DH Instruments GB-152 User manual

©1998-2008 DH Instruments, a Fluke Company
GB-75™ and GB-152™
Gas Booster With Control Kit
Operation and Maintenance Manual

GB-75™ AND GB-152™ OPERATION AND MAINTENANCE MANUAL
©1998-2008 DH Instruments, a Fluke Company
This equipment described in this manual is designed and manufactured for the intended purpose of
generating high pressure gas. Certain precautions need to be followed during installation and operation of this
device. Reading and understanding the material is essential to the safe and correct operation of the unit.
Pressurized equipment is potentially dangerous. The equipment described in this manual generates and
controls very high gas pressures. It should not be operated by anyone who has not become thoroughly familiar
with this manual. Additional training in general and pressure specific safety procedures will help assure
protection from harm or damage to personnel or property. Responsibility for the proper and safe operation of
this instrument rests with the user.
High pressure liquids and gases are potentially hazardous. Energy stored in these liquids and gases can
be released unexpectedly and with extreme force. High pressure systems should be assembled and operated
only by personnel who have been instructed in proper safety practices.
© 1998-2008, DH Instruments, a Fluke Company All rights reserved.
Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted in any
form or by any means, electronic or mechanical, for any purpose, without the express written permission of DH Instruments, a
Fluke Company 4765 East Beautiful Lane Phoenix AZ 85044-5318 USA.
DH Instruments makes every effort to ensure the accuracy and quality of its published materials; however, no warranty, expressed
or implied, is provided. DH Instruments disclaims any responsibility or liability for any direct or indirect damages resulting from the
use of the information in this manual or products described in it. Mention of any product or brand does not constitute an
endorsement by DH Instruments of that product or brand. This manual was originally composed in English and was subsequently
translated into other languages. The fidelity of subsequent translations cannot be guaranteed. In case of conflict between the
English version and another language version, the English version takes precedence.
DH Instruments, DH Instruments, a Fluke Company, DH, DHI, PPC, PPCK+, PPC2+, RPM3 and CalTool are trademarks,
registered and otherwise.
Swagelok is a registered trademark of the Swagelok Company.
Teflon is a registered trademark of the 3M Corporation.
Document No. 550038c
080423
Printed in the USA.

TABLE OF CONTENTS
Page I ©1998-2008 DH Instruments, a Fluke Company
T
TA
AB
BL
LE
E
O
OF
F
C
CO
ON
NT
TE
EN
NT
TS
S
TABLE OF CONTENTS...............................................................I
TABLES.................................................................................III
FIGURES................................................................................III
ABOUT THIS MANUAL .............................................................IV
1. INTRODUCTION .................................................................1
1.1 PRODUCT OVERVIEW ...........................................................................................................................1
1.2 LOCATION AND DESCRIPTION OF THE COMPONENTS....................................................................2
1.2.1 GAS BOOSTER.............................................................................................................................................2
1.2.2 DRIVE AIR CONTROL KIT............................................................................................................................3
1.3 SPECIFICATIONS ...................................................................................................................................5
2. INSTALLATION ..................................................................7
2.1 UNPACKING AND INSPECTION ............................................................................................................7
2.2 SITE REQUIREMENTS............................................................................................................................8
2.2.1 SITE................................................................................................................................................................8
2.2.2 GAS SUPPLIES.............................................................................................................................................8
2.2.2.1 DRIVE AIR SUPPLY ..................................................................................................................................8
2.2.2.2 HIGH PRESSURE INSTRUMENT GAS.....................................................................................................9
2.3 INSTALLATION AND SETUP..................................................................................................................9
2.3.1 GENERAL CONSIDERATIONS.....................................................................................................................9
2.3.2 INSTALLATION ...........................................................................................................................................10
2.3.2.1 BOOSTER INSTALLATION .....................................................................................................................10
2.3.2.2 DRIVE AIR CONTROL KIT INSTALLATION............................................................................................10
2.3.2.3 BOOSTER HIGH PRESSURE INSTRUMENT GAS SUPPPLY CONNECTION.....................................11
2.3.2.4 BOOSTER HIGH PRESSURE OUTPUT CONNECTION........................................................................11
2.4 INITIAL START UP................................................................................................................................12
3. OPERATION..................................................................... 13
3.1 GENERAL OPERATING PRINCIPLE AND INFORMATION.................................................................13
3.2 OPERATION..........................................................................................................................................13
3.2.1 SETTING THE BOOSTER OUTPUT PRESSURE ......................................................................................14
3.2.2 TURNING THE BOOSTER ON AND OFF...................................................................................................14
4. MAINTENANCE AND ADJUSTMENTS .................................. 15
4.1 MAINTENANCE.....................................................................................................................................15

