Cypress Diagnostics CYANHemato User manual

Register now at: https://diagnostics.be/warranty
Rev.(0.0),202110 Page1
INSTALLATIONREPORTANDCHECKLIST
Foranalyzer CY006CYANHemato
Distributor
(Companyname)
Coordinates
(Country,city)
(TelephoneandEmail)
Maintenanceengineer
(Name)
(TelephoneandEmail)
Institution
(Laboratoryname)
Coordinates
(Country,city)
(TelephoneandEmail)
Labmanager
(Name)
(TelephoneandEmail)
Installationreport
Serialnumber
Deliverydate
Installationdate
Contentpackinglistverified □Yes□No
Installationchecklistcompleted
(seepage2) □Yes□No
Usertraininggiven □Yes□No
Warrantyactivated
(seepage2)
□Yes□No
Warrantyenddate:
Theinstallationreportofthelaboratoryequipmentisapprovedandsignedby:
Laboratory
manager
Maintenanceengineer
Firstandlastname:
Date:
Signature:
This form is a compulsory requirement for ISO certification, laboratory accreditation, good
laboratorypracticesandthelegalrequirementsinmanycountries.Fillitoutandbefutureready.
P.S.Keepacopy

Register now at: https://diagnostics.be/warranty

Register now at: https://diagnostics.be/warranty
Rev.(0.0),202110 Page2
Checklistforinstallation(detailedinstructionsinchapter3oftheusermanual)
Ensureoptimaloperation
□ Verifythelocationspecifications
□ Checktheelectronicalcurrentrequirements.Ifnecessary,use:
□ Electronicstabilizer
□ UPS
InstallationofaccessoriesontheCYANHematoanalyzer
□ Connectthe3reagents:diluent,lyseandcleanerandthewastetubetothewastebottle
□ Connectanyperipherals:externalprinter,keyboard,mouse
□ Installprintingpaper
□ Installthecorrecttubeadaptor
□ Connectthepowercordtoasuitedpowersupply
VerificationoftheCYANHematoanalyzer
□ Performaprimingcycleforallreagents
□ Runaselftesttochecktheinstrument’sperformance
□ Runblankmeasurementstoexcludebackgroundnoise(threeblankmeasurements)
□ Performacalibrationwithnormalcontrol(atleastthreemeasurements)
□ Performaqualitycontrolwithlow,normaland/orhighcontrolblood
Adaptationofthesystemtothecustomer
□ Configurelanguage,heading,measurementunits,normalranges,dateandtime
□ Createuserprofiles
□ Giveextensiveinformationondailyuseandmaintenancetotheuser
□ Registertheinstrumenttoactivateitswarranty
Warrantyactivation:
Link:https://diagnostics.be/warranty

Register now at: https://diagnostics.be/warranty

Register now at: https://diagnostics.be/warranty
Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium • www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be
Rev.(0.0),202110 Page1
MAINTENANCEINSTRUCTIONS
Analyzer CY006–CYANHemato
Serialnumber
TobeperformedbyPeriod Code Mustlog Activitydescription
LaboratoryTechnician Weekly D /
Cleanneedlewashinghead
Checktubingsystemforleakage
Cleanoutsideofinstrument
Periodically W Desinfectionandhardcleaningoftheneedle
MaintenanceEngineer
Upgradenotice UN Installthelatestsoftwareversion
6months
M1 / Fullselftestanddevicestatistics
M2 / Removedust/dirtfromtheinsideoftheinstrument
M3 Greasingofthedilutorblock(cogwheelmachinegreaseandsilicongrease)
M4 Hardcleaningoftheneedle
M5 Emergencycleaning
1year Y1 Replaceneedlewashinghead
Every23years Y2
Replacetubing: Reagenttubingkit
Pressuretubes(2x)
Internalhydraulictubes(10x)
Replacerubbers/Orings/beltsof: Microdilutor(3x)
Aperture
Referenceelectrode
Needlebelt
Every35years Y3
Replace: Needle
Microdilutor
Aperture
Countingchamber
Samplerotorbelt
Pumptubekit
Small(4x)andbig(1x)Tconnector
Steelreagentinletconnector(4x)
Problems P Performreparation

