truFreeze® Passive Venting Set-up Instructions
CSA Medical, Inc.
For service, call: 866.481.7786
www.csamedical.com
© 2014 CSA Medical, Inc. All rights reserved.
truFreeze is a registered trademark of CSA Medical, Inc.
15-00293-00
11. If using the remote, ensure the remote control
has two (2) working AA batteries installed by
selecting the run tab and attempting to adjust
the site counter. Reset site to “1”.
10. Locate the sterile truFreeze catheters.
Make sure the catheters have not passed their expiration dates.
8. Select the “run” screen.
9
9. Confirm that the device has sufficient
liquid nitrogen via the indicator on screen
(in the setup tab).
If the tank indicator displays orange or red,
the device must be filled prior to use.
Filling is recommended but not required if the
indicator is yellow.
P See separate tank filling instructions.
7. If the BIT fails, a red “x” and “fail – complete”
will display. Press the test button at the home
screen and let the system repeat the BIT.
If the system fails BIT two (2) times with the
same error, please call service at 866.481.7786
Please read the Instructions For Use and Operator’s Manual for complete
information about the truFreeze System.