MANUALP/4000_SERVICE_0712 July2012 10
WARNING: To prevent FIRE or SHOCK hazard, DO NOT expose this
product to rain or any type of moisture.
FOR YOUR SAFETY
To ensure safe operation the three-pin plug supplied must be inserted
ONLY into a standard three-pin power outlet which is effectively earthed
through the normal building wiring.
Extension cords are NOT recommended.
The fact that the equipment operates satisfactorily does NOT imply that
the power outlet is earthed and that the installation is completely safe. For
your safety, if in any doubt about the effective earthing of the power outlet,
consult a qualified electrician.
This appliance should be regularly tested (or after any service work on
the appliance) according to the procedures prescribed in AS/NZS 3760.
The basic safety checks and tests on electrical appliances required by
AS/NZS 3760 are:
a) A visual check to ensure that there is no mechanical damage, that
controls etc. are in good working order and that no parts are
missing.
b) An earth continuity test. (Maximum allowed resistance is 1 )
An insulation resistance test. (Minimum insulation resistance is
1 M)
WARNING: This cabinet is fitted with a UV Sterilising Lamp
UVC refers to shorter UV wavelengths, usually 200 to 280 nM. Wavelengths
in the UVC range, especially from the low 200's to about 275 nM, are
especially damaging to exposed cells. Such shortwave UV is often used for
germ killing purposes. The typical wavelength for ‘sterilising’ purposes is
253nM.
CAUTION: This chamber has been fitted with a SPECIAL UVC Sterilising lamp
located behind the stainless steel rear duct. Note that this ULTRA-VIOLET
lamp is HARMFUL to the eyes and MUST NOT be viewed directly or indirectly
for any period.
The UV lamp is automatically turned OFF when the outer door is opened and
turns on for 5 minutes when the door is reclosed.