
9. Pump change, 10. Exchange of receptalce, 11. Use of emergency operation lever
21
Lift the faceshield.
When opening, approach the
facshield always from above
in order to prevent unwanted
pumping action.
After opening, an acoustic
signal sounds.
b
e
e
p
b
e
e
p
b
e
e
p
Reinsertion: Push in the pump
from the front side into the in-
tended guidance until it snaps
in. Reclose the faceshield.
9.1 Change of the
dosage
The dosage of the pump is
adjusted to the maximum
of approx. 1,5ml / stroke. If
required, the dosage can be
reduced in stages by turning
the dosage piece (part of the
pump) up to approx. 0,75 ml
/ stroke.
For this remove the pump
from the dispenser.
Push and hold the pump catch
(“PUSH” button) with the
finger and pull out the pump
towards the front.
10. Exchange of receptacle
• Lift up the empty receptacle from below, draw it to the front and remove it from the dispenser.
• Insert the filled-up receptacle from below into the dispenser. Please pay attention to the correct
position of the receptacle.
• If necessary, remove the locking plate before.
• Please take care that only receptacle with an opening width of max. 45mm is used. Make sure
that the bottle adapter sits tight onto the opening, so that no gas can exhaust from the receptacle.
• In order to avoid uninteded pumping actions, take care that the faceshield is opened.
11. Use of the emergency operating lever
In case of a failure of the automatic funtions (for instance batteries low) an operation by means of
the emergency operating lever is possible.
Operation: For this purpose pull down the lever to the lower limit stop and return it again manually
back up in the initial position.
• Only actuate the emergency operating lever in case of a
failure of the automatic funtion !
• Use only pumps with the silicone bottle adapter! Please also take
care that the bottle adapter sits always tight onto the bottle’s
opening so that no gas can exhaust.
EN