DH Instruments PPC3 User manual

© 2002-2007 DH Instruments, a Fluke Company
PPC3™
Pressure Controller/Calibrator
Operation and Maintenance Manual

© 2002-2007 DH Instruments, a Fluke Company
High pressure liquids and gases are potentially hazardous. Energy stored in these liquids and gases
can be released unexpectedly and with extreme force. High pressure systems should be assembled
and operated only by personnel who have been instructed in proper safety practices.
This instrument is not to be operated in any other manner than that specified by the manufacturer.
© 2002 - 2007 DH Instruments, a Fluke Company All rights reserved.
Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted in any
form or by any means, electronic or mechanical, for any purpose, without the express written permission of DH Instruments, a
Fluke Company 4765 East Beautiful Lane, Phoenix AZ 85044-5318, USA.
DH Instruments makes sincere efforts to ensure accuracy and quality of its’ published materials; however, no warranty, expressed
or implied, is provided. DH Instruments disclaims any responsibility or liability for any direct or indirect damages resulting from the
use of the information in this manual or products described in it. Mention of any product or brand does not constitute an
endorsement by DH Instruments of that product or brand. This manual was originally composed in English and was subsequently
translated into other languages. The fidelity of the translation cannot be guaranteed. In case of conflict between the English version
and other language versions, the English version predominates.
Products described in this manual are manufactured under international patents and one or more of the following U.S. patents:
5,142,483; 5,257,640; 5,331,838; 5,445,035. Other U.S. and international patents pending.
AutoRange, AutoZ, DH Instruments,DH,DHI, COMPASS, PPC, PPC3, QDUT, Q-RPT, RPM, RPM4, SDS and SPLT are
trademarks, registered and otherwise, of DH Instruments, a Fluke Company.
Document No. 550128b-02
050815
Printed in the USA.

TABLE OF CONTENTS
Page I © 2002-2007 DH Instruments, a Fluke Company
T
TA
AB
BL
LE
E
O
OF
F
C
CO
ON
NT
TE
EN
NT
TS
S
TABLE OF CONTENTS...............................................................I
TABLES..................................................................................V
FIGURES................................................................................VI
ABOUT THIS MANUAL ............................................................VII
1. INTRODUCTION .................................................................1
1.1 PRODUCT OVERVIEW ...........................................................................................................................1
1.2 SPECIFICATIONS ...................................................................................................................................1
1.2.1 GENERAL SPECIFICATIONS.......................................................................................................................1
1.2.2 PRESSURE MEASUREMENT SPECIFICATIONS........................................................................................2
1.2.2.1 QUARTZ REFERENCE PRESSURE TRANSDUCER (Q-RPT)................................................................2
1.2.2.2 UTILITY SENSOR......................................................................................................................................4
1.2.2.3 ON-BOARD BAROMETER ........................................................................................................................4
1.2.3 PRESSURE CONTROL SPECIFICATIONS..................................................................................................4
2. INSTALLATION ..................................................................5
2.1 UNPACKING AND INSPECTION ............................................................................................................5
2.1.1 REMOVING FROM PACKAGING..................................................................................................................5
2.1.2 INSPECTING CONTENTS.............................................................................................................................5
2.2 SITE REQUIREMENTS............................................................................................................................6
2.3 SETUP .....................................................................................................................................................6
2.3.1 PREPARING FOR OPERATION ...................................................................................................................6
2.3.2 FRONT AND REAR PANELS........................................................................................................................7
2.3.2.1 FRONT PANEL ..........................................................................................................................................7
2.3.2.2 REAR PANEL.............................................................................................................................................7
2.3.3 POWER CONNECTION.................................................................................................................................8
2.3.4 REMOTE [ENTER/SET P] CONNECTION (FOOTSWITCH OR OTHER SWITCH) .....................................8
2.3.5 CONNECTING TO A PRESSURE SUPPLY (SUPPLY PORT).....................................................................8
2.3.6 CONNECTING A VACUUM PUMP (EXHAUST PORT) ................................................................................8
2.3.7 CONNECTING EXTERNAL Q-RPTS IN RPM4 REFERENCE PRESSURE MONITORS.............................9
2.3.8 CONNECTING TO THE DEVICE UNDER TEST (TEST(+) AND TEST(-) PORTS)....................................10
2.3.8.1 INSTALLING A SELF PURGING LIQUID TRAP (SPLT) .........................................................................11
2.3.8.2 INSTALLING A DUAL VOLUME UNIT (DVU), G15K AND BG15K Q-RPTS...........................................12
2.3.9 THE VENT PORT.........................................................................................................................................12
2.3.10 CHECK/SET SECURITY LEVEL.................................................................................................................12
2.3.11 TURN OFF ABSOLUTE AND NEGATIVE GAUGE MODE (AXXX RPT)....................................................12
2.4 POWER-UP AND VERIFICATION.........................................................................................................13
2.4.1 SWITCH POWER ON ..................................................................................................................................13
2.4.2 CHECK PRESSURE MEASUREMENT OPERATION.................................................................................13
2.4.2.1 CHECKING ABSOLUTE MODE PRESSURE MEASUREMENT.............................................................13
2.4.2.2 CHECKING GAUGE MODE PRESSURE MEASUREMENT...................................................................13
2.4.3 LEAK CHECK..............................................................................................................................................14
2.4.4 PURGE.........................................................................................................................................................14
2.4.5 CHECK PRESSURE CONTROL OPERATION...........................................................................................14
2.5 SHORT TERM STORAGE.....................................................................................................................14
3. OPERATION..................................................................... 15
3.1 USER INTERFACE................................................................................................................................15
3.1.1 MAIN RUN SCREEN....................................................................................................................................15
3.1.2 FUNCTION / DATA KEYPAD LAYOUT AND PROTOCOL.........................................................................17
3.1.3 DIRECT PRESSURE CONTROL KEYS......................................................................................................18

