dacell SM110 User manual

DACELL CO., LTD
http://www.dacell.com
TEL:+82-43-2602242, FAX : +82-43-2602245
1
DIGITAL INDICATOR
SM110, 150
- INSTRUCTION MANUAL -
DACELL CO.,LTD
TEL:+82-43-260-2242
FAX:+82-43-260-2242
WEBSITE:http://www.dacell.com

DACELL CO., LTD
http://www.dacell.com
TEL:+82-43-2602242, FAX : +82-43-2602245
2
Checks install place and connection prior to setting, caution the following
instruction.
* No power ON/OFF switches.
Cautions to operate from the voltage supply the moment.
* Input/output part are not isolated electrically. Use the specific cable to protect
indicator against over voltage and current and indicator get to the ground connection
* Setting lock mode for system and data protection
* Do not internal disjointing and adjustment voluntarily because the indicator is
composed of electron component and P.C.B circuit to sharp at the static electricity .
DACELL is not liability for mistake to use that change a property voluntarily, change
design, adjustment voluntarily, user carelessness cause defect.
1. Installation
1) Product Spec.
- Spec : 96 ㎜×48 ㎜/3.78“ × 1.89"(1/8 DIN)
depth : 126 ㎜/4.96"
- SIZE : 92 ㎜(+1.0/-0) × 45 ㎜(+0.8/-0)
3.62“(+0.04/-0)× 1.77“(+0.03/-0)
* Mount the indicator using Bracket.
* Get a indicator into front of the panel and fixing guide push in a terminal direction
and volt fix.
* Thickness of mounting plane : 1 ㎜~8㎜
2) Install place
Install the indicator separately from switching device and relay, avoid
installation as follows.
* MC, Switching device, relay * Thyristor unit
* Motor * Dust, moisture, corrosion gas, heat line exist place.
* Operating temperature : 0 ~50℃
3) Power characteristics
* Install line filter if harmonic source and arc welder exist around.
* Separate sensor and control line from power line

DACELL CO., LTD
http://www.dacell.com
TEL:+82-43-2602242, FAX : +82-43-2602245
3
2. Connection and wiring
1) AC input : 1 and 2 ports, 110/ 220V (50Hz/60Hz).
Separate power cable from sensor cable.
Refer to 5-1).
2) HOLD : 3 and 4 ports.
short state : current display value holding
open state : current value display(normal operation mode)
3) ANALOG OUTPUT : 4 →-(GND), 5 →+
Analog output is as following voltage or current
DIP switch 5 on : voltage output.
DIP switch 5 off : current output.
ON Voltage Output ( 0-5V, 1-5V ,0-10V, 2-10V)
OFF Current Output ( 0-10 ㎃, 0-20 ㎃, 4-20 ㎃)
Adjust zero point and maximum point of analog output using
zero and span VR
signal 0 : zero VR →0V, signal max : span VR →10V
zero range : -3V~ +3V, span : 10.5V
0~10V : Switch 5 off →current output mode (0 ~ 20 ㎃)
4) SENSOR connection port : 6,7,8,9 ports
Set the hardware and software as following conditions.

DACELL CO., LTD
http://www.dacell.com
TEL:+82-43-2602242, FAX : +82-43-2602245
4
6 and 7 port : excitation voltage (power for sensor)
* EXC+ : +2.5Vdc * EXC- : -2.5Vdc
* Load current : Max 120mA ( EXC = 5Vdc )
Max 60mA ( EXC = 10Vdc )
8 and 9 port : signal connection ports(output of sensor).
----- HARDWARE setting -----
* DIP S/W 1 on : sensor output is 0.700mV/V - 1.249mV/V.
* DIP S/W 2 on : sensor output is 1.250mV/V - 1.749mV/V.
* DIP S/W 3 on : sensor output is 1.750mV/V - 2.499mV/V.
* DIP S/W 4 on : sensor output is 3.000mV/V - 3.499mV/V.
☞Must be among four switched on only one in accordance with kind of sensor.
☞Under 0.700mV/V or over 3.500mV/V are option.
------ SOFTWARE setting-----
* In cal mode, input as follows according to sensor output.
5) RELAY OUTPUT : A(NO) and B(NC) contact.
* Capacity AC -- 250V/5A Max, 125V/10A Max,
DC -- 30V/5A Max,
* Output of transistor NPN open collector is option.
* LOW OUT : A contact ON ( ≤setting data)
10-NO, 11-COM, 11-NC
* HIGH OUT : A contact ON ( > setting data)
13-NO, 14-COM, 15-NC.
6) ZERO : External reset port. When (16) and GND port are
shorted, display is zero. maximum measurement data of
indicator is removed.
7) PEAK : Display maximum measurement data when (18) and
GND port are shorted. Display present measurement data in case of open.
Refer to 6-3) (auto peak function).

