manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Dostmann Electronic
  6. •
  7. Measuring Instrument
  8. •
  9. Dostmann Electronic P795 User manual

Dostmann Electronic P795 User manual

DOSTMANN electronic
DOSTMANN electronic GmbH
SERIEP700
P795
Precision 2-channel hand-held measuring instrument
Operation Manual
1.4.4 Channel activation (only 2-channel instruments) / [Chnl]
Chnl = Channel
To activate or deactivate a measuring channel push [ENTER/
MENUE]. Use the up and down keys
[

]
to select Chnl. Push
again [ENTER/MENUE]. On the left corner of the display appears a
small 1, which indicates the selected channel. To change the chan-
nel use the up and down keys
[

]
. Push [ENTER/MENUE] to con-
firm. Use the up and down keys
[

]
again to activate on or deacti-
vate off the requested measuring channel and push [ENTER/MENUE]
to confirm. Push [ESC] to revert back to the measuring mode.
Note: As a minimum one channel is active!
1.4.5 Memory Setup Lo6
Push [ENTER/MENUE] and use the up and down keys
[

]
to select
Lo6. Push [ENTER/MENUE] to confirm. Use the up and down keys
[

]
again to start [on] or Stop [off] the logger. Push [ENTER/
MENUE] to confirm. Use the up and down keys
[

]
again to select
between automatic storage [Auto] or manually operated storage
[SPot]. Push [ENTER/MENUE] to confirm]. Use the up and down keys
[

]
again to select between to add on data [Add] and creating a
new file [nLo6]. Push [ENTER/MENUE] to confirm. By selected auto-
matic storage at the end you have to select the time interval:
1 S 1 second
5 S 5 seconds
10 S 10 seconds
20 S 20 seconds
30 S 30 seconds
1 M 1 minute
2 M 2 minutes
5 M 5 minutes
10 M 10 minutes
20 M 20 minutes
By selected manually operated storage you are able to save the
measurement by pressing ESC by each time.
Push [ENTER/MENUE] to confirm. Push [ESC] to revert back to the
measuring mode.
E.g. of the instrument’s LCD by activated logging mode:
On the bottom you can see the percentage of the
occupied memory (0..99%). If a calibration option
is activated the display is alternating between
displaying memory status and calibration infor-
mation.
1.5 Recalling the memory data (HOLD MAX MIN AVE)
After pushing first time the key [HOLD MAX MIN AVE] the actual val-
ues will be held on the display. Pushing again the key [HOLD MAX
MIN AVE], the saved maximum-, minimum and average values will be
displayed.
Note: During the recall of the memory data the extremes (MAX MIN)
and the average value (AVE) will not be calculated or carried on.
Clearing the memory (MAX MIN AVE)
Press [CLEAR] key once to erase the stored maximum, minimum and
average from memory. On the display appears Clr. – After erasing the
memory the instrument automatically reverts back to measuring mode
indicating the actual measured value again.
1.6 Measuring rate
(Normal-Mode / FAST-Mode / Filter-Mode)
The instrument has three different response times to select:
Normal-Mode: high resolution
(0.001 from -199.999 to +199.999)
FAST-Mode: reduced resolution
(0.01 from -199.99 to +199.99)
Filter-Mode: high resolution
(0.001 from -199.999 to +199.999)
Press [FAST/

] key to change the measuring rate.
By using the Filter-Mode the instrument is performing a moving aver-
age to stabilize the dispayed measurements. After turning on the
instrument it is in the Normal-Mode. After pressing once the [FAST/