GB-75™ AND GB-152™ OPERATION AND MAINTENANCE MANUAL
©1998-2008 DH Instruments, a Fluke Company Page II
5. TROUBLESHOOTING ........................................................ 17
5.1 GENERAL INFORMATION....................................................................................................................17
5.2 BOOSTER WILL NOT RUN...................................................................................................................17
5.3 BOOSTER RUNS TOO SLOWLY .........................................................................................................18
5.4 PRESSURE GENERATES TOO SLOWLY OR NOT AT ALL...............................................................18
5.5 BOOSTER RUNS (CYCLES) CONTINUOUSLY...................................................................................19
5.6 CANNOT ACHIEVE DESIRED PRESSURE..........................................................................................19
5.7 LEAKS...................................................................................................................................................20
5.8 GAS CONTINUOUSLY VENTS THROUGH EXHAUST MUFFLER......................................................20
6. WARRANTY STATEMENT .................................................. 21
6.1 WARRANTY STATEMENT....................................................................................................................21

TABLES AND FIGURES
Page III ©1998-2008 DH Instruments, a Fluke Company
T
TA
AB
BL
LE
ES
S
Table 1. GB-75 or GB-152 Parts List...........................................................................................................7
Table 2. DHI Authorized Service Providers...............................................................................................21
F
FI
IG
GU
UR
RE
ES
S
Figure 1. GB-152 Gas Booster....................................................................................................................2
Figure 2. Drive Air Control Kit .....................................................................................................................3
Figure 3. System Schematic .......................................................................................................................4

GB-75™ AND GB-152™ OPERATION AND MAINTENANCE MANUAL
©1998-2008 DH Instruments, a Fluke Company Page IV
A
AB
BO
OU
UT
T
T
TH
HI
IS
S
M
MA
AN
NU
UA
AL
L
Manual Conventions
(CAUTION) is used in throughout the manual to identify user warnings and cautions.
(NOTE) is used throughout the manual to identify operating and applications advice and
additional explanations.

1. INTRODUCTION
Page 1 ©1998-2008 DH Instruments, a Fluke Company
1
1.
.
I
IN
NT
TR
RO
OD
DU
UC
CT
TI
IO
ON
N
1.1 PRODUCT OVERVIEW
GB-75 and GB-152 are gas boosters with control kits intended to be used to provide a very high pressure
gas supply to systems using a GPC1 High Pressure Gas Controller and other high gas pressure test and
calibration systems.
Both models include a pneumatically driven, piston type, self-cycling gas booster. The gas booster
boosts a lower pressure, generally supplied from a bottle (2 MPa [300 psi] minimum), to higher pressure.
The booster is powered by drive air. The high pressure output is equal to the drive air input times the
boosting ratio of 75 or 152. The high pressure output is connected to the SUPPLY port of a GPC1 or
other system to which high pressure gas is to be supplied.
Both gas booster models include a drive air control kit and interconnecting tubing. The kit is on a bracket
that can be mounted at a convenient location remote from the booster itself.
GB-72 and GB-152 are not a good choice for supplying pressure to a PPCK+ high gas pressure
controller/calibrator. A different model, GB-K-75 and GB-K-175, with features specifically intended to support
PPCK+ is available for this application.

