Cecomp®Battery Powered Digital Pressure Gauges with Min/Max & Selectable Units F16B, F16BN
Ranges and Resolution
See table below. Select range and default engineering units.
Consult factory for special engineering units.
Units may be changed to any listed under the same sensor range.
Resolution is fixed and limited to available display digits.
Accuracy
Accuracy includes linearity, hysteresis, repeatability
Standard accuracy: ±0.25% of full scale ±1 least significant digit
HA accuracy option: ±0.1% FS ±1 LSD, see ranges for availability
Sensor hysteresis: ±0.015% FS, included in accuracy
Sensor repeatability: ±0.01% FS, included in accuracy
Display
3 readings per second nominal display update rate
4 digit LCD, 0.5" H and 5 character 0.25" H alphanumeric
BL: Red LED backlight
Batteries, Battery Life, Low Battery Indication
2 AA alkaline included
B: Approx. 2000 hours
BL: Approx. 150-1500 hours depending on backlight usage
Low battery symbol on display
Controls & Functions
Front button turns gauge on or off, zeros gauge reference gauges,
and cycles through min./max. functions
Internal buttons for engineering unit selection, auto shutoff time,
min/max setup, calibration. Pass code protected setup & calibration.
BL: Front button activates backlighting for 1 minute
Min./Max. Functions
Minimum and maximum readings stored 3 times per second
Front button cycles through min. display, max. display, clear
Configurable for min. only, max. only, min./max., or none
Configure to clear or retain min./max. values at power off
O±0.25% Test Gauge Accuracy
O316L Stainless Steel Wetted Parts
OUser Configurable Min/Max, Units, Shutoff Time
Calibration
Pass code protected calibration
Non-interactive zero, span, and linearity, ±10% of range
Auto Shutoff
User selectable 1 minute to 8 hours or front button on/off
Factory default 5 minutes, unless other time is specified
Weight
Gauge: 9 ounces , shipping: 1 pound (approximately)
Housing
F16B: Extruded aluminum case, epoxy powder coated, ABS/ poly-
carbonate bezel (aluminum bezel optional), front and rear gaskets,
polycarbonate label
F16BN: ABS/polycarbonate NEMA 4X case, rear gasket, polycar-
bonate label
Connection and Material
1/4" NPT male fitting, all wetted parts are 316L stainless steel
Overpressure, Burst, Vacuum
Ranges using 3000 psig sensor: 5000 psig
Ranges using 5000 psig sensor: 7500 psig
All others: 2 X pressure range
3000 psi, 5000 psi, and 4 digit ranges 112.5% full scale out-of-
range display: 1– – – or I –.–.–.–
4 X sensor burst pressure rating, or 10,000 psi, whichever is less
Vacuum service: 15 psia, ±15 psig, 15 psig, 30 psia,
100 psig, 100 psia, 200 psig sensors
Under-range display (non-vacuum sensors): –Err
Environmental
Storage temperature: –40 to 203°F (–40 to 95°C)
Operating temperature: –4 to 185°F (–20 to 85°C)
Compensated temperature: 32 to 158°F (0 to 70°C)
F16B
F16BN
NEMA 4X
How to Specify Type
F16B range - time - options Standard housing
F16BBL range - time - options Std. housing, backlit display
F16BN range - time - options NEMA 4X housing
F16BNBL range - time - options NEMA 4X, backlit display
HHHHHH
HHHHH
HHHHHH
HHHHH
HHHHHH
HHHHH
HHHHHH
HHHHH
HHHHHH
Made in USA
Sensor Ranges and Engineering Units * -HA option not available Range codes are rounded off
3 psig *Res 15 psig vac *Res 30 psia Res –15V100psig *Res 300 psig Res
3PSIG .001 100KPAVAC .1 2KGCMA .001 –15V100PSIG .1 300PSIG .1
6INHGG .001 0.1MPAVAC .0001 2ATMA .001 –30INHG/100PSIG .1 610INHGG .1
85INH2OG .1 1BARVAC .001 30 psig Res –30V200INHGG .1 4800ZING 1
50ZING .01 1KGCMVAC .001 30PSIG .01 –400V2770INH2OG 1700FTH2O .1
210GCMG .1 1ATMVAC .001 60INHGG .01 –240V1600ZING 12000KPAG 1
150MMHGG .1 15 psig Res 850INH2OG 1–760V5200MMHGG 12MPAG .001
150TORRG .1 15PSIG .01 480ZING .1 –760V5200TORRG 120BARG .01
200MBARG .1 30INHGG .01 2100GCMG 1–100V700KPAG 120KGCMG .01
200CMH2OG .1 400INH2OG .1 1600MMHGG 1–0.1V0.7MPAG .001 20ATMG .01
2000MMH2OG 1240ZING .1 1600TORRG 1–1V7BARG .01 500 psig Res
7FTH2O .001 1000GCMG 12000MBARG 1–1V7KGCMG .01 500PSIG .1
20KPAG .01 760MMHGG .1 2100CMH2OG 1–1V7ATMG .01 1020INHGG 1
5 psig *Res 760TORRG .1 70FTH2O .01 100 psig Res 1150FTH2O 1
5PSIG .001 1000MBARG 1200KPAG .1 100PSIG .1 3500KPAG 1
10INHGG .01 1000CMH2OG 10.2MPAG .0001 200INHGG .1 3.5MPAG .001
140INH2OG .1 35FTH2O .01 2BARG .001 2770INH2OG 135BARG .01
80ZING .1 100KPAG .1 2KGCMG .001 1600ZING 135KGCMG .01
350GCMG .1 0.1MPAG .0001 2ATMG .001 7000GCMG 135ATMG .01
260MMHGG .1 1BARG .001 60 psig Res 5200MMHGG 11000 psig Res
260TORRG .1 1KGCMG .001 60PSIG .01 5200TORRG 11000PSIG 1
350MBARG .1 1ATMG .001 120INHGG .1 7000MBARG 12040INHGG 1
350CMH2OG .1 ±15 psig *Res 1660INH2OG 17000CMH2OG 12300FTH2O 1
3500MMH2OG 1±15PSIG .01 960ZING 1230FTH2O .1 7000KPAG 1
12FTH2O .01 –30INHG/15PSIG .01 4200GCMG 1700KPAG .1 7MPAG .001