]
key the instrument switches to the Fast-Mode. Pressing the same key
once again the instrument switches to the Filter-Mode.
E.g. Display with activated Filter-Mode:
The arrow on the top indicates that the instru-
ment is working in the Filter-Mode.
Note: After switching off the instrument, this function is automatically
deactivated.
1.7 AUTO-OFF-function
dAoF = Disable Auto-off
EAoF = Enable Auto-off
Press [ESC/AUTO-OFF] key once. On the display appears EAoF.
Now the instrument switches off automatically after app. 30 minutes.
Press [ESC/AUTO-OFF] key again. On the display appears dAoF.
Now the Auto-Off-function is deactivated.
Attention: This setting will be deactivated when the instrument turns
off. (Standard setting is EAoF).
1.8 TARE-Function (ZERO-Mode F1)
The Instrument has a special Tare/Zero-button. Pressing the [F1] key
once, the instrument will subtract the last measured value from the
actual measured value. So if the measurement does not change
zero appears on the instrument’s display. If you press the [F1] key
once again the instrument move to the Normal-Mode.
E.g. Display with activated Zero-Mode:
The arrow on the left indicates that the instru-
ment is working in the Zero-Mode [F1].
Note:
After switching off the instrument, this function is automatically
deactivated.
2. Power supply
•For the power supply of the instrument a 9 Volt dry battery is used.
To exchange the battery switch of the instrument and open the
rear battery cover. Remove the battery from the instrument and
replace with a new battery.
•The “BAT” symbol in the display indicates that the battery needs
to be exchanged. After displaying the “BAT” symbol, the instru-
ment allows app. 1 hour of further operation.
•The battery symbol indicates according to the battery status
between 1 to 3 segments.
Attention:
Please do not dispose of old electronic devices and empty batteries
in household waste. To protect the environment, take them to your
retail store or to appropriate collection sites according to national
or local regulations.
3. Error Codes
By displaying the following error codes the instrument support the
operation of the instrument:
Error Meaning
oPEn no probe
or wrong probe is connected
|
-oLo „too low“
underflow of the measuring range
|
-oHi „too high“
exceeding of the measuring range
|
-Er1 temperature of the cold junction
above the measuring range
|
-Er2 temperature of the cold junction
above the measuring range
E15 low battery
(battery is completely exhausted)
E19 low battery
(battery is completely exhausted)
E1d low battery
(battery is completely exhausted)
E16 time out / auto-off-function
E1o time out / auto-off-function
E1e time out / auto-off-function
E23 EE-prom destroyed or
EE-prom index is wrong
E25 EE-prom destroyed or
EE-prom index is wrong
E31 temperature of cold junction
out of range
LoLi internal buffer battery (CR2032) empty
Unit L in2 CAL Chnl Lo6

 MENUE
Unit L in2 CAL Chnl Lo6

 MENUE
DE_P795_Anleitung_engl_10_12 09.10.2012 9:57 Uhr Seite 1
DOSTMANN electronic
DOSTMANN electronic GmbH
Summary
1. Handling
1.1. General advices
1.2. Setting to work
1.3. Switch on/off
1.4. Menu
1.4.1. Measuring unit switching (Unit)
1.4.2. Difference temperature (L in2)
1.4.3. Calibration (CAL)
1.4.4. Channel activation (Chnl)
1.4.5. Memory setup (Lo6)
1.5. Recalling memory data (HOLD/MAX/MIN/AVG)
1.6. Measuring rate (FAST-mode)
1.7. AUTO-OFF-function
1.8. TARE-Function (ZERO-Mode F1)
2. Power supply / Changing the battery
3. Error codes / troubleshooting
1. Handling
1.1 General advices
•For cleaning the instrument please do not use abrasive cleaner but
a dry or wet piece of cloth.
•Please store the measuring instrument in a dry and clean place.
•Avoid any force like shocks or pressure to the instrument.
•Do not use force to connect the probe or interface plugs in. The
interface plug is different from the probe plug.
•If no sensor is connected to the instrument while switching on
“open” shows on the display (Please refer to chapter error codes /
troubleshooting).
•A retractable stand on the back of the instrument allows it to be
used as a bench top instrument.
1.2 Operation
Before switching on the instrument, connect the probe/s to the instru-
ment and insert the battery (Please refer to chapter 2. Power
supply/changing the battery). A number on the instrument’s housing
marks each port.
1.3 Switching on and off
By operating the ON/OFF-key the instrument switched on or off. After
switching on the instrument indicates a full segment test for 15
seconds, then it starts to function in measurement mode indicating the
actual measurement value.
1.4 Menu
The adjustments of the instruments function: measurement value,
calibration of probes, deactivation of channels, for example, are
selected from the menu structure. Enter into the main menu by push-
ing [ENTER/ MENUE]. Use the up and down keys [

