VistaCool is a patent-pending system designed to eliminate the
need for autoclave condenser/waste bottles and condensation
tanks by cooling the exhaust water and automatically sending it
directly down the drain. With the use of a simple adapter kit, the
VistaCool Model V7502 is designed to enable the Midmark/Ritter
M9 and M11 autoclaves to use fresh deionized/distilled water for
each sterilizing cycle—converting these chambered autoclaves to
operate like quick-turnaround, cassette systems.
Introduction
VistaCool can be used with:
Midmark/Ritter M3
Midmark/Ritter M9*
Midmark/Ritter M11*
SciCan STATIM 2000
SciCan STATIM 5000
We recommend that a professional technician or plumber familiar with dental/medical oces perform the
installation since interface with a cold water line and drain is involved. He/she should be familiar with local
plumbing codes and techniques for successful dental/medical equipment installations. Failure to install
correctly could result in water leaks, leading to possible water damage. It is the responsibility of the system
owner to validate the use of all equipment within its installed environment.
Please keep the Installation Guide & Owner’s Manual handy for future reference and ensure that anyone
responsible for operation and maintenance of the system is familiar with all details contained in this manual.
Please return the Warranty Registration form or register online immediately upon installation.
IMPORTANT INFORMATION
For System Owners
Please read this entire manual before proceeding with installation and operation, and always follow local
plumbing codes.
VistaCool systems have only been verified for installation on the compatible autoclaves (listed above).
Please check with the manufacturer of other brands of autoclaves or contact Vista Research Group before
connecting to other than SciCan or Midmark/Ritter autoclaves. See page 13 for autoclave-specific instructions.
VistaCool Model V7502 has been verified for installation on Midmark/Ritter M9 & M11 units with the
use of the S7560 adapter kit. Only M9, M9D, M11 and M11D (-02X, -03X) units with green displays can
be connected to VistaCool. First generation models with red displays (-0XX) are not compatible.
The VistaCool tank must be installed so that the top of the water tank is ABOVE the connection to the
drain adapter. Please see page 12 for important tank positioning details.
Check for leaks and run the autoclave(s) for at least two consecutive cycles after installation to
ensure the system is operating properly.
IMPORTANT INFORMATION
For System Installers
* Requires Model V7502 and S7560 adapter kit