Care of Sweden CuroCell IQ Cirrus User manual

Instructions for use
2020-08-14
CuroCell® IQ Cirrus
Automatic active airflow controlled mattress system
Instructions for use item number: 95-001450-EN0000

2
Table of contents
WARNINGS AND SAFETY PRECAUTIONS .......................... 3
1.
Introduction ............................................................... 4
2.
Assembly/installation................................................. 6
3.
Operation................................................................... 7
4.
Product description.................................................. 12
5.
Reuse, cleaning and reconditioning........................... 14
6.
Storage ....................................................................... 14
7.
Maintenance............................................................... 15
8.
Troubleshooting ......................................................... 16
9.
Technical description................................................ 17
10.
Spare parts ............................................................. 21
11.
Options and accessories ......................................... 22
12.
Warranty................................................................ 22
13.
Other information .................................................. 23

3
WARNINGS AND SAFETY PRECAUTIONS
Read all instructions before use or repair
WARNING! To minimize the risk of fire, personal injury and equipment/property damage adhere to the following
instructions:
1.
The product must only be installed and used for its intended purpose according to the instructions in this manual and/
or other documentation from Care of Sweden. The product must not be combined, assembled or repaired with parts
(e.g. control unit and mattress), accessories or spare parts other than those described in this manual or other
documentation from Care of Sweden. The product must not be modified in any way.
2.
The product must be placed and used so that it does not become trapped or damaged. Note: Be particularly aware of
trapping damage when using side rails. Always make sure that the mattress is the correct size for the bed.
3.
Regularly check product functionality by performing a hand check (see section 3.3).
4.
When the product is used for individuals needing special supervision, such as children, continuous monitoring is required.
5.
The mattress is protected by a hygiene cover; avoid using multiple hygiene covers as this can affect the vapor
permeability of the mattress.
6.
The hygiene cover does not allow liquid or air to penetrate, but is vapour permeable. Make sure that the user is
positioned correctly to avoid the risk of suffocation.
7.
Be careful with sharp objects to prevent damage to the hygiene cover.
8.
Do not open the control unit housing – risk for electric shock. Servicing and maintenance must be performed by
Care of Sweden or one of its authorized service technicians.
9.
Route the power cable to the control unit carefully to avoid tripping. Also make sure that the user is lying correctly on
the mattress according to the instructions and use a cable holder if possible.
10.
To avoid the risk of strangulation, make sure that the cable and tubes are routed to prevent someone getting caught up in
them.
Note: The responsible healthcare provider must inform the user/operator of the following:
11.
Do not use the product in bathrooms or other area where there is a risk of the control unit coming into contact with
water or other liquids. Except for specified cleaning, never handle a product that has come into contact with
water/liquid. Pull the plug out of the electrical socket immediately and send the product to an authorized service
technician for servicing
12.
Do not use close to or in contact with fire sources/hot surfaces, such as fire, burning cigarettes, hot lamps, heating fans
or heating stoves/open fires as this could damage the product.
13.
Do not store or use the product in direct sunlight. The product may be damaged by the elevated temperature and UV
light.
14.
Strong magnetic fields or wireless communication equipment (e.g. wireless home network products, mobile phones,
walkie-talkies, cordless phones and their base stations, radio transmitters, etc.) may affect the product’s functionality
and should be kept at a distance of at least 1 meters from the control unit.
15.
Never use the product if the power cable, plug of the control unit or power supply housing is defective, if the control
unit housing is damaged, or if it is not functioning properly. Contact an authorized service technician for examination
and repair.
16.
Never connect anything other than the Care of Sweden supplied power supply to the control unit power cable
connector.
17.
Never use the external communication input (3,5mm connector), this input should only be used by the manufacturer.
18.
If the hygiene cover is equipped with side handles, these are intended for managing or relocating the mattress. Do
not use the handles to lift the mattress with an user lying on it. All other use takes place under your own liability and
is not covered by the product warranty.
19.
To prevent the power supply from being pulled out, exercise caution when there are children and pets in the
environment around the equipment.
20.
Use of this product adjacent to or stacked with other equipment should be avoided because it could result in improper
operation. If such use is necessary, this product and the other equipment should be observed to verify that they are
operating normally.
21.
In order to minimize the risk of wounds occurring on the feet, make sure that the user doesn’t come into contact with
the hangers of the control unit.

