
Secon 7
Integra 3™ Temperature Control
Research Instruments Ltd
7
Secon 7
Integra 3™ Temperature Control
Research Instruments Ltd
7
When Integra 3 is powered up, the display briey shows RI and the
product logo. Aer start up the screen shows the actual temperatures
of the devices connected. The heang systems are connually
monitored for safe operaon.
Various heang opons are available for the Integra 3.
A brief descripon and how to idenfy each opon is detailed below.
Refer to the Integra 3 Installaon and Service Manual (6-54-701IM) for
further informaon about changing the heang opon (mode).
To ensure embryo safety, the temperature setpoint should
be set to maintain the specimen at the correct temperature.
See ‘Adjusng Specimen Temperature’.
In the event of a temperature sensor malfuncon, the setpoint
temperature may be exceeded and the surface may become hot
to the touch. It is recommended that temperatures are checked
on a regular basis.
The Thermosafe Air Heang System passes a steady stream of warm air
across the central hole of a Heated Metal Insert. Air passes through a
lter, fan and heater before exing from the nozzle located close to the
central hole of the Heated Metal Insert.
It is normal for the air heater enclosure to become warm when in
use.
The temperature of air exing the nozzle is controlled to a high degree of
accuracy, with the quanty of air preset to an opmum value to give the
most consistent heang across the central hole of the metal insert. As
with most microscope based heang systems, sample temperature will
be aected slightly by ambient condions.
The system has been designed to not interfere with the normal operaon
of the manipulaon system. Consequently, all components of this system
are located underneath the baseplate, with the excepon of the lter
which is located at the rear of the baseplate for ease of access.
Refer to the Care and Maintenance secon for requirements to change
the Thermosafe Air Heang System lter (Figure 11-2 on page 67).
The Thermosafe Air Heang System has been designed to
minimise audible noise and vibraon, which should not be
noceable under normal operang condions. In the event of
audible noise or vibraon from the system, contact RI for support.
The Heated Stage Plate and, when ed, the Heated Metal Insert
are shown on the main screen. The setpoint of each can be changed
individually, however, in pracce the setpoint of each will normally be
set to a similar value.
The Heated Stage Plate and when ed, the Heated Glass Insert
are shown on the main screen. The setpoint of each can be changed
individually. In pracce the setpoint temperature of the Heated Glass
Insert will normally be set slightly higher than the Heated Metal Stage.
The Heated Stage Plate and Heated Metal Insert can be used with the
Thermosafe Air Heang System. The setpoint of each can be changed
individually, however in pracce the setpoint of each will normally be
set to a similar value.
The Integra 3 can be congured to operate only the Heated Stage Plate.
This is typically only required when using the Integra 3 with an externally
controlled stage insert not manufactured by RI.
When using an externally controlled heang stage insert Mode 4 enables
the Heated Stage Plate setpoint temperature to be adjusted between 30
and 45°C.
When using an externally controlled heang/cooling stage insert, Mode
5 enables the Heated Stage Plate setpoint temperature to be adjusted
between 30 and 45°C or switched o.