COMPASS HEALTH NEB-SOCCER User manual

User Manual
Compressor Nebulizer
NEB-SOCCER
SOCCER-TRU
SOCCER-TRUWB-PB
CAUTION–Federal (U.S.A.) law restricts this device to sale by, or on the order of a
physician.
Manufactured for Compass Health Brands.
Table of Contents
IEC SYMBOLS.........................................................................................................................2
IMPORTANT SAFEGUARDS.................................................................................................3
INTRODUCTION .....................................................................................................................2
WORK PRINCIPLE..................................................................................................................4
COMPONENTS AND SUGGESTED USAGE LIFE...............................................................5
GETTING STARTED...............................................................................................................6
COMPRESSOR NEBULIZER CLEANING............................................................................ 9
FILTER CHANGE..................................................................................................................
MAINTENANCE ....................................................................................................................
SPECIFICATIONS..................................................................................................................
TROUBLESHOOTING GUIDE ............................................................................................ 12
10
10
11
WARRANTY...........................................................................................................................
18
ELECTROMAGNETIC COMPATIBILITY ......................................................................... 13
Guidance and manufacturer´s declaration –electromagnetic emission.................................. 14
Emissions test ......................................................................................................................... 14
Group 1................................................................................................................................... 14
Guidance and manufacturer´s declaration –electromagnetic immunity............................... 14
Immunity test ...................................................................................................................... 14
Immunity test ...................................................................................................................... 15
Recommended separation distance ..................................................................................... 17

2
IEC SYMBOLS
INTRODUCTION
The Soccer Ball Nebulizer Compressor System is a medical device designed to
deliver medicine prescribed by a physician to the airways of the lungs. The Nebulizer
Set and the Air Compressor combined provides aerosol mist for inhalation by the
patient for medical treatment only.
Statement of Indication for Use
The Soccer Ball Nebulizer Compressor is including an AC powered air compressor
that provides a source of compressed air for use with a jet nebulizer in a home health
care setting. The compressor provides a maximum flow rate of 8 Lpm that can be
connected to is used in conjunction with a jet (pneumatic) nebulizer with the same
form for inhalation by a patient. The patient population is determined by the
nebulizer to which it is connected.
The disposable nebulizer cup is used to administer various aerosol treatments to adult and
pediatric patients in both the homecare and healthcare settings. Its use is indicated
whenever a physician or healthcare professional administers or prescribes medical aerosol
products to a patient using a Small Volume Nebulizer. The disposable nebulizer cup is
disposable, single-patient-use and intended for use with pediatric (5 years and older) and
adult patient.
The Reusable Jet Nebulizer is a handheld, pneumatic nebulizer designed to aerosolize
prescription drugs for inhalation by a patient. Its use is indicated whenever a healthcare
professional administers or prescribes medical aerosol products to a patient using a small
volume nebulizer. The Reusable Jet Nebulizer is intended for use with pediatric (2 years
and older) and as defined by the prescribed medication and adult patients in hospital/
institutional settings, home care use, schools, and long term care facilities.
performance specifications to convert certain inhalable drugs into an aerosol

3
DANGER
To reduce the risk of electrocution:
1. Always unplug this product immediately after using.
2. Do not use while bathing.
3. Do not place or store product where it can fall or be pulled into a tub or sink.
4. Do not place in or drop into water or other liquid.
5. Do not reach for a product that has fallen into water. Unplug immediately.
WARNING
To reduce the risk of burns, electrocution, fire or injury to persons:
1. The product should never be left unattended when plugged in.
2. Close supervision is necessary when this product is used by, on, or near children or
physically incapacitated individuals.
3. Use this product only for its intended use as described in this guide. Do not use
attachments not recommended by the manufacturer.
4. Never operate this product if:
a. It has a damaged power cord or plug.
b. It is not working properly.
c. It has been dropped or damaged.
d. It has been dropped into water.
Return the product to an authorized Dealer or Service Provider for examination and
repair.
IMPORTANT SAFEGUARDS
When using electrical products, especially when children are present, basic safety
precautions should always be followed.
Read all instructions before using. Important information is highlighted by these
terms:
DANGER–Urgent safety information for hazards that will cause serious injury or
death.
WARNING–Important safety information for hazards that might cause serious injury.
CAUTION–Information for preventing damage to the product.
NOTE–Information to which you should pay special attention.
READ ALL INSTRUCTIONS BEFORE USING

