Covidien VNUS RFG2 User manual

VNUS™
Radio Frequency Generator
RFG
PLUS
Model RFG2 Operator’s Manual

en
Table of Contents
Introduction ............................................................... 1
Intended Use ...................................................................1
General Safety Guidelines ........................................................1
Precautions/Operation Warnings and Cautions.......................................2
RF Treatment Warnings and Cautions ...............................................2
Environmental Conditions ........................................................3
Unpacking and Inspection of Components ..........................................3
Setup and Installation............................................................4
Mechanical Specications ........................................................4
Equipment Type.................................................................4
Equipment Description ...................................................... 5
Labeling Symbols ...............................................................5
Visual Indicators ................................................................6
Audible Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Front Panel .....................................................................7
Rear Panel .....................................................................8
LCD Screen – ClosureFAST .........................................................9
LCD Screen – ClosureRFS.........................................................10
Gauges – ClosureFAST ..........................................................11
Gauges – ClosureRFS............................................................12
Instructions for Use ........................................................ 13
AC Power......................................................................13
Changing Settings..............................................................13
Connecting a Device ............................................................14
Measure Mode .................................................................15
RF Treatment - ClosureFAST ......................................................16
RF Treatment - ClosureRFS .......................................................18
Using MultiMediaCards (MMCs) ..................................................19
Functional and Advisory Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Operating Modes .......................................................... 21
Power On Self Test (POST) Mode ..................................................21
Identication Mode.............................................................21
Measure Mode .................................................................22
Saline Test Mode - ClosureRFS only ................................................22
Body Test Mode - ClosureRFS only.................................................23
Alarm Acknowledge Mode.......................................................23
Ready Mode - ClosureFAST only...................................................23
RF Treatment Mode .............................................................23
Error Mode ....................................................................24
RFG Setup Mode ...............................................................24
Device Setup Mode .............................................................25
Maintenance.............................................................. 26
Cleaning and Disinfection .......................................................26
Disposal ......................................................................26
Troubleshooting........................................................... 27
Technical Assistance ............................................................27
Error Codes ....................................................................27
RF Treatment Diculties.........................................................27
Electromagnetic Interference (EMI) ...............................................30
Glossary of Terms .......................................................... 31

2
Français
Operator’s Manual
1
Français
VNUS Radio Frequency Generator
Introduction
Prior to using the RF Generator, review the warnings and cautions in this manual for important safety
information.
The VNUS Radiofrequency Generator (RFG
PLUS
) is a product of VNUS Medical Technologies, Inc. It is
designed to provide controlled delivery of radiofrequency (RF) energy to RF Devices (e.g., catheters)
marketed by VNUS.
The term “Device,” used throughout this manual, refers to compatible VNUS RF Devices that can be
attached to the RF Generator.
Please refer to the Instructions for Use included with the RF Device for information related to the Device such
as clinical procedures, cautions, warnings, precautions, potential complications, and contraindications.
Intended Use
The VNUS RF Generator model RFG2 is intended for use with radiofrequency devices intended for vessel
and tissue coagulation.
The RF Generator measures and displays RF output Power, load Impedance (ClosureRFS only), and
elapsed time of RF delivery. The RF Generator also interfaces with a sensor in the Device to provide a
continuous display of measured Temperature during RF delivery.
General Safety Guidelines
The safe and effective use of RF energy is highly dependent on factors under the operator’s control. There
is no substitute for properly trained staff. It is important to read, understand, and follow these directions.
The RF Generator is classified as an electrosurgical product. It must be operated in accordance with the
safety guidelines described in this section to ensure a safe environment and safe delivery of RF energy to
the patient or to the Device. Failure to adhere to these safety guidelines might result in damage to the RF
Generator and/or injury to the patient.
The RF Generator is for prescription use only. It is intended for use only by trained clinicians (operators) in
a hospital or clinical environment.
Caution: US federal law restricts this device to sale by or on the order of a physician.
The RF Generator contains no operator-serviceable parts. The RF Generator must be either returned to
the factory or repaired by qualified service personnel.
Precautions/Operation Warnings and Cautions
Before operating the RF Generator, review and adhere to these guidelines:
• DonotusetheRFGeneratoriftheACPowercordisdamaged.
• Inspectallcordsandcablesregularlyforwearordamage.Discontinueuseanddiscardifdamaged.
• Donotuse3–to–2pronggroundingadapters.
• AlwaysuseanACPowercenter-tapconfigurationfor240VsystemsintheU.S.
• DonotoperatetheRFGeneratorincontactwithflammablematerials,chemicals,orsubstances.
• ProtecttheRFGeneratorfromexposuretoextrememoisture,suchasrain.
• VerifythattheRFGenerator’sventopening,locatedatthebottomoftheunit,isnotobstructed.
Covering the speaker opening might prevent the operator from detecting auditory tones.
• DonotremovethecoveroftheRFGenerator.Thereisapotentialforelectricalshock.
• Alwaysrefertoauthorizedpersonnelforservice.
• AnRFGeneratorfailurecouldresultinanunwantedincreaseinRFPower.
WARNING: The RF Generator produces high voltages on the Device.
RF Treatment Warnings and Cautions
Before performing RF Treatment, review and adhere to these guidelines:
• DANGER,EXPLOSIONHAZARD:Donotuseinthepresenceofaflammableanestheticmixturewithair
orwithoxygenornitrousoxide.
• UsetheVNUSRFGeneratoronlywithVNUSClosureFAST or ClosureRFS. DO NOT use with Devices from
any other manufacturer with the RFG2 as this may increase the possibility of increased emissions or
decreased immunity.
• Avoidcontactofcordsandcableswithpatient,leads,orotherequipment.
• DonotwraptheDevicecablearoundmetalobjects,asthismightinducehazardouscurrentsintothe
patient.
• Bealertforpotentialinterferencewithpacemakersandotheractiveimplants.
• Topreventpatientinjury,alwayschecktodetermineifthemeasuredTemperature,Impedance(not
displayed for ClosureFAST), and RF Power shown in the Data Display area are within safe ranges.
• WhileusingtheRFGeneratorduringaprocedure,donotallowthepatienttocomeintodirectcontact
with grounded metal objects.
• WhentheRFGeneratorisactivated,theconductedandradiatedelectricalfieldsmightinterferewith
other medical electrical equipment. Refer to the“Electromagnetic Interference (EMI)”section of this
manual for more information.
• DonotactivateRFPoweruntiltheDeviceisproperlypositionedinthepatient.
• RFPoweractivationtonesandlightsareimportantsafetyfeatures.Donotobstructlights.Donot
disable auditory tones.