GB-75™ AND GB-152™ OPERATION AND MAINTENANCE MANUAL
©1998-2008 DH Instruments, a Fluke Company Page 2
1.2 LOCATION AND DESCRIPTION OF THE COMPONENTS
1.2.1 GAS BOOSTER
1. High Pressure Instrument Gas Filter
2. High Pressure Instrument Gas Supply Port
3. Drive Air Supply Port (from Drive Air Control Kit)
4. Breather Port
5. Low Pressure Piston Body (50 % shorter in GB-75)
6. Muffler
7. High Pressure Piston Body
8. High Pressure Output Port
Figure 1. GB-152 Gas Booster

1. INTRODUCTION
Page 3 ©1998-2008 DH Instruments, a Fluke Company
1.2.2 DRIVE AIR CONTROL KIT
1. Drive Air Supply Connection (not visible)
2. Drive Pressure Adjust Regulator
3. Drive Pressure Indication Gauge
4. Drive Air Shut Off Valve
5. Drive Air Out Connection, to Booster (not visible)
Figure 2. Drive Air Control Kit

GB-75™ AND GB-152™ OPERATION AND MAINTENANCE MANUAL
©1998-2008 DH Instruments, a Fluke Company Page 4
Figure 3. System Schematic
1. Drive Air Regulator 5. High Pressure Outlet Port
2. Drive Air Gauge 6. Booster Pump
3. Drive Air Shutoff Valve 7. 1/4 in. PFA Tubing
4. High Pressure Instrument
Gas Supply Port 8. Shop Drive Air Inlet Port

1. INTRODUCTION
Page 5 ©1998-2008 DH Instruments, a Fluke Company
1.3 SPECIFICATIONS
Dimensions: GB 75 Booster:
GB 152 Booster:
432 mm H x 241 mm W x 241 mm D (17 in. x 9.5 in. x 9.5 in.)
584 mm H x 241 mm W x 241 mm D (23 in. x 9.5 in. x 9.5 in.)
Weight: GB 75 Booster:
GB 152 Booster:
Drive Air Control Kit:
13.4 kg (29.4 lbs)
16.4 kg (36 lbs)
1.8 kg (4 lbs)
Pressure Supply
Pressure Range: Shop Drive Air:
High Pressure Supply Gas:
0.15 to 1 MPa (20 to 150 psi)
Maximum high pressure output is drive air pressure
times 75 (GB-75) or 152 (GB-152)
2 to 20 MPa (300 to 3 000 psi)
Maximum high pressure output is high pressure supply
times 25
Supply Flow
Rates: Shop Drive Air:
High Pressure Supply Gas:
425 to 2 125 slm (15 to 75 scfm)
140 to 560 slm (5 to 20 scfm)
Pressure
Connections: Drive Air Supply Inlet:
High Pressure Gas
Supply Inlet:
High Pressure Outlet:
1/4 in. NPT female
1/4 in. NPT female
DH 500 female (DH500 F is a gland and collar type fitting for
coned and left hand threaded 1/4 in. (6.35 mm) OD tube,
equivalent to AE F250C, HIP HF4, etc.)
High Pressure
Output: GB 75 Booster:
GB 152 Booster:
Drive air setting times 75, to 70 MPa maximum
Drive air setting times 152, to 110 MPa maximum
Booster Piston
Ratio: GB 75 Booster:
GB 152 Booster:
75:1
152:1
Due to a policy of continuous product improvement, all specifications are subject to change
without notice.

GB-75™ AND GB-152™ OPERATION AND MAINTENANCE MANUAL
©1998-2008 DH Instruments, a Fluke Company Page 6
N
NO
OT
TE
ES
S

2. INSTALLATION
Page 7 ©1998-2008 DH Instruments, a Fluke Company
2
2.
.
I
IN
NS
ST
TA
AL
LL
LA
AT
TI
IO
ON
N
2.1 UNPACKING AND INSPECTION
GB-75 and GB-152 are delivered enclosed in plastic film and secured by foam in place in a
corrugated box. The Drive Air Control Kit and Interconnections Kit are included in the same box.
Remove all parts from the shipping box and plastic bag. Be sure not to lose or discard the Drive Air
Control Kit and Interconnections Kit.
Inspect all parts for damage. If damage is noted, report it to your Receiving Department for
appropriate action.
Inspect for any missing components or accessories using Table 1. Should any items be missing, contact
DHI or you local supplier.
Table 1. GB-75 or GB-152 Parts List
DESCRIPTION PART #
GB-75 Gas Booster 401002
GB-152 Gas Booster 400509
GB-75 Gas Booster, Hydrocarbon Free 401816
GB-152 Gas Booster, Hydrocarbon Free 401815
Accessories including:
1 ea. Operation and Maintenance Manual 550038
1 ea. Drive Air Control Kit 400998
Interconnections Kit: 401817
2 m (6 ft) 1/4 in. PFA tubing 101450-Z
1 ea. Tube weld, 1.5 m (60 in.) x 1/8 in. (3 mm) with DH500 tips (coiled) 402506