Register now at: https://diagnostics.be/warranty

Register now at: https://diagnostics.be/warranty
Cypress Diagnostics: Nijverheidsstraat 8 • 2235 Hulshout • Belgium • www.diagnostics.be • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be
Rev.(0.0),202110 Page_
MAINTENANCELOGBOOK
Date Code ActivityperformedPerformedby
Needmoreprints?
Link:https://www.diagnostics.be/Log

Register now at: https://diagnostics.be/warranty

Register now at: https://diagnostics.be/warranty
Layout 20181024
ISO 13485-2016
www.diagnostics.be • Belgium • Tel: ++ 32 15 67 67 68 • e-mail: cypress@diagnostics.be
CYANHemato
User Manual
For a clear and precise diagnose
CY006
English

Register now at: https://diagnostics.be/warranty

Register now at: https://diagnostics.be/warranty
CYANHematousermanual Rev.(7.0),202111 Page1
INDEX
1 INTRODUCTION......................................................................................................................4
1.1 INTENDEDUSE....................................................................................................................................4
1.2 MEASUREMENTPARAMETERS.............................................................................................................4
1.3 REAGENTS..........................................................................................................................................5
1.4 QUALITYCONTROLBLOOD................................................................................................................5
1.5 CALIBRATION.....................................................................................................................................5
1.6 BLOODSAMPLES................................................................................................................................5
1.6.1 Samplecollection..................................................................................................................................................5
1.6.2 Samplestorage.......................................................................................................................................................6
1.6.3 Samplepreparation..............................................................................................................................................7
1.6.4 Samplehandling....................................................................................................................................................7
1.7 SYSTEMDESCRIPTION........................................................................................................................7
1.7.1 STARTbutton..........................................................................................................................................................8
1.7.2 Statusindicator......................................................................................................................................................8
1.7.3 LCDTouchscreen..................................................................................................................................................8
1.7.4 Buildinthermalprinter.......................................................................................................................................9
1.7.5 Powersupply........................................................................................................................................................10
1.7.6 Backlabel...............................................................................................................................................................10
1.7.7 Warninglabelsontheanalyzer.....................................................................................................................11
1.8 TECHNICAL&OPERATIVESPECIFICATIONS........................................................................................11
1.9 LIMITATIONSOFUSE........................................................................................................................12
1.10 HANDLINGINCASEOFEMERGENCY..................................................................................................12
1.11 REGULATORYCOMPLIANCES.............................................................................................................12
2 CHARACTERISTICSOFTHEMEASUREMENT..............................................................13
2.1 THEMEASUREMENTPROCESS...........................................................................................................13
2.2 PRINCIPLEOFTHEIMPEDANCEMETHOD...........................................................................................14
2.3 PRINCIPLEOFTHEHEMOGLOBINMEASUREMENT...............................................................................13
2.4 PARAMETERS....................................................................................................................................14
2.5 CHARACTERISTICSOFTHEPARAMETERS............................................................................................15
2.6 LINEARITYRANGESOFTHEPARAMETERS...........................................................................................16
3 INSTALLATION......................................................................................................................17
3.1 TRANSPORTANDSTORAGE...............................................................................................................17
3.2 UNPACKING.....................................................................................................................................17
3.3 INSTALLATION..................................................................................................................................17
3.3.1 Locationspecifications.....................................................................................................................................18
3.3.2 Installthereagentsandthewastebottle..................................................................................................18
3.3.3 Installanyperipherals.......................................................................................................................................20
3.3.4 Installorchangetheprintpaperforthebuildinprinter....................................................................21
3.3.5 Installtherequiredsampleholder...............................................................................................................21
3.3.6 Connectthepowersupply..............................................................................................................................22
3.3.7 TurnONtheinstrument...................................................................................................................................23
3.3.8 Verifytheinstrument........................................................................................................................................24
3.3.9 Adaptthesystemtotheuser.........................................................................................................................24
3.3.10 Giveextensiveexplanationtotheuser!................................................................................................24
3.3.11 Shutdown.........................................................................................................................................................25