35KPAG .01 ±30INHG .01 3100MMHGG 10.7MPAG .0001 70BARG .01
15 psia Res ±400INH2OG 13100TORRG 17BARG .001 70KGCMG .01
15PSIA .01 ±240ZING .1 4100MBARG 17KGCMG .001 70ATMG .01
30INHGA .01 ±1000GCMG 14200CMH2OG 17ATMG .001 2000 psig Res
400INH2OA .1 ±760MMHG 1140FTH2O .1 –15V200 psig *Res 2000PSIG 1
240ZINA .1 ±760TORRG 1400KPAG .1 –15V200PSIG .1 4070INHGG 1
1000GCMA 1±1000MBARG 10.4MPAG .0001 –30INHG/200PSIG .1 4600FTH2O 1
760MMHGA .1 ±1000CMH2OG 14BARG .001 –30V400INHGG .1 14MPAG .01
760TORRA .1 ±100KPAG .1 4KGCMG .001 –400V5500INH2OG 1140BARG .1
1000MBARA 1±0.1MPAG .0001 4ATMG .001 –240V3200ZING 1140KGCMG .1
1000CMH2OA 1±1BARG .001 100 psia Res –100V1400KPAG 1140ATMG .1
100KPAA .1 ±1KGCMG .001 100PSIA .1 –0.1V1.4MPAG .001 3000 psig Res
0.1MPAA .0001 ±1ATMG .001 200INHGA .1 –1V14BARG .01 3000PSIG 1
1BARA .001 30 psia Res 2770INH2OA 1–1V14KGCMG .01 6100INHGG 1
1KGCMA .001 30PSIA .01 1600ZINA 1–1V14ATMG .01 6900FTH2O 1
1ATMA .001 60INHGA .01 7000GCMA 1200 psig Res 20MPAG .01
15 psig vac *Res 850INH2OA 15200MMHGA 1200PSIG .1 200BARG .1
15PSIVAC .01 480ZINA .1 5200TORRA 1400INHGG .1 200KGCMG .1
30INHGVAC .01 2100GCMA 17000MBARA 15500INH2OG 1200ATMG .1
400INH2OVAC .1 1600MMHGA 17000CMH2OA 13200ZING 15000 psig Res
240ZINVAC .1 1600TORRA 1700KPAA .1 480FTH2O .1 5000PSIG 1
1000GCMVAC 12000MBARA 10.7MPAA .0001 1400KPAG 135MPAG .01
760MMHGVAC .1 2100CMH2OA 17BARA .001 1.4MPAG .001 350BARG .1
760TORRVAC .1 200KPAA .1 7KGCMA .001 14BARG .01 350KGCMG .1
1000MBARVAC 10.2MPAA .0001 7ATMA .001 14KGCMG .01 340ATMG .1
1000CMH2OVAC 12BARA .001 14ATMG .01
G gauge reference pressure
CPD inHg vac / psig pressure
VAC gauge reference vacuum
A absolute reference
psi = PSI
inHg = INHG
oz/in2= ZIN
inH2O = INH2O
ftH2O = FTH2O
mmHg = MMHG
torr = TORR
mmH2O = MMH2O
kg/cm2= KGCM
g/cm2= GCM
kPa = KPA
MPa = MPA
mbar = MBAR
bar = BAR
cmH2O = CMH2O
atm = ATM
Range—See table at left. Select a range code for default units.
Please specify if vacuum gauge requires a minus sign.
Accessories—order separately
CON14SS
Quick connector to install or remove gauge with-
out tools. 304 stainless steel, urethane seal.
SCR14SS
Filter screen fitting keeps debris out of gauge
sensor. For food vacuum packaging applications.
303SS body, 100 micron 304SS screen.
RB
High visibility orange rubber boot protects gauge
for portable applications. Not for NEMA 4X models.
TP
Top gauge port. Primarily used with tire pressure
applications. Not available with NEMA 4X models.
Quick Link: cecomp.com/bat
Time—auto shutoff time.
55 minutes. Default if not specified.
10 10 minutes
30 30 minutes
ON No auto shutoff. On/off via front button.
Options—add to end of model number. See price list for details.
HA High accuracy, ±0.1% FS ±1 LSD. See range table.
PM Panel mount, 4.1" x 4.1", n/a NEMA 4X
FP Sealed housing and CC for high humidity food process-
ing applications. F16BN in absolute ref. ranges only.
MC Metal front cover instead of plastic, n/a NEMA 4X
CS Case bottom stiffener plate, n/a NEMA 4X
CC Moisture resistant circuit board conformal coating
TP Top port, gauge port on top of case, n/a NEMA 4X
SM Surface mount plate, n/a NEMA 4X
CD Calibration data; 5 test points and date
NC NIST traceability documentation, 5 points and date
© 10-18
cecomp.com
1220 American Way Libertyville, IL 60048
800-942-0315
CECOMPDiv. of
ABSOLUTE PROCESS INSTRUMENTS
Pb Lead
Free
4Read these instructions before using the gauge. Configuration
may be easier before installation. Contact the factory for assis-
tance.
4These products do not contain user-serviceable parts. Contact
us for repairs, service, or refurbishment.
4Gauges must be operated within specified ambient temperature
ranges.
4Outdoor or wash down applications require a NEMA 4X gauge or
installation in a NEMA 4X housing.
4Use a pressure or vacuum range appropriate for the application.
4Use fittings appropriate for the pressure range of the gauge.
4Due to the hardness of 316 stainless steel, it is recommended
that a thread sealant be used to ensure leak-free operation.
4For contaminated media use an appropriate screen or filter to
keep debris out of gauge port.
4Remove system pressures before removing or installing gauge.
4Install or remove gauge using a wrench on the hex fitting only.
Do not attempt to turn gauge by forcing the housing.
4Good design practice dictates that positive displacement liquid
pumps include protection devices to prevent sensor damage
from pressure spikes, acceleration head, and vacuum extremes.
p Avoid permanent sensor damage! Do not apply vacuum to non-
vacuum gauges or hydraulic vacuum to any gauges.
p Avoid permanent sensor damage! NEVER insert objects into
gauge port or blow out with compressed air.
Gauges are not for oxygen service. Accidental rupture of sensor
