CI-203 Instruction Manual rev. 8/9/2016
1554 NE 3rd Ave, Camas, WA 98607, USA
Phone (360) 833-8835 Fax (360) 833-1914 sales@cid-inc.com http://www.cid-inc.com
The CI-203 automatically adjusts an internal threshold to a level that is optimum for detecting
the presence of a leaf in the scanner objective. If out-of-the-ordinary measurements are being
recorded or if the instrument has suffered from abuse or excessive aging, the manual threshold
setting may need to be adjusted. If the instrument appears to be making incorrect
measurements, make sure the instrument is clean and the reflective tape and the objective
window appears to be undamaged and in good condition. If there is something blocking the
optical path in the scanner, no amount of threshold adjustment will correct the problem. Refer
to the section on cleaning and maintenance for more information on insuring that the instrument
is clean. Please do not hesitate to consult with the manufacturer or with the nearest
manufacturer’s representative if there is a problem and the threshold needs adjustment.
Use the UP/DOWN arrow to highlight Scanner on the Setup Menu display. Next, press the
RIGHT arrow to enter the Setup Scanner menu. The options on the Setup Scanner menu
include HiThld (High Threshold), LoThld (Low Threshold), Manual, Filter and Adjustment. Use
the UP/DOWN key to switch between options and the RIGHT arrow to select.
Threshold Adjustment (Width Calibration)
In the event you need to alter the normal threshold settings, perform the following steps. First,
check the threshold of the CI-203 alone, with its arm in place. To do so, access the Setup
Scanner menu, select “HiThld” and press the RIGHT arrow key. The instrument’s motor will
turn on and a “t”and “w”value will appear and count down on the display. Next, the display will
show the high threshold value briefly before returning to the Scanner Setup menu. Make note
of the HiThld value.
Next, switch to “LoThld” menu and place a piece of white paper under the arm and press the
RIGHT arrow key. Let the scanner run. The display will show a t value that counts down and
w=0. When the t value is done counting down, the final Low Threshold value (t= ) will display
briefly before exiting back to the Scanner Setup menu. Make note of the LoThld value.
Remove the paper from under the arm. Next, select ‘Manual’from the Scanner Setup menu by
pressing the RIGHT arrow key. Make sure that the width reads 0 when there is no object under
the arm. Stop the Manual scan and the motor by pressing the STOP key. On the scanner
menu, scroll down to ‘Adjustment’. Here is where the user can adjust the width measurement
based on the high and low threshold values. After increasing or decreasing the Manual
threshold value, press STOP. You will be prompted to STOP (discard changes) or SAVE
changes if you made any.
The next step is very important: every time the threshold is changed in any mode, switch to the
“Measure” menu and press the LEFT and RIGHT key at the same time. The instrument will
automatically compute and save the timing parameters. The user should repeat this
computation a few times allowing the instrument to generate an accurate leaf width reading.