GB-75™ AND GB-152™ OPERATION AND MAINTENANCE MANUAL
©1998-2008 DH Instruments, a Fluke Company Page 8
2.2 SITE REQUIREMENTS
2.2.1 SITE
There are no special site requirements.
The GB gas booster system is divided into two parts: the gas booster itself and the Drive Air
Control Kit. The two parts are separated so that they may be installed separately.
The gas booster, which is large and relatively noisy, does not need to be accessed regularly.
It is generally mounted out of the way, for example behind and under a work bench.
The booster may be set horizontally or vertically.
Connections of gas supplies and output need to be considered carefully when selecting the
installation site.
See Section 2.3.1 for additional recommendations.
2.2.2 GAS SUPPLIES
Two sources of compressed gas are required to operate the gas booster. They are drive air
(see Section 2.2.2.1) and high pressure instrument gas (see Section 2.2.2.2).
2.2.2.1 DRIVE AIR SUPPLY
The drive air supply provides power to operate the booster. The booster’s very
high pressure output is the drive air multiplied by either 75 (BG-75) or
152 (GB-152). This supply is usually “shop” or “factory” air.
Drive air requirements are:
Flow rate: 425 slm (15 scfm) minimum
Cleanliness: Not critical, use 60 micron filter
Humidity: 20 to 50 % RH. Do not use dry gas.
Approximate drive air pressures required to reach different high pressure
outputs are:
MODEL
BOSTER DRIVE AIR
[kPa (psi)] OUTPUT PRESSURE
[MPa (psi)]
GB-75 400 (60) 25 (3 500)
GB-75 700 (100
)
50 (7 500)
GB-75 1 000 (150
)
70 (10 000)
GB-152 450 (60) 50 (7 500)
GB-152 550 (80) 70 (10 000)
GB-152 900 (130
)
110 (16 000)

2. INSTALLATION
Page 9 ©1998-2008 DH Instruments, a Fluke Company
2.2.2.2 HIGH PRESSURE INSTRUMENT GAS
High pressure instrument gas is boosted by the booster and output from the
booster high pressure OUT port.
High pressure instrument gas supply requirements are:
Flow rate: 140 slm (5 scfm) minimum
Cleanliness: Use clean, dry instrument grade gases only.
Pressure: 2 MPa (300 psig) minimum, 20 MPa (3 000 psig) maximum.
Maximum boosted pressure output is high pressure supply
times 25.
2.3 INSTALLATION AND SETUP
2.3.1 GENERAL CONSIDERATIONS
The GB gas booster system is divided into two parts: the gas booster itself and the Drive Air
Control Kit. The two parts are separate so that they may be installed separately.
The gas booster, which is large and relatively noisy, does not need to be accessed regularly.
It is generally mounted out of the way, for example behind and under a work bench.
The booster may be set horizontally or vertically.
The Drive Air Control Kit allows local control of booster operation. The output pressure of the
booster can be set by adjustment of the drive air pressure and the SHUTOFF valve turns the
booster ON and OFF. The Drive Air Control Kit is presented on a bracket meant to be
mounted at a location convenient to the operator, for example on the front of a work bench.
When selecting the site for installation of the GB gas booster system, carefully consider
access for the connections that need to be made (drive air to the Drive Air Control Kit, high
pressure instrument gas to the booster, very high pressure gas output from the booster to the
test or calibration system).
The orientation of the booster is of no consequence to its operation. It may be installed
vertically, horizontally or any combination of the two with no affect on performance
or maintenance.
A variety of factors must be considered when determining where to locate the gas booster
and the Drive Air Control Kit. Factors include, but are not limited to:
•If control over booster operation by the operator is desired, locate the Drive Air Control
Kit where it can be easily accessed;
•The high pressure being generated and associated safety concerns;
•The source of gas supplies (drive air supply and high pressure instrument
gas supply);
•Point of use of output pressures;
•Noise levels. The booster operation is noisy;
•Vibration during use;
•Access to the unit for operation of the regulators and valve;
•Vents to atmosphere.