4
Français
Operator’s Manual
3
Français
VNUS Radio Frequency Generator
Environmental Conditions
Transportation and Storage
During transportation and storage, adhere to these guidelines:
• StoretheRFGeneratorattemperaturesbetween–20°Cand70°Candin
non-condensing relative humidity ≤ 90%.
• DonotstackitemsontopoftheRFGenerator.
• WhentheRFGeneratorisnotinuse,unplugtheunitfromtheACPoweroutletandwrapthecord
around the AC Power cord wrap area.
• DonotcarryorlifttheRFGeneratorbytheACPowercord.
• ThechemicalsfromabrokenLCDscreenaretoxicwheningested;handlewithcare,especiallyifthe
LCD screen is broken. If chemicals from a broken LCD screen are accidentally ingested, obtain medical
attention immediately.
Operation
Conditions for operating the RF Generator include the following:
• Temperaturebetween10ºCand40ºC
• Relativehumidity(non-condensing)between0%and90%
Caution: After removing the RF Generator from a storage environment, allow the unit to fully acclimate to
the new environment before turning on the unit. Operate the RF Generator only at temperatures between
10°C and 40°C. Operate the RF Generator only in non-condensing humidity conditions.
Unpacking and Inspection of Components
Upon receiving the RF Generator, carefully unpack and inspect the items listed below:
• RFGeneratorwithoutphysicaldamagetothefrontorbackpanels,cover,orscreen
• Approvedhospital-gradeACPowercordwithoutcracks,frays,oranyvisiblecordorplugdamage
• CD-ROMcontainingthisOperator’sManualandtheServiceManual
All damaged items must be repacked and returned to VNUS Customer Service Department with prior
approval. See the back of this manual for contact information.
Setup and Installation
The RF Generator must be installed and put into service according to the guidance provided in this document
to ensure its electromagnetic compatibility. If in doubt, consult VNUS Customer Service or a local distributor.
TheRFGeneratorshouldbeusedonlywithaHospitalGradepowercordandpluggedintoonlygrounded,
HospitalGradeACsources.
The VNUS RF Generator can be placed on any stable cart, table, or platform. It is recommended that the
cart have conductive wheels and be rated to hold at least 11.5 kg (25 lbs.). Refer to facility protocols and
local ordinances for more information.
Provide at least 10 to 15 cm (4–6 in.) of space around the sides and top of the RF Generator for air
circulationandcooling.Aftercontinuoususeforextendedperiodsoftime,itisnormalforthetopand
rear panel to be warm. Do not block vent openings.
Warning: Do not stack any items on top of the RF Generator. Doing so could damage the
unit. The RF Generator should not be used adjacent to other equipment. If adjacent or
stacked use is necessary, the RF Generator should be observed to verify normal operation in
the configuration in which it will be used.
Mechanical Specifications
• Size:38cmx38cmx19cm(15in.x15in.x7.5in.)maximum
• Weight:9kg(20lbs.)maximum
Equipment Type
The Class I RF Generator is designed to work with Type CF, Defibrillator-Proof RF Devices.
TheRFGeneratorisdesignedtowithstandanapplicationofanexternaldefibrillatorwhile
the RF Device is in use.

6
Français
Operator’s Manual
5
Français
VNUS Radio Frequency Generator
Equipment Description
Labeling Symbols
There are several symbols and icons used throughout this manual that are used by the RF Generator and
its displays. The operator should be familiar with these symbols and their meanings.
Table 1: Data Display Area Symbols and Icons
ΩOhms (Impedance) Information, Question, or
Verication
°C Degrees Celsius
(Temperature)
Warning: Consult
alarm/advisory message
WWatts (Power) Data Logging Indicator
OR
ClosureFAST icon:
RF not actively
powered OR
ClosureFAST icon:
RF actively powered to cause
tissue heating
Table 2: RF Generator Symbols and Icons
AC Power ON Non-IonizingRadiation
AC Power OFF
Earth Ground
RF Power ON MMC Card MultiMediaCard (MMC) Slot
RF Power OFF Increment/UP
Attention: Consult
accompanying documents Decrement/DOWN
Fuses Patient Isolated
Alternating Current Electrostatic Sensitive Device
(referring to MMC card)
Visual Indicators
The RF Generator includes the following visual indicators, as depicted in Figure 1:
• AC Power: Indicates when the RF Generator is turned on.
• RF Power: Indicates RF Power status by changing color and illumination, as listed in Table 3.
Table 3: RF Power Button Status
Light Color Status Mode
OFF None RF Power is disabled. Identication or
Measure
ON Green RF Power is enabled
(ready for use). Measure
ON Flashing Green
RF Power is enabled,
ready for a catheter
or button press
to activate RF
treatment
Ready
ON White
RF Power is activated
(in use) and currently
delivering RF Power.
RF Treatment
Treatment RF Power cannot be enabled until a Device is connected.
Audible Indicators
The RF Generator has several audible indicators designed to alert the operator:
• Alarm: Set of three rapid, high-frequency tones. Sound pressure level of alarm tones is
> 45dB (procedural tone volume at quietest setting) and
> 80dB (procedural tone volume at loudest setting).
• Informational: Single short, low-frequency tone
• Invalid: Single short, low-frequency tone
• Power On: Three ascending-scale tones
• RF Power On – Treatment: Two descending-scale tones
• RF Power On – Alert: Two rapid groups of two high-frequency tones
• RF Start: Single, long low-frequency tone
• RF Stop: Single, long low-frequency tone
• Valid: Single short, high-frequency tone
Each button press produces a valid or invalid tone. If a button fails to sound, this indicates button
malfunction. Should a button malfunction, have the unit serviced.

8
Français
Operator’s Manual
7
Français
VNUS Radio Frequency Generator
Front Panel
LCD Screen
RF Power
Button and
Indicator
Delivery
Device
Receptacle
Increment/UP and
Decrement/DOWN Buttons
Soft Key Menu
Buttons
AC Power
Indicator
Vent Slots
Figure 1: Front Panel
The RF Generator includes the following buttons on its front panel, as depicted in Figure 1:
• SoftKeyMenu:ActivatetheSoftKeyMenusdisplayeddirectlyabovethebuttonsontheLCDdisplay.
• Increment/UP(p) and Decrement/DOWN (q): Cycle up or down through the available menu
choices. After selecting a choice, press these buttons to increase, decrease, or toggle the value of that
choice.
• RFPower:ActivatesordeactivatesRFPower.
Rear Panel
AC Power Cord Wrap
AC Power Inlet
Equipotential
Grounding Lug
Fuse Holder
MMC
Slot
Handle
Vent
Openings
Service
Port
AC Power Switch
Figure 2: Rear Panel
The RF Generator includes these features, as depicted in Figure 2:
• AC Power Switch: Turns on or off the RF Generator’s AC Power.
• AC Power Cord Wrap: When the RF Generator is not in use, unplug the unit from the AC Power
source and wrap the cord around the AC Power cord wrap area.
• AC Power Inlet: Connects the AC Power cord to the RF Generator.
• Equipotential Grounding Lug: ProvidesapointforanauxiliaryEarthGroundconnection.
• Fuse Holder:Includestwo4A/250V,slo-blo,5x20mmfuses.
• Handle: The RF Generator has a handle on each side. Use these handles while carrying or lifting the
unit.
• MultiMediaCard (MMC) Slot: Accepts a MultiMediaCard that can be used to store procedure data
and to perform software upgrades.
• Service Port: The RF Generator includes a port that is used for calibration and repairs. This port is for
use by qualified service personnel only.
• Vent Openings: The RF Generator includes vent openings on the rear panel. Do not block these
openings.

