TECHNICAL INFORMATION
Specification
Other important information
Do not use this equipment near explosive gases (e.g. anaesthetics).
Please note Oxygen is NOT an explosive gas.
Interference
Interference should not occur between this and other equipment, such as televisions, radios and
other electronic equipment. If you have any interference involving your Quoddy Bed, try relocating
or adjusting the position of the Quoddy Bed and or the other equipment.
Also try connecting them on different mains circuits.
Recycling
This equipment contains electronic components. Check with your local authority if there are any
special requirement on disposal and follow the Local WEEE regulations.
The life expectancy of this equipment is 5 years.
Centrobed offer a scrapage service under which we will collect and safely dispose of this equipment.
Please contact the office for details
Country of Origin England
Name of Equipment Quoddy Bed
Description Electric Adjustable Bed
Voltage 230V (AC)
Power Frequency 50 Hz
Power Consumption (In Use) 2 A Peak
Power Consumption (Stand by) 55 mA
Fuse (mains plug) 5 A
Class II (Double Insulated)
Type BF
Electrical Safety Standard EN 60601-1 : EN 60601-2-52
EMC Standard EN 60601-1-2
Duty Cycle (Bed) 1
Duty Cycle (per Motor) 20%
Page 8
Symbols
Handset Symbols see Page 6
CE Mark
Maximum patient weight
Combined weight limit