Troubleshooting
Measurement
Quick Reference Guide
Maintenance
• Open front cover and check if all electrodes and the right side sample
sensor have O-rings.
• Make sure maintenance has been performed in accordance with
the maintenance schedule.
• Check the FLUID PACK level by going to menu [PACK STATUS]
• Check for clots and crystals that may be forming in the tubing
or in the measuring chamber.
• Pull pump windings from pump roller and check for kinked tubing
and tears in the tubing.
• Start a calibration pressing [CALIBRATION]
LOAD SAMPLE FROM
TUBE/SYRINGE/CUP
4
WIPE PROBE
CLOSE SAMPLE DOOR
5
3
WAIT FOR PROMPT
“LOAD SAMPLE”
• Perform cleaning cycle at end of the day
• Perform conditioning cycle at end of the day
• Check printer paper supply
• Clean sample probe and fill port
• Clean exterior analyzer surfaces
• Clean the reference electrode housing
(see chapter 5 of Operator Manual)
• Replace peristaltic pump tubing
• Replace main tubing harness
Daily
Weekly
Monthly
Semi-annually
Annually
Always wear gloves when
handling blood or parts
potentially contaminated
with blood.
Keep the main door
closed during
measurement.
1
PRESS
SAMPLE TYPE
2
OPEN
SAMPLE DOOR
ENSURE INSTRUMENT
IS CALIBRATED
ENTER PATIENT ID
(OPTIONAL)
6
RESULTS DISPLAYED
& PRINTED
7