SPECIFICATIONS
Range: –76 to 932°F / –60 to 500°C
Resolution: 0.1° (-9.9 to 199.9)
1° (otherwise)
OPERATION
Press and continue to hold down the
trigger to take temperature readings.
The unit will take temperature readings
while the trigger is pressed and held
(SCAN appearing on the display indicates
measurements are being taken).
After the trigger is released, the last
temperature measured will be displayed
for approximately 15 seconds and the
display will turn off automatically. HOLD
will appear on the display when the
trigger has been released to indicate that
the unit is no longer taking temperature
measurements.
SELECTING UNITS OF MEASURE
Each press of the °C/°F button will toggle
between °C and °F.
LASER TARGETING
Laser targeting is always active when
the trigger is being pressed. Each time
the trigger is pressed and held, the laser
beam will be emitted.
CAUTION: Eye damage may result from
direct exposure to laser light.
MAXIMUM TEMPERATURE
While taking measurements, the unit
will automatically record the maximum
temperature reading that has been
measured. This value will appear on the
display to the right of MAX.
DISTANCE SPOT/RATIO
There are other factors that may affect
measurement accuracy. The target
must completely ll the spot diameter
seen by the infrared sensor; otherwise
readings will be inuenced by the
surface surrounding the target. The ratio
of the distance to the size of the spot
being measured is 11:1. For example,
an object’s diameter of 100mm can be
measured from 1100mm.
DISPLAY MESSAGES
“Er2” will appear on the display when
the unit is exposed to rapid changes in
ambient temperature.
“Er3” will appear on the display when
the ambient temperature is below 32° F
(0° C) or above 122° F (50° C).
“Er” will appear on the display when the
unit needs to be reset. To reset the unit,
remove the batteries, wait 2 minutes and
then re-install the batteries.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
©2017 Traceable®Products. 92-4470-10 Rev. 3 122017
ALL OPERATIONAL DIFFICULTIES
If this thermometer does not function
properly for any reason, replace the
batteries with new high quality batteries
(see Battery Replacement section). Low
battery power can occasionally cause
any number of “apparent” operational
difculties. Replacing the batteries
with new fresh batteries will solve most
difculties.
BATTERY REPLACEMENT
Erratic readings, faint readings, no display,
or appearing on the display are all
indications that the batteries must be
replaced. Open the battery compartment
by sliding the battery cover in the direction
indicated by the arrow and then lift the
battery cover. Remove the exhausted
batteries and replace them with two (2)
newAAAalkaline batteries. Make certain
to install the new batteries in the directions
indicated in the battery compartment.
Close the battery cover.
WARRANTY or SERVICE
For warranty or service, contact:
TRACEABLE®PRODUCTS
Tr aceable®Products is ISO 9001:2018 Quality-
Certi ied by DNV and ISO/IEC 17025:2017
accredited as a Calibration Laboratory byA2LA.
TRACEABLE
INFRARED
THERMOMETER
GUN
INSTRUCTIONS
®
Traceable®is a registered trademark of Cole-Parmer.
Note: Traceable Infrared thermometers, are
NOT approved for Medical usage, and are
not FDAapproved.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Other manuals for Traceable 4470

1

Other Cole Parmer Thermometer manuals

Cole Parmer Traceable 4470 User manual

Cole Parmer

Cole Parmer Traceable 4470 User manual

Cole Parmer TRACEABLE LOLLIPOP 6419 User manual

Cole Parmer

Cole Parmer TRACEABLE LOLLIPOP 6419 User manual

Cole Parmer Digi-Sense J User manual

Cole Parmer

Cole Parmer Digi-Sense J User manual

Cole Parmer 8533-40 User manual

Cole Parmer

Cole Parmer 8533-40 User manual

Cole Parmer Traceable User manual

Cole Parmer

Cole Parmer Traceable User manual

Popular Thermometer manuals by other brands

Emos M0503 manual

Emos

Emos M0503 manual

Weber Style 32910 user manual

Weber

Weber Style 32910 user manual

Sygonix SY-4904438 operating instructions

Sygonix

Sygonix SY-4904438 operating instructions

Omron MC-110 manual

Omron

Omron MC-110 manual

TFA 30.1017 instruction manual

TFA

TFA 30.1017 instruction manual

Anritsu AP-400 Series user manual

Anritsu

Anritsu AP-400 Series user manual

Omega Engineering HH147 user guide

Omega Engineering

Omega Engineering HH147 user guide

Steren DT-01 user manual

Steren

Steren DT-01 user manual

CDN DTTC operating instructions

CDN

CDN DTTC operating instructions

Bresser Climate Smile instruction manual

Bresser

Bresser Climate Smile instruction manual

BST BT-66 Operation manual

BST

BST BT-66 Operation manual

Traceable 90000-39 operating instructions

Traceable

Traceable 90000-39 operating instructions

Global 695746 instruction manual

Global

Global 695746 instruction manual

Velleman TA25 user manual

Velleman

Velleman TA25 user manual

Silvercrest KH 8105 Operation & safety manual

Silvercrest

Silvercrest KH 8105 Operation & safety manual

Omega Engineering OSXL450 user guide

Omega Engineering

Omega Engineering OSXL450 user guide

PCE Instruments PCE-TC 34N user manual

PCE Instruments

PCE Instruments PCE-TC 34N user manual

Kaz V971CFN-CAN quick start guide

Kaz

Kaz V971CFN-CAN quick start guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.