8200-0820-0M Rev A •225D Firmware Update Instructions
Page 1
225D INDICATOR
FIRMWARE UPDATE
INSTRUCTIONS
Items Required
•PC with Microsoft Windows OS and USB port
•AVR-ISP program (Either AVRISP-U or AVRUSB)
•AVR USB interface cable (Dongle)
•Firmware HEX file downloaded from the dealer-protected software section of
CardinalScale.com. Download the file from CardinalScale.com by navigating to 225
Navigator > Resource Type > Software >.
1. Install the AVR software on your computer if not already installed. Please note that either the
AVRISP-U or AVRUSB software/hardware will work with these instructions. Otherwise,
proceed to step 2.
•Refer to Figure 1 on the next page for the AVRISP-U program.
•Refer to Figure 2 on the next page for the AVRUSB program.
IMPORTANT! When loading the AVR program on your computer, do NOT
connect the USB cable to the computer until the AVR program has been loaded
and is operating.
Failure to do so could allow the computer to assign an incorrect driver to the
AVR dongle.
8200-0820-0M Rev A •225D Firmware Update Instructions
Page 2
Figure 1
Figure 2
8200-0820-0M Rev A •225D Firmware Update Instructions
Page 3
IMPORTANT! If the message in Figure 3 appears after connecting the AVR dongle interface
cable to the PC, follow the update wizard to update the firmware. Note that this will require an
internet connection.
Warning!Do NOT connect the ribbon cable to the indicator during this process.
Connecting the ribbon cable to the indicator while updating the firmware may cause the indicator
to fail!
Figure 3
8200-0820-0M Rev A •225D Firmware Update Instructions
Page 4
2. Save the indicator firmware HEX file to a location on your computer such as My Documents.
IMPORTANT! Confirm that the indicator firmware is in a hex file format where the file name
ends with .hex such as 225D.hex or if custom firmware such as B12345.hex or SO123456.hex.
Should the file be compressed, the file name will end with .zip such as 225D.zip. Zip files must
be decompressed before they can be loaded into the indicator. To decompress the file, save
the .zip file to a location on your computer that will be easy to locate like My Documents, then
open the file. The window will have an option to Extract all files. See Figure 4. Click on this and
use the folder extraction wizard to decompress and save the file(s). Ensure that you select a
location to save the file to such as My Documents so you can locate the file when using the
AVR programmer. See Figure 5.
Figure 4
Figure 5
8200-0820-0M Rev A •225D Firmware Update Instructions
Page 5
3. Power the 225D on.
4. When the 225D is in the operation mode, connect the AVR dongle ribbon cable to the port
labeled P8 2560 ISP on the 225D circuit board. See Figure 6.
IMPORTANT! Connecting the dongle to the ISP port before the indictor is fully operating
will cause connection problems.
Figure 6
5. Connect the dongle to a USB port of the computer and open the AVR program.
IMPORTANT! If the message “Your ISP requires a firmware update” is displayed, see
page 3 for instructions.
8200-0820-0M Rev A •225D Firmware Update Instructions
Page 6
6. Click on the Reset button in the AVR program. A message will be displayed at the bottom left
of the AVR screen in the Status Box indicating the condition of the connection. See Figure 7.
•RED light indicates a problem with the connection.
•GREEN light indicates a good read along with the processor detected.
NOTE: The correct processor for a 225D will be “Detected device: ATmega2560”.
Figure 7
Device shown must match device
detected in Status Box
IE: ATmega2560
Status Box
Red light indicates no connection with processor
Green light indicates a good read
Device detected is displayed IE: ATmega2560
8200-0820-0M Rev A •225D Firmware Update Instructions
Page 7
7. The detected device shown in the Status Box must match the Device window at the top of the
screen. See Figure 7. If they do not match use the drop-down list of the Device window to
select the correct device.
NOTE: The correct device for the 225D Main PC Board is ATmega2560.
Note that some revisions of the AVR software will automatically detect that the device is
incorrect and ask to change to the correct device. See Figure 8.
Figure 8
Device selected in Device
window “ATmega128” does
not match the detected device
“ATmega2560”. Click on YES
to swap Device selected to
“ATmega2560”
8200-0820-0M Rev A •225D Firmware Update Instructions
Page 8
8. Next, read the Fuse settings from the 225D Main PC Board. From the menu list at the top of
the screen select Device, Read,and then Fuses. See Figure 9.
Figure 9
8200-0820-0M Rev A •225D Firmware Update Instructions
Page 9
9. The Fuses and Lock Bits screen will open after it reads the fuses of the indicator. The correct
settings are shown in Figure 10.
IMPORTANT! Ensure that the values read, match the values shown in Figure10.
If necessary, change the values that do not match.
Failure to do so will cause the 225D to fail!
Figure 10
8200-0820-0M Rev A •225D Firmware Update Instructions
Page 10
10. From the menu list at the top of the screen select Device and then Auto Program Options.
See Figure 11.
Figure 11
8200-0820-0M Rev A •225D Firmware Update Instructions
Page 11
11. This will open the Auto Program Options window. See Figure 12. Select the options shown in
Figure 12 then click OK to save them. Generally, when using a notebook PC, it is best to
select Medium Programmer ISP Speed.
IMPORTANT: Make sure the selected options match those shown in Figure 12.
Failure to do so will cause the 225D to fail.
Figure 12
8200-0820-0M Rev A •225D Firmware Update Instructions
Page 12
12. Next, load the hex file into the AVR so that it can be transferred to the 225D. From the menu
list at the top of the screen select File, Load,and then Flash. See Figure 13.
Figure 13
8200-0820-0M Rev A •225D Firmware Update Instructions
Page 13
13. Browse to the location where you saved the hex file (.hex). Then click on the file to highlight it
and click Open to select it. The suggested location to store this file in Step 2 was in My
Documents. See Figures 14 and 15. Figure 14 shows a standard 225D hex file while Figure
15 shows a custom firmware hex file. NOTE:Some custom firmware files could also start
with the letter “B” such as “B12345”.
IMPORTANT: It is critical to select a file that ends in .hex.
Failure to do so will cause the 225D to fail.
Figure 14
Figure 15
8200-0820-0M Rev A •225D Firmware Update Instructions
Page 14
14. Now that the flash file is successfully loaded, a confirmation message will be displayed in the
Status Box at the bottom of the AVR screen. NOTE: A good hex file will always start with OC
94 for the first two bits. See Figure 16.
IMPORTANT: Do NOT load a program that starts with anything other than OC 94 as in
Figure 16.
Failure to do so will cause the 225D to fail.
Figure 16
8200-0820-0M Rev A •225D Firmware Update Instructions
Page 15
15. Once the flash file has been successfully loaded into the AVR and the Fuse settings and Auto
Program Options have been verified, either press F5 on the computer keyboard or select
Auto Program from the Device menu to begin programming. See Figure 17.
Figure 17
8200-0820-0M Rev A •225D Firmware Update Instructions
Page 16
16. While programming, a progress bar will be displayed in the Status Box located at the bottom
right of the AVR screen along with a message indicating the current activity being performed.
See Figure 18.
Figure 18
8200-0820-0M Rev A •225D Firmware Update Instructions
Page 17
17. When programming is complete, the Status Box located at the bottom of the AVR screen will
display a confirmation message “Auto Program Complete” along with a GREEN light. See
Figure 19.
If an error occurs during programming a RED light will be displayed along with an error in the
Status Box. If this occurs, first confirm that all settings are correct, save a copy of the EPROM
from AVR to your hard drive, go to Auto Program options and remove the check next to Read
EEPROM, and try to program again.
IMPORTANT: Indicator setup and calibration data are stored in the EEPROM.
If the EEPROM file is lost or corrupted the indicator will require setup and calibration.
Figure 19
18. The update of the 225D indicator is now complete.
19. Disconnect the dongle from the circuit board and confirm the indicator operation.
8200-0820-0M Rev A •225D Firmware Update Instructions
Page 18
8200
-0820-0M Rev A 09/21
Cardinal Scale Mfg. Co.
102
E. Daugherty, Webb City, MO 64870 USA
Ph: 417
-673-4631 or 1-800-641-2008
Fax: 417
-673-2153
www.cardinalscale.com
Technical
Support: 1-866-254-8261
E
-mail: tech@cardet.com

