Applying
electrodes incorrectly may cause faulty analysis of heart rhythm and / or faulty or
inefficient defibrillation.
Caution!
In order to prevent faulty interpretation of ECG data, make sure that the patient is lying
still, not being touched or moved while CardiAid is analysing heart rhythm. Do not perform
basic life support (CPR) during rhythm analysis.
Warning!
Before delivering the electroshock, make sure the patient is disconnected from other
medical devices which do not have defibrillation protection.
Warning!
Although CardiAid is one of the safest devices in its class, remember that wrong
interpretation of heart rhythm may be possible.
Warning!
Make sure that the electrodes are placed firmly to the patient’s chest. If not, air between
the patient’s skin and electrodes can cause burns.
Electrodes
Caution!
Only use original electrodes supplied with CardiAid. CardiAid CT0207RS and CardiAid
CT0207RF should be used with CardiAid CA-10ES Adult Defibrillation Electrodes or
CardiAid CR-13P Paediatric Electrodes.
Caution!
CardiAid CA-10ES Adult Defibrillation Electrodes should be used for patients older than 8
years or weighing more than 25 kg. CardiAid CR-13P Paediatric Electrodes should be used
if the patient is 1 to 8 years old or weighs less than 25 kg. Therapy should not be delayed to determine
the exact age or weight of the patient.
Caution!
Never use electrodes which have damage on the package and / or on the pads. Do not use
electrodes after the expiry date written on the electrode package.

14
ENGLISH
2. User Manual
Caution!
Open the electrode package only in case of an emergency and just before use.
Caution!
If the patient has an implanted pacemaker, do not stick the electrodes on the pacemaker.
Using the defibrillator on a patient with implanted pacemakers can lead to incorrect
analysis of the heart rhythm and to irreversible damage of the myocardium in case the electrodes are
placed too close to the pacemaker.
Caution!
Do not place the electrode pad on the nipple.
Caution!
If there is excessive hair on chest, use the razor in the emergency kit of CardiAid (present
inside the cover of the device) to clear the chest before applying electrode pads.
Caution!
Remove all clothing from upper body before applying electrode pads. Clothes or
undergarments with metal parts may cause burns on skin.
Caution!
CardiAid defibrillation electrodes are for single-use only. After using CardiAid, immediately
contact Cardia International or an authorized service provider for replacement of the electrodes.
Caution!
Pay attention to the operating and storage conditions of the device and its accessories,
which are indicated in technical specifications.
The storage outside the specified temperature range will impact the electrode gel contact and
therefore sometimes a 2nd or 3rd shock delivery maybe neccessary.
Caution!
Keep and store the device and its accessories away from children. Electrode cables may
cause strangulation and suffocation.
2.4 Side Effects
The following adverse side effects may occur when CardiAid is used:
• Burns on skin
• Rashes on skin
• Delivering electroshock to a patient who has implanted pacemaker or is connected to other
electronic devices can cause damage to these devices.
• Delivering electroshock to a patient having a non-shockable rhythm may cause fibrillation.

15
ENGLISH
Unpacking the Device
Remove CardiAid from its packaging carefully. Check whether all parts are present according to
“Content of Delivery” on Section 11. Check all components for any sign of damage. Contact your sales
representative or Cardia International directly if there are any missing or damaged components.
Opening the Cover
Open the cover of CardiAid. Device will switch on automatically.
Connecting Electrode Plug
CardiAid is delivered with electrodes pre-connected to the device. Always keep
the device in this condition to help save time in an emergency. If not connected
already, connect the electrode plug to the socket on the device. Special design
of the plug prevents user mistake. It can only be connected as required.
Placing Emergency Kit
CardiAid is delivered with an emergency kit placed inside the cover of the device and electrodes
placed inside the package in front of the emergency kit. Always keep the device in this condition to
help save time in an emergency.
Emergency kit includes a single-use razor, scissors, respiration mask, gloves, and alcohol pad. The
items in the emergency kit are for single-use only. After using CardiAid, contact Cardia International or
an authorized service provider for replacement.
Closing the Cover
Close the cover of CardiAid carefully. The device will switch off automatically.
Danger!
Make sure that the electrode cable is not wedged while closing the cover of the device. It
may damage the cable.
Installation
Different storage options are available for CardiAid. You can choose the product which suits your
needs the best:
• CardiAid Wallmount: It provides practical storage for CardiAid AED. CardiAid Wallmount also
provides storage for spare electrodes, if necessary.
• CardiAid Indoor Cabinets: Specially designed for CardiAid, they ensure that CardiAid AED is
noticeable and easily reachable in case of an emergency while providing its safety.
• CardiAid Outdoor Cabinet: It provides climate protection and high visibility outdoors.
Installation instructions and necessary parts are included in the packaging of the products.
3. Preparing CardiAid for Use