10
Français
Operator’s Manual
9
Français
VNUS Radio Frequency Generator
LCD Screen – ClosureFAST
TheLCDscreenconsistsofthreeareas:theDataDisplayarea,OperatorMessagearea,andSoftKeyMenuarea.
Device
Temperature
Data Logging
Indicator Timer Device ID
Operator
Message
Area
Data Display
Area
Soft Key
Menu
Area
Progress Meter RF Power
Gauges
Figure 3: ClosureFAST Main Screen
The Data Display area displays information about the RF Generator’s status. Depending on the status of
the RF Generator, information in this area changes to show the following items:
• Data Logging Indicator: Indicates that a MultiMediaCard has been inserted and data is being
logged.
• Device ID: Indicates type of Device connected to the RF Generator.
• Device Temperature: Indicates current Temperature of connected Device.
• Gauges: Indicates Temperature and RF Power with an acceptable range shown for Temperature.
• Progress Meter: Indicates treatment time remaining
• Timer: The timer counts down starting from the selected treatment time.
•RF Power: Indicates current RF Power delivered to the Device.
LCD Screen – ClosureRFS
Device
Temperature
Temperature
Set Point
Data Logging
indicator Timer Device ID
Operator
Message
Area
Data Display
Area
Soft Key
Menu
Area
Device
Impedance
RF Power
Gauges
Figure 4: ClosureRFS Main Screen
The Data Display area displays information about the RF Generator’s status. Depending on the status of
the RF Generator, information in this area changes to show the following items:
• Data Logging Indicator: Indicates that a MultiMediaCard has been inserted and data is being
logged.
• Device ID: Indicates type of Device connected to the RF Generator.
• Device Impedance: Indicates current Impedance of connected Device
• Device Temperature:IndicatescurrentTemperatureofconnectedDevice;itisnotintendedto
measure patient body temperature.
• Gauges: Indicates whetherTemperature, Impedance, and RF Power are within or outside of preset limits.
• Timer: Indicates the amount of time RF Power has been delivered during the current RF Power On
cycle.
• RF Power: Indicates current RF Power delivered through the Device.

12
Français
Operator’s Manual
11
Français
VNUS Radio Frequency Generator
Gauges – ClosureFAST
The RF Generator’s Data Display area includes colored gauges when a Device is connected. These gauges
provide a quick and easy way to determine if the Device’s parameters (Temperature and RF Power) are
withinacceptableranges.Ifaparameterisnotwithinanexpectedrange,theRFGeneratordisplaysan
advisory message and sounds a tone.
Gauge within
normal range
Upper Advisory
Limit violated
Power Gauge with reference
ticks at 10, 20, 30 and 40 Watts
Operator
Message
Area
Lower
Advisory
Limit
violated
Figure 5: ClosureFAST Data Display with Gauges
The position of the Temperature gauge’s needle is based on the current value. It points to one of three
colored areas, as described below:
Green: Within normal range
Red: Temperature above Advisory Limit
Blue: Temperature below Advisory Limit
TheextentofilluminationofthePowergaugeisbasedonthecurrentPowervalue.
Green: Within normal range
Red: Power above Advisory Limit
For more information about these Limits, see the “Functional and Advisory Limits” section of this manual.
Gauges – ClosureRFS
The RF Generator’s Data Display area includes colored gauges when a Device is connected and not in
Measure Mode. These gauges provide a quick and easy way to determine if the Device’s parameters
(Temperature, Impedance, and RF Power) are within acceptable ranges. If a parameter is not within an
expectedrange,theRFGeneratordisplaysanadvisorymessageandsoundsatone.
Gauge within
normal range
Upper Advisory
Limit violated
Gauge within
normal range
Operator
Message
Figure 6: ClosureRFS Data Display with Gauges
The position of the gauge’s needle is based on the current value of the corresponding parameter
(Temperature, Impedance, or RF Power). It points to one of three colored areas, as described below:
• Green: Within normal range
• Yellow: Violation of Advisory Limits
• Red: Violation of Functional Limits
Red
(lower)
Yellow
(upper)
Red
(upper)
Yellow
(lower) Green
Gauge
needle
Figure 7: Gauge
For more information about these Limits, see the “Functional and Advisory Limits” section of this manual.

14
Français
Operator’s Manual
13
Français
VNUS Radio Frequency Generator
Instructions for Use
To ensure safe RF delivery, it is important to be familiar with how the RF Generator works, how to change
desired settings, and how to interpret displays.
AC Power
The VNUS RF Generator includes an approved hospital-grade AC Power cord.
Prior to each use, inspect the AC Power cord, the AC Power inlet, and the AC Power outlet that the unit
plugs into.
To turn on the RF Generator:
1. Plug the AC Power cord into the AC Power inlet at the back of the unit.
2. Plug the AC Power cord into a grounded hospital-grade AC Power outlet.
3. Press the AC Power switch UP to turn on the unit.
4. Verify the AC Power indicator is lit when the unit turns on.
If the indicator is not lit, check the AC Power cord, fuses, and AC Power outlet.
5. Verifyallpixelsontheunit’sLCDscreenilluminatewhitewhenturningontheunit.
Caution: Do not plug the RF Generator into the AC Power outlet if any of its surfaces appear to be cracked or
damaged. Do not use the RF Generator if the AC Power cord is damaged. Do not use extension cords and/or
adapter plugs.
Changing Settings
The RF Generator lets the operator change the values of Device and RF Generator settings. The operator
should be familiar with how to change setting values:
1. Use the Soft Key Menu buttons located below the Data Display area to choose the DEVICE SETUP or
RFG SETUPSoftKeyMenu.
The Operator Message area displays a sub-menu of settings.
TheSoftKeyMenuareadisplaysrelatedSoftKeyMenus:
• DEFAULTS:ReturnsallsettingstotheRFGenerator’sfactorydefaults(exceptforDate,Time,and
Language settings).
• SELECT: Indicates the options for the highlighted setting.
• EXIT: Returns to the previous screen.
2. Use the Increment/UP (p) and Decrement/DOWN (q) buttons to highlight a setting.
3. Press the SELECTSoftKeyMenubuttontoselectthehighlightedsetting.
The Operator Message area indicates the value(s) for the highlighted setting.
TheSoftKeyMenuareadisplaysrelatedSoftKeyMenus:
• DEFAULT:Restorestheselectedsettingtothefactorydefaultvalue(exceptforDate,Time,and
Language settings).
• ACCEPT: Saves the setting’s value(s) as the default value(s).
• CANCEL: Cancels the selection and returns to the previous screen.
4. Use the Increment/UP (p) and Decrement/DOWN (q) buttons to increment, decrement, or
toggle the value(s) of the selected setting.
5. Press the ACCEPTSoftKeyMenubuttontosavethechoiceasadefaultsettingorpresstheCANCEL
SoftKeyMenubuttontoreturntothepreviousscreenwithoutsavinganychanges.
6. After changing the desired setting(s), press the EXITSoftKeyMenubuttontoreturntotheRF
Generator screen.
Connecting a Device
To connect a Device, insert the Device’s connector into the receptacle on the RF Generator. If the unit is
turned on before a Device is attached, the RF Generator prompts the operator to connect a Device.
Notice the small red dot at the top of the Device receptacle. This dot must be lined up with the small red mark
and/or raised line on the Device connector being connected to the RF Generator.
Device
connector
Device
cable
Device
receptacle
Figure 8: Connecting a Device
When a Device is connected, the RF Generator’s LCD screen displays RF Treatment information, as shown
in Figures 9 and 10.