Register now at: https://diagnostics.be/warranty
CYANHematousermanual Rev.(7.0),202111 Page2
4 SOFTWAREDESCRIPTION................................................................................................26
4.1 GENERALSTRUCTURE.......................................................................................................................26
4.2 MAINMENU....................................................................................................................................27
4.3 MEASURE.........................................................................................................................................28
4.3.1 START:Runninga(new)sample....................................................................................................................28
4.3.2 NEW:Enteringpatientinformation..............................................................................................................29
4.3.3 RERUN:Runthesamesampleagain............................................................................................................34
4.3.4 BLANK:Runningablankmeasurement.....................................................................................................35
4.3.5 PRINT.......................................................................................................................................................................37
4.3.6 DISCARD................................................................................................................................................................37
4.3.7 EXIT..........................................................................................................................................................................37
4.4 DATABASE.......................................................................................................................................37
4.4.1 DETAIL....................................................................................................................................................................38
4.4.2 PRINT.......................................................................................................................................................................39
4.4.3 FILTER......................................................................................................................................................................39
4.4.4 TRENDS...................................................................................................................................................................41
4.4.5 MANAG...................................................................................................................................................................42
4.4.6 EXIT..........................................................................................................................................................................43
4.5 MAINTENANCE.................................................................................................................................43
4.5.1 Cleaning.................................................................................................................................................................43
4.5.2 Calibration.............................................................................................................................................................44
4.5.3 Qualitycontrol.....................................................................................................................................................50
4.5.4 Diagnostics...........................................................................................................................................................56
4.5.5 Reagentstatus.....................................................................................................................................................57
4.6 SETTINGS.........................................................................................................................................60
4.6.1 Printer.....................................................................................................................................................................60
4.6.2 General...................................................................................................................................................................64
4.6.3 Measurement.......................................................................................................................................................64
4.6.4 DateandTime.....................................................................................................................................................68
4.7 EXIT.................................................................................................................................................69
4.7.1 Logout....................................................................................................................................................................69
4.7.2 Shutdown..............................................................................................................................................................69
4.7.3 Preparationforshipment................................................................................................................................70
4.7.4 Usermanagement..............................................................................................................................................71
5 DAILYROUTINE....................................................................................................................73
5.1 PREPARATIVEWORK........................................................................................................................73
5.1.1 Preparationoftheanalyzer.............................................................................................................................73
5.1.2 Preparationofcontrolbloodforcalibrationand/orqualitycontrol................................................73
5.1.3 Preparationofbloodsamplesforanalysis................................................................................................73
5.2 STARTANALYSIS...............................................................................................................................74
5.2.1 Blankmeasurement...........................................................................................................................................74
5.2.2 Qualitycontrolmeasurement........................................................................................................................75
5.2.3 Samplemeasurement.......................................................................................................................................76

Register now at: https://diagnostics.be/warranty
CYANHematousermanual Rev.(7.0),202111 Page3
5.3 EVALUATETHERESULTS...................................................................................................................76
5.4 SHUTDOWN–ENDOFTHEDAY.......................................................................................................76
5.5 WEEKLY...........................................................................................................................................77
5.6 DAILYROUTINESCHEMATICS............................................................................................................78
SAMPLEMEASUREMENT..............................................................................................................78
RESULTS:ANALYSIS,PRINT&MANAGE..................................................................................78
6 MAINTENANCE.....................................................................................................................79
6.1 DAILY..............................................................................................................................................79
6.2 WEEKLY...........................................................................................................................................79
6.2.1 Cleaningneedlewashinghead.....................................................................................................................79
6.2.2 Checkofhydraulicsystem...............................................................................................................................79
6.2.3 Cleaningoftheoutsideoftheinstrument................................................................................................80
6.3 PERIODICALLY..................................................................................................................................80
6.3.1 Disinfection/Emergencycleaning..............................................................................................................80
6.3.2 Hardcleaningoftheneedle...........................................................................................................................82
7 TROUBLESHOOTING..........................................................................................................83
7.1 OPERATIONAL..................................................................................................................................83
7.2 TECHNICAL......................................................................................................................................88
8 HYDRAULICDIAGRAM......................................................................................................90
REAGENTINSERTS
CONSUMABLES

