pH Measurement –
Hang the pH electrode in the water or nutrient solution in such a manner that it is out of
harms way and the currents in the tank will ow by. The monitor will constantly display the pH
value of the solution.
When not in use, store the electrode in this manner. Clean the electrode with Rejuve it! elec-
trode cleaner, rinse with distilled water and blot dry with a clean cloth. Fill the plastic storage
bottle with Store it! electrode storage solution. Insert the electrode through the storage bot-
tle’s cap and tighten.
Occasionally clean the electrode with Rejuve It! electrode cleaner. This will rejuvenate the
glass bulb by safely cleaning build-up from the glass.
pH Calibration –
If this is a new meter, skip step 1. Allow the electrode’s glass bulb to sit in CWP’s brand of
electrode cleaner called Rejuve it! for 15 minutes. Using a soft toothbrush lightly scrub the
glass bulb. Gently rinse with distilled water and blot the bulb dry. Be careful as the glass bulb
is fragile. This procedure will help remove contaminates from the electrode’s surface so that a
true and proper calibration can be done.
Dip the pH electrode into a pH 7 buffer solution. Wait for the meters display to stabilize. Gently
turn the pH 7 calibration screw until the display reads pH 7.00. Do not turn calibration screw
past its stop.
Rinse the electrode with distilled water and blot it dry with a clean cloth.
Dip the electrode into a pH 4 buffer solution. Gently turn the pH 4 calibration screw until the
display reads pH 4.00. Do not turn calibration screw past its stop.
If necessary, repeat procedure 1 to 4 until the display shows the correct values. If calibration
is still not possible see the troubleshooting section. Rinse and blot the electrode dry each time
before inserting the electrode in a buffer solution.
Troubleshooting –
If the meter fails to calibrate pH properly after using the Rejuve it! cleaner on the electrode, the
pH calibration screws may need to be repositioned to center travel position before calibrating again.
To do this, gently turn each pH calibration screw counterclockwise until it stops and then back clock-
wise until it stops. This will show you the full travel of the calibration screw. Now gently turn each
calibration screw to the center or to its halfway position and start the calibration steps 2-4 again. If
this does not allow for enough calibration screw travel during calibration, gently turn each calibra-
tion screw (from full clockwise position) counterclockwise to the 1/4 position and calibrate the meter
again using steps 2-4. The starting position of the calibration screws is very important.
Note: If the meter fails to calibrate pH properly the pH electrode’s life may be over and will need re-
placed. The item number for this part is CW505.
•
•
•
1.
2.
3.
4.
5.