4
WORK PRINCIPLE
lubricating, single-cylinder piston pump to create air pressure. The air flow created by
the Air Compressor comes out of the jet outlet and then enters the Nebulizer Set. The
main air flow into and out of the air compressor unit is illustrated in Figure 1 below.
5. Keep the power cord away from heated surfaces.
6. Never block the air openings of the product or place it on a soft surface, such as a
bed or couch, where the air openings may be blocked. Keep the air openings free of
lint, hair, and the like.
7. Never use while drowsy or asleep.
8. Never drop or insert any object into any opening or hose.
9. Do not use outdoors. This product is for household use only.
10. Do not use in an oxygen enriched environment.
11. Do not operate unit in a backpack or carrying bag.
12. Care should be taken when this product is used in the vicinity of other electronic
equipment concerning the possible interference between them. They may need to be
moved apart.
This is a home use desktop device and it is not intended to be used as a portable
device.
This unit is compact, simple to operate and convenient to carry It uses a non-oil

5
ILLUSTRATION OF PARTS
NEB-SOCCER Components
(Disposable Nebulizer components
and Older)
1. Mouthpiece
2. Medication Cup
3. Connection Tubing
4. Tube Connector
5. Air Openings
6. Power Switch
7. Filter & Cover
8. Resevoir Tube
9. T-Adapter
10. Aerosol Mask
46
1
7
5
2
3
8
9
10
Figure 1: Disposable Nebulizer
Figure 2: Reusable Jet Nebulizer
1
6
4
2
3
1. Mouthpiece
2. Cap with Valve
3. Nebulizer Insert
4. Medicine Cup
5. Connection Tubing
6. Aerosol Mask
SOCCER-TRU &
SOCCER-TRUWB-PB
Reusable Jet Nebulizer Components
(For Adults and Children
5
for Adults and Children 5Years
2Years and Older)

6
GETTING STARTED
1. Check to ensure the following components of the nebulizer are enclosed:
• Compressor unit
2.
on hand at all times in case original parts become lost, damaged,
It is recommended to keep extra tubing, accessories and air filters
or soiled.
CAUTION
If you did not receive the components listed in this manual, please contact
the dealer where you purchased your nebulizer.
SETTING UP THE NEBULIZER COMPRESSOR AND NEBULIZER ACCESSORIES
CAUTION
Tubing and accessories are NOT sterile from package.
Prior to initial operation, the nebulizer should be thoroughly cleaned.
1. Wash hands with soap and warm water and dry completely with
a clean towel.
2.
or wet to ensure proper operation.
3. Make sure tubing and accessories have been cleaned and/or disinfected
Disinfecting Tubing and Accessories section on page 9.
4. Remove the nebulizer cap by twisting counter-clockwise until it stops and
lift up (Figure 3-1). Be sure that the cone is put inside the bottom section
5. Fill nebulizer cup with medication from unit dose vial, eye dropper, or syringe.
Do not fill more than 5 ml per treatment (Figure 3-2).
6. Close the nebulizer cap and turn clockwise until it is tightened (Figure 3-3).
Place the compressor on a clean, dust-free, stable and flat surface. Check
the filter monthly or every 10 – 12 treatments and replace if dirty, clogged,
before first use and after last daily treatment. Refer to the Cleaning and
Disposable Nebulizer Cap
Prescribed
Medication
Figure 3-1 Figure 3-2 Figure 3-3
• Tubing and accessories (see pages 5 for illustration of parts)

