Ambulatory Infusion Pump
BodyGuardTM 323
Clinician Quick Reference
Priming the Administration Set
1. Make sure that patient is not connected to administration set.
2. Turn pump power ON and press
3. Press
4. Use pump keypad to enter volume of fluid required to prime set.
For priming volume, check tubing package.
5. Press to start priming procedure.
Display screen shows priming progress.
6. For additional priming volume, repeat steps 1 – 5.
Note:
• To stop priming at any time, press
• If line must be primed, it can only be primed on pump.
• When pump is locked, patient can prime set in all operating modes.
For 24/7 clinical and technical support, call: 877.263.0111
CAUTION: Before use, the user must become thoroughly familiar with the information contained in the device operating instructions.
Note: This Clinician Quick Reference provides a convenient information summary. For complete information on proper use of the BodyGuard
pump and administration set, refer to the device operating instructions supplied with the product. © 2008 by CME America, LLC. All rights
reserved. The following are trademarks of CME America, LLC: BodyGuard, BodyComm, TSyringePump, the CME America logotype with
“swirl” logomark and the tag line “Drug Delivery Designed for the Way You Deliver Care.” SM-0015 07/2008
PRIME
Disconnect Patient
Press OK to Start
Prime Vol ml
Press OK to Continue
Prime Vol ___ml
Press OK to Continue
5
Prime 2 ml
05
STOP
NO
START
OK
PRIME
PIGGY
START
OK