Care Pump LITE4 User manual

UM_EN_CPLITE4_04
Lymph Drainage Therapy Device
(Pressotherapy Device)
User Manual

2
Table of contents
1. Safety warnings 4
1.1. Energy security 5
1.2. Possibility of injuries 5
1.3. Safe use and maintenance 6
1.4. Correct use of the cuffs 6
2. Basic characteristic of the device 7
2.1. CarePump pressotherapy devices 7
2.2. Pressotherapy (lymphatic drainage massage)
and its function 9
2.3. Indications for pressotherapy 9
2.4. Contraindications to pressotherapy 10
3. User manual 11
3.1. CarePump Lite4 device – front view 11
3.2. CarePump Lite4 device – back view 11
4. Installation and starting the device 12
4.1. How to connect the power supply 13
4.2. How to connect the air ducts 14
5. Basic device configuration 16
5.1. Main screen 16
5.2. Work mode setting 16
5.3. Pressure level 17
5.4. Treatment time 18
5.5. Speed of filling the chambers 18
5.6. Turning off individual chambers 19
5.7. Information/Language settings 19
5.8. How to start/pause/stop the treatment 20
5.9. Provided programs 21
Mode A 21
Mode B 21
Mode C 22
Mode D 22
6. Technical parameters 23
7. Contents of the package 24
8. Additional accessories 25

4
1. Safety
warnings
Before starting work with the device, please read the follow-
ing manual. It is recommended to use the device after con-
sulting a doctor.
CarePump series of devices has been designed in compli-
ance with all safety standards and norms that allow them
to be used. Please read the following chapter carefully, be-
cause it will allow you for safe and hygienic use of the device
in accordance with its intended purpose.
Attention!
The equipment can be used by children at least 8 years old
and by people with reduced physical and mental abilities
and people with inexperience and knowledge of the equip-
ment, if supervision or instruction on the use of the equip-
ment in a safe manner is provided so that the risks associ-
ated with it are understandable. Children should not play
with the equipment. Children without supervision should
not clean and maintain the equipment.
!

5
CarePump series is powered by the mains voltage of 100-
240V. Before connecting the device, make sure that the pa-
rameters of the transmission network are consistent with the
data provided on the rating plate at the bottom of the case.
The manufacturer recommends connecting the devices to
an installation equipped with a residual current device. This
will reduce the risk of shock, fire, permanent injury or damage
of the product. In case of problems with the power supply,
malfunctioning of the device, sparking or a burning smell
emanating from the machine, immediately disconnect the
device from the power supply by pulling the plug from the
wall socket. It is forbidden to leave the device switched on
without the operator’s supervision or with the power plug in-
serted into the wall socket. In case of atmospheric discharges
and storms, immediately stop the treatment and disconnect
the device’s power supply as mentioned above.
When disconnecting the power supply, pull it by the plug,
not the cable, as this may damage the cable and result in
electric shock or fire. Make sure if the cable is not tightly
coiled or twisted, as this may damage it. Do not use dam-
aged or stripped cable.
Be sure to disconnect the device and operate it with dry hands.
It is forbidden to use the device on patients:
▶ with an implanted artificial heart and other heart
prostheses,
▶ with metal or joint implants, as the procedure may
cause inflammation or chronic pain around them,
▶ with heart disease,
▶ with decompensated blood pressure,
▶ with skin diseases and its inflammations,
▶ with a fever,
▶ with neoplastic disease,
▶ with vascular diseases,
▶ with acute dermatitis and festering wounds,
▶ deep vein inflammation and venous thrombosis,
▶ after surgical operations until full recovery,
▶ with diseases of the nervous system and brain,
▶ pregnant,
▶ extremely tired,
▶ under the influence of alcohol and other stimulants,
▶ with diseases that give the feeling of constant fatigue.
1.2. Possibility of injuries
1.1. Energy security