16
Français
Operator’s Manual
15
Français
VNUS Radio Frequency Generator
Measure Mode
RF Power
Device
Temperature
Data Logging
Indicator Timer Device ID
Soft Key
Menu Area
Figure 9: ClosureFAST Measure Mode
RF Power
Device
Temperature
Device
Impedance
Data Logging
Indicator Timer Device ID
Soft Key
Menu Area
Figure 10: ClosureRFS Measure Mode
RF Treatment - ClosureFAST
WARNING: Review this section and the “General Safety Guidelines” section of this manual
before starting RF Treatment.
Before starting RF Treatment, review the Device’s Instructions For Use. Confirm that the connected Device
matches the Device ID shown in the Data Display area.
Entering Ready Mode
Press the RF Power button to enter Ready Mode. The RF Power button will change from non-
illuminated to flashing green illumination.
Initiating RF Treatment
Press the START RF button or the Device handle button to start RF Treatment. While delivering RF
Power, the RF Power button illuminates white. The Device will increase to the set temperature and
hold at that temperature until the treatment time has completed. After the treatment time has counted
downtozero,RFPowerdeliverywillstopautomaticallyandtheRF Power button will change to
flashing green illumination to indicate Ready Mode. Pressing the START RF button or the Device handle
buttonwillstartthenextRFTreatment.
WARNING: Do not activate RF Power until the Device is properly positioned in the patient.
Refer to the “General Safety Guidelines” section of this manual before starting RF Treatment.
Data Logging
Indicator Countdown Timer Device ID
Soft Key
Menu
Gauges
during
RF
Treatment
Mode
Operator
Message
Area
Figure 11: ClosureFAST RF Treatment Started

18
Français
Operator’s Manual
17
Français
VNUS Radio Frequency Generator
Temperature and Impedance must be within Functional Limits before RF Treatment can start. This
temperature display is not intended to measure patient temperature but reflects Device temperature within
a relative range.
Halting RF Treatment
During RF Delivery, press the STOP RF button or the Device handle button to halt RF Treatment (pressing
the RF Power button will also halt RF Treatment and disable Ready mode).
RF Treatment halts automatically if:
• ThetimesetforRFTreatmenthaselapsed(forDeviceswhereRFTreatmenttimecanbeset).
• AFunctionalLimithasbeenviolated.
• TheRFGeneratordetectsanerror.
• TheDeviceisdisconnected.
WARNING: If RF Treatment does not halt automatically, disconnect the Device.
Restarting RF Treatment
The operator should correct any alarm condition before restarting RF Treatment. (See the “RF Treatment
Difficulties” section of this manual for more information.)
If the operator unplugs a Device and then reconnects a Device of the same model, the unit prompts the
operator to either retain or discard the Timer Value. Press the YES or NOSoftKeyMenubuttontoretain
or discard, respectively, the sum of all times from all previous RF Power On cycles.
Iftheunithasbeenturnedoff,anewTimerValuewillstartwiththetimeresettozero(0:00)whenthe
unit is turned on again.
The Total Treatment Time displays in the Operator Message area immediately after RF Power halts.
RF Treatment - ClosureRFS
WARNING: Review this section and the “General Safety Guidelines” section of this manual
before starting RF Treatment.
Before starting RF Treatment, review the Device’s Instructions For Use. Confirm that the connected Device
matches the Device ID shown in the Data Display area.
It is also highly recommended that the operator perform the optional Saline and Body Tests. These tests
verify the Temperature and Impedance measurements of the Device. For information on performing the
Saline Test and Body Test, refer to the “Saline Test Mode” and “Body Test Mode” sections of this manual.
When the RF Generator and connected Device are ready to start RF Treatment, the RF Power button
illuminates green. This indicates that Temperature and Impedance measurements are within the
Functional range for the connected Device, and RF Treatment can start.
Temperature and Impedance must be within Functional Limits before RF Treatment can start.
Initiating RF Treatment
Press the RF Power button to start RF Treatment. While delivering RF Power, the RF Power button
illuminates white. The Device will increase to the set temperature and hold at that temperature until the
RF Treatment is halted.
WARNING: Do not activate RF Power until the Device is properly positioned in the patient.
Refer to the “General Safety Guidelines” section of this manual before starting RF Treatment.
Data Logging
Indicator Timer Device ID
Gauges
during
Saline Test
and RF
Treatment
Modes
Operator
Message
Area Soft Key
Menu
Commands
Figure 12: ClosureRFS RF Treatment Started