PPC3™ OPERATION AND MAINTENANCE MANUAL
© 2002-2007 DH Instruments, a Fluke Company Page II
3.1.4 REMOTE [ENT/SET P] FOOTSWITCH.......................................................................................................18
3.1.5 SOUNDS......................................................................................................................................................19
3.2 GENERAL OPERATING PRINCIPLES .................................................................................................19
3.2.1 AUTOMATED PRESSURE CONTROL.......................................................................................................19
3.2.2 PRESSURE READY/NOT READY..............................................................................................................21
3.2.3 GAUGE AND NEGATIVE GAUGE MODES WITH AN AXXX (ABSOLUTE) Q-RPT, DYNAMIC
COMPENSATION FOR ATMOSPHERIC PRESSURE ...............................................................................22
3.2.4 MULTIPLE INTERNAL AND EXTERNAL Q-RPTS.....................................................................................23
3.2.5 MULTIPE RANGES (Q-RPTS, AUTORANGE AND INFINITE RANGING).................................................24
3.2.6 OPERATION WITH A PG7000 PISTON GAUGE........................................................................................25
3.2.7 DIRECT FUNCTION KEYS SUMMARY......................................................................................................25
3.3 DIRECT FUNCTION KEYS....................................................................................................................26
3.3.1 [RANGE] ......................................................................................................................................................26
3.3.2 [UNIT]...........................................................................................................................................................28
3.3.3 [MODE] ........................................................................................................................................................29
3.3.4 [AUTORANGE] ............................................................................................................................................30
3.3.5 [RPT]............................................................................................................................................................33
3.3.6 [AUTOTEST]................................................................................................................................................34
3.3.6.1 QUICK AUTOTEST..................................................................................................................................35
3.3.6.2 QDUT AUTOTEST...................................................................................................................................38
3.3.7 [HEAD].........................................................................................................................................................42
3.3.8 [PURGE] ......................................................................................................................................................43
3.3.9 [LEAK CK] ...................................................................................................................................................45
3.3.10 [ENT/SET P] (SET PRESSURE AUTOMATICALLY)..................................................................................47
3.3.10.1 INTERRUPTING AUTOMATED PRESSURE CONTROL........................................................................48
3.3.10.2 AUTOMATED PRESSURE COMMANDS FOR ZERO PRESSURE....................................................48
3.4 [SETUP].................................................................................................................................................49
3.4.1 <1RANGE> ..................................................................................................................................................49
3.4.1.1 SAVING AN AUTORANGE RANGE ........................................................................................................50
3.4.1.2 DELETING AUTORANGE RANGES........................................................................................................50
3.4.2 <2RES> (RESOLUTION).............................................................................................................................50
3.4.3 <3JOG>........................................................................................................................................................51
3.4.4 <4UL> (UPPER LIMIT) ................................................................................................................................52
3.4.4.1 OVER PRESSURE FUNCTION...............................................................................................................53
3.4.5 <5ATEST>....................................................................................................................................................54
3.4.6 <6CONTROL>..............................................................................................................................................55
3.4.6.1 <LIMITS> (CUSTOM CONTROL PARAMETERS) ..................................................................................56
3.4.6.2 TURNING-OFF CUSTOM CONTROL PARAMETERS............................................................................57
3.4.7 <7DRV> (DRIVERS) ....................................................................................................................................57
3.5 [SPECIAL] .............................................................................................................................................58
3.5.1 <1AUTOZ>...................................................................................................................................................58
3.5.1.1 EDIT AUTOZ............................................................................................................................................62
3.5.1.2 RUN AUTOZ.............................................................................................................................................62
3.5.2 <2REMOTE>................................................................................................................................................63
3.5.2.1 <1COM1> AND <2COM2>.......................................................................................................................64
3.5.2.2 <3IEEE-488>............................................................................................................................................64
3.5.2.3 <4FORMAT>............................................................................................................................................64
3.5.2.4 <5RS232 SELF-TEST>............................................................................................................................65
3.5.3 <3HEAD> .....................................................................................................................................................65
3.5.4 <4RESET> ...................................................................................................................................................66
3.5.4.1 <1SET> ....................................................................................................................................................66
3.5.4.2 <2UNITS...................................................................................................................................................67
3.5.4.3 <3ATEST>................................................................................................................................................67
3.5.4.4 <4CAL> ....................................................................................................................................................67
3.5.4.5 <5ALL>.....................................................................................................................................................68
3.5.5 <5PREFS> ...................................................................................................................................................68
3.5.5.1 <1SCRSVR> ............................................................................................................................................69
3.5.5.2 <2SOUND> ..............................................................................................................................................69
3.5.5.3 <3TIME>...................................................................................................................................................69
3.5.5.4 <4ID>........................................................................................................................................................70
3.5.5.5 <5LEVEL> (SECURITY)...........................................................................................................................70
3.5.6 <6PUNIT>.....................................................................................................................................................73
3.5.7 <7INTERNAL> .............................................................................................................................................75
3.5.7.1 <1CONFIG> .............................................................................................................................................75
3.5.7.2 <2CONTROL REF>..................................................................................................................................75
3.5.7.3 <3BARO>.................................................................................................................................................76
3.5.7.4 <4PURGE> ..............................................................................................................................................77
3.5.7.5 <5LO VNT> ..............................................................................................................................................77
3.5.8 <8CAL>........................................................................................................................................................79
3.5.9 <9LOG>........................................................................................................................................................79