7
7. If using the mask (optional), press it directly into the nebulizer cap (Figure 4).
If using the mouthpiece, attach it firmly into the nebulizer cap. The cap is
designed to fit tightly (Figure 4).
8. If using reusable jet nebulizer reference Figure 5 for instructions.
Figure 4
Figure 5
Mask (optional)

8
Figure 6-1 Figure 6-2
Nebulizer
Air Outlet
9.
compressor unit. It is important that the tubing be pushed all the way onto the
Attach one end of the tubing firmly onto the air outlet on the back of the
air outlet. The tubing is designed to fit tightly (Figure 6-1).
10. Firmly press the other end of the tubing onto the bottom of the nebulizer cup
(Figure 6-2).
11. Plug the power cord into an appropriate electrical outlet. (
When treatment is finished, shut off the unit and unplug it from the electrical
DO NOT use with
power strips or extension cords). Make sure at this stage, the power switch is
at “OFF” status.
12. Press power switch to “ON” status to begin your prescribed treatment.
(DO NOT run unattended). When treating children, adult supervision required.
13. outlet.
WARNING!
• The exterior surfaces of the unit and the air outlet connector may become slightly
warm to the touch during use. Warm air can also be felt exiting the vents on the
sides of the case of the unit. This is normal.
• Never operate if this unit has a damaged cord or plug, has been dropped
in any way, immersed into water or if it does not work properly. Return it to a
service center for repair.
• If any abnormality occurs, discontinue use immediately until the unit has been
examined and repaired.
• Always unplug the product immediately after use.
• Never allow the power supply cord to be pulled, jerked, strained, twisted or
severely bent, especially at the plug connections. Damage will occur at the high
cord carefully for longer life.
• Never block the air openings of the main unit or place it where the air openings
may be obstructed.
• Safety overload: DO NOT exceed 30 minutes of continuous operation.
Connection
Tubing

9
materials.
–Nebulizer Set with DISPOSABLE accessory for single patient single use
only.
COMPRESSOR NEBULIZER CLEANING
1. With power switch in the “Off” position, unplug power cord from wall outlet.
2. Wipe outside of the compressor cabinet with a clean damp cloth every few days to
keep dust-free.
DANGER–Do not submerge in water; doing so will result in compressor damage.
WARNING
OPERATION
To operate the compressor Nebulizer, do the following:
1. Plug in theCompressor, but do not connect the Nebulizer Set.
2. Make sure that the air filter is attached to the Compressor.
3. Turn on the power switch.
4. After the Compressor runs properly for 1-2 minutes, connect the jet outlet of
the Compressor to the inlet of the Nebulizer Set
5. Begin treatment according to your physician’s instructions.
6. Inhale and exhale naturally. Be sure to stop when the medicine is used up.
7. When you have finished, turn the air compressor unit off before you disconnect
the air hose.
WARNING –Do not use near flammable objects or locations near explosive

10
FILTER CHANGE
Compass Part No. for Filter is CIF-NEB-5
1. Filter should be changed every 6 months or sooner if filter discolors.
2. Remove filter by grasping it firmly and pulling out of the unit. Discard
old filter.
3. Replace with a new filter. Additional filters should be purchased from your
Compass provider.
4. Push new filter into position.
CAUTION–Do not reuse or wash filter or substitute any other material such as
cotton for an air-inlet filter. It will result in compressor damage.
MAINTENANCE
To keep your Compressor Nebulizer working properly, please do follow the
If experiencing a problem regarding the device, such as the device cannot be turned on,
noise being too loud, no medicine being dispersed, or water drops forming in the air hose,
instructions in above mentioned cleaning and disinfection processes.
please proceed to the "TROUBLESHOOTING" guide on the page 12.

