F.2 DISPLAY/SEGMENT CHECK
1) Press “MODE” and hold. At the same time, quickly press and release “ON”. Release
“MODE”.
2) Display should read “F.1” (if the display reads “F.0”, press “MODE” again, which will display “F.1”).
3) Press “MODE” until display reads “F.2”.
4) Press “HOLD” to enter DISPLAY/SEGMENT CHECK function. Display will illuminate
all segments of the display to check for failure or damage.
5) Return to NORMAL operation by pressing “ON”.
F.3 OFF-SET VALUE
1) Press and hold “MODE”. At the same time, quickly press and release “ON”. Release
“MODE”.
2) Display should read “F.1” (if the display reads “F.0”, press “MODE” again, which will display “F.1”).
3) To get to “F.3”, press “MODE” twice and then press “HOLD”
4) Display will read current value. If the unit is operating properly, this value will be be-
tween 5,000 and 15,000. If the value is not within this range, you must adjust the off-
set potentiometer on the rear PC board. If, upon adjustment, the value remains outside
the proper range, contact CCi Scale Company Inc. for instructions.
5) Return to NORMAL operation by pressing “ON”
F.4 AUTOMATIC SHUT OFF FUNCTION
1) Press and hold “MODE”. At the same time, quickly press and release “ON”. Release
“MODE”.
2) Display should read “F.1” (if the display reads “F.0”, press “MODE” again, which will display “F.1”).
3) To get to “F.4”, press “MODE” three times and then press “HOLD”
4) The display will indicate either:
a) “4 OFF”, which indicates the unit will automatically shut itself off after four (4) min-
utes on non-operation.
b) “0 OFF”, which indicates the unit will remain in the ON until physically turned OFF.
5) To select the mode that is not displayed, press “HOLD” one time. This should change the
display to the next mode. When the display reaches your selection press “ON”. This returns
the scale to normal operation and your choice is implemented during future operations.
TLEC