TABLE OF CONTENTS
Page III © 2002-2007 DH Instruments, a Fluke Company
4. REMOTE OPERATION ....................................................... 81
4.1 OVERVIEW............................................................................................................................................81
4.2 INTERFACING.......................................................................................................................................81
4.2.1 RS232 INTERFACE.....................................................................................................................................81
4.2.1.1 COM1.......................................................................................................................................................81
4.2.1.2 IEEE-488..................................................................................................................................................82
4.2.1.3 COM2.......................................................................................................................................................82
4.3 PROGRAMMING FORMATS.................................................................................................................82
4.3.1 CLASSIC PROGRAM MESSAGE FORMAT...............................................................................................83
4.3.2 ENHANCED PROGRAM MESSAGE FORMAT ..........................................................................................83
4.3.2.1 USING COMMAND TYPE COMMANDS .................................................................................................83
4.3.2.2 USING QUERY TYPE COMMANDS...................................................................................................84
4.4 COMMANDS..........................................................................................................................................85
4.4.1 PROGRAMMING MESSAGES....................................................................................................................85
4.4.2 ERROR MESSAGES ...................................................................................................................................86
4.4.3 PROGRAM MESSAGE DESCRIPTION OVERVIEW..................................................................................87
4.4.4 PROGRAM MESSAGE DESCRIPTIONS....................................................................................................88
4.5 STATUS REPORTING SYSTEM.........................................................................................................114
4.5.1 ERROR QUEUE.........................................................................................................................................114
4.5.2 STATUS BYTE REGISTER .......................................................................................................................114
4.5.3 STANDARD EVENT REGISTER...............................................................................................................115
4.5.4 READY STATUS REGISTER ....................................................................................................................116
4.6 IEEE STD. 488.2 COMMON AND STATUS PROGRAM MESSAGES................................................117
4.6.1 PROGRAM MESSAGE DESCRIPTIONS..................................................................................................117
5. MAINTENANCE, ADJUSTMENTS AND CALIBRATION...........121
5.1 OVERVIEW..........................................................................................................................................121
5.2 CALIBRATION OF QUARTZ REFERENCE PRESSURE TRANSDUCERS (Q-RPTS) ......................121
5.2.1 PRINCIPLE ................................................................................................................................................121
5.2.1.1 PA AND PM COEFFICIENTS ................................................................................................................122
5.2.1.2 AS RECEIVED AND AS LEFT DATA ....................................................................................................122
5.2.2 EQUIPMENT REQUIRED..........................................................................................................................123
5.2.3 SET-UP AND PREPARATION...................................................................................................................124
5.2.4 RECOMMENDED CALIBRATION POINT SEQUENCE............................................................................124
5.2.4.1 STANDARD CLASS Q-RPTS ................................................................................................................125
5.2.4.2 PREMIUM CLASS Q-RPTS...................................................................................................................126
5.2.5 TURNING OFF ABSOLUTE AND NEGATIVE GAUGE MEASUREMENT MODES FOR AXXX
(ABSOLUTE) Q-RPTS .........................................................................................................................127
5.2.6 Q-RPT CALIBRATION USING CALTOOL FOR RPTS SOFTWARE........................................................128
5.2.7 EDITING AND VIEWING Q-RPT CALIBRATION INFORMATION............................................................128
5.2.8 Q-RPT CALIBRATION/ADJUSTMENT WITHOUT CALTOOL FOR RPTS SOFTWARE.........................130
5.3 ADJUSTMENT OF ON-BOARD BAROMETER...................................................................................131
5.4 ADJUSTMENT OF UTILITY SENSOR ................................................................................................131
5.5 PNEUMATIC CONTROL MODULE CONFIGURATION (<CONFIG>) ................................................132
5.6 OVERHAUL .........................................................................................................................................133
5.7 RELOADING EMBEDDED SOFTWARE INTO FLASH MEMORY......................................................134
5.8 SUBASSEMBLY DESCRIPTION AND LOCATION ............................................................................135
5.8.1 POWER SUPPLY MODULE......................................................................................................................135
5.8.2 MINI MICRO BOARD.................................................................................................................................135
5.8.3 Q-RPT MODULE........................................................................................................................................136
5.8.3.1 HI Q-RPT OR UTILITY SENSOR...........................................................................................................136
5.8.3.2 LO Q-RPT...............................................................................................................................................136
5.8.4 ON-BOARD BAROMETER........................................................................................................................136
5.8.5 VACUUM SENSOR....................................................................................................................................136
5.8.6 PRESSURE CONTROL MODULE ............................................................................................................137
5.8.7 DISPLAY....................................................................................................................................................137
5.8.8 DRIVER BOARD........................................................................................................................................137
5.9 PNEUMATIC SCHEMATICS................................................................................................................137
5.9.1 PRESSURE CONTROL MODULE ............................................................................................................137
5.9.2 Q-RPT MODULE CONFIGURATIONS (PRESSURE MEASUREMENT)..................................................138

PPC3™ OPERATION AND MAINTENANCE MANUAL
© 2002-2007 DH Instruments, a Fluke Company Page IV
6. TROUBLESHOOTING .......................................................141
7. APPENDIX ......................................................................145
7.1 DRIVERS .............................................................................................................................................145
7.2 UNIT CONVERSION............................................................................................................................146
7.2.1 PRESSURE................................................................................................................................................146
7.3 REMOTE [ENT] ...................................................................................................................................146
8. WARRANTY ....................................................................147
8.1 OVERVIEW..........................................................................................................................................147
9. GLOSSARY.....................................................................149

TABLES & FIGURES
Page V © 2002-2007 DH Instruments, a Fluke Company
T
TA
AB
BL
LE
ES
S
Table 1. Reference Pressure Transducer (Q-RPT) Designations and Ranges..........................................2
Table 2. PPC3 Packing List .........................................................................................................................5
Table 3. Position Designators of Q-RPTs in a PPC3 System ...................................................................24
Table 4. Settings and What They Are Specific To (Range, Measurement Mode, Q-RPT, System).........25
Table 5. Summary of PPC3 Function Key Operation................................................................................26
Table 6. Settings Made by AutoRange......................................................................................................31
Table 7. Default Pressure Control Parameters..........................................................................................55
Table 8. AutoZ ON and OFF......................................................................................................................60
Table 9. Reset – Sets.................................................................................................................................67
Table 10. Reset – Cal ................................................................................................................................68
Table 11. Reset – All..................................................................................................................................68
Table 12. Security Levels...........................................................................................................................72
Table 13. UNIT Function - Available Units of Measure .............................................................................74
Table 14. COM1 Pin Designations and Connections................................................................................81
Table 15. COM2 DB-9F Pin Designations.................................................................................................82
Table 16. Program Message List...............................................................................................................85
Table 17. Error #s and Descriptions ..........................................................................................................86
Table 18. 8 Bit Status Byte Register........................................................................................................114
Table 19. 8 Bit Standard Event Register..................................................................................................115
Table 20. 8 Bit Ready Status Register.....................................................................................................116
Table 21. Program Message List.............................................................................................................117
Table 22. Calibration Point Sequence, Standard Class, Axxx and Gxxx Q-RPTs..................................125
Table 23. Calibration Point Sequence, Standard Class, BGxxx Q-RPTs................................................125
Table 24. Calibration Point Sequence, Standard Class, BA100K Q-RPT...............................................126
Table 25. Calibration Point Sequence, Premium Class, Axxx and Gxxx Q-RPTs ..................................126
Table 26. Calibration Point Sequence, Premium Class, BGxxx Q-RPTs................................................127
Table 27. Troubleshooting Guide.............................................................................................................141
Table 28. External Drivers Current Output ..............................................................................................145
Table 29. External Drivers Pin Outs.........................................................................................................145
Table 30. Pressure Unit of Measure Conversion Coefficients.................................................................146
Table 31. DHI Authorized Service Providers...........................................................................................147