GB-75™ AND GB-152™ OPERATION AND MAINTENANCE MANUAL
©1998-2008 DH Instruments, a Fluke Company Page 10
2.3.2 INSTALLATION
The installation of the gas booster system is broken down into four parts:
•Installation of the gas booster (see Section 2.3.2.1)
•Installation and connection of the Drive Air Control Kit (see Section 2.3.2.2)
•Connection of the booster high pressure instrument gas supply (see Section 2.3.2.3)
•Connection of the booster high pressure output (see Section 2.3.2.4)
2.3.2.1 BOOSTER INSTALLATION
To install the gas booster system, follow the steps below:
Place the gas booster in the appropriate location.
Consider the booster high pressure instrument gas supply and high pressure
output connections. The booster may sit vertically or horizontally.
Orientation does not affect operation.
Use the attached mounting brackets to secure the booster to a fixed location
if desired.
Due to the reciprocating nature of the booster, it is advised that shock mounts
be used when rigidly mounting the booster.
2.3.2.2 DRIVE AIR CONTROL KIT INSTALLATION
To install the Drive Air Control Kit, follow the steps below:
nIdentify an appropriate location for the Drive Air Control Kit bracket. If the
booster will regularly be turned ON/OFF and/or frequent adjustment of
booster output pressure is desired, the Drive Air Control Kit Bracket should
be readily accessible. Consider the routing of the PFA tubing connection
from the Control Kit bracket to the booster when selecting a location. The kit
bracket is often mounted on the front bottom surface of a work bench.
oMount the Drive Air Control Kit bracket onto the desired location using the
bracket mounting holes provided
pBack off (rotate counterclockwise) the DRIVE ADJUST regulator and put
the SHUTOFF valve in the OFF position.
qConnect the drive air supply to the 1/4 in. NPT F connection on the DRIVE
ADJUST regulator. Use tubing rated for at least 1 MPa (150 psig)
working pressure. See Section 2.2.2.1 for information on drive air
supply requirements.
Maximum input pressure to the DRIVE ADJUST regulator is 1 MPa (150 psig).
Pressures above this level may result in a failure that could damage the instrument
and/or cause personal injury.
rConnect the Drive Air Supply Kit SHUTOFF valve output to the booster
DRIVE AIR INPUT port using the 2 m (6 ft.) length of 1/4 in. PFA tubing
provided in the GB accessories. Both fittings are 1/4 in. swage.

2. INSTALLATION
Page 11 ©1998-2008 DH Instruments, a Fluke Company
2.3.2.3 BOOSTER HIGH PRESSURE INSTRUMENT GAS
SUPPPLY CONNECTION
To connect the booster high pressure instrument gas supply, follow the
steps below:
nConnect a high pressure gas supply to the gas booster high pressure
IN port. The IN port is a 1/4 in. NPT F fitting on the high pressure
gas filter. Use a thread sealer to seal the connection. The high pressure
gas supply should not exceed 200 MPa (3 000 psi). See Section 2.2.2.2 for
additional information on high pressure gas supply requirements.
Pressure present at the High Pressure IN port will be present at the high
pressure OUT port. Do not supply pressure to the high pressure IN port unless you
are prepared for there to be pressure at the high pressure OUT port.
2.3.2.4 BOOSTER HIGH PRESSURE OUTPUT CONNECTION
The booster high pressure OUT port is usually connected to the high pressure
inlet of a test or calibration system to which high pressure is to be supplied, such
as the DHI GPC1 Gas Pressure Controller. The booster high pressure OUT port
is connected to the SUPPLY port of the GPC1 (see the GPC1 Operation and
Maintenance Manual).
To connect the booster high pressure OUT port, proceed as follows:
-Connect the 1.5 m, 1/8 in. nipple with DH500 tips between the HIGH
PRESSURE OUT port and the GPC1 SUPPLY port. The DH500 glands and
collars are delivered installed in the female ports. Remove the orange plastic
plug, slip the gland onto the DH500 tip, install the collar onto the DH500 tip and
thread the gland into the female DH500 fitting. Torque gland to 15 Nm (11 lb ft).
Do not bend tubing with radius less than 50 mm (2 in.) or pressure capacity
may be reduced.
Depending on the drive pressure set by the DRIVE ADJUST regulator, GB-75 may
generate pressure over 70 MPa (10 000 psi) and GB-152 may generate pressure over
110 MPa (16 000 psi). Be sure to use fittings and tubing rated for the pressure that
is to be output from the booster. Always use a DH500 or equivalent tube on the
DH500 F high pressure OUT port connection. To limit the maximum pressure that
can be generated by the booster, limit the drive air supply to the Drive Air Control Kit
(see Section 2.2.2.1).