diaphragm may cause silicone oil inside sensor to react with
oxygen.
Bipolar ranges read positive pressure and vacuum in the same
units, and zero with the gauge port open.
1000 psi and higher sensor are a sealed reference type. They read
zero with the gauge port open are internally referenced to 14.7 psi.
Functionally similar to gauge reference sensors.
Absolute reference gauges read zero at full vacuum and
atmospheric pressure with the gauge port open. With an open
gauge port the readings will vary continuously due to the effects of
barometric pressure.
Range Codes
The range code is part of the gauge model number and indicates
the default range when the gauge is ordered.
Selectable Ranges
Engineering units may be changed to any of those listed in the same
sensor range as shown in the range table.
Conversion
Engineering units are calculated from the factory default unit to the
newly selected units. The ranges listed under the psi sensor range
are rounded off.
Power-Up
Press and hold the front button for approximately 1 second.
The display is tested, the default full-scale range is indicated, the
full-scale range in the selected units are indicated, the display is
tested again, then the actual pressure and units are displayed.
Power-Up With Zero
This applies to gauge reference models only. Absolute reference
gauges do not use the zero feature since they read atmospheric
pressure under normal conditions.
Be sure the gauge port is exposed to normal atmospheric pressure
and no pressure is applied. The zeroing function is only activated
at each power-up and the stored zero correction is erased when
the gauge is shut off.
Press and hold the front button. The display is tested and then oooo
is displayed. The gauge is now zeroed. Release the button and the
full-scale range in the selected units are indicated, the display is
tested again, then the actual pressure and units are displayed.
Attempting to zero the gauge with pressure greater than approxi-
mately 3% of full-scale pressure or vacuum applied will result in an
error condition, and the display will alternately indicate Err 0 and
the actual measured pressure. The gauge must be powered down
to reset the error condition.
Normal Operation
Following the start-up initialization, the display indicates the pres-
sure reading updated approximately 3 times per second. The auto
shutoff timer starts when the gauge is powered up or whenever the
button is pushed, unless the gauge shutoff time was set to zero for
on/off operation.
If excessive vacuum is applied to a pressure-only gauge, the display
will indicate -Err until the vacuum is released. Applying vacuum to
a gauge designed for pressure may damage the pressure sensor.
If excessive pressure is applied (112.5% over range), an out-of-
range indication of 1 – – – or 1.–.–.– will be displayed depending
on model.
Display Backlighting (BL models only)
Display backlighting can be turned on by momentarily pressing the
front button whenever the gauge is on. The backlighting will turn
on for one minute and then automatically shut off. This also restarts
the auto shutoff timer. The display backlighting will not be apparent
under bright lighting conditions.
Minimum and Maximum Readings
Gauges are normally configured with minimum and maximum
capture functions enabled. One or both can be disabled in the
Advanced Configuration mode.
Minimum and maximum readings are continuously stored and
updated whenever the gauge is on. The stored readings can be
manually cleared if desired. The MAX and MIN memory is also
cleared whenever the gauge is off unless configured to save the
readings.
Press and hold the button for about 1 second until MAX is displayed
alternating with the units. The maximum reading will be continu-
ously updated. The gauge may be left in this mode.
After MAX is displayed, press and hold the button for about 1 sec-
ond until MIN is displayed alternating with the units. The minimum
reading will be continuously updated. The gauge may be left in this
mode. If excessive vacuum is applied to a pressure-only gauge
while in this mode, the display will indicate -Err until the MAX/MIN
readings are cleared.
After MIN is displayed, press and hold the button again for about 1
second until * * * * is displayed. The MAX and MIN memory is not
erased and the gauge returns to normal operation with the display
indicating the current reading.
Press and continue to hold the button until the display indicates clr
MX/MN (about 3 seconds total) and then release the button. Both
maximum and minimum values are cleared and the gauge returns
to the normal mode and displays the current pressure.
Shut-Down
To shut off the gauge manually at any time, press and hold the button
until the display indicates OFF (about 5 seconds) and then release.
