Cordless floor mat is placed on floor in
doorway or beside the bed.
Alarm signals sent
to Monitor
QUICK START GUIDE
2700 - 6 Channel Cordless Falls Monitor
3
2
2719CordLess
PadTransmitter Insert
Status/ In Use
Connect 2719 Transmitter to
Cordless Pad or Mat. Ensure
that it is switched OFF.
Test falls monitor prior to every use.
1
PROGRAM
Melody
123
Volume
Decrease / Increase
Tone
LO HI Scale
Back View
Insert 3 X AA alkaline batteries
into battery compartment.
Step
Insert 3 x Alkaline AA Batteries.
PROGRAM
Use screwdriver, (accepts both
Phillips or flat head) to undo
screw & slide battery cover UP
to open.
Step
Pair with cordless Pad/s or floor Mat.
Step
Place pad on mattress or position mat on floor & test prior to using.
Cordless bed
pad is placed
on top of
the mattress,
underneath
bedding.
Mattress
Bed Sheets
Bed Pad
Bottom
Sheet
1. Place up to six cordless pads or mats within reach. Each
pad or mat must have a 2719 Cordless Pad Transmitter
Insert connected & turned OFF.
3. The display will briefly display “PP” and then “01”,
where “0” indicates the memory slot is open and “1” is the
device number from 1 to 6. Turn ON the first Pad
Transmitter.
4. The Monitor will beep 3 times and the display will briefly
change to “F1” where “F” indicates that memory slot “1” is
now full. The display will change to the next available
empty slot e.g. “02”.
5. Turn ON the next device to add additional devices to the
monitor.
6. To cycle through the memory slots press the Pause
button. To clear a memory slot press and hold the Pause
button so that the “F” changes back to a “0” The unit will
beep 3 times when the slot is cleared.
2. Press & hold the Program button on back of the Monitor
until you hear 3 beeps.
7. Press the Program button or press and hold the Reset
button to exit programming mode. If “t0” or “S0” is
displayed, continue pressing the Reset button until the
memory slots are displayed.
The 2700 6 Channel Cordless Falls
Monitor is designed to work in a facility on
alkaline batteries equivalent to those
originally supplied with the device.
Operable alkaline batteries must be
installed in the device at all times during
operation. Batteries must be removed from
the device while it’s in storage.
Use of an optional 2406 Plug Pack will
extend the usable battery life. Essentially a
Plug Pack may not be required in most
instances.