PPC3™ OPERATION AND MAINTENANCE MANUAL
© 2002-2007 DH Instruments, a Fluke Company Page VI
F
FI
IG
GU
UR
RE
ES
S
Figure 1. Front Panel ...................................................................................................................................7
Figure 2. Rear Panel....................................................................................................................................7
Figure 3. MAIN RUN Screen Display Fields..............................................................................................16
Figure 4. Keypad Layout............................................................................................................................17
Figure 5. Direct Pressure Control Keys .....................................................................................................18
Figure 6. Dynamic Pressure Control Operation.........................................................................................20
Figure 7. Static Pressure Control Operation..............................................................................................20
Figure 8. Ready/Not Ready in Dynamic Pressure Control Mode..............................................................21
Figure 9. Ready/Not Ready in Static Control Mode...................................................................................22
Figure 10. Status Register Schematic .....................................................................................................114
Figure 11. Internal View...........................................................................................................................135
Figure 12. Pressure Control Module Schematic......................................................................................137
Figure 13. Q-RPT Module Schematics for Various Configurations.........................................................139
Figure 14. Drivers Connector Schematic.................................................................................................145
Figure 15. Remote [ENT/SET P] Connector Schematic..........................................................................146

ABOUT THIS MANUAL
Page VII © 2002-2007 DH Instruments, a Fluke Company
A
AB
BO
OU
UT
T
T
TH
HI
IS
S
M
MA
AN
NU
UA
AL
L
This manual is intended to provide the user with the basic information necessary to operate a PPC3
pressure controller/calibrator. It also includes a great deal of additional information provided to allow you
to optimize PPC3 use and take full advantage of its many features and functions.
Before using the manual, take a moment to familiarize yourself with the Table of Contents structure:
Sections 1, 2 and 3 should be read by all first time PPC3 users. Section 3 is most important for those
using the local front panel interface but should be read over by all users to familiarize themselves with
general PPC3 operating principles. Section 4 is for remote operation from an external computer. Section 5
provides maintenance and calibration information. Section 6 is a quick troubleshooting guide. Use it to
troubleshoot unexpected PPC3 behavior based on the symptom of that behavior. Certain words and
expressions have specific meaning as they pertain to PPC3. The Glossary, Section 6, is useful as a
quick reference for exact definition of specific words and expressions as they are used in the manual.
For those of you who “don’t read manuals”, go directly to Section 2.3 to set up your PPC3 and then go
to Section 2.4 for power-up and verification. This will get you up and running quickly with a minimal
risk of causing damage to yourself or your new PPC3. THEN… when you have questions or start to
wonder about all the great features you might be missing, get into the manual!
Manual Conventions
(CAUTION) is used in throughout the manual to identify user warnings and cautions.
(NOTE) is used throughout the manual to identify operating and applications advice and
additional explanations.
[ ] indicates direct function keys (e.g., [RANGE]).
< > indicates PG7000 screen displays (e.g., <1yes>).

PPC3™ OPERATION AND MAINTENANCE MANUAL
© 2002-2007 DH Instruments, a Fluke Company Page VIII
N
NO
OT
TE
ES
S

1. INTRODUCTION
Page 1 © 2002-2007 DH Instruments, a Fluke Company
1
1.
.
I
IN
NT
TR
RO
OD
DU
UC
CT
TI
IO
ON
N
1.1 PRODUCT OVERVIEW
PPC3 is a stand-alone, pressure controller intended to precisely set and control gas pressure into a
closed volume as is commonly needed for the calibration and testing of pressure measuring instruments.
It has been designed to provide very high performance combined with maximum versatility and ease of use.
PPC3 can be equipped with a low cost utility sensor for pressure monitoring or one or two Quartz Reference
Pressure Transducers (Q-RPTs) to allow it to set and measure pressure with very low measurement
uncertainty. Up to four external Q-RPTs in one or two external Reference Pressure Monitors (RPM4s),
can also be integrated into a PPC3 system. In some cases, a barometer is also included.
Pressure control is achieved by a patented pneumatic module based on digitally controlled solenoid
valves and differential pressure regulators.
PPC3 is controlled locally by the operator using a front panel display, keypad and function keys or
remotely by a computer using ASCII character command strings over an RS232 or IEEE-488.2 interface.
PPC3 models are available to measure and control pressure in ranges from as low as - 3 to 3 kPa (0.4 psi)
to as high as 0 to 10 MPa (0 to 1 500 psi) in absolute, gauge and compound gauge measurement modes.
1.2 SPECIFICATIONS
1.2.1 GENERAL SPECIFICATIONS
Power Requirements 85 to 264 VAC, 50/60 Hz, 30 VA max consumption
Operating Temperature Range 15 to 35 °C
Storage Temperature Range - 20 to 70 °C
Vibration Meets MIL-T-28800D
Weight 12.7 kg (28.2 lb) approx
Dimensions
Ventilation 18 cm H x 32 cm W x 40 cm D (7.1 in. x 12.6 in. x 15.8 in.)
To prevent product overheating, provide proper ventilation. Allow 10 cm
(4 in.) clearance from rear panel cooling fan.
Microprocessors Motorola 68302, 16 MHz
Communication Ports
Fuses RS232 (COM1), RS232 (COM2), IEEE-488.2
1 A, 250 VAC fuse, 5 x 20 mm, time lag type fuse
Internal power supply fuse not replaceable by operator: 2A, 250 V (UV 440-2
power supply), 3.15A, 250 V (NFS40-7612 power supply)
Pressure Ranges Six controller models from 200 kPa (30 psi) max to 10 MPa (1 500 psi) max.
Low uncertainty measurement provided by selection of quartz reference
pressure transducer(s) (Q-RPTs)
Operating Medium Any clean, dry, non-corrosive gas
Pressure Connections SUPPLY: 1/8 in. NPT F
TEST(+): 1/8 in. NPT F
TEST(-): 1/8 in. NPT F
ATM (Vent): 10-32 UNF
EXHAUST: 1/4 in. NPT F
Pressure Limits Maximum Working:Pressure: Controller or Hi Q-RPT maximum
Maximum Test Pressure:w/out Damage: 110 % controller or Hi Q-RPT
maximum
Recommended Supply:Pressure: Maximum control pressure + 10 %
Maximum Supply Pressure:w/out Damage: PPC3-10M: 14 MPa (2 000 psi)
PPC3 -7M: 9.6 MPa (1 400 psi)
PPC3-2M, -700K, -200K: 3.5 MPa (500 psi)