When an auto shutoff timer is used, the display indicates OFF five
seconds prior to auto shutoff. The button can be pressed to keep
the gauge on. The auto shutoff and backlight (if equipped) timers
are reset whenever the button is pressed and released.
If the gauge is set up without auto shutoff (on/off operation) it will
stay on until manually shut off or until the batteries are depleted.
Turn gauge off when not in use to conserve battery life.
Instructions F16B, F16BN
Operation—continued
Types of Gauges
Function Button Prompt (Release Button)
On Press for 1
second
Display test > default units > selected
units FUll scale > display test >
actual reading
Zero
(gauge ref.
only)
Press and
hold
Display test >
oooo
> default units >
selected units FUll scale >
display test > actual reading
Hi reading Press/hold Max > max. reading & units
Lo reading Press/hold MIN > min. reading & units
Exit Hi/Lo Press/hold
SSSS
> actual reading
Clear Hi/Lo Press/hold Max > Clr mx/mn > actual reading
Clear, off Press/hold Max > Clr mx/mn > OFF
Installation Precautions
¼" NPT
2.88"
3.38" Standard
3.5" NEMA 4X
0.75"
2.0"
Cecomp Electronics / Made in USA
1.65"
NEMA 4X
3.0"
Standard
Turn at hex
fitting only
XXXXX
8
8
8
8
82
Low battery
voltage
indicator
Dimensions
Operation
Selectable Ranges
cecomp.com
1220 American Way Libertyville, IL 60048
800-942-0315
CECOMPDiv. of
ABSOLUTE PROCESS INSTRUMENTS
Cecomp maintains a constant effort to upgrade and improve its products.
Specifications are subject to change without notice. See cecomp.com for
latest product information. Consult factory for your specific requirements.
WARNING: This product can expose you to chemicals including
nickel and chromium, which are known to the State of Califor-
nia to cause cancer or birth defects or other reproductive harm.
For more information go to www.P65Warnings.ca.gov
Accessing the Internal Buttons
1. Remove the 6 Phillips screws on the back of the unit.
2. Lift up the battery holder.
3. Move the battery holder to the side.
4. For gauges with the standad metal houseing, remove the battery
cradle if necessary to gain access to the two internal buttons
located near the lower right and left corners of the circuit board.
Auto Shutoff Time Selection
Auto shutoff time selection is done via the internal buttons to help
prevent accidental or unauthorized changes. The selected shut off
time is stored in non-volatile memory and will be retained even with
the battery off or batteries removed.
Remove the rear cover to gain access to the two internal buttons
located near the lower right and left corners of the circuit board.
With the gauge powered up, press and hold the DOWN button.
Release the button when the auto shutoff time is displayed on the
upper display.
The lower display will indicate AST M if the time displayed is in
minutes, and AST H if it in hours.
An auto shutoff time of 0 signifies that the auto shutoff feature is
disabled and the front button turns the gauge on and off.
Use the UP and DOWN buttons to select 0, 1, 2, 5, 10, 15, 20 or 30
minutes, or 1, 2, 4, or 8 hours.
When the desired time is displayed, press and release the front but-
ton to save the selection and return to normal operation.
Note: The gauge will automatically revert to normal operation if no
buttons are operated for approximately 15 seconds.
Replace the battery holder assembly and rear cover taking care not
to pinch the power wires between the cover and the case.
Engineering Unit Selection
Engineering unit selection (except compound ranges) is done
via internal buttons to help prevent accidental or unauthorized
changes. The selected engineering unit is stored in non-volatile
memory and will be retained even with the gauge off or batteries
removed. The available engineering units depend on the sensor
range and display resolution.
Compound (inHg/PSIG) gauges must be changed to display single-
unit vacuum/pressure readings in the Advanced Configuration mode
before different engineering units can be selected.
The default engineering units are mathematically converted to the
newly selected engineering unit. When the gauge is powered up,
the originally configured range is displayed and then the conversion
with the selected engineering unit is displayed.
Powered the gauge up by holding the front button for 1 second.
Press and hold the internal UP button.
Release the button when the engineering units begin to flash.
Use the UP and DOWN buttons to scroll through the list of engineer-
ing units available for the pressure range of the sensor.
When the desired units are displayed, press and release the front
button to save the selection and return to normal operation.
Note: The gauge will automatically revert to normal operation if no