6
The device’s operating environment should meet the follow-
ing conditions. Non-compliance with them may result in per-
sonal injury or permanent damage of the device.
Working conditions
▶ closed room,
▶ ambient temperature 0°C - 40°C,
▶ ambient humidity approx. 60%,
▶ 100-240V mains supply, appropriately supervised by
technical services,
▶ the distance from other electromagnetic receivers/
transmitters should be min. 1,5 m,
▶ the distance from devices emitting heat should be
min. 3 m,
▶ the distance from other electrical devices should be
min. 1,5 m,
▶ the floor should be made of non-conductive materials,
▶ the ground on which the device is located should be
stable, made of non-conductive and non-slip materials,
▶ the device should not be exposed to direct sunlight.
Safe use of the device cuffs contains the following principles:
▶ Before putting on the cuffs, remove your watch, jew-
ellery, and empty the pockets to avoid any injury to
the body or damage of the cuff.
▶ Do not place the cuffs directly on the body.
▶ The cuffs should be placed on thin, non-pressing
clothing.
▶ Do not store cuffs near sharp objects or in wet, sun-
ny and hot places.
▶ Do not expose the cuffs to staining with oil, gasoline,
alcohol, corrosive chemicals.
1.4. Correct use of the cuffs
1.3. Safe use and maintenance

7
2.
Basic characteristic
of the device
CarePump is a series of pressotherapy devices designed
and manufactured as a result of many years of experience
and research conducted by physiotherapy masters. Consul-
tations with our clients regarding necessary functions, ex-
ploitation and comfort of use allowed us to create a diversi-
fied line of devices adjusted to every need of medical staff,
patients, athletes and all other users.
CarePump devices are used for regeneration, relaxation
and improving your well-being. The device works very sim-
ilar to manual massage, which can be performed on your
own body. This is executed on the principle of a pressure
massage. Massage is performed by air-filled chambers
placed in the cuffs, to which air is supplied through a sys-
tem of flexible pipes from the central unit containing the
compressor. Each cuff contains a plurality of independent
chambers, which are sequentially inflated according to se-
lected program.
2.1. CarePump pressotherapy
devices

8
The basic set that allows you to perform the treat-
ment with CarePump device includes:
Central unit (device) – used to set parameters such as:
time, program, speed of pumping the chambers in the
cuffs or pressure.
Air ducts – used to supply air through independent chan-
nels to the individual cuff chambers. They connect the de-
vice with the cuffs.
Cuffs – used on individual parts of the body (arms, legs,
etc.). By sequential inflation of the cuff chambers, a part of
the body (where the cuff is placed) is being compressed to
allow achieving a massage effect.
A
B
C
A
B
C
B

9
It is recommended to perform pressotherapy every day.
Work mode, time of the treatment and drainage / massage
pressure depend on the individual preferences and possi-
bilities, the aim of the treatment and applied prevention.
▶ increase the body’s resistance,
▶ stimulation of metabolism and restoration of its
function,
▶ prevention and treatment of overweight and obesity,
▶ reduction of swelling,
▶ improving regeneration of tired and sore muscles,
▶ exudates,
▶ cellulite,
▶ heavy legs
▶ and others the same as in classic manual massage.
Pressotherapy consists of sequential and directional pres-
sure on a part of the body, where dedicated cuff is placed.
The air-filled cuff chamber (by pneumatic compression) in-
duces a mechanical massage effect. Particular chambers fill
in with air in specified time and accordingly to programmed
cycle. This causes a sequential pressure effect which turns
into a rhythmic wave. Lymphatic drainage massage is pos-
itively perceived as a very pleasant and effective way to
regenerate and relax. The patient feels relief in the areas
covered by the massage.
During this type of massage similar physiological processes
take place as in manual massage. It mainly stimulates the
activity of the body, restores and stimulates its functioning,
improves the flow and accelerates the exchange of fluids
in the system. The benefits are also the same as in classic
manual massage:
▶ regeneration,
▶ relaxation,
▶ improved well-being,
▶ improved skin elasticity,
▶ preventive healthcare.
2.3. Indications for pressotherapy
2.2. Pressotherapy
(lymphatic drainage massage)
and its function
Possible side effects of pressotherapy include:
▶ short-term pain exacerbation,
▶ petechiae,
▶ hematomas and bruises.