Register now at: https://diagnostics.be/warranty
CYANHematousermanual Rev.(7.0),202111 Page4
1 INTRODUCTION
1.1 INTENDEDUSE
TheCYANHematoisaneasytooperate,fullyautomated,benchtophematologycellcounterdesigned
for
invitro
diagnosticuse.Itcombinespowerfulfeaturesinacompact,independentunitmakingitideal
forsmallclinics,andpointofcarelaboffices.ItimplementstheCoultermethodforcountingcellspassing
throughasmallaperture.Theprincipleofspectrophotometryisusedtomeasurethehemoglobincontent
ofredbloodcells.
TheanalyzerfeaturesacolorgraphicalLCDdisplaymodulewithtouchscreen,andhasaseparateSTART
button.Thesoftwareallowssendingresultstoanexternalprinter(viaUSBport),orthebuiltinthermal
printer module. Its internal memory is capable of storing 10.000 records with full histograms, and
individualpatientdata.QCmeasurementsarestoredinseparatedatabase.Thesoftwareoperatingthe
instrumentiseasytoupgradeusingaUSBflashpendrive.Makingbackups,exportingdataforprocessing
onacomputerandrestoringrecordstoandfromaUSBpendriveisalsopossible.
! WARNING! Please read the user manual before operating the CYANHemato and make sure the
installationandmaintenanceareperformedcorrectly.
Note: If the installation, usage and maintenance directions are not followed correctly and/or safety
indications are not respected, Cypress Diagnostics cannot guarantee correct functioning of the
instrument.Misusecancompromisethesafetyoftheoperatorandhissurroundingsandwillvoidthe
warranty.Consumablesarenotincludedinthewarranty.
BIOHAZARD:Wearpersonalprotectiveequipment.Useuniversalprecautionsrecommended
whenworkingwithbiohazardmaterials.
1.2 MEASUREMENTPARAMETERS
Theanalyzercanprocess30samplesperhourinthe3partWBCdifferentialmode.Samplescanhave
individualsampledata,and separatetestparameters.Theusercan customizethe reportformat.The
analyzerdeterminesthefollowing20hematologyparametersfroma25µLwholebloodsample:
WBC totalwhitebloodcellcount
LYM lymphocytescount
MID midsizecellcount
GRA granulocytescount
LYM% lymphocytespercentage
MID% midsizecellpercentage
GRA% granulocytespercentage
HGB hemoglobin
RBC redbloodcellcount
HCT hematocrit
MCV meancorpuscularvolume
MCH meancorpuscularhemoglobin
MCHC meancorpuscularhemoglobinconcentration
RDW* redbloodcelldistributionwidth
PLT plateletcount
PCT plateletpercentage
MPV meanplateletvolume
PDW* plateletdistributionwidth
PLCC plateletlargecellcount
PLCR plateletlargecellratio
*RDWandPDWparametershavetwoformsofrepresentation:CVorSD.Bothparametersdescribethe
distributionwidth,butfromdifferentaspects.Theusercanselecttheunitstouse.