buttons are operated for approximately 15 seconds.
Replace the battery holder assembly and rear cover taking care not
to pinch the power wires between the cover and the case.
Advanced configuration requires a pass code for access and allows
more features to be configured.
Remove the rear cover to gain access to the buttons located near
the lower right and left corners of the circuit board.
With the gauge off, press and hold the UP button. Then press the
front button. Release all buttons when the display indicates CFG
and the program version then the full-scale range is indicated and
the display is tested.
The display then indicates _ _ _ _ with the first underscore blink-
ing, and with CFGPC (configuration pass code) on the character
segments.
Note: The gauge will automatically revert to normal operation if
no buttons are operated for approximately 15 seconds. To cancel
and return to normal operation, press and release the front button
without entering any pass code characters.
Pass Code Entry
The factory default is 3510, but this may be changed by the user
under the Pass Code Configuration section.
1. Use the UP or DOWN buttons to set the left-most digit to 3.
2. Press and release the front button to move to the next position.
The 3 will remain, and the second position will be blinking.
3. Use the UP or DOWN buttons to select 5.
4. Press and release the front button to index to the next position.
35 will remain, and the third position will be blinking.
5. Use the UP or DOWN buttons to select 1.
6. Press and release the front button to index to the next position.
351 will remain, and the fourth position will be blinking.
7. Use the UP or DOWN buttons to select 0.
8. Press and release the front button to proceed with configuration
procedures.
If an incorrect pass code is entered, the gauge will return to the
start of the pass code entry sequence.
Factory/User Configuration
The upper display section will be blank, and the lower section will
display either USER_ or FCTRY.
If USER_ is selected, the user configuration can be modified as
described in the following steps.
To select USER_, press and release the DOWN button.
The lower display will indicate USER_.
Press and release the front button to continue.
If FCTRY is selected, the existing user configuration will be replaced
by the original factory configuration.
To select FCTRY, press and release the UP button.
The lower display will indicate FCTRY.
Press and release the front button to restore the factory configura-
tion and restart the gauge.
Max/Min Configuration
Use the UP and DOWN buttons to select from the following:
MX/MN Both highest and lowest values will be captured
MX/--- Only highest value will be captured
--/MN Only lowest value will be captured
--/-- Capture feature is disabled
Press and release the front button to move to the next parameter.
Max/Min Memory
The upper display section will indicate clr.
Use the UP and DOWN buttons to select from the following:
AUTO Automatically clear max. and min. values when the
gauge is powered off
MAN Manually clear max. and min. values
Press and release the front button to move to the next parameter.
Gauge Type Configuration
This will only appear with 15, 100, or 200 psig ranges that were
originally ordered as compound gauges.
Use the UP and DOWN buttons to select from the following:
-/+EU Vacuum is indicated as negative pressure in the
selected engineering units
CMPND Vacuum is negative INHG, pressure is PSIG. This setting
will disable engineering unit selection.
Press and release the front button to save the user configuration
and restart the gauge.
Replace the rear cover taking care not to pinch the power wires
between the cover and the case.
A low battery indication will be shown in the upper left-
hand corner of the display when the battery voltage falls
sufficiently. The batteries should be replaced soon after the
indicator comes on or unreliable readings may result.
1. Remove the 6 Phillips screws on the back of the unit.
2. Lift up the battery holder.
3. Remove batteries by lifting up the positive end of the battery
(opposite the spring) taking care not to bend the battery holder
spring.
4. Discard old batteries properly, do not discard into fire, sources
of extreme heat, or in any hazardous manner.
5. Always replace both batteries at the same time with high quality
alkaline batteries.
6. Install batteries with correct orientation. Incorrect polarity will
damage the gauge. The negative (flat) end of each battery
should be inserted first facing the battery holder spring.
7. Replace battery holder and back cover, including the rubber
gasket and reinstall the six screws.
Instructions F16B, F16BN
Advanced ConfigurationBasic Configuration Battery Replacement
cecomp.com
1220 American Way Libertyville, IL 60048