4
CuroCell® IQ Cirrus is a fully automated and modern hybrid
mattress system designed to provide excellent comfort while
effectively aiding in the prevention and treatment of
pressure ulcers/pressure injuries. The system identifies the
user and automatically manages all settings to optimize the
pressure according to the user’s needs.
Unprevented/untreated pressure ulcers/pressure injuries
can very quickly deteriorate resulting in higher risk of
complications, pain and suffering for the user.
Always read the instructions for use
prior to use.
For further questions regarding installation, use or
maintenance please contact Care of Sweden or your local
distributor.
1.1
General information
This system is a medical device with CE marking in
accordance with Medical Device Regulation MDR (2017/745).
The system has been fully or partially designed and verified in
accordance with the standards listed in section 9.2.
Standards.
According to statutory regulations made by the authorities
regarding medical devices, the manufacturer is required to
report all accidents or incidents involving the products. We
would be very grateful for all information involving accidents
or incidents relating to our products, by reporting them
immediately to us, at Care of Sweden.
1.2
Intended use
The systems consist of a control unit actively controlling and
managing the air flow in the mattress, enabling them to be
used for prevention and as an aid in the treatment of
pressure ulcers/pressure injuries (PU/PI).
1.3
Indications
For users with pressure ulcers/pressure injuries (PU/PI) up to
and including category 4 and unclassified PU/PI (unstageable
and suspected deep tissue injury¹). As well as for users in risk
of PU/PI.
1.4
Contraindications
There are no known contraindications.
1.5
User application
The systems can be used in all kinds of health care
environments.
The mattress is intended for use by persons of a
recommended minimum length of 120 cm. The
recommended user weight is up to 230 kg.
We recommend that position changes are made regularly. The
time interval must be evaluated by the responsible personnel,
depending on the user’s status, diagnosis and general
condition.
The mattress should only be used lying
in the lengthwise direction on the mat-
tress with the feet at the end, marked
with the feet symbol.
No training or special skills are required for the use of
CuroCell® IQ Cirrus. It is designed to be easy to handle and
understand, reading the quick guide and the instructions for
use before use is recommended.
Prescription shall be made by persons authorized for
prescriptions and with clinical education. Note that the
user also may be the operator.
(1) European Pressure Ulcer Advisory Panel, National Pressure Injury
Advisory Panel and Pan Pacific Pressure Injury Alliance. Prevention and
Treatment of Pressure Ulcers/Injuries: Clinical Practice Guideline. The
International Guideline. Emily Haesler (Ed). EPUAP/NPIAP/PPPIA: 2019
1.6
Other information
Action plan for pressure ulcers/pressure injuries
For the best possible results when using this product, we
recommend a structured and planned approach. Examples
that can be applied to the use of our products can be found
in the ‘Action plan for pressure ulcers’ guidelines.
The Mattress Guide
This guide provides guidance to personnel and prescribers
when choosing a mattress from Care of Sweden. The
CuroCell® IQ Cirrus is a mattress in function group C.
This information can be obtained from our website, www.
careofsweden.com, or ordered from Customer Services (see
contact information on the final page).
1. Introduction
NOTE:
1. For certain users, e.g. amputees, the recommended length
measurement may not be reached. Users in these groups may,
however, require other settings, as the entire surface is not
under load. For function controls, refer to section 3.2.
2. When used together with positioning pillows, read the pil-
low instructions for use carefully for correct positioning of
the user.
3. Whenever this product is used with evacuation equipment
(evacuation cover or sheet), it is the responsibility of the
authorised personnel to ensure that evacuation can be
performed safely.
4. The mattress may be inappropriate for use during x-ray
examinations because of the risk of blurred images or
artefacts that may lead to diagnostic errors.
5. In the event of a power cut or similar, the mattress will
retain air for at least 12 hours.