Register now at: https://diagnostics.be/warranty
CYANHematousermanual Rev.(7.0),202111 Page5
1.3 REAGENTS
TheCYANHematoisaclosedsystem,onlyreagentssuppliedbyCypressDiagnosticscanbeusedonthe
CYANHematoandviceversa.IfotherreagentsareusedontheCYANHematoorifthereagentsareused
onadifferentanalyzerthantheCYANHematothentheaccuracycannotbeguaranteed.
1. Diluent(Code:CY006R01):Isotonicsalinesolution,usedtodilutewholebloodsamplesandtorinse
thefluidicsystembetweenmeasuringprocedures.
2. Lyse(Code:CY006R02):Createshemolysatefor3partWBCdifferentialandforHGBmeasurement.
3. Cleaner(Code:CY006R03):Forcleaningprocessofthefluidics.
CypressDiagnosticsalsooffersemergency cleaners.Theseemergencycleanersare not designedfor
regularcleaning,butshouldbeusedincaseofproblematicblankmeasurements.
4. Hypoclean(Code:CY006R04):Hardcleaner(bleachbased).
5. Clenz(Code:CY006R05):Enzymaticcleaner.
6. HypocleanCC(Code:CY006R06):Hardcleaner,usedtocleanthesampleneedle.
Note:theinsertsofthedifferentreagentscanbefoundintheappendixofthisusermanual.
1.4 QUALITYCONTROLBLOOD
The analyzer allows the continuous monitoring of measurement performance with control blood
(low/normal/high). Their specifications (assay values, allowed tolerances and expiry date) are always
includedinthecontrolpackage.
ITEM PRODUCTNUMBER
ControlPackageNormal(1x2.5mL) CY006C01A
ControlPackageLow/Normal/High(3x2.5mL) CY006C02A
1.5 CALIBRATION
TheCYANHematoarrivestoyourlaboratoryfactorycalibrated.However,calibrationisworkingcondition
dependent (temperature, humidity, reagent lot …) and thus should be repeated at instrument
installation.Use normal control bloodfor calibration.Thetargetvalues arementionedontheinsert
providedwiththecontrolblood.Thetargetvaluesforcontrolbloodislotdependent.
1.6 BLOODSAMPLES
TheCYANHematomeasureswholebloodhumansamplestreatedwithK3EDTA.Thebloodneedsto
betreatedwithanticoagulanttopreventlargegroupsofcellsformingintoclotsthatwillclogthecell
counter.Thechoiceofanticoagulantisveryimportant,assomeanticoagulantswillaffecttheshapeand
sizeofbloodcells.IngeneralK3EDTA(orK2EDTA),aretheonlyanticoagulantsrecommendedforuse
withelectronicbloodcounters.
1.6.1 Samplecollection
Alwaysfollowtheinstructionsforusethatarespecificforthesamplecollectiontubethatisused.Meeting
therequirementsofthetubemanufacturerwillguaranteegoodqualitysampling.
Thepremanufacturedsamplingtubeshavetheanticoagulantpreparedinsidethevialinformofliquid
orpowder.Toensurethecorrectamountofbloodthatcanbetreatedwiththepredosedanticoagulant,
samplevialsareshippedwithvacuuminsidethatensuresaspirationoftherightamountofbloodduring
theoriginalsamplingprocess.Itisthusveryimportanttoneveropenthepremanufacturedsampling
tubesbeforesamplecollection,asthiswouldeliminatethevacuum.Thevialsthemselvesalsohaveamark
indicatingtheoptimalsamplelevelfortheanticoagulantalreadyincludedinthesamplevial.
Whenusing“homemade”tubespredosedwithEDTA,itisimportantthattheratioofEDTAtoblood
doesnotexceed3mg/mL.Ifthisratioisexceeded,thiswillresultinosmotictransferfromtheRBCswhich
shrinks them and thereby affect the measurements. Generally, we suggest using premanufactured
sampletubescontainingthenecessaryamountofEDTA.
Aftercollectingthecorrectamountofblood,itisveryimportanttoimmediatelymixthesample.Thevial,
afterremovalfromthesamplingadapterorsamplingneedlemustbegentlyinvertedseveraltimesto
allowhomogenousmixingofthebloodandthepredosedanticoagulantalreadypresentinthesampling
vial.