DACELL CO., LTD
http://www.dacell.com
TEL:+82-43-2602242, FAX : +82-43-2602245
5
3. DISPLAY and KEY operation
1)DISPLAY : 4½ Digit LED
2) L.E.D Display
* Key operation start : press to MODE key.
Key operation exit : press to ENT key.
* Display : setting mode and operation mode.
1. Setting mode : data confirmation or correction.
2. Operation mode : normal operation state.
3. Press to MODE key : setting mode.
Press to ENT key : operation mode.
LED state Description
operation Display operation mode of high relay.
HI
setting Confirm and adjust high data
operation Display operation mode of low relay.
LO
setting Confirm and adjust low data
FS setting Confirm and adjust full scale data.
Input the full range or capacity of sensor.
DP setting Confirm and adjust decimal point.
CAL setting key in sensor output data ( mV/V )* refer to 6 - 1)
FS+CAL setting auto calibration execution mode.
cal data is adjusted automatically.* refer to 3 - 6)
FS+ DP setting Adjust display time and unit of
measurement data* refer to 6 - 2)
DP+CAL setting Software adjustment mode.* refer to 6 - 3)
FS+ DP+CAL operation Set or reset data lock mode* refer to 3 - 4)

DACELL CO., LTD
http://www.dacell.com
TEL:+82-43-2602242, FAX : +82-43-2602245
6
3) KEY operation
Key result Description
Mode
select
1. HI→LO→FS→DP→CAL→FS+CAL→FS+ DP→DP+CAL
2. All FND of selected mode will blink.
Position
select
3. Press key : 10000→1000→100→10→1
Selected digit position will blink
4. DP(decimal point mode) : decimal point
position movement key.
10.000→100.00→1000.0→10000
N
umbe
r
select
5. Data correction key.
6. 5 digit →1, -1, -, non
7. The others 1→2→3→4→5→6→7→8→9→0
Input 8. Input data. After setting up must press
key therefore become input.
+
Auto zero 9. Sensor : No load state(0V, 4 ㎃)
10. Display : 0
4) PROTECTION mode
Use this mode for protection and security of System.
high, low data is possible to adjust irrespectively.
Key result Description
+
Lock
mode
11. Protect setting data.
12. Reset lock mode to adjust data
Mode confirm and setting
Unlock
* Press and key during 5 sec at the same time
FS, DP, CAL LED will light up and reset lock mode (data correction enable)
Lock
* Press and key during 5 sec at the same
time in unlock mode : FS, DP, CAL LED turn off (lock mode)
* power off : Set lock mode automatically.

DACELL CO., LTD
http://www.dacell.com
TEL:+82-43-2602242, FAX : +82-43-2602245
7
5) AUTO ZERO mode
Initialize sensor and indicator.
Execute this mode at unlock mode.
Execution process (display : operation mode)
(1) Sensor : no load.
(2) Voluntary data display.
(3) Press key and key at the same time.
Zero display (ex : 0, 0.0, 0.00, 0.000)
(4) This data is memorized to indicator though power turn off.
(5) If operate mode with input signal to remain, this point is memorized zero point.
6) AUTO CALIBRATION mode
Convenience to use when do not know calibration data of sensor or
displayed the measurement value differently.
(1) Adjust zero point of indicator.
(2) Apply known load to sensor. Large load is high safety.
(3) Press key: F.S and CAL L.E.D turn on at the same time.
(4) As example table, 100.05 or 99.05 will blink.
(5) After correction of 100.00 using key and key, press key.
(6) 100.05 is display correction of 100.00.
(7) At this time, cal data is corrected minutely.
(8) If corrected cal data is over limits of 2-4), correct of DIP S/W setting and again
set the auto calibration.
EX) At sensor Supply : 100.00 kg/㎠
Display : 100.05 or 99.05 kg/㎠
Adjustment : 100.00 kg/㎠

