
- -
Upon turning on the instrument automatically detects the probes fitted with SICRAM
module: the Probe Type menu item is configured by the instrument as AUTO and cannot be
modified by the user.
Upon turning on the temperature probes direct 4-wire Pt100, the Pt1000 and Ni1000 display
the message "NO_PRBE_SER_NUM". In this case the probe type must be entered
manually. Select Probe type using the MENU key and then select the type of probe with
the arrow keys; confirm using ENTER.
4) Management of memorized data (only HD21...4.2): the message
“>>>_LOG_DUMP_or_ERAS” (Transfer data or erase) is scrolled in the comment line.
The center figure reports the number of free memory pages (FREE). Pressing
SERIAL/EraseLOG permanently erases all memory data. By pressing LOG/DumpLOG, the
data transfer of the logged data on the serial port is started: the “BAUD-RATE” must have
previously been set to the maximum value (please see the menu items described below and
the paragraph "STORING AND TRANSFERRING DATA TO A PERSONAL
COMPUTER" on page 25).
5) Print and log interval (only HD21...4.2): sets the interval in seconds between two loggings
or data transfers to the serial port. The interval can be set at 0, 1s, 5s, 10s, 15s, 30s, 60s
(1min), 120s (2min), 300s (5min), 600s (10min), 900s (15min), 1200s (20min), 1800s
(30min) and 3600s (1hour). If the value 0 is set, SERIAL works on command: the
sending of data to the serial port is performed each time the key is pressed. Recording
(LOG) is performed with one-second intervals even if the interval is set to 0. With an
interval from 1 to 3600s, continuous data transfer is started when the SERIAL key is
pressed. To end the recording (LOG) and continuous data transfer operations (SERIAL with
an interval greater than 0), press the same key again.
6) Sleep_Mode_LOG (Automatic turning off during recording) (only HD21...4.2): this
function controls the instrument's automatic turning off during logging, occurring between
the capture of a sample and the next one. When the interval is lower than 60 seconds, the
instrument will always remain on. With intervals greater than or equal to 60 seconds, it is
possible to turn off the instrument between loggings: it will turn on at the moment of
sampling and will turn off immediately afterwards, thus increasing the battery life. Using the
arrows select YES and confirm using ENTER in order to enable the automatic turning off,
select NO and confirm to disable it and keep the instrument on continuously.
Note: even if Sleep_Mode_LOG=YES is selected, the instrument does not turn off for less
than one minute intervals.
7) YEAR (only HD21...4.2): to set the current year. Use the arrows to modify this parameter
and confirm using ENTER.
8) MNTH (month) (only HD21...4.2): to set the current month. Use the arrows to modify this
parameter and confirm using ENTER.
9) DAY (only HD21...4.2): to set the current day. Use the arrows to modify this parameter and
confirm using ENTER.
10) HOUR (only HD21...4.2): to set the current hour. Use the arrows to modify this parameter
and confirm using ENTER.
11) MIN (only HD21...4.2): to set the current minutes. In order to correctly synchronize the
minute, it is possible to reset the seconds by pressing the UNIT key. Use the arrows to set
the current minute plus one, and as soon as that minute is reached press UNIT: this
synchronizes the time to the second. Press ENTER to go onto the next item.