PPC3™ OPERATION AND MAINTENANCE MANUAL
© 2002-2007 DH Instruments, a Fluke Company Page 2
1.2.2 PRESSURE MEASUREMENT SPECIFICATIONS
1.2.2.1 QUARTZ REFERENCE PRESSURE TRANSDUCER (Q-RPT)
Quartz reference pressure transducers (Q-RPTs) can be installed in PPC3 to obtain
low uncertainty pressure measurement. One or two Q-RPTs can be included in the
PPC3 and/or additional Q-RPTs can be used externally mounted in DHI RPM4
Reference Pressure Monitors (see Section 3.2.4).
The type (Axxx, Gxxx, BGxxx, BAxxx) and range of the Q-RPT module(s)
determines the PPC3 measurement specifications.
All Q-RPTs whose maximum pressure is over 200 kPa (30 psi) are of the absolute
pressure type (Axxx) using an evacuated, permanently sealed reference. Axxx Q-RPTs
can measure absolute, gauge and negative gauge pressure. Gauge pressure with
an Axxx (absolute) Q-RPT is obtained by offsetting atmospheric pressure and
applying dynamic compensation for atmospheric changes using the on-board
barometer (see Section 3.2.3). Gxxx (gauge) Q-RPTs can measure positive gauge
pressure only. BGxxx (bi-directional gauge) Q-RPTs can measure gauge and
negative gauge pressure. See Section 3.3.3, PRINCIPLE, for additional information on
absolute,gauge and negative gauge measurement modes.
PPC3s configured with two Q-RPT modules have only one TEST(+) and TEST(-) port.
PPC3 internal valves and logic handle switching between the two Q-RPTs as needed.
Q-RPTs are available with two different performance levels, STANDARD class
and PREMIUM class. See the product label on the PPC3 rear panel, the label
on the Q-RPT module on the rear panel and/or the product calibration
reports to determine the class of the Q-RPTs installed in PPC3.
Table 1. Reference Pressure Transducer (Q-RPT) Designations and Ranges
SI VERSION US VERSION
RPT
DESIGNATION MAXIMUM
PRESSURE
Absolute
[kPa]
MAXIMUM
PRESSURE
Gauge
[kPa]
MAXIMUM
PRESSURE
Absolute
[psi]
MAXIMUM
PRESSURE
Gauge
[psi]
A10M110 000 10 000 1 500 1 500
A7M17 000 7 000 1 000 1 000
A3.5M13 500 3 500 500 500
A2M12 000 2 000 300 300
A1.4M11 400 1 400 200 200
A700K1700 700 100 100
A350K1350 250 50 35
A200K1200 100 30 15
A160K1 160 60 23 8
A100K1110 10 16 1.5
BA100K4110 - - 16 - -
G200K2- - 200 - - 30
G100K2- - 100 - - 15
G15K2- - 15 - - 2.2
BG15K3- - ±15 - - ±2.2
1. All AXXXX RPTs support absolute, gauge and compound (negative) gauge modes.
2. All GXXXX RPTs are positive gauge mode only.
3. BG15K is bi-directional gauge from - 15 to + 15 kPa (- 2.2 to + 2.2 psi).
4. BA100K is a barometric range whose low point is 70 kPa absolute (10 psi).