] to select the
required menu. Push [ESC] to revert back to the measuring mode.
Menu structure
1.4.1 Measuring unit switching °C, °F and Resistance (Ohm’s) [Unit]
Unit = Measuring unit
Measuring unit temperature (°C=Celsius, °F=Fahrenheit; o=Ohm)
To change the measuring unit push [ENTER/MENUE]. Use the up and
down keys
[

]
to select Unit. Push again [ENTER/MENUE]. On the
right corner of the display appears °C, °F or o. Use the up and down
keys
[

]
again to adjust the requested measuring unit and push
[ENTER/MENUE] to confirm. Push [ESC] to be back in the measuring
mode.
1.4.2 Temperature difference (only 2-channel instruments) / [L in2]
To display the temperature difference push [ENTER/MENUE]. Use
the up and down keys
[

]
to select L in2. Push again
[ENTER/MENUE]. On the right corner of the display appears a T1-T2.
Use the up and down keys
[

]
to adjust the requested selection.
Push [ENTER/MENUE] to confirm. Push [ESC] to revert back to the
measuring mode.
Note: Both channels have to be activated for displaying temperature
difference.
1.4.3 Calibration function / [CAL]
Despite high quality manufacturing techniques, each probe is slightly
different from specified standards. To eliminate inaccuracies
caused by exchanging or ageing of probes, the instrument offer
easy calibration functions which guarantee that the system accura-
cy is always as good as if the instrument was specifically calibrated
to the individual probes in our laboratory.
The instruments offer five different calibration options:
1) [CoFF]: Standard characteristic curve
(e.g. Pt100-resistance according EN 60751)
2) [oP 1]: Calibration by code
(2 x four digit code) is equivalent to a 2-point calibration.
The code is marked clearly by a label on each standard
probe.
3) [oP 2]: Calibration by physical standard references
(1-point, 2-point or 3-point calibration)
4) [oP 3]: Calibration according to Coefficients of EN60751 (R0, ABC)
5) [oP 4]: Smart EEprom probes with internal calibration
(AUTO-Detection)
CAL = calibration
Push [ENTER/MENUE] to calibrate the instrument with sensor. Use
the up and down keys
[

]
to select CAL. Push again [ENTER/
MENUE]. On the left corner of the display appears a small 1, which
indicates the selected channel. To change the channel (1or 2) use
the up and down keys
[

]
. Push [ENTER/MENUE] to confirm.
Use the up and down keys
[

]
to select the requested calibration
option. Push [ENTER/MENUE] to confirm.
1) Standard calibration according to DIN EN 60751 / [CoFF]
Use the up and down keys
[

]
to select [CoFF
]
. Push
[ENTER/MENUE] to confirm. Push [ESC] to revert back to the
measuring mode.
2) Calibrationby code / oP1
Use the up and down keys
[

]
to select oP 1. Push [ENTER/
MENUE] to confirm. On the bottom of the display appears a small
1, after this number a four-digit number (Hex-Code/0..F) is dis-
played. For changing the number use the up key
[

]
. To step to
the next number use the down key
[

]
.
If the requested number is complete then push [ENTER/MENUE] to
confirm. At the bottom of the display a very small 2appears, after
which a second four-digit number is displayed. For changing the
number please follow the manual as before. Push [ESC] to revert
back to the measuring mode.
Note:
After confirming oP 1 by pushing [ENTER/MENUE] the function oP 1
(calibration by code) is activated, even though you leave the menu by
pressing [ESC].
Display-indication with active calibration code (oP 1):
The CAL-segment and the small 1indicates to the
user that oP 1 is activated.
3) Calibration by physical standard references / oP 2
Use the up and down keys
[

]
to select oP 2. Push [ENTER/
MENUE] to confirm. On the bottom of the display appears 1 P. For
changing between a 1-Point [1 P], 2-Point [2 P] or 3-Point [3 P] - cali-
bration use the up and down keys
[