10
▶ pain or numbness of unknown origin,
▶ severe atherosclerosis or ischemia,
▶ vascular diseases are a relative contraindication to
pressotherapy,
▶ pregnancy,
▶ hypertension,
▶ pacemaker,
▶ thrombophlebitis,
▶ skin diseases, moles,
▶ lymphangitis,
▶ dermatitis, wounds,
▶ arthritis,
▶ high body temperature,
▶ thick, external, visible varicose veins,
▶ cardiac rhythm disturbances (arrhythmia),
▶ blood pressure disorders,
▶ myocardial ischemia,
▶ asthma,
▶ tumors, infiltrates, neoplasms,
▶ cardiac and respiratory failure,
▶ kidney failure
▶ and others the same as in classic manual massage.
2.4. Contraindications
to pressotherapy

11
3. User manual
1
2
3
4
7
It is absolutely necessary to read following user manual be-
fore starting work with the device.
3.1. CarePump Lite4 device –
front view
main power switch
input to connect the air ducts
adjustment knob
touch screen display
3.2. CarePump Lite4 device –
back view
power cord input
holes of cooling fans
safety button input
5
6
7
3
2
6
1
4
5

12
4.
Installation
and starting the device
Before starting the device, check the completeness of the
set and make sure if it has no defects or mechanical dam-
age. In case of any noticed damage, contact the distribu-
tor or manufacturer of the device. Do not proceed with the
installation and starting the machine if the packaging is
damaged. Unpack the device and place it on a stable sur-
face adequate for its weight. Remember not to place the
device too close to the wall, because the holes of cooling
fans are located on the back of the case. Do not place on
or near the device other electrical devices that may gener-
ate electromagnetic field or high temperature and objects
containing water or other liquids.
When the device is delivered in autumn or winter months,
when the ambient temperature is below 10ºC, wait about
2hours before starting the device to equilibrate the tem-
perature of the room and the device. The condensation of
water vapor may cause an electric shock or fire.

13
4.1. How to connect
the power supply
Connect the power cord into the input on the back of the
device. The cable should be fully inserted, please make
sure if it is firmly connected. Plug the other end of the cord
into a power socket.
Attention!
▶ Use the supplied power cord or one that complies
with electrical standards.
▶ Make sure that the voltage in the installation is 100-
240V 50/60Hz.
▶ If the cable is mechanically damaged, replace it with
another one.
▶ Do not place the machine close to the wall, because
the holes of cooling fans and power cord input are
located on the back of the case.
!

14
A
4.2. How to connect
the air ducts
Connect the plug of a single air duct or the air duct splitter
(in case you need to connect two cuffs at the same time) to
the input located on the front of the case. When connect-
ing the air duct or the splitter, pay attention to the notch
which indicates proper connection of them.
To connect the air duct, push it into the correct input un-
til you hear a click of both side buttons. To disconnect the
air duct, press buttons on both sides of the plug at the
same time.
When connecting two single air ducts to the splitter, pay
attention to the notch which indicates proper connection
of them.
A
B
B
▶
▶

15
The next step is connecting the air duct to the cuff. The sep-
arate ends of the air duct should be clipped into the individ-
ual chambers of the cuff with correct order. The proper con-
nection is indicated by a dark gray plug. The separate ends
should be connected in following order: from the longest to
the shortest one. Correctly plugged end in the cuff cham-
ber will make a characteristic click.
After performing all the above steps, the device is ready for use.
CC
1
3
2
4

16
0
50
5
1
2
3
4
mmHg
Ai
min
level
pressure A
ENG
i
time
speed
play/pause
mode
stop
1
2
3
4
0
50
5
mmHg A
B
C
D
i
min
level
1
2
3
4
5.
Basic device
configuration
To select one of the four provided programs, expand the al-
phabetical list in the upper right corner of the display A.
The particular work modes are marked with the letters A, B,
C, D. After pressing each letter, an animation will start show-
ing how the chambers are pumping Band the default
treatment time for the current mode will change. To select
a work mode, press the corresponding letter, which will be
highlighted in white. The selected work mode will appear in
the window with the list of all provided programs A.
5.1. Main screen
5.2. Work mode setting
A
To start the device, press the power button on the left side
of the case. The blue diode will light up when the button is
pushed. The device will start up and CarePump logo will ap-
pear on the display. Then the main screen view will appear,
where you can set basic treatment optionse.
0
50
5
1
2
3
4
mmHg
Ai
min
level
pressure A
ENG
i
time
speed
play/pause
mode
stop
1
2
3
4
0
50
5
mmHg A
B
C
D
i
min
level
1
2
3
4
B