5
1.7 Evacuation instructions
Evac is a mattress cover with a built-in function for
evacuation of bed-bound users. In case of fire or other
emergency, there is no time to think about what to do next.
Therefore, it is very important that the safety equipment, as
well as the evacuation procedures, are well planned and
known in advance.
The Evac cover is equipped with three handles on the head
and foot end. On the sides, there are handles with velcro
straps. These straps are to be used to fixate the user to the
mattress before commencing the evacuation. The straps are
folded in the protective pockets on the side of the mattress
with the ends sticking out to ensure that they are easy to pull
out.
Evacuation instructions
Down from the bed
• Inform the user about what will happen.
• Lower and lock the bed in position.
• Disconnect the tubes from the control unit.
• Pull out and fasten the velcro straps.
• Use the outer handles at the foot end of the mattress and turn the
mattress out from the bed.
• Slowly pull the user down on to the floor.
On an even surface
• Pull on the outer handles at the foot end
of the mattress.
• Working position should be a little bit
leaning back.
Down the stairs
• Walk backwards until the main part of the mattress has slide pass
the top stair.
• Turn the back to the user and look in the moving direction.
• Hold one hand on the stair handle and
pull with the other hand
on the middle handle of the mattress.
• Walk down the stairs. It is important to
maintain speed, without running.
Training/practise
To bear in mind when training.
Pulling down from the bed:
• The user in the bed should not be at the very top of the mattress.
There could be a risk of hitting the head in the bed when pulling
down.
Pulling in the stairs:
• Start with pulling an empty mattress down the stairs, in order to
practice the technique.
• When training with a person on the mattress, a string or rope can be
fasten in the handles at the head end. Another person can then walk
behind the mattress and help out if problems should occur.
• It is advisable to practice in the same stairs where an evacuation can
be necessary.
General:
• The main purpose of the training is not to reach high speed, but to
learn the correct technique.
Training covers
• Because of the heavy strain on the equipment when evacuating,
always check the seams and handles at the end of training. We
recommend that equipment used for training is marked with
“Training equipment” and is only used for this purpose.
Explanations
• If the velcro straps are not used, there is a risk of the user
sliding/rolling down from the mattress.
• Pulling at the foot end reduces the strain on the puller, which means
faster and safer evacuation is possible.
• Pulling in the stairs is the most difficult part of the evacuation and
should be practiced extra carefully.
• The arms of the user shall be inside of the velcro straps. This way
reduces the risk of the user grabbing hold to something out of fear.
NOTE!
When using wider mattresses, make sure that the evacuation
routes are clear (size of doors etc.). Responsible personnel
have to ensure that the evacuation can be performed safely. To
ensure proper function of the Evac cover, it is important to
regularly check the quality of the mattress. Check for damages
that could cause problems when evacuating.
WARNING!
Person to perform an evacuation needs to have the necessary
training. Always check the equipment, cover and mattress
before commencing, to make sure an evacuation can be
carried out safely. It is the user’s responsibility to ensure that
it is safe to use for evacuation.

6
When unpacking, check that no parts are damaged. If any damage is found, contact Care of Sweden or your local distributor
before using the product.
1. Place the mattress on the bed
base. Secure the mattress to
the bed using the fastening
straps on the underside of the
mattress.
Note:
•Make sure that the mattress is the correct size for the
bed.
•Make the bed with sheets for added comfort.
•If an adjustable bed is used the mattress should be fixed
only to the moving parts of the bed.
•Check the cells and press studs to ensure they are
correctly assembled.
2. Hang the control unit on the
foot end of the bed or place it
on a level, steady surface.
3. Place the power cable in the
cable holder by opening the
press studs, placing the cable
in the gap and closing the press studs again. If the cable
holder is not used: place the power cable so there is no risk
of stumbling over it, running over it with the bed wheels, or
getting it jammed when raising or lowering the bed.
4. Open the lid on the air tube
connector (marked ‘CPR’) and
connect it to the side of the control unit.
5. A click is heard and felt
when correctly connected.
Secure that both sides of
the connection are closed.
6. Check that the switch on
the side of the control unit
is set to ’0’ (off). Plug the
power supply into an approved
electrical socket (100– 240 V).
7. Check that the power cable
has been correctly connected
to the control unit and that the
correct power supply has been
used. (See section 9.1 Technical
specification).
The correct P/N must be shown on the label on the power
supply. The power supply is part of the equipment and may
not be replaced.
If the control unit has been stored in its minimum or
maximum storage temperature (-25°C to 70°C), wait at least
1 hour before starting it. This time is based on an ambient
temperature of 20°C.
2. Assembly/installation
.
NOTE:
.
Do not hold the 12 V plug on the power supply while touch-
ing the user.
Plug the power supply into an easily accessible socket if
there will be a need to unplug the power supply and com-
pletely cut the power to the control unit.