20
Français
Operator’s Manual
19
Français
VNUS Radio Frequency Generator
Halting RF Treatment
During RF Delivery, press the RF Power button to halt RF Treatment.
RF Treatment halts automatically if:
• AFunctionalLimithasbeenviolated.
• TheRFGeneratordetectsanerror.
• TheDeviceisdisconnected.
Restarting RF Treatment
The operator should correct any alarm condition before restarting RF Treatment. See the “RF Treatment
Difficulties” section of this manual for information.
If the operator unplugs a Device and then reconnects a Device of the same model, the unit prompts the
operator to either retain or discard the Timer Value. Press the YES or NOSoftKeyMenubuttontoretain
or discard, respectively, the sum of all times from all previous RF Power On cycles.
Iftheunithasbeenturnedoff,anewTimerValuewillstartwiththetimeresettozero(0:00)whenthe
unit is turned on again.
The Total Treatment Time displays in the Operator Message area immediately after RF Power halts.
Using MultiMediaCards
MultiMediaCards (MMCs) can store data from recent procedures, log generator errors, and perform
software updates. The operator must use VNUS part number MMC.
WARNING: Proper ESD precautionary procedures should be used when handling,
inserting, or removing the MMC card.
Before handling the MMC card, touch the metal frame of the RFG2 to discharge any static electricity.
Alternatively, you may use an ESD wrist strap to bond yourself to the RFG2 or to earth. This is important
to avoid electrostatic discharge damage to the MMC card.
AllpersonnelwhohandletheMMCcardshouldreceiveanexplanationoftheESDwarningsymboland
training in ESD precautionary procedures. This includes clinical/biomedical engineering and health-care
staff. At minimum, this training should include:
• an introduction to the physics of electrostatic charge
• the voltage levels that can occur in normal practice
• the damage that can be done to electronic components if equipment are touched by an operator
who is electrostatically charged
• methods to prevent build-up of electrostatic charge,
• how and why to discharge one’s body to earth or to the frame of the RFG2, or bond oneself by means
of a wrist strap to the RFG2 or to earth prior to handling the MMC card
Functional and Advisory Limits
While a Device is connected, the RF Generator monitors the Device’s parameters (Temperature,
Impedance, and RF Power). Parameters must be within Functional Limits before RF Treatment can start.
The Limits are determined by predefined settings for the connected Device. Advisory Limits are within
the range of Functional Limits.
WARNING: Operation of the RF Generator below these values may cause inaccurate results.
The RF Generator will not allow treatment if measurements persist outside of these limits.
The limits can also be affected by Set Points that the operator defines in the Device Setup menu. See the
“Device Setup Mode” section of this manual for information about defining Set Points.
Functional Limits
Functional limits for ClosureFAST:
• PatientTemperature:10°Cto130°C
Functional limits for ClosureRFS:
• PatientTemperature:10°Cto110°C
• DeviceImpedanceinTreatment,Saline,orBodyTestModes:25-1000Ω
• DeviceImpedanceinMeasureMode:≥10Ω
If the RF Generator detects that a parameter violates Functional Limits, the gauge displays within the red
range. The unit also:
• HaltsRFTreatment.
• SoundsanAlarmTone(forTemperatureorRFPowerparameters)orsoundsanInformationalTone
(for the Impedance parameter).
• EntersAlarmAcknowledgeMode.
• DisplaysanAlarmMessage.
Advisory Limits - ClosureFAST
During RF treatment:
• Ifthesettemperatureisnotreachedwithinapredeterminedtime,RFpowerwillbestoppedbefore
the countdown has completed. This indicates a condition of poor contact between the Device and the
tissue to be treated.
• IftheRFGeneratordetectsthattemperatureisbelowtheAdvisoryLimit,thegaugedisplayswithin
the blue range. The unit sounds an Informational Tone and displays a message.
Advisory Limits - ClosureRFS
If the RF Generator detects that a parameter violates Advisory Limits, the gauge displays within the
yellow range. The unit sounds an Informational Tone and displays a message.

22
Français
Operator’s Manual
21
Français
VNUS Radio Frequency Generator
Operating Modes
The RF Generator has several operating modes. The operator must pay special attention to the Data
Display area and Operator Message area during the Body Test, Saline Test, Measure, and RF Treatment
modes, in which some level of RF Power is delivered.
Failure during any operating mode results in the RF Generator either not starting RF Treatment or
automatically halting RF Treatment.
Power On Self Test (POST) Mode
During the Power On Self Test (POST), the RF Generator performs a number of internal self-tests of its
software:
• Verifiesintegrityofsystemrandomaccessmemory(RAM).
• Verifiescontentsofprogramread-onlymemory(ROM).
• Activatesallvisualindicators,includingtheLCDscreenandthebacklitSoftKeyMenubuttonsaround
the screen.
• SoundsPowerOntone.
• Verifiesoperationofradiofrequency(RF)generationandmeasurementcircuits.
• DisplaysanoticeifPOSTmustrestorethefactorydefaultsettingsduetoacorruptionofpersistent
settings, such as screen brightness or tone volume settings.
Prior to using the RF Generator, observe that the screen, lights, and audible alarms cycle.
Caution: The RF Generator enters Error Mode if it detects a failure during the Power On Self Test (POST)
Mode. If any error messages occur indicating that the RF Generator is not safe to use, power off the unit and
contact VNUS Customer Service.
Identification Mode
During Identification Mode, the RF Generator checks for a connected Device:
• IfnoDeviceisconnected,theOperatorMessageareaaskstheoperatorto“Pleaseconnectdevice.”
• IftheconnectedDeviceisunsupported,theOperatorMessageareasays,“Theconnecteddeviceis
unsupported. Please connect another device.”
• IftheconnectedDeviceisunknownorinvalid,theOperatorMessageareasays,“Thedeviceisinvalid.”
The Data Display area displays the current software version and the VNUS logo, as shown in Figure 13.
RFG2 - VERSION 4.1.0
Data Display
Area
Operator
Message
Press the Soft Key
Menu button
below a menu
command to
view or change
settings.
Figure 13: Identification Mode
Measure Mode
Measure Mode starts whenever a Device is attached to the RF Generator. During Measure Mode, the Data
Display area displays the current parameters for the connected Device.
For ClosureFAST, violating Functional Limits while in Measure Mode results in an advisory message. The
operator cannot start RF Treatment if the Device violates Functional Limits.
Caution: For ClosureRFS the Measure, Saline Test, and Body Test modes use a small amount of RF energy (2.3
mW
RMS
) to measure Impedance at the Device tip. During these modes, output power is insufficient to cause
tissue heating. However, the operator should be alert for interference to nearby equipment. For more
information about interference, please consult the “Electromagnetic Interference (EMI)” section of this
manual.
Saline Test Mode - ClosureRFS only
The Saline Test performs Temperature and Impedance measurements using 0.9% saline solution.
When the operator presses the SALINE TESTSoftKeyMenubutton,theSalineTestdisplaysfor
approximately10secondsbeforetimingout.Pressingthebuttonagainwillrestartthetest.TheSaline
Test displays the relationship between:
• ThemeasuredImpedanceandtheexpectedImpedancerange.
• ThemeasuredTemperatureandtheexpectedTemperature.
Values can vary due to actual saline temperature and should be verified before RF Treatment starts.
Caution: If a Saline Test indicates advisory conditions, this might indicate a faulty Device. Consult the Device’s
Instructions for Use for more information.

