Creative PC-60NW-1 User manual

User Manual for Fingertip Oximeter
I
Instructions to User
Dear Users,
Thank you very much for purchasing our product. Please
read the manual very carefully before using this device.
Failure to follow these instructions can cause measuring
abnormality or damage to the oximeter.
The manual is published in English and we have the
ultimate right to explain the Manual. No part of this
manual may be photocopied, reproduced or translated into
another language without the prior written consent. We
reserve the right to improve and amend it at any time
without prior notice.
Version of the Manual: Ver 1.0
Revised date: October 24, 2012
All rights reserved.
3502-2290052

User Manual for Fingertip Oximeter
II
Warnings
Check the device before using it to make sure
that there is no visible damage which may affect
user’s safety or its performance with regard to
sensors and clips. It is recommended that the
device should be inspected minimally once a
week. When there is obvious damage, stop using
the device.
Special attention should be paid while the
oximeter is used constantly under the ambient
temperature over 37°C, burning hurt may occur
because of over-heating of the sensor at this
situation.
An uncomfortable or painful feeling may appear
if the oximeter is used continuously on the same
place for a long time, especially for poor
microcirculation patients.
It is recommended that the oximeter should not
be applied to the same location for longer than 2
hours. If any abnormal condition is found, please
change the placing site of oximeter.
Avoid placing the device on the same limb which
is wrapped with a cuff for blood pressure
measurement or during venous infusion.

User Manual for Fingertip Oximeter
III
DO NOT clip this device on edema or tender
tissue.
The light (the infrared light is invisible) emitting
from the device is harmful to the eyes, so service
technician or testee should not stare at the light.
The oximeter is not a treatment device.
When disposing of this device and its
accessories, the local law should be followed.
Instructions for Operation
The finger should be put in properly and correctly.
Do not shake the finger. Keep at ease during
measurement.
Do not put wet finger directly into sensor.
Do not let anything block the emitting light from
the device.
Ensure that there is artery vessel within
measuring site where the light transmits through.
Vigorous exercise and the interference from the
electrosurgical device may affect the measuring
accuracy.
Nail polish may affect the measuring accuracy,
and too long fingernail may cause failure of
measurement or inaccurate result.

User Manual for Fingertip Oximeter
IV
Keep the oximeter away from dust, vibration,
corrosive substances, explosive materials, high
temperature and moisture.
Existence of high intensive light sources, such as
fluorescence light, ruby lamb, infrared heater or
strong sunshine, etc. may cause inaccuracy of
measurement result. Please put an opaque cover
on the sensor or change the measuring site if
necessary.
If the first reading appears with poor waveform
(irregular or not smooth), then the reading is
unlikely true, the more stable value is expected
by waiting for a while, or a restart is needed when
necessary.

User Manual for Fingertip Oximeter
V
Declaration of Conformity:
The manufacturer hereby declares that this
device complies with the following standards:
IEC 60601-1
IEC 60601-1-2
ISO 9919
and follows the provisions of the council
directive
MDD 93/42 EEC
Caution:
U.S. federal law restricts this device to sale or
use by or on the order of a physician.

User Manual for Fingertip Oximeter
VI
Content
1 Overview.....................................................................1
1.1 Appearance...........................................................1
1.2 Name and Model ..................................................2
1.3 Intended Use.........................................................2
1.4 Structure and Conformation ...............................2
1.5 Features.................................................................2
2 Battery Installation.....................................................3
3 Operation....................................................................4
3.1 Directly Measurement..........................................4
3.2 Over-limit indication and Beep Silence...............8
3.3 Setup Menu Screen...............................................9
3.4 Data transmission...............................................10
4. Technical Specifications .......................................... 11
5. Accessories ...............................................................13
6. Repair and Maintenance.........................................13
6.1 Oximeter Maintenance....................................... 13

User Manual for Fingertip Oximeter
VII
6.2 Cleaning and Disinfecting Instruction..............15
7. Troubleshooting.......................................................16
Appendix......................................................................17
A. Key of Symbols.....................................................17
B. SpO2Common Knowledge.................................... 18

User Manual for Fingertip Oximeter
1
1 Overview
1.1 Appearance
Figure 1 Front/Rear View