800-942-0315
CECOMPDiv. of
ABSOLUTE PROCESS INSTRUMENTS
Gauges are calibrated at the factory using equipment traceable to
NIST. There is no need to calibrate the gauge before putting it into
service. Calibration intervals depend on your quality control pro-
gram requirements, although many customers calibrate annually.
Calibration should only be performed by qualified individuals using
appropriate calibration standards and procedures. The calibration
equipment should be at least four times more accurate than the
gauge being calibrated.
The calibration system must be able to generate and measure pres-
sure/vacuum over the full range of the gauge.
A vacuum pump able to produce a vacuum of 100 microns (0.1
torr or 100 millitorr) or lower is required for vacuum and absolute
gauges.
Warning: application of vacuum to non-vacuum models may result
in irreparable damage to the sensor.
Allow the gauge to acclimate to ambient temperature for 20
minutes.
Remove the rear cover to gain access to the UP and DOWN buttons
located near the lower right and left corners of the circuit board.
Install fresh batteries.
Entering Calibration Mode
With the gauge off, press and hold the DOWN button. Then press
the front button. Release all buttons when the display indicates CAL.
The display begins by indicating the full-scale positive pressure
rating of the gauge in the engineering units as configured by the
factory, and then shows all display.
Before the gauge enters the Calibration Mode, the display initially
indicates _ _ _ _ with the first underscore blinking, and with CALPC
(calibration pass code) on the lower display.
Note: The gauge will automatically revert to normal operation if
no buttons are operated for approximately 15 seconds. To cancel
and return to normal operation, press and release the front button
without entering any pass code characters.
Enter the User-Modifiable Pass Code
The factory default is 3510, but this is user changeable.
1. Use the UP or DOWN buttons to set the left-most digit to 3.
2. Press and release the front button to move to the next position.
The 3 will remain, and the second position will be blinking.
3. Use the UP or DOWN buttons to select 5.
4. Press and release the front button to index to the next position.
35 will remain, and the third position will be blinking.
5. Use the UP or DOWN buttons to select 1.
6. Press and release the front button to index to the next position.
351 will remain, and the fourth position will be blinking.
7. Use the UP or DOWN buttons to select 0.
8. Press and release the front button to proceed with configuration
procedures.
If an incorrect pass code is entered, the gauge will return to the
start of the pass code entry sequence.
Calibration Mode
The gauge enters and remains in the calibration mode until
restarted manually or power is removed. Features not related to
calibration are disabled and compound range models are set for the
same engineering units for pressure and for vacuum.
The calibration may be performed in any of the available engineer-
ing units as well as percent (PCT). For greatest accuracy, use the UP
and DOWN buttons to select engineering units for calibration with
highest resolution (highest number of display counts). Press and
Instructions F16B, F16BN
release the front button when the appropriate engineering units are
displayed. Suggested units are listed below.
Sensor Suggested units for calibration
5 PSI 5.000 PSI
15 PSI 775.7 MMHG (TORR)
30 PSI 61.08 INHG
50 PSI 50.00 PSI
60 PSI 60.00 PSI
100 PSI 7.031 KG/CM2
200 PSI 407.2 INHG
300 PSI 610.8 INHG
500 PSI 500.0 PSI
1000 PSI 70.31 KG/CM2
2000 PSI 4072 INHG
3000 PSI 6108 INHG
5000 PSI 5000 PSI
Any 100.00 PCT (percent)
The display will then indicate the currently applied pressure in the
engineering units selected for calibration.
UP and DOWN Button Operation
Each time one of the calibration buttons is pressed and released
quickly, a small change is made to the digitized pressure signal. It
may take more than one of these small changes to result in a single
digit change on the display.
To make larger changes, press and hold the appropriate calibration
button. After about one second, the display will begin to change
continuously. Release the button to stop. Then make fine adjust-
ments by pressing and quickly releasing the calibration buttons as
previously described.
Gauge Reference Pressure Gauges
Apply zero pressure by venting the gauge port to atmosphere. The
character display will alternate between ZERO and CAL. Adjust for
a display indication of zero using the UP and the DOWN buttons.
Apply full-scale pressure. The character display will alternate
between +SPAN and CAL. Adjust for a display indication of full-
scale pressure using the UP and the DOWN buttons.