11
NOTE–If the problem persists after the above basic maintenance, any further
maintenance must be performed by a qualified Compass provider or authorized
service center. Please refer to the contact information in the last page of this manual.
DANGER–Electric shock hazard. Do not remove compressor cabinet. All
disassembly and maintenance must be done by a qualified Compass provider.
SPECIFICATIONS
Air Compressor Nebulizer Machine Features
Power
120VAC, 60Hz
Electrical outlet
Power Consumption
200VA
Storage/Transport Temperature Range
Storage/Transport Humidity
10 - 95%
Operating Humidity
10 - 95% non-condensing
Operating Temperature Range
Atmospheric Pressure
From 700 hPa –1060hPa
Max Compressor Flow
8 LPM
Compressor Flow with nebulizer set
5~8 LPM
Max Pressure
30 psi or greater / 206 Kpa or greater
Operating Pressure Range
8~13 psi / 55Kpa~90Kpa
Nebulizer MMAD
5 microns or less
Noise Level
below 63dBA
Dimension (L x W x H)
222 x 173x 100mm
Product Weight
3.2 lbs
Protection against electric shock
Class II
Degree of protection against electric shock
Type BF
Mode of operation
Recommended shelf life for Nebulizer Set
5 Years
Recommended shelf life for Compressor
2 Years
-25C to 70C (-13C to 158C )
10C to 40C (50C to 104C )
30 min on 30 min off

12
WARNING: Equipment not suitable for use in the presence of flammable anesthetic
mixture with air or with oxygen or nitrous oxide.
This product complies with the following electromagnetic compatibility standard:
IEC 60601-1-2:2nd Edition.
This is a home use desktop device and it is not intended to be used as a portable
device.
TROUBLESHOOTING GUIDE
Symptom
Probable Cause
Solution
Too much
noise
No filter in the unit
Put a filter in the unit
The
Nebulizer
does not
run
Main compressor switch is off
Turn the unit on
Nebulizer was not cleaned after its
previous use
Clean the unit
Air hose is bent
Remove any bend or kink in
the air hose
The filter is clogged
Replace the filter
No medicine is left
Add the appropriate amount of
medicine prescribed by your
physician to the Medicine
Container
The cap is not attached properly
Reattach the cap
Water
Drops are
formed in
the air
hose
The medicine was over-full
Adjust the amount of medicine
in the Medicine Container to
the appropriate level, connect
the air hose to the Air
Compressor, and turn the unit
on
The Nebulizer Kit was washed but
not dried
Open and close the Nebulizer
Kit repeatedly until the drops
disappear
NOTE–Please contact your COMPASS provider for assistance in case you have any
other difficulties.

13
ELECTROMAGNETIC COMPATIBILITY
The information contained in this section (such as separation distances) is in general
specifically written with regard to Soccer Ball Nebulizer. The numbers provided
will not guarantee faultless operation but should provide reasonable assurance of such.
This information may not be applicable to other medical electrical equipment;
older equipment may be particularly susceptible to interference.
General Notes:
Medical electrical equipment requires special precautions regarding electromagnetic
compatibility (EMC) and needs to be installed and put into service according to the
EMC information provided in this document and the remainder of the instructions for
use this device.
Portable and mobile RF communications equipment can affect medical electrical
equipment.
Cables and accessories not specified within the instructions for use are not authorized.
Using other cables and/or accessories may adversely impact safety, performance and
electromagnetic compatibility (increased emission and decreased immunity).
Care should be taken if the equipment is used adjacent to or stacked with other
equipment; if adjacent or stacked use is inevitable, the equipment should be observed
to verify normal operation in the configuration in which it will be used.
The MODEL will continue to provide air to the nebulizer when exposed to the
electromagnetic disturbance levels described below.

