Dräger Babylog V Series Quick start guide

D-5755-2018
Disassembly/Assembly Neonatal Expiratory Valve
Open the ap by lifting the lower
edge upwards.
Close the ap.
Turn the locking ring to the left
end position.
Turn the locking ring to the right
until it clicks into place.
Remove and empty the water
trap container.
Remove the muer from the
expiratory valve.
Turn the locking ring as far as
possible to the left and push the
expiratory valve into the tting.
Remove the expiratory valve from
the tting.
Remove the diaphragm.
Do not disassemble further.
91 03 108 | 19.11-1 | AW | LE | Subject to modications | © 2019 Drägerwerk AG & Co. KGaA
Attach the diaphragm onto the edge
of the expiratory valve housing.
Make sure that it tts properly.
Attach the muer to the
expiratory valve.
Connect the water trap container.
Dräger Babylog®V-Series
Neonatal Expiratory Valve
Disassembly/Assembly
Disassembly/Assembly Neonatal Expiratory Valve
The muer is only used with the Babylog® V-Series.
6
65
This guideline does not replace or amend the instructions for use. Please always refer to the
reprocessing instruction in the latest instruction for use of the components. The instructions
for use and the hospital guidelines and policies always have to be observed.
D-5755-2018
Disassembly/Assembly Neonatal Flow Sensor
Remove the ow sensor housing
from the ypiece.
Fit the breathing hoses to
the neonatal ow sensor.
Disconnect plug of the ow
sensor cable from the
neonatal ow sensor.
Connect plug of the ow
sensor cable to the
neonatal ow sensor.
Fit the insert into
the ow sensor.
Gently press the knobs on
both sides while removing the
insert out of the neonatal
ow sensor.
Dräger Babylog®V-Series
Neonatal Expiratory Valve
Machine cleaning
Dräger Neonatal Flow Sensor
Disassembly/Assembly
Connect the hose to the nozzle
of the expiratory valve.
Connect the other end of the
hose to a suitable nozzle on the
load carrier.
Machine cleaning of the Neonatal Expiratory Valve
This guideline does not replace or amend the instructions for use. Please always refer to the
reprocessing instruction in the latest instruction for use of the components. The instructions
for use and the hospital guidelines and policies always have to be observed.
Position the components
(muer, diaphragm, water trap
container) to be stable.
Prepare the cleaning agent and the
disinfectant in accordance with the
manufacturer’s instructions.
Place the components in the
cleaning agent.
Ensure that the solution ushes
all surfaces and interior spaces.
Swirl the components back and
forth at least 3 times. Make sure
the cleaning agent reaches all
surfaces and interior spaces.
Fill a syringe that can hold at least 50 mL with the cleaning agent. Remove the
ow sensor housing from the cleaning agent and spray at least 50 mL through
each opening. Put the ow sensor housing back into the cleaning agent.
* The Neonatal Flow Sensor can also be used with Evita® V-Series.
Reprocessing of ISO 15 Neonatal Flow Sensor
and Y-piece Neonatal Flow Sensor*
D-5755-2018
Dräger Babylog® V-Series
Dräger Neonatal Flow Sensor
Reprocessing instructions
Manual cleaning
Sensor housing of the ow sensor Y-piece Sensor housing of the ISO 15 ow sensor
At the end of the specied contact time, swirl the components
back and forth again at least 3 times.
Sensor housing of the ow sensor Y-piece Sensor housing of the ISO 15 ow sensor
This guideline does not replace or amend the instructions for use. Please always refer to the
reprocessing instruction in the latest instruction for use of the components. The instructions
for use and the hospital guidelines and policies always have to be observed.
D-5755-2018
Dräger Babylog®V-Series
Dräger Neonatal Flow Sensor
Reprocessing instructions
Manual cleaning
Fill a syringe that can hold at least 50 mL with the cleaning agent again.
Remove the ow sensor housing from the cleaning agent again and
spray 50 mL through each opening.
* The Neonatal Flow Sensor can also be used with Evita® V-Series.
Reprocessing of ISO 15 Neonatal Flow Sensor
and Y-piece Neonatal Flow Sensor*
Check the components for visible soiling and repeat steps, if necessary.
Check the components for visible damage, paying particular attention to the
measuring wires and their pins, and replace if necessary.
Allow the components to dry completely.
Sensor housing of the ow sensor Y-piece Sensor housing of the ISO 15 ow sensor
After the contact time, rinse the components in the water bath (at least drinking water quality)
until no cleaning agent residue is visible.
Shake out the residual water carefully.
5
6
7
This guideline does not replace or amend the instructions for use. Please always refer to the
reprocessing instruction in the latest instruction for use of the components. The instructions
for use and the hospital guidelines and policies always have to be observed.
D-5755-2018
Dräger Babylog®V-Series
Dräger Neonatal Flow Sensor
Disinfection
Disinfection
At the beginning of the contact time, swirl the components back and forth at least 3 times. Make sure the disinfectant
reaches all surfaces and interiors. At the end of the contact time, swirl the components back and forth again
at least 3 times.
After the contact time, rinse the components in the water bath (at least drinking water quality) until no disinfectant residue is visible.
Place the components in the disinfectant. Observe the specied contact time.
Shake out residual water carefully.
Allow the components to dry com-
pletely.
Manufacturer:
Drägerwerk AG & Co. KGaA
Moislinger Allee 53–55
23558 Lübeck, Germany
This guideline does not replace or amend the instructions for use. Please always refer to the
reprocessing instruction in the latest instruction for use of the components. The instructions
for use and the hospital guidelines and policies always have to be observed.