Apply 50% full-scale pressure. The character display will alternate
between +MID and CAL. Adjust for a display indication equal to
50% of full-scale pressure using the UP and the DOWN buttons.
Gauge Reference Vacuum Gauges
Apply zero pressure by venting the gauge port to atmosphere. The
character display will alternate between ZERO and CAL. Adjust for
a display indication of zero using the UP and the DOWN buttons.
Apply full-scale vacuum. The character display will alternate
between +SPAN and CAL. Adjust for a display indication of full-
scale vacuum using the UP and the DOWN buttons.
Apply 50% full-scale vacuum. The character display will alternate
between +MID and CAL. Adjust for a display indication equal to
50% of full-scale vacuum using the UP and the DOWN buttons.
Absolute Reference Gauges
Apply full vacuum to the gauge. The character display will alternate
between ZERO and CAL. Press the UP and DOWN buttons to obtain
a display indication of zero.
Apply full-scale pressure. The character display will alternate
between +SPAN and CAL. Press the UP and DOWN buttons to obtain
a display indication equal to full-scale pressure.
Apply 50% of full-scale pressure. The lower display will alternate
between +MID and CAL. Press the UP and DOWN buttons to obtain
an indication equal to 50% of full-scale pressure.
Compound and Bipolar Gauges
In addition to the steps described above for pressure gauges, apply
full-scale vacuum. The character display will alternate between
–SPAN and CAL. Adjust for a display indication of actual applied
vacuum using the UP and the DOWN buttons.
For bipolar and –30.00inHg/+15.00psig compound range models
only, apply 50% full-scale vacuum. The character display will alter-
nate between –MID and CAL. Adjust for a display indication equal
to 50% of full-scale vacuum using the UP and the DOWN buttons.
Save Calibration
Once the adjustments are complete, press and hold the front but-
ton until the display indicates – – – – then release the button to
store the calibration parameters in non-volatile memory and restart
the gauge.
Verify the pressure indications at 0%, 25%, 50%, 75% and 100%
of full scale.
Replace the rear cover taking care not to pinch the wires between
the cover and the case.
Remove the rear cover to access the buttons located near the lower
right and left corners of the circuit board.
View Or Change User Configuration Pass Code
With the unit off, press and hold the UP button, then press the
front button.
Release all buttons when the display indicates CFG.
View Or Change User Calibration Pass Code
With the unit off, press and hold the DOWN button, then press the
front button.
Release all buttons when the display indicates CAL.
Enter Access Code 1220
Before the unit enters the view or change pass code mode, the
display initially indicates _ _ _ _ with the first underscore blinking,
and with CFGPC or CALPC on the character display.
Note: The gauge will automatically revert to normal operation if no
buttons are operated for approximately 15 seconds.
To cancel and return to normal operation, press and release the
front button without entering any pass code characters.
1. Use the UP and DOWN buttons to set the left-most digit to 1.
2. Press and release the front button to move to the next position.
The 1 will remain, and the second position will be blinking.
3. Use the UP and DOWN buttons to select 2.
4. Press and release the front button to index to the next position.
12 will remain, and the third position will be blinking.
5. Use the UP and DOWN buttons to select 2.
6. Press and release the front button to move to the next position.
1 2 2 will remain, and the fourth position will be blinking.
7. Use the UP and DOWN buttons to select 0.
8. Press and release the front button to proceed.
Note: If an incorrect access code was entered, the gauge will return
to the start of the access code entry sequence.
Once the access code has been entered correctly, the display will
indicate the existing user-defined pass code with CFGPC or CALPC
on the character segments.
1. Operate the UP or DOWN button to select the first character of
the new pass code.
2. When the correct first character is being displayed, press and
release the front button to proceed to the next pass code char-
acter.
3. Repeat above until the entire pass code is complete.
4. To exit, press and hold the front button. Release the button
when the display indicates – – – – to restart the gauge.
5. Replace the rear cover taking care not to pinch the power wires
between the cover and the case.
User-Defined Pass Code ConfigurationCalibration Preparation
Calibration
Calibration—continued
cecomp.com
1220 American Way Libertyville, IL 60048
800-942-0315
CECOMPDiv. of
ABSOLUTE PROCESS INSTRUMENTS

