Segmented Calibration
Segmented Calibration allows you to calibration between any two temperatures along the full range of the probe,
providing a more accurate reading between those temperatures. For example, placing all six (6) calibration points
between 0.0 °C and 600 °C will yield more accurate temperature readings between those temperatures than the
full-range calibration would.
WARNING: Measurements outside the calibrated segment may be notably skewed.
Default Calibration Points
Default calibration points come preloaded in the controller (also see Table 2). These points have been selected
and tested to provide accurate temperature readings across the full range. While these calibration points can be
adjusted for segmented calibration, it is not suggested that they be changed for full-range calibration due to nonlin-
earity of temperature probes.
SENSOR
TYPE
DEFAULT CALIBRATION POINTS (°C)
P1 P2 P3 P4 P5 P6
THERMOCOUPLES
Type J -185 -85 0230 660 995
Type K -195 -95 0230 660 1367
Type N -195 -95 0230 660 1295
Type R 050 250 650 1350 1750
Type S 0100 250 650 1120 1750
Type T -195 -95 0195 290 395
Type B 600 800 1000 1200 1500 1795
Type E -195 -95 0260 660 950
PLATINUM RTD
0.003850 -200 -150 -100 0400 800
0.003916 -200 -150 -100 0400 630
THERMISTOR
YSI 400 -30
(39.86KΩ)
-15
(16.43KΩ)
0
(7,355Ω)
20
(2,814Ω)
50
(811.300Ω)
95
(177.100Ω)
YSI 700 -15
(43.78KΩ)
-10
(33.20K )
0
(19.59KΩ)
20
(7,496Ω)
50
(2,162Ω)
95
(472.4Ω)
Table 2. Default Calibraon Points