1. INTRODUCTION
Page 3 © 2002-2007 DH Instruments, a Fluke Company
Warm Up Time None required, 30 minute temperature stabilization
recommended for best performance from cold power up.
Resolution To 1 ppm, user adjustable
Compensated Temperature Range 5 to 35 °C
Acceleration Affect
Allows operation at ± 20° from
significant effect
Predicted One Year Stability1
± 0.008 % /g maximum, worst axis
Allows operation at ± 20° from reference plane without
significant effect
± 0.005% of reading
STANDARD CLASS PREMIUM CLASS
Precision2± 0.008% of reading or
0.0024% of Q-RPT span,
whichever is greater5
± 0.005% of reading, 0.0015%
of AutoRanged span, or
0.0005% of Q-RPT span,
whichever is greater6
Measurement Uncertainty3± 0.01% of reading or 0.0030%
of Q-RPT span, whichever is
greater5
± 0.008 % of reading, 0.0024%
of AutoRanged span, or
0.0007% of Q-RPT span,
whichever is greater6
Delivered Pressure Uncertainty4± 0.011 % of reading or
0.0033% of Q-RPT span,
whichever is greater5
± 0.009 % of reading,
0.0027% of AutoRanged span,
or 0.0008% of Q-RPT span,
whichever is greater6
1. Predicted Q-RPT measurement stability limit (k=2) over one year assuming regular use of AutoZero function. AutoZero
occurs automatically in gauge mode whenever vented, by comparison with barometric reference in absolute mode.
Absolute mode predicted one year stability without AutoZ is ± (0.005 % Q-RPT span + 0.005 % of reading).
2. Combined linearity, hysteresis, repeatability. Add + 1 Pa (0.00015 psi) in gauge mode with an Axxx (absolute) Q-RPT for
the resolution and short term stability of the on-board barometer.
3. Maximum deviation of the Q-RPT indication from the true value of applied pressure including precision, predicted one year
stability limit, temperature effect and calibration uncertainty, combined and expanded (k=2) following the ISO “Guide to the
Expression of Uncertainty in Measurement.”
4. Maximum deviation of the PPC3 controlled pressure from the true value including measurement uncertainty and dynamic
control hold limit.
5. % of reading value times measured pressure from 100 to 30 % of Q-RPT span. Under 30 % of Q-RPT span, % of reading
value times 30 % of Q-RPT span. For example, if the Q-RPT is a Standard A160K, the Measurement Uncertainty in
pressure is 0.010% times the measured pressure to 48 kPa (160 kPa span x 30%) and 0.0048 kPa (160 kPa span x 30%
x 0.01%) under 48 kPa.
6. % of reading value times measured pressure from 100 to 30 % of AutoRanged span. Under 30% of AutoRanged span, %
of reading value times 30% of AutoRanged span. If AutoRanged span is less then 30% of maximum Q-RPT span, % of
reading values times measured pressure, or % of reading times 9% of Q-RPT span, whichever is greater. For example, if
the Q-RPT is a Premium A160K and AutoRanged span is 160 kPa, the Measurement Uncertainty in pressure is measured
pressure x 0.008% to 48 kPa (160 kPa AutoRanged span x 30%) and 0.0038 kPa (160 kPa span x 30% x 0.008%) under
48 kPa. If the AutoRanged span is 100 kPa (greater than 30% of 160 kPa maximum Q-RPT span), the measurement
uncertainty in pressure is measured pressure x 0.008% to 30 kPa (100 kPa AutoRanged span x 30%) and 0.0024 kPa
(100 kPa span x 30% x 0.008%) under 30 kPa. If the AutoRanged span is 30 kPa (less than 30% of the 160 kPa
maximum Q-RPT span), the measurement uncertainty in pressure is measured pressure x 0.008% to 14.4 kPa (160 kPa
maximum Q-RPT span x 9%) and 0.0012 kPa (160 kPa maximum Q-RPT span x 9% x 0.008%) under 14.4 kPa.

PPC3™ OPERATION AND MAINTENANCE MANUAL
© 2002-2007 DH Instruments, a Fluke Company Page 4
1.2.2.2 UTILITY SENSOR
Base PPC3 pressure controllers include a utility sensor. If a Hi Q-RPT
(see Section 3.2.4) is installed, there is no utility sensor.
Utility sensors are used for pressure control, system monitoring and safety functions.
They are intended for indication only, not to provide reference pressure
measurement. Q-RPTs (see Section 1.2.2.1) are used for reference measurement.
Warm Up Time None required
Range PPC3-10M: 10 MPa (1 500 psi) absolute and gauge
PPC3-7M: 7 MPa (1 000 psi) absolute and gauge
PPC3-2M: 2 MPa (300 psi) absolute and gauge
PPC3-700K: 700 kPa (100 psi) absolute and gauge
PPC3-200K: 300 kPa (45 psi) absolute;
200 kPa (30 psi) gauge
Resolution 0.001 % of span
Precision 0.1 % of span
1.2.2.3 ON-BOARD BAROMETER
The on-board barometer is used only to measure changes in atmospheric
pressure to provide dynamic compensation of the Q-RPT’s atmospheric pressure
offset when using an Axxx (absolute) Q-RPT to make gauge pressure
measurements.
Warm Up Time None required
Range 70 to 110 kPa (10 to 16 psi)
Resolution 0.001 % of span
Precision
Predicted Stability
0.1 % of span
0.05 kPa/yr (0.008 psi/yr)
The on-board barometer is only present in PPC3s that have an Axxx
(absolute) or BGxxx (bi-directional gauge) internal Q-RPT.
1.2.3 PRESSURE CONTROL SPECIFICATIONS
PPC3-200K PPC3-700K PPC3-2M PPC3-7M PPC3-10M
Control Precision ± 0.002 % of reading or of 2 % of controller span, whichever is greater.*
Q-RPT used must provide resolution higher than expected
control precision.
Lowest Controllable Pressure
(gauge mode) Zero set by automated venting. Lowest point above or below zero
limited only by Q-RPT resolution and control precision.
Lowest Controllable Pressure
(absolute, negative gauge modes) 1.5 kPa
(0.2 psia) 3 kPa
(0.4 psia) 5 kPa
(0.75 psia) 5 kPa
(0.75 psia) 10 kPa
(1.5 psia)
Ultimate Pressure
(absolute, negative gauge)
TEST(+) port is isolated and connected to vacuum
supply. Typically < 50 Pa (0.008 psia) depending on
vacuum source and test volume configuration.
5 kPa
(0.75 psia)
Typical Pressure Setting ReadyTime 15 to 30 seconds.
Optimum TEST Volume 500 cc 500 cc 500 cc 250 cc 250 cc
Default Dynamic Control Hold Limit ± 0.005 % of current range
* Control precision is reduced in absolute mode under 7 kPa (1 psi) absolute.

2. INSTALLATION
Page 5 © 2002-2007 DH Instruments, a Fluke Company
2
2.
.
I
IN
NS
ST
TA
AL
LL
LA
AT
TI
IO
ON
N
2.1 UNPACKING AND INSPECTION
2.1.1 REMOVING FROM PACKAGING
PPC3 is delivered in a corrugated container with high density polyethylene inserts to hold it in
place; or in the optional molded, medium density polyethylene shipping case with a custom
foam insert.
Remove the PPC3 and its accessories from the shipping container and remove each element
from its protective plastic bag.
2.1.2 INSPECTING CONTENTS
Check that all items are present and have no visible damage.
A standard PPC3 includes all items indicated in Table 2.
Table 2. PPC3 Packing List
DESCRIPTION PART #
1 ea. PPC3 Pressure Controller/Calibrator FAM007
1 ea. Calibration Report 550100
1 ea. Test Report 550200
ACCESSORIES: 401918 (401918-CE)
1 ea. Operation and Maintenance Manual 550128
1 ea. Power Cord (7.5 ft.) 100770 (100770-CE)
1 set (6) Rubber Feet Caps 400203
1 ea. Drivers Connector 401382
1 ea. General Accessories Disk (white CD) 102987