24
Français
Operator’s Manual
23
Français
VNUS Radio Frequency Generator
Body Test Mode - ClosureRFS only
The Body Test performs Temperature and Impedance measurements using the Device positioned in the
patient for RF Treatment.
When the operator presses the BODY TESTSoftKeyMenubutton,theBodyTestdisplaysfor
approximately15secondsbeforetimingout.Pressingthebuttonagainwillrestartthetest.TheBody
Test displays the relationship between:
• ThemeasuredImpedanceandtheexpectedImpedancerange.
• ThemeasuredTemperatureandtheexpectedTemperaturerange.
When the Body Test has completed, the most recent Temperature and Impedance measurements display
in the Operator Message area.
Caution: If a Body Test indicates advisory conditions, this might indicate a faulty Device. Consult the Device’s
Instructions for Use for more information.
Alarm Acknowledge Mode
The Alarm Acknowledge Mode starts when RF Treatment halts due to an alarm or unrecoverable error.
See the “RF Treatment Difficulties” section of this manual for information about unrecoverable errors. To
end the Alarm Acknowledge Mode, press the OKSoftKeyMenubuttonorwait30secondsforthemode
to complete.
Once the Alarm Acknowledge Mode has completed, the RF Generator enters Measure Mode if a Device is
connected or Identification Mode if a Device is not connected.
Ready Mode - ClosureFAST only
Ready Mode starts after a ClosureFAST Device is attached to the RF Generator and the RF Power button
is depressed as prompted on the screen. During Ready Mode, the Data Display area displays the current
parameters for the connected Device. During Ready Mode, the RF Generator is enabled and will deliver
RF power when the START RF button or the Device handle button is pressed.
RF Treatment Mode
During the RF Treatment Mode, the RF Generator delivers RF Power. The RF Treatment Mode becomes
active when the following are true:
ClosureFast:
• TemperatureandImpedancemeasurementsarewithinFunctionalrangefortheconnectedDevice.
• TheRF Power button has been pressed to enter ready mode.
• TheoperatorpressestheSTART RF button or the Device handle button (note that if body
temperaturehasnotbeenmeasured,analertwilldisplay;thisalertcanbeclearedbypressingOK to
proceed).
ClosureRFS:
• TheoperatorpressestheRF Power button.
When RF Treatment has completed, the RF Generator enters Measure Mode or Alarm Acknowledge
Mode. For more information about RF Treatment, read the “RF Treatment” sections of this manual.
Error Mode
Error messages display when the RF Generator encounters an unrecoverable (unable to continue) error.
The Error Mode becomes active and an error code or message displays if any of the following occurs:
• TheRFGeneratordetectsanunexpectedinternalerror.
• ThePowerOnSelfTest(POST)fails.
When the Error Mode occurs, RF Treatment halts immediately and the RF Generator logs an entry in the
Error Log.
When an error message indicates a non-recoverable error has occurred, turn off the unit, wait a few
seconds, and then turn on the unit.
If an error persists, make a note of it and then contact VNUS Customer Service or a local distributor.
RFG Setup Mode
The RFG Setup menu provides a list of settings for the RF Generator. When the operator presses the RFG
SETUPSoftKeyMenubutton,theOperatorMessageareadisplaysthesettingsavailablefortheRF
Generator.
To change a setting’s value(s):
1. Press the RFG SETUPSoftKeyMenubuttontodisplaytheRFGSetupsettings.
2. Use the Increment/UP (p) and Decrement/DOWN (q) button to highlight a setting.
3. Press the SELECTSoftKeyMenubuttontoselectthehighlightedsetting.
4. Use the Increment/UP (p) and Decrement/DOWN (q) button to cycle through the values until
the desired value displays.
5. Press the ACCEPTSoftKeyMenubuttontoacceptthechange.
6. Press the EXITSoftKeyMenubuttontoreturntotheRFGeneratorscreen.
Proc Tone Vol
The Proc Tone Vol setting adjusts the Procedural Tone Volume from 2 (quietest) to 10 (loudest) The
Procedural Tone Volume controls the volume of alarm tones and tones associated with RF Power
delivery, such as RF Power start and stop tones. When the volume setting is increased or decreased,
a single tone sounds to denote the change.
The operator is prevented from adjusting the Procedural Tone Volume to be quieter than the
Informational Tone Volume. If the operator adjusts the Procedural Tone Volume to be equal to or less
than the Informational Tone Volume, the Informational Tone Volume automatically adjusts to one level
lower than the Procedural Tone Volume.
Info Tone Vol
The Info Tone Vol setting adjusts the Informational Tone Volume from 1 (quietest) to 9 (loudest).
The Informational Tone Volume controls the volume of button acknowledgement tones and
informational tones. When the volume setting is increased or decreased, a single tone sounds to
denote the change.

26
Français
Operator’s Manual
25
Français
VNUS Radio Frequency Generator
Gauges - ClosureRFS only
The Gauges setting adjusts the Parameter Range Indication gauges as either ON or OFF.
Brightness
The Brightness setting adjusts screen Brightness from 1 (dimmest) to 10 (brightest).
Language
The Language setting selects the Language displayed on the RF Generator’s screen. The default
language is English.
The operator can change the default language from the RFG Setup Mode. The operator can also
change the language during the Power On Self Test (POST) Mode:
1. Press the AC Power switch UP to turn on the unit.
2. When the Power On Self Test (POST) starts, simultaneously press the first two (starting at left)
SoftKeyMenubuttons.
3. Changethesetting’svalueusingtheSoftKeyMenubuttonsandtheIncrement/UP (p) and
Decrement/DOWN (q) buttons.
Date
The Date setting adjusts the current Date.
Time
The Time setting adjusts the current Time.
Device Setup Mode
The Device Setup menu provides a list of settings for the connected Device. When the operator presses
the DEVICE SETUPSoftKeyMenubutton,theOperatorMessageareadisplayssettingsfortheconnected
Device.
To change a setting’s value(s):
1. Press the DEVICE SETUPSoftKeyMenubuttontodisplaytheDeviceSetupsettings.
2. Use the Increment/UP (p) and Decrement/DOWN (q) button to highlight a setting.
3. Press the SELECTSoftKeyMenubuttontoselectthehighlightedsetting.
4. Use the Increment/UP (p) and Decrement/DOWN (q)button to cycle through the values until
the desired value displays.
5. Press the ACCEPT SoftKeyMenubuttontoacceptthechange.
6. Press the EXITSoftKeyMenubuttontoreturntotheRFGeneratorscreen.
Settings
Device Setup settings differ, depending on which Device the operator connects. The Device
determinesthemaximumandminimumvaluesallowedforeachsetting.Availablesettings
include:
• Phase (1 or 2): ClosureFAST only. Adjusts the RF Treatment time (in seconds). Note that there are
twophasedsettingstoallowahighermaximumpowerduringinitialheatingfollowedbyalower
power to maintain temperature. Total treatment time is the sum of the Phase 1 and Phase 2 values,
after which treatment will stop automatically.
• Maximum Power (W):AdjuststhemaximumRFPower(inWatts).
• Temperature (ºC): Adjusts the desired target Temperature.
• RFS Advisory Interval (s): ClosureRFS only. Adjusts the time between advisory tones.
Maintenance
Cleaning and Disinfection
Caution: Always unplug the AC Power cord prior to cleaning the RF Generator.
Avoid spilling liquids into the RF Generator during cleaning.
Avoid getting cleaning materials inside Device receptacle.
TheRFGeneratorrequiresnoscheduledmaintenanceotherthancleaningexternalsurfaces.Itis
recommended that the operator clean the unit prior to each use or as needed.
The recommended agents for cleaning RF Generator surfaces (not including the LCD screen) to prevent
surface degrading or discoloration include the following:
• Fivepercentsolutionofhouseholdbleach(approximately2,500ppmSodiumHypochlorite)
• Sporicidin®
• IPA70/30
The recommended agent for cleaning the transparent, protective cover of the LCD screen is IPA 70/30.
RFGeneratorcannotbesterilizedandmustnotenterasterilesurgicalfield.
Disposal
Follow local governing ordinances and recycling plans regarding disposal or recycling of components.
WARNING: The RF Generator includes a lithium battery. The battery holder
inside the RF Generator is marked with the symbol at left.
Do not incinerate or dispose of lithium batteries in general trash collection.
Check state and local regulations dealing with the disposal of these materials.