7
3.1 Start
1. Set the On/Off switch on the side of the control unit to position 1 (On).
2. Air will start to be pumped into the mattress. During the inflation
period, the mute diode and the notification diode to the right will
remain orange. When the orange lights are extinguished, the mattress
system is fully inflated.
3. The diode over the checkmark will blink during the inflation. Depending
on its size, it takes approximately 20–30 minutes for the mattress to
inflate fully.
4. When the green diode over the checkmark is constant, the system has
adapted the settings to the user’s individual needs and the system can be taken
into use. This takes approximately 20-30 minutes.
3. Operation
NOTE:
•During the automatic setting, try to avoid larger movements on
the mattress, otherwise, the automatic setting time will be
extended.

8
3.2 Control unit functions
Button
Function
Mute the information signal
Pack & Go. Automatic deflation of the system
Maximum pressure (caring mode)
Information signal
Incorrect connection of the air connector (CPR)
System is ready to use
Notification diodes

9
3.2.1 Automatic Setting
The control unit is equipped with a function that
automatically sets the pressure based on the user’s weight,
posture and position. This function works in the following
three ways:
1. The control unit carries out an automatic setting
immediately after switching the system on.
2. During usage, if any significant change occurs, the control
unit will perform an automatic setting.
3. When in use, the control unit will carry out automatic
settings at fixed intervals to ensure correct control of the
pressure in the mattress at all times.
Once the automatic setting has completed, the control unit
will start in the Pulsating mode.
3.2.2 Programs
CuroCell®IQ Cirrus is preset with a Pulsating mode, a
comfort mode that combines Constant low pressure
(CLP) with Alternating mode. A cycle period of 10
minutes is default.
3.2.3 Maximum pressure (caring mode)
With this function, the entire mattress is
inflated and provides firmed support. This
function automatically reverts to the
previous setting after approx. 20 minutes.
The function should be used when caring the user, shifting
the user’s position or moving the user in or out of bed.
3.3 Hand check (function control)
The hand check is used to ensure that the mattress is
working properly. This must be performed regularly; we
recommend once per shift, or every eight hours.
Open the cover and insert a hand between the upper cells
beneath the user’s sacrum (center of the mattress). Check
to ensure that there is a gap to the underlying surface so
that the user does not ‘bottom out’. If you can feel the
user’s sacrum resting in the palm of your hand, the gap is
too small. Make a new automatic setting or see section 8
Troubleshooting.
3.4 Sitting positioning in bed
When raising the head end of the bed into a sitting position,
always secure the user´s position. To ensure the product
functionality, we always recommend to perform a hand-
check (see section 3.3). For additional support, positioning
pillows can be used. We recommend that a sitting position
only shall be used for short periods of time.
3.5 CPR (Cardiopulmonary resuscitation)
In case of an emergency
where CPR
(cardiopulmonary
resuscitation) is
necessary, remove the
connection (marked ‘CPR’)
from the control unit and
leave the lid open in order
to empty the mattress of
air quickly.
3.6 Notifications
Different notifications exist based on how serious
the warning is. With a malfunction or an error, a
notification will be given by way of a flashing
warning triangle. To mute the warning signal,
press the mute button.
The notification code is shown on the
four different diodes above the wrench
symbol. For more information about
the different notification codes, see section 3.6.1.

10
3.6.1 Table of notifications
Information about each notification is shown in the
notification table:
•Notifications from 1-10 are both audible and visual.
The error code will be displayed until the error has been
rectified. If the mute button is pressed the audible
warning will crease for a period of 5 minutes and will
return until the error has been rectified.
•The notifications 11-13 have no audible alarm. The error
code is shown until the system is restarted.
Notification
Description and
troubleshooting
1
High temperature. Valves and
compressors are turned off.
If the control unit is in direct
sunlight, relocate it.
Otherwise, contact the
support.
2
Default settings are not
completed. Contact the
support.
3
Incorrect input voltage. Make
sure that the correct power
supply is used, otherwise
contact the support.
4
Low pressure. Secure the
CPR, mattress, air tubes and
air filter. If the problem
remains, contact the support.
5
Automatic setting failure. The
correct pressure has not been
reached within the time limit.
If the problem remains,
contact the support.
6
Leakage under the automatic
setting period. The mattress
is leaking too much for the
weighing to be accomplished.
Control the mattress and the
air connections. If the
problem remains, contact the
support.
7
High pressure. The pressure
cannot be reduced to the
desired value within the time
limit. Contact the support.
8
The automatic setting has
been restarted too many
times during the automatic
setting period. Contact the
support.
9
The mattress control
parameters have not been
read. Connect the CPR or
contact the support.
10
The mattress control
parameters have been
changed during the use.
Restart the system. If the
problem remains, contact the
support.
11
Leakage in one of the cell
sections. Secure the CPR,
mattress and connection
tubes. If the problem
remains, contact the support.
12
Leakage in one of the cell
sections. Secure the CPR,
mattress and connection
tubes. If the problem
remains, contact the support.
13
Leakage in one of the cell
sections. Secure the CPR,
mattress and connection
tubes. If the problem
remains, contact the support.
3.6.2 Maximum pressure notification
When the function Maximum pressure has
been used for a long time, the Maximum
pressure diode will blink. If the use is
intentional, ignore the notification.
3.7 Transport function
If the user needs to be moved in bed, there are two ways
to perform the transportation:
•Unplug the CPR connection, close the lid, place the CPR
connection at the end of the bed and remove the
control unit from the bed. The mattress will retain air
for at least 12 hours.
or
•Remove the power supply from the wall socket and
leave the control unit hanging on the bed during
transport. The mattress will retain air for at least 12
hours.
We recommend using this function for short periods only.