User Manual for Fingertip Oximeter
2
1.2 Name and Model
Name: Fingertip Oximeter
Model: PC-60NW
1.3 Intended Use
This Fingertip Oximeter is intended for measuring the
pulse rate and functional oxygen saturation (SpO2)
through patient’s finger. It is applicable for spot-checking
SpO2and pulse rate of adult and pediatric patients in
homes and clinics.
1.4 Structure and Conformation
It consists of main unit and photoelectric sensor.
1.5 Features
• Wireless data transmission can communicate with
PC/mobile phone/PDA.
• Large true color OLED display of SpO2, PR Pulse
Bar, PI & Plethysmogram
• Innovative 4 directions display
• Automatic power on/off
• Audible & visible over-limit indication, high & low
limits are adjustable
• Shift parameter display between PR and PI

User Manual for Fingertip Oximeter
3
• 2AAA alkaline batteries with low power consumption
• Low battery voltage indication
2 Battery Installation
Figure 2 Battery Installation
1. Refer to Figure 2, insert two AAA size batteries into
the battery compartment properly.
2. Replace the cover.
Please make sure that the batteries are correctly
installed since the incorrect installation may
cause the device inoperable.

User Manual for Fingertip Oximeter
4
3 Operation
3.1 Directly Measurement
1. Open the clip as shown in Figure 3.
Figure 3 Put finger into the Oximeter
2. Put finger into the rubber cushions of the clip (make
sure the finger is in the correct position), and then
clip the finger.
3. The device will power on automatically in 2 seconds,
and start to display software version number.
4. Next enter into data display screen (as shown in
Figure 4). The user can read the values and view the
waveform from display screen.
Shift parameter display between PR and PI during
measurement:
Short time press Display Key, the display orientation will
rotate 90 degree, at the same time, PR value shift to PI
value automatically and there is no waveform displayed as
shown in 4B. But if no operation is performed during the
first 20 seconds, the PI value will change to PR value

User Manual for Fingertip Oximeter
5
automatically as shown in figure 4C. Press the key again,
the screen will change its direction as shown in 4D. It
obviously shows that the alternative display between
parameter PR and PI is under the condition that the
direction of screen is rotated to the left and right side,
which is accompanied by no waveform displayed as
shown in 4B and 4E. After 20 seconds without operation,
the PI will automatically shift back to PR, as shown in 4C
and 4F.
Figure4A Figure4B
Figure4C Figure4D
Figure4E Figure4F

User Manual for Fingertip Oximeter
6
Screen Description:
“%SpO2”: SpO2symbol; “99”: SpO2 value;
“PR”: Pulse Rate icon; “65”: Pulse Rate value;
“”: Pulse beat symbol;
“”: Pulse intensity bar graph;
“PI%”: Perfusion index icon; “1.4”: Perfusion
Index value;
“”: Wireless symbol.
Wireless icon“ ”:
The color of “ ”
Definition
“ ” flashes blue
The device is being to establish
a wireless connection with the
surrounding host.
“ ” long lights blue
Successful wireless connection
between the device and a host is
established.

User Manual for Fingertip Oximeter
7
No display“ ”icon 1.
"Wireless" function is
disabled;
2.
The device fails to setup a
wireless connect
ion with the
surrounding host
within 3
minutes;
3. Hardware failure of wireless
transmission function while the
“Wireless” function is enabled.

User Manual for Fingertip Oximeter
8
3.2 Over-limit indication and Beep Silence
When measuring, if SpO2value or pulse rate value
exceeds the preset alarm limit, the device will beep
automatically and the value which exceeds limit on the
screen will flash. The detailed information refers to
chapter 4.
When the beep sound is activated by over-limit, it will
become silent at the following situations:
1. The SpO2and PR value return to normal range.
2. Press Display Key to mute. If this over-limit
event continues, the oximeter will resume beeping
automatically later in 2 minutes.
3. Remove the finger from the oximeter or SpO2
probe.