Register now at: https://diagnostics.be/warranty
CYANHematousermanual Rev.(7.0),202111 Page6
Asamplewithclotsinsidewillnotonlyhavefalsereadings(duetothefactthatthecoagulationprocess
decreasesthenumberofbloodcellssuspendedinthebloodsamplebycollectingthemintoclots)but
canalsocauseaclogginginsidethedelicatetubingsystemofahematologyanalyzer.Additionally,the
samplemustbedrawnagainfromthepatient,asclottedsamplesarenotsuitableforanalysis.
1.6.2 Samplestorage
Theanticoagulantrequiressometime(1530minutes)tostabilizeitseffectandcompletelystopthe
coagulationprocessfromstarting.BloodsamplesshouldberunonCYANHematoonlyafterthisreaction
timehaselapsed.
Blood samples can either be collected and analyzed at the same spot (making sure that the anti
coagulant’sreactiontimeisguaranteed)orcanbeanalyzedatanoffsitelocation.
Thecollectedsamplescanbekeptatroomtemperature(15–30°C)forupto8hours.Ifthesamples
cannotbeanalyzedwithinafewhoursaftercollection,samplesshouldbestoredatatemperatureof2
8°Ctodelaythedecayofthebloodcells.Whenthesamplesarestoredat28°Cimmediatelyaftersample
collection,thesamplescanbeanalyzedupto24hoursaftercollection.
Samplesmustbeprocessedwithinthetimeperiodsmentionedabove(timeisdependingonthestorage
temperature).Readingsmadeafterthesetimeperiodscanresultinbadornoninterpretablereadingsdue
tothedecayingeffectsofbloodcells.
Notes:
a. Thebloodsamplemustneverfreeze,asthis(freezingoftheliquidincludedinbloodplasma)damages
cellstructuresandcellmembranes,resultinginbadornoninterpretablereadings.
b. Thetimeperiodsmentionedabovecanonlybeobtainedwhenthebloodsamplesarestoredona
rollermixer.Therollermixerisidealforstoringbloodsamples.Themixingissoftandthesurface
betweentheliquidandtheairislarge,sothatthecellsremainincontactwithoxygen.Thiswillkeep
thecellsalivemuchlongerandthecellskeepalltheircharacteristics.
Ifthevialsareleftuprightinarack,thebloodsamplesinvialstendtosedimentwith
time. The cells sink down to the bottomof the vial leaving blood plasma in the top.This makes
diffusionfromoxygenfromtheairtothecellsimpossibleandthecellswilllosetheircharacteristics
and die. A wellhomogenized sample is a homogenous red color liquid. A sample that has been
sedimented,isdividedintosections:PLT(thrombocyte)richplasmafloatsonthetop,andathinwhite
layerisalsoformed(WBC,leukocytes–“buffycoat”)abovethedarkredbottomhalfmadeupofRBC
(erythrocytes).

Register now at: https://diagnostics.be/warranty
CYANHematousermanual Rev.(7.0),202111 Page7
1.6.3 Samplepreparation
1. Beforeanalysis,thesamplesneedtobehomogenized.
a. Theidealwaytohomogenizesamplestakenoutisputtingthemonarollermixerfor2030
minutesensuringagentlemixingofthecellsandplasma.
b. Immediatelybeforesampling,itisalsorecommendedtogentlyinverttheclosedsampletubeat
least8times.
c. Avoidshakingoraccidentallydroppingthesamplesasmeansofmixing,asthisactioncanbreak
thebloodcellsandmightintroducemicrobubblesthatwillalsoinfluencereadings.
2. Theanalysistemperaturedependsonthestoragetimeperiodofthesamples.
a. Ifstoredforlessthan6hoursthesamplesshouldbetakenfromthesamplecooler(orrefrigerator)
andgentlywarmeduptoroomtemperature.Avoidwarmingthemupimmediatelyinyourhand
asthismayintroduceaheatshocktothecellsandinfluencereadings
b. Ifstoredformorethan6hours,itisessentialthattheanalysisisperformedoncoldsamples.
Homogenize the samples and analyze them immediately. Do not let them warm to room
temperature,thiswouldresultitrapiddecayofthecellsandinfluencetheanalysis.
3. Alwaystakeoffthecapofthetubes!Otherwise,theneedlewillcrash!Theneedlecannotpiercecaps.
Note:
Ifthesamesampleneedstobemeasuredmultipletimes,pleasemixthesamplegentlybeforeevery
sampling(invertthetubeatleast8times).Makesuretowipethecapcleanbeforeplacingitagainon
thevial.
1.6.4 Samplehandling
Bloodsamplesmustalwaysbeconsideredasbiohazardandinfectiousbiologicalmaterial.
Itisrecommendedtowearprotectiveglovesandgoggles.
Spilledbloodmustbecleanedupassoonaspossible,andthesurfaceincontactwithit
mustalsobecleanedwithasuitablesanitizer.
1.7 SYSTEMDESCRIPTION
TheCYANHematoisamicrocomputerbasedindependentunit,composedofthreemainunits:
1. FluidicSystem
Performssampling,diluting,mixingandlysingoperations.Generatestheregulatedvacuumusedfor
movingcellsthroughtheapertureduringcounting.
2. DataProcessing
Systemcounts,measuresandcalculatesbloodcellparameters,generatesandstoresnumericalresults
andhistograms.
3. ControlPanel
FeaturesanLCDtouchscreenSTARTbutton,statusLED,andUSBportinterfaces.