11
3.8 Pack&Go®function (deflation)
After use, the product can easily be packed as
follows:
•Ensure that no-one is lying on the
mattress.
•Press the Pack&Go® button and hold
it down for 2 seconds.
The Pack&Go®diode will flash during deflation. The mattress
will empty of air and be ready to be simply folded together
within 20 minutes. The control unit gives an audio signal
once deflation is complete.
Carefully fold the mattress together, place the control unit
between the folds of the mattress and place the system in a
transportation bag (accessory) or equivalent for protective
storage. Ensure that the power supply is packed complete.
3.9 Restart
If a restart is
required, set the
On/Off switch on
the side of the
control unit to
position 0 (Off).
Wait for approx. 10
seconds and restart
the control unit.
The control unit
will now start up. See section 3.1.

12
4.1 Control unit
1
1. Control panel
2. Tube/CPR connection
3. Power switch, On/Off
4. 3,5 mm plug input (only for use by manufacturer)
5. Connection power cable
6. Air filter
7. Hangers
8. Power supply
2
3
8
4
5
6
7
4.2 Mattress
Heel function
The CuroCell® Cirrus mattress have a built-in heel
function, designed to reduce pressure on sensitive
heels.
Covers
The CuroCell® Cirrus mattress is supplied with a hygiene
cover. It is easy to handle and maintain and conforms to
the stringent requirements in respect of cleaning and
hygiene. The hygiene cover is vapour-permeable, i.e.
the vapour is transported away, thus reducing the risk
of skin maceration. The following covers are available
for the CuroCell® Cirrus mattress:
Cover Stone/Evac – A removable cover with a drip-
proof zip fasterner and welded seams. Divisible (top
and bottom part).
The Evac cover is equipped with three handles on the
head and foot end. On the sides, there are handles with
velcro straps. These straps are to be used to fixate the
user to the mattress before commencing the
evacuation. The straps are folded in the protective
pockets on the side of the mattress with the ends
sticking out to ensure that they are easy to pull out.
Colour: dark grey/black.
4. Product description

13
CuroCell®IQ Cirrus
1. Mattress
2. Tube connection & CPR (quick deflation)
3. Control unit
1 2 3 CuroCell® IQ Cirrus with cover Stone
Hygiene cover
Cells
Foam core
Inner cover
Tubing set
Bottom part