GB-75™ AND GB-152™ OPERATION AND MAINTENANCE MANUAL
©1998-2008 DH Instruments, a Fluke Company Page 12
2.4 INITIAL START UP
After installing the gas booster (see Section 2.3.2), perform the initial start up as follows:
nSupply drive air: Check that the Drive Air Control Kit SHUTOFF valve is in the OFF position and
that the DRIVE ADJUST regulator is set to zero (fully counterclockwise). Apply drive pressure to the
SHUTOFF valve (see Section 2.2.2.1 for information on drive air supply requirements).
oSupply high pressure instrument gas: Check that all connections to the high pressure IN and OUT
ports are secure. Apply high pressure gas to the high pressure IN port (see Section 2.2.2.2 for
information on high pressure instrument gas).
Ensure the high pressure supply does not exceed 20 MPa (3 000 psi). Pressures above this range may
result in a failure that could damage the instrument and/or cause personal injury.
pAdjust the DRIVE ADJUST regulator: Rotate the DRIVE ADJUST regulator knob (clockwise)
while observing the set pressure on the DRIVE gauge. Set the drive pressure to the maximum
desired high pressure output divided by the booster ratio (75 for GB-75, 152 for GB-152).
For example, if the desired high pressure output of a GB-152 is 70 MPa (10 000 psi), set the drive air
pressure to 70 MPa/152 = 460 kPa (67 psi). Remember that when the SHUTOFF valve is in the ON
position, the booster will operate until the high pressure output is equal to the drive air setting times
the booster ratio.
The DRIVE ADJUST regulator is NOT self venting so it cannot be used to set pressure lower than the
current pressure when the SHUTOFF valve is OFF. If you overshoot the desired set point, set the SHUTOFF
valve to ON until the DRIVE pressure is lower than your desired setting. Then set the SHUTOFF valve to OFF
and use the regulator to adjust the pressure in the ascending direction.
qGenerate high pressure: Set the SHUTOFF valve to the ON position. The booster will begin to cycle
and will continue to cycle until it stalls. When it stalls, the high pressure output is roughly equal to the
drive pressure setting times the booster's ratio (75 or 152).

3. OPERATION
Page 13 ©1998-2008 DH Instruments, a Fluke Company
3
3.
.
O
OP
PE
ER
RA
AT
TI
IO
ON
N
3.1 GENERAL OPERATING PRINCIPLE AND INFORMATION
Refer to Figure 3.
The purpose of the GB booster is to automatically boost gas pressure to very high pressure.
The value of very high pressure that is generated depends on the drive air pressure setting. The unit
generates high pressure using a Pneumatically Operated Gas Booster Pump and a Drive Air
Control Kit.
GAS BOOSTER PUMP
The booster pump is a Pascal press utilizing two pistons connected together on the same axis having a
normal area ratio of 75:1 (GB-75) or 152:1 (GB-152). The booster is a two-stroke, single stage
reciprocating pump that generates gas pressures 75 or 152 times greater than the shop drive air applied
to the pump. The pump operates automatically, provided drive air is supplied to the
DRIVE ADJUST regulator and the regulator is set to at least 140 kPa (20 psig). Operation is continuous
until the outlet pressure is 75 (GB-75) or 152 (GB-152) times the shop drive air pressure.
At this point, the opposing forces within the pump reach equilibrium and the pump stalls. For example,
using a GB-75 with the drive air pressure set to 500 kPa (73 psig), the pump will run until the high
pressure output reaches 75 x 0.5 = 37.5 MPa (5 450 psig).
1. The Shop Drive Air section (low pressure) consists of a piston, cylinder, air cycling valve, pilot valve
and vent section. This section provides the reciprocating action and compression force needed to
operate the booster and generate the high pressure gas. Drive air is channeled to the appropriate
side of the piston (compression or suction stroke) by the air cycling valve. When the piston reaches
full stroke, a pilot valve is mechanically activated causing the air cycling valve to change position.
Shop drive air is routed to the opposite side of the piston reversing piston direction where a second
pilot valve is activated repeating the process.
2. The high pressure section (high pressure) consists of a small piston and an inlet/outlet check valve
assembly. The small piston moves forward and backward with the air drive piston. During the
suction stroke (backward movement), the outlet check valve closes and the inlet check valve opens
letting supply high pressure gas enter the compression chamber. During the compression stroke, the
inlet check valve closes and the outlet check valve opens letting boosted gas out of the pump.
The compression ratio of the high pressure piston is 25:1 (for both models). Maximum output
pressure is limited by the instrument gas supply pressure. For example, with 2 MPa (300 psi)
applied, maximum output pressure cannot exceed 2 x 25 = 50 MPa (7 500 psi).
3.2 OPERATION
Once the gas booster has been set up and the initial start up is completed, operation is very simple.
There are two possible operator actions:
•Setting the booster output pressure (see Section 3.2.1).
•Turning the booster ON and OFF (see Section 3.2.2).