This manual suits for next models

3

Other Cecomp Measuring Instrument manuals

Cecomp CTP1B User manual

Cecomp

Cecomp CTP1B User manual

Cecomp CTP3B User manual

Cecomp

Cecomp CTP3B User manual

Cecomp F16AD User manual

Cecomp

Cecomp F16AD User manual

Cecomp DPG2000B D4 User manual

Cecomp

Cecomp DPG2000B D4 User manual

Cecomp F20B User manual

Cecomp

Cecomp F20B User manual

Popular Measuring Instrument manuals by other brands

Agilent Technologies u8903a user guide

Agilent Technologies

Agilent Technologies u8903a user guide

Armstrong Steam QM-1 Installation and operation manual

Armstrong

Armstrong Steam QM-1 Installation and operation manual

YOKOGAWA FLXA402T user manual

YOKOGAWA

YOKOGAWA FLXA402T user manual

Nidec DT-361 instruction manual

Nidec

Nidec DT-361 instruction manual

TESTO 340 instruction manual

TESTO

TESTO 340 instruction manual

CHY 831 Operator's manual

CHY

CHY 831 Operator's manual

Extech Instruments EC150 user guide

Extech Instruments

Extech Instruments EC150 user guide

Somogyi Elektronic home SMA 830 instruction manual

Somogyi Elektronic

Somogyi Elektronic home SMA 830 instruction manual

Hanna Instruments HI 93711 user manual

Hanna Instruments

Hanna Instruments HI 93711 user manual

Mazurczak NT Series Installation and operation instruction

Mazurczak

Mazurczak NT Series Installation and operation instruction

Kett PM-600 operating manual

Kett

Kett PM-600 operating manual

Arenal QA03-SDA User & maintenance manual

Arenal

Arenal QA03-SDA User & maintenance manual

Skydrop 2102 user guide

Skydrop

Skydrop 2102 user guide

Newport PHH-60 Operator's manual

Newport

Newport PHH-60 Operator's manual

mru OPTIMA user manual

mru

mru OPTIMA user manual

TESA TESA-hite 400 instruction manual

TESA

TESA TESA-hite 400 instruction manual

Hameg hm5510 user manual

Hameg

Hameg hm5510 user manual

SMC Networks PFMV3 Series Operation manual

SMC Networks

SMC Networks PFMV3 Series Operation manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.