14
The product is reusable. When reusing, it is important to
follow the instructions below for cleaning and
reconditioning. Before using the product again for a new
user, it is recommended to clean the mattress in
accordance with the instructions below. Consult your
hygiene manager or Care of Sweden for help and
instructions if you are unsure.
Cleaning and reconditioning should otherwise be performed
as required.
5.1 Cleaning
CONTROL UNIT
Wipe off the control unit with a damp cloth and
mild cleaning solution for this purpose: such as
alcohol with or without tenside surfactants,
oxidizing solutions, chlorine max 1 % or hydrogen
peroxide max 1,5 %.
If using a different agent, choose one that does not affect the
control unit.
MATTRESS COVER
Covers can be wiped off with a mild cleaning solu-
tion for this purpose such as: alcohol with or with-
out tenside surfactants, Isopropanol 70 %, oxidizing
solutions, chlorine max 1 % (max 10% for cover Stone) or
hydrogen peroxide max 1,5 %.
Excrement and blood stains must be removed as soon as
possible using cold water. Carefully follow local instructions
and the instructions for the detergent.
Multi-layer covers should be divided before washing. The
parts should be washed with similar colours.
Chlorine and phenol-based cleaners could adversely
affect the PU surface and should be avoided. If chlo-
rine is used, we recommend a mixture of max 1 %
(max 10% for cover Stone).
INNER COVER
Clean the affected area with a mild cleaning solution for this
purpose such as: alcohol with or without tenside surfactants,
Isopropanol 70 %, oxidizing solutions, chlorine max 1 % or
hydrogen peroxide max 1,5 %.
5.2 Reconditioning
CONTROL UNIT
See section 5.1 for instructions.
MATTRESS
1. Disconnect the tube connector from the control unit
and remove the air from the mattress.
2. Remove the covers. For more information of cleaning
the covers, see section 5.1.
3. Move the mattress to a clean area that is suitable for
cleaning.
4. Wipe off the cells, all tubes and the CPR module with a
mild cleaning solution for this purpose such as: alcohol with
or without tenside surfactants, Isopropanol 70 %, oxidizing
solutions, chlorine max 1 % or hydrogen peroxide max 1,5
%. Let it dry.
5. Wipe off the work surfaces using a suitable detergent and/
or disinfectant.
6. Place the air cells back into the foam core. If cells for any
reason have become loose from the tubes, these must be
replaced according to the drawing in section 4.2.
FOAM CORE
Clean the affected area with a mild detergent (such as
washing-up liquid) and water or with an alcohol-based
disinfectant (cleaner intended for this purpose).
Gently squeeze out any water.
It is advisable to store the mattress and control unit in the
product bag (accessory), original package or equivalent for
protective storage. Handle the packaged product with
caution. Do not place any heavy objects on top of it. For
additional information about storage temperature, see
section 9.
5. Reuse, cleaning and reconditioning
6. Storage
NOTE:
Check the hygiene cover, cells and hoses each time the product is
cleaned.If damaged, it must be replaced or repaired. Also check
the control unit, tube connectors and power cable during
cleaning. Damaged parts must be replaced or repaired.
Do not wring or roll the foam core to extract the water. Let it
dry in a warm, ventilated area (not in direct sunlight). The foam
core must be completely dry before it is used again.

15
7.1 General
The CuroCell® IQ Cirrus hybrid mattress is a medical
device, in accordance with Medical Device Regulation
MDR (2017/745), that is used as an aid in preventing and
treating pressure ulcers/pressure injuries. For this reason,
we recommend that the control unit be regularly serviced
and inspected to maintain functionality and performance.
Like other technical devices, a control unit that is properly
cared for works better and lasts longer.
Servicing and maintenance must always be performed by
Care of Sweden or one of its authorized technicians. Only
use spare parts that have been approved by Care of
Sweden. Use of non-approved spare parts invalidates the
warranty. For more information, see the CuroCell® IQ
Cirrus service manual.
Faults and defects regarded as warranty faults are fixed
free of charge.
A warranty claim must be submitted to Care of Sweden be-
fore the product is returned
After the product is used, check that:
1. The power cable and power supply are undamaged.
2. The connecting tubes (marked CPR) on the side of the
control unit are positioned correctly and not leaking.
3. The hygiene cover is intact and the cover and cells are
correctly assembled.
4. No tubes or connectors are damaged or jammed.
Contact Care of Sweden or your local distributor if any spare
parts are required.
7.2 Replacing the air filter
Before any maintenances are done, make sure that the
control unit is turned off. Services shall not be done while
using the product.
7. Maintenance
To replace the air filter:
1. Loosen the small protective plate on the rear of the
control unit using a size T10 Torx screwdriver.
2. Remove the filter from the holder.
3. Place the new filter in the holder with the pink side facing
outwards. Put the protective plate back in place and secure
using the screws.
If the control unit is used in a
very dirty environment the
filter should be checked regularly.