Other manuals for 225D SmartCell

2

Other Cardinal Measuring Instrument manuals

Cardinal 200 User manual

Cardinal

Cardinal 200 User manual

Cardinal 225 Manual instruction

Cardinal

Cardinal 225 Manual instruction

Cardinal 190 Series User manual

Cardinal

Cardinal 190 Series User manual

Cardinal 212 Series User manual

Cardinal

Cardinal 212 Series User manual

Popular Measuring Instrument manuals by other brands

eDAQ mmb16 user manual

eDAQ

eDAQ mmb16 user manual

Electrex ZEPTO D6 installation instructions

Electrex

Electrex ZEPTO D6 installation instructions

Viavi WiFi Advisor WFED-300AC quick start guide

Viavi

Viavi WiFi Advisor WFED-300AC quick start guide

JAVAC JAV-1041 Safety instructions & operation manual

JAVAC

JAVAC JAV-1041 Safety instructions & operation manual

JUMO 202710/30 operating manual

JUMO

JUMO 202710/30 operating manual

Circutor DH96-CPP manual

Circutor

Circutor DH96-CPP manual

TESTO 570 instruction manual

TESTO

TESTO 570 instruction manual

Ingersoll-Rand THERMO KING TriPac EVOLUTION installation instructions

Ingersoll-Rand

Ingersoll-Rand THERMO KING TriPac EVOLUTION installation instructions

QUBINO Smart meter user manual

QUBINO

QUBINO Smart meter user manual

Hioki P9000-01 instruction manual

Hioki

Hioki P9000-01 instruction manual

Riftek RF089 Series user manual

Riftek

Riftek RF089 Series user manual

Instantel Micromate 721A2401 instruction sheet

Instantel

Instantel Micromate 721A2401 instruction sheet

Edwards AIM-S-NW25 instruction manual

Edwards

Edwards AIM-S-NW25 instruction manual

Omron K3MA-L manual

Omron

Omron K3MA-L manual

Vega VEGAPULS 67 Quick setup guide

Vega

Vega VEGAPULS 67 Quick setup guide

YOKOGAWA wt1600 user manual

YOKOGAWA

YOKOGAWA wt1600 user manual

Tense TPM-04SH-DL user manual

Tense

Tense TPM-04SH-DL user manual

Sanwa MT-5 user manual

Sanwa

Sanwa MT-5 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.