DynasafeECOManagementControlSystem
Configuration Functions
FUNCTION ALARM INPUTS/OUTPUTS
1. “Fire Alarm” AL1 External alarm (volt free input) NO Relay
2. “Spare” AL2 Not Used (volt free input) NO Relay
3. “Remote Alarm” AL3 Used for external alarm (volt free input) NO Relay
4. “Remote Start” AL4 Used for external start (volt free input) NO Relay
5. “Low Level” AL5 Scrubber low level (volt free input) NO Relay
6. “Isolator 1” Emergency Isolator (Front Panel)
7. “Pressure Alarm” Measure duct press N/A) 4 to 20
8. “Isolator 2” AL6 External isolator (volt free input) NO Relay
9. “Low Air Flow” AL7 Low air flow indicator (volt free input) NO Relay
10. “Battery Alarm” Low battery alarm
11. “Mains Alarm” Mains Failure
12. “Pump Configuration” Scrubber pump (240v output)
13. “Syncroflow” Variable Volume Sash Control
14. “Spray Bar” Fume cupboard spray bar (240v output)
15. “GPO Outlets” Fume cupboard GPO’s (Output to Contactor) 240 Volt
16. “Lighting” Fume cupboard light
17. “Aux Relay” Control board relays
18. “Alarm Relay” General alarm relays
19. “Sash Input” Require to calibrate sash(s)
20. “Air Sensor” Variable Hot wire airflow sensor (N/A)
21. Commissioning” Activate commissioning (setup)
22. “Auto Start” Auto start after power failure (Allows Services)
23. “RS485” BMS interface (High Level) N/A
24. “Remote Panel” Additional control panel (D/S Fume Cupboards)
25. “PH Meter” Monitor PH Control (alarms only)
26. “Air Con” Start/Stop output (NO or NC 240v)
27. “Sash Warning” Exceed sash opening (Duel Sash)
28. “Dual Sash” Allows twin sash to operate
29. “Green Sash” Pre-set alarm to close sash
30. “BMS Interface” BMS Interface (Low Level) outputs Volt Free
1. Power Failure
2. Start
3. Low Level
4. Airflow Failure
5. Emergency Stop
31. “Green Lighting” Turn light “OFF” at set Time
32. “Service Meter” Set Next Service Schedule
33. “Pump 1” Waste System only
34. “Pump 2” Waste System only
35. “Valve 1” Waste System only
36. “Valve 2” Waste System only
37. “VSD Direct” Adjust VSD or Control Valve from the Fume Cupboard
38. “FAN Disable” Deactivates “FAN” Button, FC Runs Continually
39. “AUTO SASH” Automatic Sash Movement (Fire Mode)
40. “Services” Auto start after power failure (NO Services)
40 Programmable Functions