GB-75™ AND GB-152™ OPERATION AND MAINTENANCE MANUAL
©1998-2008 DH Instruments, a Fluke Company Page 14
3.2.1 SETTING THE BOOSTER OUTPUT PRESSURE
The booster output pressure is determined by the value of the drive air pressure. The
approximate booster output will be the drive air pressure times 75 (GB-75) or 152 (GB-152).
To set the drive air pressure, adjust the DRIVE ADJUST regulator:
nWith the SHUTOFF valve in the ON position, observing the DRIVE gauge, adjust the
regulator counterclockwise to decrease pressure to a value lower than the desired
set pressure. This is necessary because the regulator is not self-venting.
oPut the SHUTOFF valve in the OFF position.
pRotate the DRIVE ADJUST regulator knob clockwise while observing the set pressure
on the DRIVE gauge. Set the drive pressure to the maximum desired high pressure
output divided by the booster ratio (75 for GB-75, 152 for GB-152). For example, if the
desired high pressure output of a GB-152 is 70 MPa (10 000 psi), set the drive air
pressure to 70 MPa/152 = 460 kPa (67 psi). Remember that when the SHUTOFF valve
is in the ON position, the booster will operate until the high pressure output is equal to the
drive air setting times the booster ratio.
To correctly set the pressure, gas flow must not occur. If flow is present in the circuit when
the regulator is adjusted, the pressure will increase when flow is reduced. If the setpoint is
exceeded, see Section 3.2.2.
The booster will run automatically, continuously until the high pressure output is equal to
the drive pressure times 75 (GB-75) or 152 (GB-152). Set the drive pressure very carefully to
avoid generating higher output pressures than desired which could be dangerous.
3.2.2 TURNING THE BOOSTER ON AND OFF
The booster is turned ON and OFF using the SHUTOFF valve of the Drive Air Control Kit.
The SHUTOFF valve controls the supply of drive pressure to the booster.
Set the SHUTOFF valve to the ON position to supply drive pressure to the booster and
cause it to operate.
Set the SHUTOFF valve to the OFF position to interrupt drive pressure to the booster and
cause it not to operate.
When the booster is OFF, the high pressure supply is still present at the high pressure gas
OUT port. This feature makes it easy to operate using the high pressure supply only when the
booster is not needed.
Emergency shut-down of the gas booster pump can be performed at any time by closing the
Drive Air Control Kit SHUTOFF Valve. This will prevent further generation of gas pressure by the
pump but WILL NOT NECESSARILY reduce pressure to high pressure point-of-use.
This manual suits for next models
1
Table of contents
Other DH Instruments Extender manuals
Popular Extender manuals by other brands

Lindy
Lindy DVI-D Dual Link Extender 500m 4x LC 50/125um Instruction

KING
KING KX1000 user manual

Extron electronics
Extron electronics FOX T Plus MM user guide

Grandbeing
Grandbeing EX0101-U56 operating instructions

Abtus
Abtus CAT-HDX11T/AP2 User's operation guide

Nikrans
Nikrans LCD-300GW installation guide