User Manual for Fingertip Oximeter
9
3.3 Setup Menu Screen
Figure 5 Menu Screen
Long time pressing display key could enter the setup
menu screen.
Setup Menu Screen Description:
“Wireless”
:
The wireless on-off button. There are 2
selectable options: “on” and “off”. The device transmits
data to PC when it is on.
“SpO2alm Lo”:SpO2low limit. The user can modify
the value within 85~99 with “1” as a step.
“PR alm Hi”
:
Pulse Rate high limit. The user can
modify the value within 100~240 with “5” as a step.
“PR alm Lo”
:
Pulse Rate low limit. The user can

User Manual for Fingertip Oximeter
10
modify the value within 30~60 with “1” as a step.
“Save, exit menu”
:
Longtime pressing this item to
store settings and exit to the normal display screen
from the setup menu.
On setup menu screen:
1. Short time press Display Key to choose the
setting item;
2. Longtime press Display Key to activate the
setting item, then short time press it to modify the
setting parameter;
3. Next, longtime press Display Key to confirm the
modification and exit from this setting item.
4. At last, move the setting item to “Save, exit
menu”, and long time pressing Display Key to
store the modification and exit from the setup
menu.
3.4 Data transmission
This oximeter has the feature of wireless data
transmission,the display reading can be transmitted to
a computer effectively via the wireless communication
function. Refer to the manual “Oximeter Data
Manager” for detailed information.

User Manual for Fingertip Oximeter
11
4. Technical Specifications
A. SpO2measurement
Technique:
dual-wavelength LED sensor with wavelength:
Red light: 663 nm, Infrared light: 890 nm.
Maximal optical output power: <= 1.5mW
Measuring range: 35%~100%
Measuring accuracy:
Arms <= 3% for SpO2range from 70% to 100%
Note: Arms is defined as root-mean-square value of
deviation according to ISO 9919.
SpO2low limit setting range: 85%~99%
B. Pulse Rate measurement
Measuring range: 30bpm~240bpm
Measuring accuracy: ±2bpm or ±2% (whichever is
greater)
Pulse Rate limit setting range:
high limit: 100~240bpm
low limit: 30~60bpm
C. Perfusion Index (PI) Display
Range: 0.2%~20%

User Manual for Fingertip Oximeter
12
D.Audible &visual over-limit indication
When measuring, if SpO2value or pulse rate value
exceeds the preset limit, the device will beep
automatically and the value which exceeds limit
on the screen will flash. The oximeter will shut
down automatically in 8 seconds with no signal.
E. Display mode
:
Color OLED Display
F. Power supply requirement:
2 x LR03 (AAA) alkaline batteries
Supply voltage: 3.0VDC
Operating current: ≤40mA
G. Environment requirement
Operating Temperature: 5 ~40°C
Operating Humidity: 30~80%
Atmospheric pressure: 70~106kPa
H. The performance under low perfusion condition
The accuracy of SpO2and PR measurement still
meets the specification described above when the
pulse modulation amplitude is as low as 0.6%.
I. Resistance to interference of surrounding light:
The difference between the SpO2 value measured
in the condition of indoor natural light and that of
darkroom is less than ±1%.

User Manual for Fingertip Oximeter
13
J. Dimensions: 60 mm (L) × 33 mm (W) × 30 mm (H)
Net Weight: 35g (including battery)
K. Classification:
The type of protection against electric shock:
Internally powered equipment.
The degree of protection against electric shock:
Type BF applied parts.
The degree of protection against harmful ingress
of liquids: Ordinary equipment without protection
against ingress of water.
Electro-Magnetic Compatibility: Group I, Class B
5. Accessories
A. A lanyard
B. Two batteries
C. Apouch
D. A User Manual
E. Quality Certificate
Note: The accessories are subject to change. Detailed
items and quantity see the Packing List.
6. Repair and Maintenance
6.1 Oximeter Maintenance
Table of contents
Other Creative Medical Equipment manuals

Creative
Creative PC 66B User manual

Creative
Creative OXY-11 User manual

Creative
Creative PC-60NW User manual

Creative
Creative PC 66B Instruction Manual

Creative
Creative POD-1 User manual

Creative
Creative OXY-110 Instruction Manual

Creative
Creative PC-68B User manual

Creative
Creative PC-68A Instruction Manual

Creative
Creative PC-66A User manual

Creative
Creative PC-60N User manual
Popular Medical Equipment manuals by other brands

ZONARE
ZONARE Z.one Ultra SP Service manual

Zeiss
Zeiss ATLAS 9000 Service manual

EGi
EGi Geodesic EEG Systems quick start guide

Ohmeda
Ohmeda BiliBlanket Plus Operation, maintenance and service manual

Wellion
Wellion MICRO-PUMP manual

Flight Medical Innovations
Flight Medical Innovations Flight 60 Ventilator Operator's manual