16
ENGLISH
4.1 Before Using CardiAid
Switching CardiAid On
Open the cover of CardiAid. Device will switch on automatically.
Self-testing
CardiAid immediately starts a self-test when switched on. During self-test, all indication and warning
LEDs light up. When self-test is completed, indication symbols show the status of the device. Observe
the status indicators before continuing to use CardiAid. Green “OK Symbol” flashing indicates that
the device is ready to use. The combination of the status indicator lights have different meanings. See
Section 7. Troubleshooting for further details.
Caution!
If the green “OK Symbol” is not flashing, the device is not ready to use. Immediately
contact Cardia International or an authorized service provider.
Caution!
If any of the red “Battery Symbol” or “Repair Symbol” is flashing, immediately contact
Cardia International or an authorized service provider. See Section 7. Troubleshooting for
further details.
Caution!
If the red “Battery Symbol” is flashing, indicating a low battery, immediately contact
Cardia International or an authorized service provider. The battery should be replaced after
max. 3 years or when the device gives low battery warning.
Warning!
Red “Repair Symbol” flashing together with the green “OK Symbol” indicates a warning
that the electrode is not plugged, rather than a problem affecting the operation of the
device. If you observe this situation, plug the electrode to the device.
Caution!
If one or more of the instruction lights does not light during self-test, the light diodes
may be faulty. The device may be used in the current incident, if there is an emergency.
Immediately contact Cardia International or an authorized service provider for repair.
4.2 Providing Reanimation
After opening the cover, verbal and visual instructions guide the user through the whole reanimation
process. In this section, you may find details on how to act on each verbal and visual instruction given.
Caution!
Note that the information in this user manual does not substitute a basic life support
training.
4. Operation

17
ENGLISH
4. Operation
Caution!
The cover of the device should not be closed during operation.
Preparation of defibrillation
1. “Check for breathing” and “Phone emergency services” (This instruction
may be differentiated to fit the national emergency number in your country.)
(These instructions can be deactivated by an authorized service provider) verbal
instructions are heard immediately after opening the cover of CardiAid. The LEDs
around the indicator of the first instruction field (on the left) light.
While starting resuscitation, make sure that the emergency number is called
without delay (preferably by other people around you).
Check the patient for the following signs of life:
• Consciousness
• Normal breathing
CardiAid should be used and defibrillation can be delivered only when the person is unconscious and
the breathing is absent or not normal.
2. “If not breathing normally, remove clothing from chest and stick on electrodes” (This
instruction can be deactivated by an authorized service provider) instruction is heard. Position the
patient on his back on a non-conductive and dry surface. Remove clothing from patient’s chest.
Patient‘s chest should be dry and not very hairy. If necessary, remove hair using the razor in the
emergency kit.
3. “Stick electrodes on patient’s bare chest.”
Open the package of the electrodes. Stick electrodes on patient’s bare chest as shown on the
electrode pads. Press the electrodes firmly to guarantee a good contact.
The instruction “Stick electrodes on patient‘s bare chest.” is repeated every 8 seconds until
electrodes are placed correctly to enable the heart rhythm analysis. If an electrode is detached or
damaged, the instruction is repeated until the contact between the electrodes and the body is re-
established.
Caution!
CardiAid CA-10ES Adult Defibrillation Electrodes should be used for patients older than 8
years or weighing more than 25 kg. CardiAid CR-13P Paediatric Electrodes should be used
if the patient is 1 to 8 years old or weighs less than 25 kg. Therapy should not be delayed to determine
the exact age or weight of the patient.

18
ENGLISH
Caution!
During the whole process, make sure that electrodes are placed firmly on chest and are not
damaged.
4. “Do not touch the patient from now on. Analysing heart rhythm”
These instructions are heard when electrodes are placed correctly, enabling the
analysis of the heart rhythm (ECG). Simultaneously, the green light and LEDs around
the second instruction field (in the middle) light, which indicates that the patient
should no longer be touched or moved.
Caution!
Patient should not be touched or moved during heart rhythm analysis. Do not perform basic
life support during the analysis. This may cause wrong interpretation of the ECG and delay
in the defibrillation process which can be life threatening.
If the patient is touched or moved during analysis which cause interruption in the analysis, a signal
tone and the warning “Movement detected.” are heard.
Caution!
When warning “Movement detected” is heard, check for the cause of the interruption. If
the patient is on a vehicle, stop the vehicle.
After the heart rhythm analysis, CardiAid decides whether shock is required. The device proceeds with
verbal instructions according to the results of the analysis. These instructions will be explained in the
following sections “Shock Necessary” and “Shock Not Advised”.
Shock Necessary
5. If a shockable rhythm (Ventricular Fibrillation (VF) or Ventricular Tachycardia (VT)) is detected,
instruction “Shock necessary. Do not touch the patient from now on!” is heard and CardiAid
starts preparing the electroshock automatically. The instruction “Preparing shock.” is heard next.
Attention!
The process by which the shock is delivered varies according to the version of CardiAid
being used:
• In the semi-automatic version (CT0207RS), the user is asked to press a button to release the shock.
• In the full-automatic version (CT0207RF), the device warns the user not to touch the patient and
then proceeds to releasing the shock automatically.
For CT0207RS:
6a. When the device is ready for defibrillation, the instruction “Press the flashing shock button”
is heard. At this moment, also the LEDs around the shock-button start flashing and the button is
activated. This instruction is repeated with a signal tone until the shock button is pressed. Press the
shock button to deliver the shock.
4. Operation
Other manuals for CT0207RS
1
This manual suits for next models
1
Table of contents
Other CardiAid Medical Equipment manuals