16
Problem Solution
The control unit does not
start
Check that the power supply has been connected to the mains supply. Check
that the LED on the power supply is showing green.
The user is ‘bottoming out’ Restart the control unit. See section 3.9. The control unit will initiate an automatic setting. Wait until the
automatic setting is complete.
Perform a further hand check (see section 3.3). If the gap is still too small, raise the comfort setting in
stages.
If the problem keeps occurring, contact Care of Sweden or your local distributor.
The mattress moves around Check that the mattress is fastened to the bed frame with the straps underneath (two at the head end and
two on each of the long sides).
Some cells have less air This is normal with a Pulsating or Alternating mode, as the air supply switches between alternating cells
for a predetermined cycle period (one cycle = 10 minutes).
The control unit makes a
noise; vibrations can be
felt
Check how the control unit is hung on the bed. Resonance can occur, vibrations are felt in parts of the bed.
Remove the control unit and listen to see whether this makes a difference. Can be resolved by putting the
control unit on a flat, steady surface or by placing a towel between the control unit and bed.
If the problem keeps occurring, contact Care of Sweden or your local distributor.
If the above information does not answer your questions, please contact Care of Sweden or your local distributor.
8. Troubleshooting

17
9.1 Technical specification
NB: Care of Sweden reserves the right to modify the product specification at any time.
CONTROL UNIT SPECIFICATION
Model
CuroCell® IQ Cirrus
Input voltage
100-240 V / 50-60 Hz
Power consumption
1.5 A
Mode of operation
Pulsating
Duty cycle 10 min
Power supply Ungrounded AC outlet, electrical safety
class II
Use only power supply with P/N WR9QE1500LRPCIMG3138
Dimensions (L x W x H)
11 cm x 30 cm x 20 cm
Weight
2.9 kg
Sound level, control unit,
max:
17 dBA (at operator position), 16,5 dBA (at head end). Specifies the
highest value in accordance with EN ISO 11201:2010 when placing
the control unit on the foot end.
Environmental
Temperature Operation: +5 to 40°C
Storage: -25 to 70°C
Transport: -25 to 70 °C
Humidity Operation: 15 % – 93 % non-condensing
Storage: < 93 % non-condensing
Atmospheric 700 hPa – 1060 hPa
Electrical classification
Class II, Type BF
IP classification
IP42
Degree of safety in
presence of inflammable
anesthetics:
The device is not intended for use with flammable anesthetic gases
Applied part
Mattress
MATTRESS SPECIFICATION
Model
Dimensions (WxLxH)
Weight
CuroCell® Cirrus 80/85/90/100/105/120x200x21 cm
18 kg (width 85 cm)
9.2 Standards
The system is tested and approved according to the following European standards where applicable requirements are met.
IEC 60601-1
EN ISO 10993
ISO 3746
IEC 60601-1-2
EN 12182
ISO 11201
IEC 60601-1-11
IEC 60601-1-6
IEC 62304
EN 597-1
EN 597-2
EN ISO 14971
EN 14126
9. Technical description

18
9.3 Marking
THE CONTROL UNIT
The control unit is marked as shown below (example). For an explanation, see section 9.4 Explanation of symbols.
POWER SUPPLY
The power supply is part of the control unit and is marked as shown below (example).
For an explanation, see section 9.4 Explanation of symbols.

19
THE MATTRESS
The mattress, cover and inner cover are marked as shown below (example).
For an explanation, see section 9.4 Explanation of symbols.

20
9.4 Symbol key
XXXX-XX-XX
Item number Manufacturer Year-Month-Day Foot placement Read the instructions
for use
Type BF
Recycling Do not dispose of with
household waste; fol-
low recycling instruc-
tions
Recommended user
weight
User information
– category
Placed directly on the
bed base
Counteracts shear Heel function Place on top of
existing mattress
Do not machine wash Machine wash at 70 °C Machine wash at 95 °C Chlorine Tumble dry
Do not tumble dry Drip dry Do not iron Do not dry clean Wipe clean
Do not rotate Do not turn around Fire requirements Minimum length Serial number
Class II equipment (Dou-
ble-insulated). Indicated
on the power supply.
IP class (Enclosure class) The mattress should be Distributed by Medical device
used with the user lying
lenghtways.
UDI
CE-marked in accordance
with Medical Device
Regulation MDR (2017/745)
Table of contents
Other Care of Sweden Personal Care Product manuals
Popular Personal Care Product manuals by other brands

Lloydspharmacy
Lloydspharmacy LCM 1 user manual

LR
LR ZeitGard instruction manual

Abbott
Abbott Agilis HisPro DS3H010-38 Instructions for use

Orliman
Orliman Pad-Fix SP110 Use and maintenance instructions

REVLON
REVLON ONE-STEP VOLUMISER PLUS Use and Care Instruction Manual

Protac
Protac SenSit Straight Directions for use