PPC3™ OPERATION AND MAINTENANCE MANUAL
© 2002-2007 DH Instruments, a Fluke Company Page 6
2.2 SITE REQUIREMENTS
The PPC3 can be installed on any flat, stable surface at a convenient height. The front feet can be
extended so that the unit can be inclined slightly for easier viewing. The PPC3 can also be mounted in a
standard 19 in. rack using the optional rack mount kit.
Minimizing the distance between the PPC3 and the device or system under test will enhance control
performance and reduce pressure setting times.
Ready access to the PPC3 rear panel should be considered to facilitate making and breaking
pressure connections.
Pneumatic and RS-232 connections to RPM4s should be considered if RPM4s are to be used as
external reference pressure measurement devices (see Section 2.3.7).
The Self Purging Liquid Trap (SPLT), if used, should be mounted vertically at the low point of the
connection between the PPC3 TEST(+) port and the test (see Section 2.3.8.1).
If you are using a G15K or BG15K Q-RPT with a Dual Volume Unit (DVU), its location and
connections should be considered (see Section 2.3.8.2).
Support facilities required include:
•An electrical power source of 85 to 264 VAC, 50 - 60 Hz.
•A continuous, regulated pressure supply of clean, dry, non-corrosive gas at PPC3 maximum
control pressure + 10 % (at least 70 kPa (10 psi) in the case of a BG15K Q-RPT) to be connected to
the PPC3 SUPPLY port. Lower gas pressure supply can be used but should exceed the maximum
desired test output pressure by 10 to 20 %.
•A vacuum source of less than 1 psia (7 kPa) and with displacement of at least 3 cfm (90 lm)
if control of pressures under 3 psi (20 kPa) gauge is desired.
2.3 SETUP
2.3.1 PREPARING FOR OPERATION
To prepare PPC3 for check out and operation:
Remove the plastic caps from the PPC3 rear panel pressure connections.
Remove the protective plastic sheet from the front panel display.
Install the rubber feet caps onto the bottom case feet, if desired.
Familiarize yourself briefly with the front and rear panel (see Section 2.3.2).
Then proceed with Sections 2.3.3 to 2.3.11

2. INSTALLATION
Page 7 © 2002-2007 DH Instruments, a Fluke Company
2.3.2 FRONT AND REAR PANELS
2.3.2.1 FRONT PANEL
1. Ready/Not Ready Indicator
2. Display
3. Remote Activity Indicator
4. Cursor Control Keys
5. Multi-Function Keypad
6. Direct Pressure Control Keys
Figure 1. Front Panel
2.3.2.2 REAR PANEL
1. Label, Product
2. VENT port
3. COM2 Connector
4. IEEE-488 Connector
5. COM1 Connector
6. Drivers (12 V) Connector
7. Power Switch
8. Fuse
9. Electrical Power Connector (IEC-320-C13)
10. Remote [ENT] Connector
11. Pressure Connection, TEST(+)
12. Pressure Connection, TEST(-)
13. Pressure Connection, EXHAUST
14. Pressure Connection, SUPPLY
Figure 2. Rear Panel

PPC3™ OPERATION AND MAINTENANCE MANUAL
© 2002-2007 DH Instruments, a Fluke Company Page 8
2.3.3 POWER CONNECTION
Check that the PPC3 power switch is OFF.
Connect the supplied power cable to the rear panel power module.
Connect the other end of the power cable to an electrical supply of 85 to 264 VAC, 50/60 Hz.
2.3.4 REMOTE [ENTER/SET P] CONNECTION (FOOTSWITCH
OR OTHER SWITCH)
Connect the optional remote ENTER footswitch, if available or a user supplied switch fitted to
the optional cable (see Section 7.3). Connect the cable to the PPC3 rear panel connection
labeled REMOTE ENTER. Activating the switch is equivalent to pressing the [ENT/SET P]
key on the front panel (see Section 3.1.4).
2.3.5 CONNECTING TO A PRESSURE SUPPLY (SUPPLY
PORT)
Using a pressure connecting hose or tube of appropriate pressure rating, connect the
pressure supply to the SUPPLY port on the rear panel of PPC3. The PPC3 SUPPLY port
connection is 1/8 in. NPT female.
The supply pressure should be equal to the maximum PPC3 control pressure + 10 % (or at
least 70 kPa (10 psi) for a BG15K Q-RPT). Lower gas pressure sources can be used but
should exceed the maximum desired test output pressure by 10 to 20 %.
Never connect a pressure supply greater than 20 % over the maximum control pressure of
the PPC3 model you are using (except models with a BG15K Q-RPT only). Be sure to
connect the pressure supply to the SUPPLY port. Connecting to another port is likely to
damage PPC3.
2.3.6 CONNECTING A VACUUM PUMP (EXHAUST PORT)
For PPC3 to set pressures under atmosphere and/or to reliably set pressure under 20 kPa (3 psi)
gauge (other than zero gauge), a vacuum supply must be connected to the EXHAUST port.
Never connect a pressure supply to or plug the PPC3 EXHAUST port.
To avoid building up pressure on the EXHAUST port or on a vacuum pump connected to
the EXHAUST port, the vacuum source should either be continuously ON or the EXHAUST
port should be bypassed to atmosphere when the vacuum source is OFF. This is because
when a supply pressure is applied to the PPC3 SUPPLY port and the PPC3 is NOT in the
vent ON condition, there may be a constant gas exhaust through the PPC3 EXHAUST port.
To assure optimum pressure control when changing the pressure applied to the EXHAUST
port from vacuum to atmosphere or vice-versa, be sure to change the control reference
setting if the setting is NOT in AUTO mode (see Section 3.5.7.2).