28
Français
Operator’s Manual
27
Français
VNUS Radio Frequency Generator
Troubleshooting
Technical Assistance
To obtain technical assistance, contact VNUS Customer Service or a local distributor. See the back of this
manual for contact information.
Error Codes
If an error occurs, the RF Generator displays an error code. An error will be one of two types:
unrecoverable or recoverable:
• Unrecoverable errors halt the RF Generator and require the operator to reset the RF Generator’s AC
Power to continue. If an unrecoverable error occurs, turn off the unit, wait a few seconds, and then
turn on the unit. If the problem persists, contact VNUS Customer Service or a local distributor.
• Recoverable errors require the operator to respond to an error message or make a clinical decision.
See Service Manual for a list of errors.
RF Treatment Difficulties
During RF Treatment, the RF Generator monitors the Device’s parameters (Temperature, Impedance, and
RF Power). Should Temperature, Impedance, or RF Power become too high or too low, as indicated by its
gauge, the RF Generator may halt RF Treatment, sound an alarm, and display an alarm message
(Impedance is not displayed for ClosureFAST).
Tables 4 and 5 provide general recommendations for some situations that the operator might encounter.
They do not include every possible situation that might occur. For more information, please refer to the
Device’s Instructions for Use.
Table 4: RF Treatment Difficulties – ClosureFAST
ClosureFAST
Diculty Possible Cause Corrective Action
Unable to activate
RF Power
Temperature and/or
Impedance parameter
outside Functional
Limits
CheckTemperaturevalue;adjustasnecessarytobring
parameters into Functional range.
Ready Mode not
entered
Press the non-illuminated RF Power button and
observe that it ashes on and o to indicate the RF
generator is in ready Mode. Then, activate RF Power
when ready.
Body temperature not
measured
Verify that the Device has been inserted into the body
and is properly positioned to begin treatment. Press
OK to proceed and then press the START RF button or
the Device handle button to activate RF Power.
Connected device is
broken
Check the Device connector to verify that it is plugged
in properly. Replace Device.
ClosureFAST
Diculty Possible Cause Corrective Action
Power high Inadequate vessel
exsanguination
Verify that the Device is properly positioned. Check for
ow, and employ or improve compression techniques
as necessary.
Temperature low Inadequate vessel
exsanguination
Verify that the Device is properly positioned. Check for
ow, and employ or improve compression techniques
as necessary.
Temperature Set
Point is set too low
Check and adjust Temperature Set Point.
Saline in Device
connector
Check for presence of saline. If present, contact VNUS
Customer Service. (Note: Take care never to plug a
wet Device connector into the receptacle, as this may
damage the RF Generator).
Device, cable, or
connector is damaged
Check all connections, cable, and the Device for visible
damage. Replace Device.
Temperature
decreases as power is
applied
Replace Device.
Temperature high Saline in Device
connector
Check for presence of saline. If present, contact VNUS
Customer Service. (Note: Take care never to plug a
wet Device connector into the receptacle, as this may
damage the RF generator).
Temperature Set
Point is set too high
Check Temperature Set Point.
Non-uniform
temperature
Uneven compression
across full length of
heating element
Note Device position with shaft markers. Check Device
and replace if damaged. If not damaged, insert
Device to desired location, Verify that the Device is
properly positioned. Employ or improve compression
techniques as necessary. Press the START RF button
or the Device handle button to activate RF Power.