DACELL CO., LTD
http://www.dacell.com
TEL:+82-43-2602242, FAX : +82-43-2602245
8
4. Operating flow chart
--Mount indicator to panel and make connection. (Check AC
POWER 110/220V)
-- Press +key(data protection unlock)
-- Set data(high, low, F.S, DP, CAL),refer to 6
--sensor output data ---DIP S/W 1, 2, 3, 4
analog output ---- DIP S/W 5
filter frequency ---- DIP S/W 6, 7, 8
-- Sensor : no load
Press +key (auto zero)
---Execute auto calibration.
---Check function(Relay output,analog output, reset, hold, peak).
--Press +key(data protection mode)
5. HARDWARE
1) Selection of AC input power
(1) Separate P.C.B from case.
(2) The following label is existed under power transformer
2) Transformation of SAMPLING TIME
15 times/sec (standard).
(1) The following label is existed on the P.C.B
Install and connection
Unlock mode
DATA Input
DIP S/W setting
Auto zero
AUTO CAL
Mode Check
Lock mode setting
①②③④110V : connect 1 and 2
110 220 110 connect 3 and 4
220V : connect 2 and 3
○○○15 times/sec : connect 1 and C
15 C 30 30 times/sec : connect 2 and C

DACELL CO., LTD
http://www.dacell.com
TEL:+82-43-2602242, FAX : +82-43-2602245
9
3) Setting of DIP switch
After connection, dip s/w must be set.
ON 1 2 3 4 5 6 7 8
1~4 S/W :One of 1,2,3 or 4 turn on according to sensor output value (mV/V)
5 S/W : Turn on or off according to analog output
6,7,8 S/W : Analog input filter frequency 6,7,8 turn off : 4 ㎑
6. SOFTWARE
1) CALIBRATION DATA
Set data as follows according to sensor output value.
Confirm calibration data after auto calibration execution.
1 mV/V range --- 0.700 ~1.249 mV/V
1.5 mV/V range --- 1.250 ~1.749 mV/V
2 mV/V range --- 1.750 ~2.499 mV/V
3 mV/V range --- 2.500 ~3.500 mV/V
2) Adjustment of display time and measurement data unit
(1) Press mode key, FS+DP LED turn on at the same time.
(2) 01.01 is displayed.
(3) Display time (set 1 times : fastest, set 15 times : slowest)
voltage
1 mV/V 1.5
mV/V
2
mV/V
3
mV/Vcurrent1Hz10Hz100Hz
0 1 . 0 1
Setting of displat time Setting of adjustment width
01 - After 1 times sampling, display
02 - After 2 times “
: :
: :
14 - After 14 times “
15 - After 15 times “
01 : display 1, 2, ……, 9
02 : display 2, 4, 6, 8, 0(even)
05 : display 5, 10, 15
10 : display 10.....90
20 : display 20, 40, 60, …
50 : display 50, 00

