Once the baseline pump timin
g is established, the humidity level of the
cabinet should be monitored on a daily basis using the humidity indicator cards available from
Follow instructions on the humidity indicator card for proper
itoring of the desiccant bags
When the indicator card shows that the desiccant bags need to be changed, r
desiccant from their holding compartment
Follow instructions on the desiccant bag for the re
activation procedure or contact
Custom Ultrasonics, Inc. at 215
1477 for further information
To maintain the function of the storage cabinet, an extra set of desiccant bags should be held in
reserve in a sealed container. The storage cabinet can t
hen be operated continuously.
filter should be replaced once a year Part # 20010.
The air pump HEPA post filter
If air is heard being released through the “relief valve” when the endoscopes are connected to the
order desiccant bags, humid
ity indicator cards, filters and repair parts, please visit our website
www.customultrasonics.com
. In the left column click on “order replacements online” then click
on “Storage Cabinet & Accessories”.
Parts may also be ordered by fax at (215) 364
Endoscope Storage Cabinet
minimum of 2 persons to support and mount the cabinet
physical damage to the equipment and surrounding area may occur.
Make certain that when bolting the cabinet to the wall:
Cabinet is located flush to desired wall
he provided mounting holes are used
ounting materials (bolts, etc.) suitable to facility’s
structure and all building