]
.
Example of a 1-Point calibration:
Push [ENTER/MENUE] to confirm. On the display appears CALC.
After the displayed measuring value is stabile push [ENTER/
MENUE]. On the first display line you can see the “frozen” measure-
ment value. On the second line as a default you can see -100.000.
Now you have to enter (instead of -100.000) the correct measure-
ment value from your reference:
By using the up key
[

]
you are able to move the decimal point to
setup the number of decimal places. Push [ENTER/MENUE] to con-
firm.
Now the algebraic sign is blinking "-". Use the up key
[

]
to toggle
for positive or negative number. Change the number using the up
and down keys
[

]
:

is changing the blinking segment

is jumping to the next segment
Push [ENTER/MENUE] to confirm, revert back to the measuring
mode.
Important:
To abort the physical calibration use the [ESC]-button.
Display-indication with active calibration code (oP 2):
The CAL-segment and the small 2indicates to
the user that oP 2 is activated.
4) Calibration according to Coefficients of EN60751 (R0, ABC) / oP 3
By using Option 3 you are able to activate coefficients according
to EN60751 (R0, A, B, C). The coefficients have to be calculated
using special software on a PC (e. g. P7_CALC). Before you are
able to activate this option you have to transmit the coefficients
from the PC to the instrument. Therefore you have to use the Soft-
ware P7_CALC, too.
Use the up and down keys
[

]
to select oP 3. Push [ENTER/
MENUE] to confirm. Now the calibration option 3 is activated!
Push [ESC] to revert back to the measuring mode.
Note: After confirming oP 3 by pushing [ENTER/MENUE] the function
oP 3 (calibration by code) is activated, even though you leave the
menu by pressing [ESC].
Display-indication with active calibration code (oP 3):
The CAL-segment and the small 3indicates to
the user that oP 3 is activated.
5) Smart EEprom-probes with internal calibration
(AUTO-Detection) / oP 4
The Option 4 will be activated automatically by using Smart
EEprom probes. This option will be activated by switching on the
instrument when the Smart EEprom probe has been connected to
the instrument.
Display-indication with active calibration code (oP 4):
The CAL-segment and the small 4indicates to
the user that oP 4is activated.
Note:
If a Smart probe will be disconnected during the instrument is work-
ing the instrument automatically switch to the CAL-menu.
❶ Up and down keys
❷ [ESC] key
❸ [ENTER/MENUE] key
❶
❷
❸
Unit
°C
°F
O
L in2 CAL Chnl Lo6
T1-T2 CoFF
oP 1
oP 2
oP 3
oP 4
oFF
on
oFF
on
Unit L in2 CAL Chnl Lo6

 MENUE
Unit L in2 CAL Chnl Lo6

 MENUE
1 P 2 P 3 P

 oP 2
CoFF oP 1 oP 2 oP 3 oP 4

 CAL
Unit L in2 CAL Chnl Lo6

 MENUE

Mounting device
for probe handle
Big
LC-Display
Measuring port 2
Retractable stand
USB PC
interface port
Measuring port 1
Keyboard
DE_P795_Anleitung_engl_10_12 09.10.2012 9:57 Uhr Seite 2