30
Français
Operator’s Manual
29
Français
VNUS Radio Frequency Generator
ClosureFAST
Diculty Possible Cause Corrective Action
Impedance low
(not displayed)
Saline in Device
connector and/or
receptacle
Check for presence of saline. If present, contact VNUS
Customer Service. (Note:Take care never to plug a
wet Device connector into the receptacle, as this may
damage the RF Generator.) If saline is not present in
Device connector or receptacle, perform a Saline Test
to verify functionality of Device. Replace Device or
instrument cable if Saline Test Impedance values are
out of range.
Short circuit Replace Device.
Impedance high
(not displayed)
Open circuit Check Device connection (disconnect and then re-
connect).
Table 5: RF Treatment Difficulties - ClosureRFS
ClosureRFS
Diculty Possible Cause Corrective Action
Impedance low Inadequate electrode-vein
wall contact
Improve or employ vein compression techniques.
Saline in Device connector
and/or receptacle
Check for presence of saline. If present, contact
VNUS Customer Service. (Note:Take care never to
plug a wet Device connector into the receptacle,
as this may damage the RF Generator.) If saline
is not present in Device connector or receptacle,
perform a Saline Test to verify functionality of
Device. Replace Device or instrument cable if
Saline Test Impedance values are out of range.
Short circuit Replace Device.
Impedance high Coagulum formation on
electrodes
Check Device tip for coagulum and remove as
required.
Device not in contact with
target tissue
Verify that Device is properly positioned.
Open circuit Check Device connection (disconnect and then
re-connect).
Power high Too much blood ow
Pullback rate too fast
Checkandimproveexsanguination.
Verify proper pullback rate.
ClosureRFS
Diculty Possible Cause Corrective Action
Temperature low Inadequate vessel
exsanguination
Verify that the Device is properly positioned. Check
for ow, and employ or improve compression
techniques as necessary.
Pullback rate too fast Verify proper pullback rate.
Temperature Set Point is set
too low
Check and adjust Temperature Set Point.
Saline in Device connector Check for presence of saline. If present, contact
VNUS Customer Service. (Note: Take care never to
plug a wet Device connector into the receptacle,
as this may damage the RF generator).
Device, cable, or connector
is damaged
Check all connections, cable, and the Device for
visible damage. Replace Device.
Temperature high Saline in Device connector Check for presence of saline. If present, contact
VNUS Customer Service. (Note: Take care never to
plug a wet Device connector into the receptacle,
as this may damage the RF generator).
Temperature Set Point is set
too high
Check Temperature Set Point.
Unable to activate
RF Power
Temperature and/or
Impedance parameter
outside Functional Limits
Check Temperature and Impedance parameter
values;adjustasnecessarytobringparameters
into Functional range.
Electromagnetic Interference (EMI)
The RF Generator might cause interference that can affect other equipment. Such interference might
occur during RF Treatment Mode or during any mode that performs low power measurements, such as
Measure, Body Test, or Saline Test Modes.
If interference affects other equipment in the treatment area, move the RF Generator to a different
location and move the RF Generator cord and the Device away from the susceptible equipment and its
cords and cables.
The RF Generator is susceptible to interference from EMI emitted by other equipment. This could result in
inaccurate RF Power delivery and possible injury to the patient. Additionally, RF Treatment might halt or
the unit might restart due to a non-repeating event, such as memory corruption related to electrical or
radiation events, power surges, or power spikes.
Possible sources of EMI interference can include, but are not limited to, cellular phones, radio
transmitters, motors, telephones, lamps and other medical equipment, such as electrosurgical products
and defibrillators. Restrict the use of this equipment in the vicinity of the RF Generator.

32
Français
Operator’s Manual
31
Français
VNUS Radio Frequency Generator
Glossary of Terms
Advisory Limits – Limits on parameters, such as Temperature, Impedance, RF Power, Time, or a
combination of these, outside of which RF Treatment is not optimal and corrective action is suggested.
Alarm Tone–TheAlarmTonesoundsiftheunitdetectsthataparameterhasexceededtheFunctional
Limits for the connected Device.
Body Test – Performs Temperature and Impedance measurements using the Device positioned in the
patient for RF Treatment.
Device – An approved VNUS RF delivery or RF powered accessory (e.g., catheter, that can be attached to
the RF Generator).
Default – A value for a setting that is assigned automatically by the RF Generator. Default values remain
in effect unless reconfigured in the RFG Setup or Device Setup menu.
EMI – Electromagnetic Interference.
Error Log – Description of, and RF delivery data surrounding, an error condition. This data is stored for
later retrieval.
Functional Limits – Limits on parameters, such as Temperature, Impedance, RF Power, Time, or a
combination of these, outside of which RF delivery halts immediately.
Impedance – Ratio of instantaneous voltage divided by instantaneous current. Both ClosureFAST and
ClosureRFSusemagnitudeimpedance(themagnitudeofthecomplexImpedancevector).
Informational Tone – The Informational Tone sounds if the unit detects that a parameter is between
the Advisory and Functional Limits for the connected Device.
LCD – Liquid-Crystal Display.
MMC – MultiMediaCard.
Parameter – The parameters that govern the RF Treatment algorithm—Temperature, Impedance, RF
Power, and/or Time, depending on the Device.
PC – IBM®-compatible Personal Computer.
Persistent Settings – Operator-adjustable settings that retain their value when the RF Generator is
turned off, such as screen brightness and tone volume settings.
POST – Power On Self Test.
Power – Real Power [heating power = I
RMS
* V
RMS
* cosθ] delivered to a Device.
Power Limit–MaximumRFPowerthattheDevicecandeliver.
Power Measurement – Current RF Power being delivered.
Power Set Point – The desired RF Power.
RAM – Random Access Memory.
Recoverable Error – An error condition that requires the operator to respond to an error message or
make a clinical decision.
ROM – Read-Only Memory.
Real Impedance–MagnitudeoftherealportionofthecomplexImpedancevector.
RF – Radiofrequency energy.
RFC – Radiofrequency Computer.
RFG – RF Generator.
RFG2 – Model number of VNUS RF Generator.
RF Treatment – Therapeutic RF delivery.
Saline Test – Performs Temperature and Impedance measurements using saline solution.
Setting – Operator-configurable controls listed within the RFG Setup and Device Setup menus.
Temperature – Temperature, as measured at the Device point of RF delivery.
Temperature Measurement – The current Temperature.
Temperature Set Point – The desired or target Temperature.
Therapeutic RF–RFPowerlevelsdeliveredbyaDevicethatexceed10mWaveragepower.Thistermis
defined to distinguish between RF Power levels used for RF Treatment and lower RF Power levels used
exclusivelyforparametermeasurement.
Unrecoverable Error – An error condition that halts the RF Generator and requires the operator to reset
the RF Generator’s AC Power to continue.
VNUS –VNUS®MedicalTechnologies,Inc.

RM 55-661-01 REV.C
5799 Fontanoso Way
San Jose, CA 95138 USA
Tel: +1-408-360-7200
Toll Free: +1–888–797–VEIN (8346)
Fax:+1-408-365-8470
Covidien Ireland Limited,
IDA Business & Technology Park,Tullamore
0123
Customer Service
Tel: +1-800-962-9888
MEDICAL EQUIPMENT MODEL RFG2
WITHRESPECTTOELECTRICALSHOCK,
FIREANDMECHANICALHAZARDSONLY
INACCORDANCEWITHUL60601-1
CAN/CSA C22.2 NO.601.1
75AK
®
Other manuals for VNUS RFG2
1
This manual suits for next models
1
Table of contents
Popular Inverter manuals by other brands

Larson Electronics
Larson Electronics HAL-SPK-2X150PNL-SCS-2X100AH-24-PLC instruction manual

Easylife
Easylife EL6058 instruction manual

Eliminator
Eliminator 011-1945-0 instruction manual

Omega
Omega OTEC IEWQ Series Service manual

Ewent
Ewent EW3991 manual

Briggs & Stratton
Briggs & Stratton 7000 WATT 206484GS Installation and operator's manual

Nidec
Nidec Leroy Somer Powerdrive MD2SL Series Installation and Maintenance

SolarEdge
SolarEdge Home Battery 48V installation guide

enphase
enphase AC Combiner Box Quick install guide

REFU
REFU REFUsol 08K Installation and operating manual

Schwarzbeck
Schwarzbeck SG 9303 manual

IBC SOLAR
IBC SOLAR ServeMaster 1600MV user manual