Calibration of the Accurate pH 3 Meter
This pH meter has a 2-point calibration system for superior
accuracy. Both calibration screws are located inside the battery
compartment. Turn the cap counterclockwise for removal. Battery
compartment removal is necessary for calibration. Control Wizard
Products recommends its brand of pH 7 and pH 4 calibration and
storage solutions, as they are high quality products. Part # NT690,
NT700, NT760, NT770
1. Turn the meter ON, remove the battery cap and pull the battery compartment out of the body. Let
it hang to the side of the body. Place the meters electrode into clean distilled water. Gently stir
and then remove the meter from the water and blot the excess from the meter’s body and glass
electrode with a tissue.
2. Dip the meter’s electrode into a 7.0 pH calibration solution. If the display reads 7.0 proceed to
step 3. If the display does not show 7.0, adjust the calibration screw positioned at the top of the
battery compartment to make it the display read 7.0.
3. Rinse the electrode with distilled water, blot dry and place it into 4.0 pH calibration solution to
make sure the display reads 4.0. If the display does not show 4.0, adjust the calibration screw
positioned at the bottom of the battery compartment to make it the display read 4.0.
4. Rinse the electrode with distilled water, blot dry and place it into 7.0 pH calibration solution to
make sure the display reads 7.0. If the display does not show 7.0, adjust the calibration screw
positioned at the top of the battery compartment to make it the display read 7.0.
5. Calibration is complete when the display reads 7.0. Rinse the electrode with distilled water and
blot the meter with a tissue before making your test.
Battery replacement: Make sure that the cap is screwed on tight after replacing batteries/calibrating
or the meter will not be waterproof.
Note: Inaccurate readings with good batteries generally mean the electrode needs to be replaced.
Contact your local supply store for a replacement. Part Number CW504