14
Electromagnetic Emissions
This equipment is intended for use in the electromagnetic environment specified below. The
user of this equipment should assure that is used in such an environment.
Emissions
Compliance
according to
Electromagnetic environment
RF emissions (CISPR 11)
Group 1
The equipment uses RF energy only
for its internal function. Therefore, its
RF emissions are very low and are
not likely to cause any interference in
nearby electronic equipment.*
CISPR Emissions Classification
Class B
The equipment is suitable for use in
all establishments including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic
purposes.
Harmonic emissions
(IEC 61000-3-2)
Class A
Voltage fluctuations / flicker (IEC
61000-3-3)
Complies
Electromagnetic Immunity
This equipment is intended for use in the electromagnetic environment specified below. The
user of this equipment should assure that is used in such an environment.
Immunity
against
IEC 60601-1-2 test
level
Complianc
e level (of
this device)
Electromagnetic
environment
electrostatic discharge,
ESD
(IEC 61000-4-2)
contact discharge: 6 kV
air discharge: 8 kV
8 kV
15 kV
The relative humidity should be
at least 5 %
electrical fast transients
/ bursts
(IEC 61000-4-4)
power supply lines: 2 kV
longer input / output lines:
1 kV
2 kV
1 kV
Mains power quality should be
that of a typical commercial,
hospital or home environment.
surges on AC mains
lines
(IEC 61000-4-5)
Common mode: 2 kV
differential mode: 1 kV
2 kV
1 kV
Mains power quality should be
that of a typical commercial,
hospital or home environment.

15
power frequency
magnetic field 50/60 Hz
(IEC 61000-4-8)
3 A/m 30 A/m Equipment which emits high
levels of power line magnetic
fields (in excess of 30A/m)
should be kept at a distance to
reduce the likelihood of
interference..
voltage dips and short
interruptions on AC
mains input lines
(IEC 61000-4-11)
dip >95%, 0.5 periods
dip 60%, 5 periods
dip 30%, 25 periods
dip >95%, 5 seconds
>95%, 0.5
per.
60%, 5 per.
30%, 25 per.
>95%, 5 sec.
Mains power should be that of a
typical commercial, hospital or
home environment. If you
requires continued operation
during power mains interruptions
it is recommended that the
Soccer Ball Nebulizer
be powered from an
uninterruptible
power supply.ensure
This equipment is intended for use in the electromagnetic environment specified below.
The customer or the user of should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance
level
Electromagnetic environment –guidance
Warning keep Soccer Ball Nebulizer
away from sources of high levels
of power line magnetic fields
(in excess of 30 A/m) to reduce the
likelihood of interference.

16
Conducted RF
rf coupled into lines
(IEC 61000-4-6)
Radiated rf
(IEC 61000-4-3)
3 V rms 150 kHz to 80
MHz outside ISM bands
6 VRMS 150 kHz to 80
MHz inside ISM bands
3 V/m
80 MHz –2.7 GHz
3 Vrms
6 Vrms
10 V/m
Portable and mobile RF
communications equipment should
be used no closer to any part of
the, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter as below.
Recommended separation
distance
d=1.2P
d=0.58/P 80 MHz to 800 MHz
d=0.70P 800 MHz to 2.5 GHz
where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site surveya,
should be less than the compliance
level in each frequency rangeb.
Interference may occur in the
vicinity of equipment marked with
the following symbol:

17
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
aField strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the equipment is used exceeds the
applicable RF compliance level above, the equipment should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the equipment.
bOver the frequency range 150kHz to 80 MHz field strengths should be less than 3 V/m
Recommended separaon distances between portable and mobile RF
communicaons equipment and the equipment
The Soccer Ball Nebulizer is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the equipment can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the equipment as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum
output power of
transmier W
Separation distance according to frequency of transmitters in meters
150 kHz – 80 MHz
d=1.2 P
80 MHz to 800MHz
d=1.2 P
800 MHz to 2.5 GHz
d=0.70 P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
√ √ √
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1At 80 MHz and 800 MHz, the separation distance for the higherfrequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
This manual suits for next models
2
Table of contents
Other COMPASS HEALTH Respiratory Product manuals