17
To set the pressure level, press the circle with the number
“0” next to the pressure unit „mmHg” C. The pressure val-
ue must be set using the adjustment knob. By turning to
the right in the direction of “+”, the pressure will increase.
By turning to the left in the direction of “-”, the pressure lev-
el will decrease. The current pressure value will appear in
acircle next to the pressure unit “mmHg” C. The pressure
range generated by the machine is 20-220 mmHg.
To confirm the pressure selection, press any icon on the
display.
Attention!
Pressure as a parameter is not defined for provided pro-
grams. It should be adapted to the needs of each patient,
based on the patient’s blood pressure and his feelings dur-
ing the treatment.
5.3. Pressure level
!
69
50
5
mmHg
Ai
min
level
1
2
3
4
69
50
5
mmHg
Ai
min
level
1
2
3
4
C

18
To set the treatment time, press the circle next to the time
unit „min” D. The treatment time is default for each of the
work modes. The treatment time range that can be set in
the device is 1-120 minutes. The treatment time should be
set by using the adjustment knob. By turning it to the right
in the direction of “+”, the treatment time will be extended
with another minute. By turning it to the left in the direc-
tion of “-”, the treatment time will be shortened. The current
treatment time will be displayed in a circle next to the time
unit “min”. To confirm the selection of the treatment time,
press any icon on the display.
5.4. Treatment time
5.5. Speed of filling
the chambers
To set the speed of filling the chambers, press the circle
next to the parameter “level” E. By default, the speed of
filling the chambers is set to level 5. The chamber inflation
speed range is 1-5, with 1 being the lowest (slowest filling)
and 5 being the highest (fastest filling). The chamber infla-
tion speed should be set using the knob. To confirm the se-
lection, press any icon on the display.
69
50
5
mmHg
Ai
min
level
1
2
3
4
69
50
5
mmHg
Ai
min
level
1
2
3
4
D
E

19
The device allows you to turn off the chambers in the cuff
that need to be bypassed (they will not be inflated with air)
during the treatment. By default, all chambers are turned
on. To turn off a particular chamber, press the correspond-
ing number on the graphic image of the cuff on the right
side of the screen. After the chamber is turned off, its num-
ber will stop being highlighted. To turn the chamber back
on, press the number of the chamber again.
5.6. Turning off
individual chambers
5 .7. Information/Language settings
After pressing “i” button, the icons explaining the settings
available on the start screen will be displayed. This view
also allows you to change the language by clicking on the
square with the abbreviation corresponding to the particu-
lar language. To return to the main screen, press any icon
on the display.
69
50
5
mmHg
Ai
min
level
1
2
3
4
69
50
5
mmHg
Ai
min
level
1
2
3
4
0
50
5
1
2
3
4
mmHg
Ai
min
level
pressure A
ENG
i
time
speed
play/pause
mode
stop
1
2
3
4
0
50
5
mmHg A
B
C
D
i
min
level
1
2
3
4

20
After setting the parameters, you can start the treatment. To
start the treatment, press the icon “play” Ain the center
of the display. After pressing it, the treatment will start and
this will be shown by the blinking chamber on the graphic
image of the cuff and the elapsing time.
To temporarily stop the treatment, press the “pause” icon
B. The treatment will stop and the treatment time will be
paused. The treatment can be resumed by pressing the
“play” icon again.
To permanently stop the treatment, press the icon “stop”
C. When treatment is completely stopped, the unit will re-
turn to its pre-treatment settings and the air will be pumped
out of the cuffs.
5.8. How to start/pause/stop
the treatment
69
50
5
mmHg
Ai
min
level
1
2
3
4
69
50
5
mmHg
Ai
min
level
1
2
3
4
69
50
5
mmHg
Ai
min
level
1
2
3
4
69
50
5
mmHg
Ai
min
level
1
2
3
4
A
B
C
Table of contents
Other Care Pump Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Thayer Medical Corporation
Thayer Medical Corporation Quake Instructions for use

otometrics
otometrics MADSEN AccuScreen OAE user manual

Symmetry Surgical
Symmetry Surgical GREENBERG manual

Arjo
Arjo Enterprise 9000X Instructions for use

Otto Bock
Otto Bock 50S16 Malleo Sprint light quick start guide

Nabtesco
Nabtesco HybridKnee NI-C3 Series instruction manual

human care
human care Lena user manual

Modsel
Modsel Contour Portare user guide

Planmeca
Planmeca Proline XC Pan Calibration manual

Medicatech
Medicatech MasterRad30 installation manual

Kalypto Medical
Kalypto Medical 140cc Dressing Application guide

LionsGate Technologies
LionsGate Technologies Kenek Edge user manual