DACELL CO., LTD
http://www.dacell.com
TEL:+82-43-2602242, FAX : +82-43-2602245
10
3) PROGRAM (SOFTWARE) OPTION
(1) Press mode key, DP+CAL LED turn on at the same time.
mode display description
0 x x x
p
eak hold on : display maximu
m
measurement data
p
eak hold off : display present measurement data..
reset : stored maximum measurement data is removed.
Adjustment of peak
hold mode
1 x x x
p
eak hold on(only) : display and refor
m
maximum measurement data
p
eak hold off : stored maximummeasurement data is
removed automatically.
mode display description
x 0 x x standard type.
ex) +20mA : display 1000.0
Parity of
Display x 1 x x
display (+) to (-)
signal and display indicate opposite.
ex) +20mA : display -760.0
x x 0 x standard type.
Display 0 : trace minute signal to 0
Tracking
mode of
zero point x x 1 x Display 0 : display minute signal.
Use in case of leakage measurement.
x x x 0 display measurement signal immediately.Average
value
display of
Signal
x x x 1
display average value of measurement
signal (7 times sampling).
7. ALARM OUTPUT (RELAY OUTPUT)
High and low relay are a and b contact respectively.
Contact capacity is maximum AC 250V/5A and DC 30V/5A.
Relay contact operating time is maximum (1/Sampling Time) mm Sec
* Output of transistor NPN open collector is option.
mode setting relay
high data > setting value ONHIGH
RELAY high data ≤setting value OFF
low data ≤setting value ONLOW
RELAY low data > setting value OFF

DACELL CO., LTD
http://www.dacell.com
TEL:+82-43-2602242, FAX : +82-43-2602245
11
8. TECHNICAL DATA
item data
allowed time temperature ―40 ~80℃
operating temperature 0℃~50℃
power of sensor 5Vdc or 10Vdc
input signal type 1 ~3mV/V
max. load current 10Vdc/60 ㎃, 5Vdc/120 ㎃
frequency response Max 5 ㎑
analog output 0~10V,0~20 ㎃,1~5V, 4~20 ㎃
relay output 220V/5A (a,b contact)
relay response Max 66 ㎜/sec
nonlinear characteristics ±0.02%/F.S
temperature coefficient ±0.02%F.S/℃
display size 14 ㎜×8 ㎜
sampling time 15 times/sec, 30 times/sec
filter frequency 1㎐, 10 ㎐, 100 ㎐,4㎑
short protection circuit protection circuit (insert)
max. display range -19999 ~19999
size w 96 ㎜×h 48 ㎜×d 126 ㎜
power 110Vac, 220Vac (50~60 ㎐)
9. Unit Label change
* Unit label is optional.
* In the case of exchanging label, refer to as follows
1, Separate P.C.B from the case.
2, Remove Conventional unit label and then adhere new label.
3, Put together indicator in reverse process.

DACELL CO., LTD
http://www.dacell.com
TEL:+82-43-2602242, FAX : +82-43-2602245
12
10. Trouble shooting and correction
problem cause correction
Display
drifts
* Defective Sensor
* AC noise in power line
* Defective connection
to sensor (Noise)
* Input filter set incorrectly
* Check sensor state and output
* Check power line and connection
* Check connection to sensor
* Check filter DIP switch
* 6-2) check display time
* 6-3) Change to AV mode
Display
19999 only
* Discordance between
sensor output and DIP
S/W setting
* Open and short of
sensor in/output line
* Check DIP switch about
sensor output
* Check sensor wiring
Display
change --
* Exchange of sensor
input line.
* data setting error
* Check sensor connection.
* refer to 6-3)
Not able to
select
FS,DP,CAL
* Data Protection * Turn off data protection
refer to 3- 4)
Analog output
Problem
* Defective voltage and current
signal
* Difficulty of setting
* When 0~5V output is
required.
* Check DIP Switch 5
* Adjust zero and span VR
refer to 2-3)
* Connect 1/4W 500Ω
accuracy resistor between
GND and OUT port
* Contact the manufacturer.
Other noise
p
rotection
* Connection noise filter with
AC input line
* Set noise filter to coil in parallel
* Check the shield.
* If you have any questions, please contact the manufacturer.
Other manuals for SM110
1
This manual suits for next models
1
Table of contents
Popular Other manuals by other brands

Rion
Rion Hobby Gardener Assembly instructions

BFT
BFT MOOVI 30-50 ALPHA BOM Installation and user manual

NEC
NEC Slot-In PC with Intel Celeron 2 x 1.6 GHz CPU STv2 (Sandy... installation instructions

Domo
Domo DO342DH Instruction booklet

DAHON
DAHON SPEED DIAL ULTIMATE installation instructions

Jegs
Jegs 81165 installation instructions