Register now at: https://diagnostics.be/warranty
CYANHematousermanual Rev.(7.0),202111 Page8
Thefiguresbelowshowthefrontviewoftheanalyzerwithcontrolsandconnectors.
1.7.1 STARTbutton
PressingandreleasingtheSTARTbuttontriggersananalysiscycle.
1.7.2 Statusindicator
A LED (light emitting diode) is located above theSTART button. Its color indicates the status of the
analyzer.
LEDCOLOR ANALYZERSTATUS
GREEN Theanalyzerisreadyforsamplemeasurement.Analysisisinitiatedbypressing
theSTARTbutton.
REDBLINKING Blood sample can be removed when the LED blinks red 3 times and the
instrumentbeeps3times.
RED Theanalyzeriscurrentlyperformingananalysis.Nonewmeasurementcanbe
started.
ORANGE Theanalyzerisperformingamaintenanceprocess.
ORANGEBLINKING Theinstrumentisinstandbyandthedisplaylightisoff.Hitthescreentowake
upfromstandby.
1.7.3 LCDTouchscreen
Thedisplayisa320x240dots,highcontrastbacklithighcolorgraphicLCDmodule,withintegratedtouch
screen.Afingerorsmallerbluntobject(styluspen)shouldbeusedtooperatethetouchscreen.Ifthe
operatorgentlytouchestheLCD,theanalyzerwillrecognizethepositionwherethescreenwaspressed;
thesoftwarewillactivatethecorrespondingfunction/menu/button.
!ATTENTIONPOINT!Thescreencanonlyreacttoonetouchpointatthesametime.Ifsomethingelseis
touchingorpressingthescreen(onanemptyposition),itwillnotrespondtoyourtouch!
1.7.3.1 KEYBOARD
Whendataneedstobeentered,akeyboardwillappear.Thekeyboardwillbeadaptedtothedatathat
canbeenteredinthatspecificarea;alphanumericornumericonly.

Register now at: https://diagnostics.be/warranty
CYANHematousermanual Rev.(7.0),202111 Page9
Press“Enter”toconfirmaninsertionandreturntothepreviousmenu,press“OK”.PressCancel”toreturn
tothepreviousmenuwithoutusingtheinsertion.
1.7.3.2 TOUCHSCREENCALIBRATION
Incasethe touch screendoesnotrespondcorrectlyto your tap(youtapa specific location,stillthe
requiredfunctionkeyisnotactivated)youhavetocalibratethetouchscreen.
Procedure:
1. Gentlytapandcontinuouslyholdanyemptylocationonthetouchscreenforatleast30seconds.
2. Afterca30seconds,acalibratingscreenappears.Tapthemiddleofthecrossofthereferencepoints
oneafteranother.
Ifyoumadeanerror,youwillhearanerrorbeep,andtheprocessrestarts.
Ifthecalibrationwassuccessful,thesoftwarewillreturntotheoriginalscreen.
!ATTENTIONPOINT!
Makesurenottopressithard,otherwiseyoucandamagethescreen
1.7.4 Buildinthermalprinter
TheCYANHematoisequippedwithabuiltinthermalprinter.The
textorgraphisprintedoncoatedthermochromicpaper.Thispaper
isselectivelyheatedwhenpassingoverthethermalprintinghead,
andtherebythecoatingontheseplacesturnsblack,producingan
image.Oneofthebigadvantagesofathermalprinteroveranormal
printer is that it does not require expensive consumables like a
toner.
Pressthefeedbuttontomovethepaperforwardwithoutprinting.
Thefiguresbelowshowthebackviewoftheanalyzer,withcontrolsandconnectors.

Register now at: https://diagnostics.be/warranty
CYANHematousermanual Rev.(7.0),202111 Page10
1.7.5 Powersupply
Theanalyzerisequippedwithanexternalpowersupply.Thepowersupplymodulehasasocalledauto
rangeinput,allowingoperationon230Vor115Vpowersystems.ThepowersupplyunitcomplieswithCE
safetycertifications.
TheinputsocketisastandardpowercableconnectionandtheoutputisaDCjack.
Note:Onlyusetheprovidedpowersupplywiththeinstrument.
1.7.6 Backlabel
Everyinstrumenthasitsownserialnumber,whichisprintedonthebacklabelandcanalsobefoundin
theDeviceInformationsubmenu.Pleasementionthisserialnumberineverycommunication.