Other Dostmann Electronic Measuring Instrument manuals

Dostmann Electronic P4000 Series User manual

Dostmann Electronic

Dostmann Electronic P4000 Series User manual

Dostmann Electronic P700-EX User manual

Dostmann Electronic

Dostmann Electronic P700-EX User manual

Dostmann Electronic HumidCheck Pro User manual

Dostmann Electronic

Dostmann Electronic HumidCheck Pro User manual

Dostmann Electronic GD 383 User manual

Dostmann Electronic

Dostmann Electronic GD 383 User manual

Dostmann Electronic MS 82 User manual

Dostmann Electronic

Dostmann Electronic MS 82 User manual

Dostmann Electronic OXY 7 User manual

Dostmann Electronic

Dostmann Electronic OXY 7 User manual

Dostmann Electronic HumidCheck Pro User manual

Dostmann Electronic

Dostmann Electronic HumidCheck Pro User manual

Dostmann Electronic P700-EX User manual

Dostmann Electronic

Dostmann Electronic P700-EX User manual

Dostmann Electronic TC 301 User manual

Dostmann Electronic

Dostmann Electronic TC 301 User manual

Dostmann Electronic P700 User manual

Dostmann Electronic

Dostmann Electronic P700 User manual

Dostmann Electronic Air CO2ntrol 3000 User manual

Dostmann Electronic

Dostmann Electronic Air CO2ntrol 3000 User manual

Dostmann Electronic P705 User manual

Dostmann Electronic

Dostmann Electronic P705 User manual

Dostmann Electronic P300 Series Instruction sheet

Dostmann Electronic

Dostmann Electronic P300 Series Instruction sheet

Dostmann Electronic P400 Series User manual

Dostmann Electronic

Dostmann Electronic P400 Series User manual

Dostmann Electronic Air Co2ntrol 5000 User manual

Dostmann Electronic

Dostmann Electronic Air Co2ntrol 5000 User manual

Dostmann Electronic DEC91 User manual

Dostmann Electronic

Dostmann Electronic DEC91 User manual

Dostmann Electronic HumidCheck User manual

Dostmann Electronic

Dostmann Electronic HumidCheck User manual

Dostmann Electronic 31.5008.02 User manual

Dostmann Electronic

Dostmann Electronic 31.5008.02 User manual

Dostmann Electronic VA893 User manual

Dostmann Electronic

Dostmann Electronic VA893 User manual

Dostmann Electronic PH CHECK 5040-0301 User manual

Dostmann Electronic

Dostmann Electronic PH CHECK 5040-0301 User manual

Dostmann Electronic MS 85 User manual

Dostmann Electronic

Dostmann Electronic MS 85 User manual

Dostmann Electronic 5020-0343 User manual

Dostmann Electronic

Dostmann Electronic 5020-0343 User manual

Dostmann Electronic MS 85 User manual

Dostmann Electronic

Dostmann Electronic MS 85 User manual

Dostmann Electronic MS 83 User manual

Dostmann Electronic

Dostmann Electronic MS 83 User manual

Popular Measuring Instrument manuals by other brands

Tractel Dynarope HF36/3/LPT Operation and maintenance manual

Tractel

Tractel Dynarope HF36/3/LPT Operation and maintenance manual

Integra Ri3 Installation and operating instructions

Integra

Integra Ri3 Installation and operating instructions

Fluidwell F103 WP manual

Fluidwell

Fluidwell F103 WP manual

KROHNE OPTISONIC 3400 quick start

KROHNE

KROHNE OPTISONIC 3400 quick start

VDO TU00-0752-5807102 installation instructions

VDO

VDO TU00-0752-5807102 installation instructions

Meriam MGF16BN instructions

Meriam

Meriam MGF16BN instructions

dynasonics UFX operating instructions

dynasonics

dynasonics UFX operating instructions

Anaheim Scientific S660 user manual

Anaheim Scientific

Anaheim Scientific S660 user manual

Tera TOM 600 manual

Tera

Tera TOM 600 manual

JPT YDFLP-E-80&100-M7-M-R user manual

JPT

JPT YDFLP-E-80&100-M7-M-R user manual

Kett PM-650 operating manual

Kett

Kett PM-650 operating manual

General MMD5NP user manual

General

General MMD5NP user manual

RTK ST 5112 Series Installation and operating instructions

RTK

RTK ST 5112 Series Installation and operating instructions

Lumel NA5PLUS user manual

Lumel

Lumel NA5PLUS user manual

RD Instruments WorkHorse user guide

RD Instruments

RD Instruments WorkHorse user guide

Hilti PR 2-HS operating instructions

Hilti

Hilti PR 2-HS operating instructions

RJG Lynx product manual

RJG

RJG Lynx product manual

BURWELL Elcometer 355 operating instructions

BURWELL

BURWELL Elcometer 355 operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.