Other Dräger Laboratory Equipment manuals

Dräger DrugCheck 3000 User manual

Dräger

Dräger DrugCheck 3000 User manual

Dräger DrugTest 5000 User manual

Dräger

Dräger DrugTest 5000 User manual

Dräger SAM 3100 User manual

Dräger

Dräger SAM 3100 User manual

Popular Laboratory Equipment manuals by other brands

Metrohm OMNIS 2.1006.00X0 product manual

Metrohm

Metrohm OMNIS 2.1006.00X0 product manual

Mocom B Futura-17 User's quick start guide

Mocom

Mocom B Futura-17 User's quick start guide

ahn myLab SM-02 instruction manual

ahn

ahn myLab SM-02 instruction manual

SPECS XR 50 user manual

SPECS

SPECS XR 50 user manual

TESTO 882 instruction manual

TESTO

TESTO 882 instruction manual

Labconco FreeZone 70020 Series user manual

Labconco

Labconco FreeZone 70020 Series user manual

Sonation HPLC user manual

Sonation

Sonation HPLC user manual

Fluke Professional Series user manual

Fluke

Fluke Professional Series user manual

Endress+Hauser CPY7 operating instructions

Endress+Hauser

Endress+Hauser CPY7 operating instructions

Agilent Technologies 1100 Series quick start guide

Agilent Technologies

Agilent Technologies 1100 Series quick start guide

Azure Biosystems chemiSOLO user manual

Azure Biosystems

Azure Biosystems chemiSOLO user manual

Witeg WISD MS-20A operating manual

Witeg

Witeg WISD MS-20A operating manual

REITEL SANDURET 2 operating instructions

REITEL

REITEL SANDURET 2 operating instructions

Hach Lange 2100AN Instrument manual

Hach

Hach Lange 2100AN Instrument manual

Huazheng HGZ-500 user manual

Huazheng

Huazheng HGZ-500 user manual

illumina HiSeq X user guide

illumina

illumina HiSeq X user guide

Agilent Technologies 5800 ICP-OES user guide

Agilent Technologies

Agilent Technologies 5800 ICP-OES user guide

illumina MiSeqDx reference guide

illumina

illumina MiSeqDx reference guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.