Security Rules
(TO BE READ IMPERATIVELY BEFORE FIRST USE)
QuickSleeper® is an electronic dental anesthesia system designed for local injections before and/or during treatment.
QuickSleeper® is a modern alternative to traditional anesthesia syringes. The practitioner should take the usual precautions
to ensure the success of every dental anesthetic injection.
WARNINGS
It is forbidden to alter your Quicksleeper without the permission
of Dental Hi Tec. Should your QuickSleeper be modified, the
modification must be performed by a service center approved
by Dental Hi Tec, or by Dental Hi Tec’s technical Service. Control
and appropriate testing must be performed to ensure that your
device can still be used safely.
Using QuickSleeper with other equipment is under the
responsibility of the user dentist.
CONTRAINDICATIONS
There are no contraindications known until now.
GENERAL RECOMMENDATIONS
· Keep mobile phones at a minimum of 1.5 metres (5 feet)
from the control box, pedal or power supply; otherwise this
could disrupt the wireless communication of QuickSleeper®.
· QuickSleeper® must only be used by a qualified dental
surgeon.
· The anesthesia act requires constant supervision of the
registered practitioner who can remove the handpiece from
the injection site at anytime.
· Do not leave QuickSleeper® within reach of children.
· Do not expose your device to the sun or leave it near a
radiator or an air-conditioner.
· Do not expose QuickSleeper® to water or chemical
products.
· The plug of the power supply serves as a cut-off and must
remain accessible.
· Imperatively wear disposable surgical gloves when
manipulating and using QuickSleeper®.
PROTECTION AGAINST ELECTRIC
SHOCKS
· The connexion must be made on an alternative single-phase
circuit.
· The external voltage should be identical to that on the
material safety data sheet located under the control unit.
· The electrical installation must have a security circuit breaker.
· Before any use, check the condition of the plug and power
cable.
· Do not pull on the cable to remove the plug from the power
socket.
SECURITY LEVELS:
· Handpiece and control box cables: very low security voltage
(VLSV-24V).
RECOMMENDATIONS
FOR USE
ANESTHETIC
· Only use 1.7ml or 1.8ml glass anesthetic cartridges, with
plastic protection film to easily withdraw the cartridge in case
of breakage.
· Refer to the cartridge manufacturers’ recommendations to
know the quantities that can be injected per session.
SECURITY
· For security reasons, do not screw or unscrew
the needle close to your patients and protect
them with safety glasses.
CONTAMINATION
· Each QuickSleeper® needle, cartridge and accessory must
be used only on one single patient to prevent risks of cross
contamination.
· Never re-use the needles and anesthetic cartridges, each
of these items must be destroyed after use in an incinerator
allocated for this purpose.
· Always replace the container by a sterile container in case of
change of needle and/or cartridge during anesthesia.
· Do not leave the needles without protection. Systematically
replace the protection after use, using the recapping system
on the handpiece holder (p. 8).
DISINFECTION
· The containers must be disinfected and sterilised before each
anesthesia (see procedure p.17).
· The holder and handpiece must be disinfected after
each patient (see the indications of disinfectant product
manufacturers).
NEEDLE BREAKAGE
· Needle breakage is the result of an incorrect manipulation.
This breakage may be caused by one or several parameters :
see p. 21
2