2. INSTALLATION
Page 9 © 2002-2007 DH Instruments, a Fluke Company
2.3.7 CONNECTING EXTERNAL Q-RPTS IN RPM4 REFERENCE
PRESSURE MONITORS
PPC3 can be connected pneumatically and by RS-232 to one or two RPM4 reference
pressure monitors to use up to four external Q-RPTs (two in each RPM4) as external
reference pressure measurement devices (see Section 3.2.4).
The PPC3 then manages communications and other RPM4 functions to integrate the RPM4
measurement capabilities into the PPC3 system. The RPM4 Q-RPT module’s Self Defense
System (SDS) is used to shut-off RPM4 Q-RPTs from the pressurized system when they are
not in use (see the RPM4 Operation and Maintenance Manual).
As a general rule, making the pneumatic connection between the PPC3 TEST port and the
remote Q-RPT as direct as possible favors good pressure control. As distance, volumes
and restrictions between the PPC3 TEST port and the remote Q-RPT are added, the
possibility of difficulty with pressure control when using the external Q-RPT increases.
To connect a PPC3 to an RPM4 to be used as part of the PPC3 system proceed as follows:
Set up the RPM4 for use as an external device to PPC3 following the instructions in
the RPM4 Operation and Maintenance Manual, Using RPM4 With a PPC3
Controller/Calibration Section.
Using tubing of appropriate pressure rating and a Tee, connect the RPM4 Q-RPT
TEST(+) port to the PPC3 TEST(+) port. The third leg of the Tee is for the connection to
the device or system under test. If there are two RPM4s, Tee the second RPM4 in series
with the first one. The order of the RPM4s on the test line is not relevant.
If an RPM4 Q-RPT will be used in a gauge pressure range of less than 100 kPa (15 psi),
consider connecting the RPM4 Q-RPT module TEST(-) port to the PPC3 TEST(-) port and
to the “low” connection of the DUT if available. This will improve pressure control and
stability when ambient pressure is unstable. If the Q-RPT is an Axxx designation Q-RPT,
this connection should be made but also be left open to atmosphere.
When external Q-RPTs are used with PPC3, the maximum set pressure is the maximum
pressure of the PPC3 controller/calibrator. The maximum pressure of the PPC3 is
determined by the PPC3 controller model or the Hi Q-RPT if one is installed (see Section 1.2.1).
Using a standard pin-to-pin DB-9M to DB-9F RS232 cable, connect COM2 of the PPC3
to COM1 of the RPM4 (see Section 4.2.1.3). If there are two RPM4s, connect COM2 of the
first RPM4 to COM1 of the second RPM4. The order of the RPM4 daisy chain is not
relevant.
Turn ON the PPC3 and RPM4(s).
Press [RPT] on the PPC3 key pad. This causes PPC3 to execute the Q-RPT search
function (see Section 3.3.5). Q-RPTs with whom communication is established will be
shown on the PPC3 display identified by their position (see Table 3).

PPC3™ OPERATION AND MAINTENANCE MANUAL
© 2002-2007 DH Instruments, a Fluke Company Page 10
For PPC3 to identify external RPM4s, the RPM4 COM1 port settings must be:
Baud rate: between 1200 and 19200
Parity: Even
Date bits: 7
Stop bits: 1
If PPC3 is not able to establish communications with RPM4s and their Q-RPTs, check
that the RPM4 COM1 port setting conform to the requirements above. If the COM1 port
settings are correct, check that the correct communications cable is being used
(standard pin-to-pin DB-9M to DB-9F RS232) and is connected to the correct
communications ports.
PPC3 COM2 > 1st RPM4 COM1; 1st RPM4 COM2 > 2nd RPM4 COM1
See the RPM4 Operation and Maintenance Manual for additional information on RPM4
RS232 communications and COM port settings.
2.3.8 CONNECTING TO THE DEVICE UNDER TEST (TEST(+)
AND TEST(-) PORTS)
If you are using a self purging liquid trap (SPLT), see Section 2.3.8.1 before proceeding to
connect the device under test.
If the PPC3 has a G15K or BG15K Q-RPT, a dual volume unit (DVU) should be installed for
very low pressure control. See Section 2.3.8.2 before proceeding to connect to the device
under test.
Using a pressure connecting hose or tube of appropriate pressure rating, connect the device or
system to be tested to the PPC3 TEST(+) port. The PPC3 TEST(+) connection is 1/8 in. NPT female.
PPC3 TEST(+) AND TEST(-) PORTS
All PPC3s have a TEST(+) and a TEST(-) port. See Figure 13 for PPC3 internal Q-RPT
TEST port configurations.
The TEST(+) port is connected to Axxx (absolute) Q-RPTs and to the high side of Gxxx or
BGxxx (gauge, bi-directional gauge) Q-RPTs.
The TEST(-) port is connected to PPC3s internal barometer, if present, and to the low side of
Gxxx or BGxxx (gauge, bi-directional gauge) Q-RPTs.
•When operating in absolute mode: The TEST(-) port is left open to atmosphere.
•When operating in gauge or negative gauge mode with a range greater than 50 kPa
(7.5 psi): The TEST(-) port is normally left open to atmosphere. A possible exception is
when the device or system under test is in an ambient pressure that may differ significantly
from the ambient pressure around the PPC3. For example, if the PPC3 is controlling
pressure into DUTs in an environmental chamber, the pressure in the environmental
chamber may be different from ambient pressure around the PPC3. In this case,
connecting a tube from the TEST(-) port to the inside of the chamber may improve
measurement results. If the Q-RPT in use is an Axxx Q-RPT, this tube must be left open to
the environment so that the pressure inside cannot deviate too far from ambient. When using
an Axxx Q-RPT, if this tube is connected to the low or reference side of DUTs, be sure to
open to the local environment as well.
•When operating in gauge or negative gauge mode with a range less than 50 kPa (7.5 psi):
As a general rule, it is preferable to connect the PPC3 TEST(-) port(s) directly to the low or
reference side of the device under test to assure that these are at the same pressure.
When using an Axxx Q-RPT, this connection must also be open to atmosphere. When using
a Gxxx or BGxxx Q-RPT, it is preferable that this connection not be open to atmosphere.
Table of contents
Other DH Instruments Test Equipment manuals
Popular Test Equipment manuals by other brands

OfiTE
OfiTE 4020-SG instruction manual

Agilent Technologies
Agilent Technologies 81250 installation guide

Teledyne
Teledyne T3EL15030 P Series user manual

Steinberg Systems
Steinberg Systems SBS-SMD-2999 user manual

Micronta
Micronta 22-204C instruction manual

Campbell
Campbell RWIS STATIONS instruction manual