Easy Vac PM60 User manual

Model No. PM60
SAVE THESE INSTRUCTIONS
CAUTION
Federal (USA) law restricts this device to sale
by or on the order of a physician.
300 Held Drive Tel: (+001) 610-262-6090
Northampton, PA 18067 USA Fax: (+001) 610-262-6080
ISO 13485 Certified www.precisionmedical.com
USER MANUAL

1
RECEIVING / INSPECTION
Remove the Precision Medical, Inc. Easy Vac Aspirator from packaging and inspect
for damage. If there is any damage, DO NOT USE and contact your Provider.
INTENDED USE
To supply a continuous vacuum source adequate within the stated operating
vacuum range, to aspirate fluids in medical suctioning procedures by trained
healthcare professionals.
READ ALL INSTRUCTIONS BEFORE USING
This manual instructs a Professional to install and operate the Easy Vac Aspirator.
This is provided for your safety and to prevent damage to the Easy Vac Aspirator.
If you do not understand this manual, DO NOT USE the Easy Vac Aspirator and
contact your Provider.
EXPLANATION OF ABBREVIATIONS
inHg Inches of Mercury
mmHg Millimeters of Mercury
SAFETY INFORMATION - WARNINGS AND CAUTIONS
DANGER Indicates an imminently hazardous situation which,
if not avoided, will result in death or serious injury.
WARNING Indicates a potentially hazardous situation which, if
not avoided, could result in death or serious injury.
CAUTION Indicates a potentially hazardous situation which, if
not avoided, may result in minor or moderate injury.
CAUTION Used without the safety alert symbol indicates a
potentially hazardous situation which, if not avoided,
may result in property damage.
CONSULT ACCOMPANYING DOCUMENTS
Type BF equipment
Class II equiprment

2
DANGER
To reduce the risk of Electrocution:
• Always unplug the Easy Vac Aspirator immediately after using.
• DO NOT use while bathing.
• DO NOT place or store Aspirator where it can fall or be pulled into a
tub or sink.
• DO NOT place or drop into water or other liquid.
• DO NOT reach for a Aspirator that has fallen into water. Unplug immediately.
• DO NOT attempt to repair Aspirator. Any attempt could result in
electrical shock.
• DO NOT use near any type of flame or flammable/explosive substances.
• DO NOT allow Aspirator to exceed Operating or Storage Temperature
specifications.
• Caution must be exercised when using this Aspirator where oxygen is
being administered.
• The Easy Vac Aspirator is not suitable for use in the presence of a
flammable anesthetic mixture with air, oxygen or nitrous oxide.
WARNING
To reduce risk of burns, electrocution, fire, or injury to persons:
• The Easy Vac Aspirator should never be left unattended when running.
• Close supervision is necessary when the Aspirator is used by, on, or
near children or invalids.
• Use the Easy Vac Aspirator only for its "Intended Use" as described in
this manual.
• DO NOT use attachments not recommended by the Precision Medical, Inc.
• DO NOT use when sleepy or drowsy.
• DO NOT use if Easy Vac Aspirator has a damaged cord or plug.
• DO NOT use if Easy Vac Aspirator has been dropped or damaged.
• DO NOT cover air vents. Restriction of air flow will cause Aspirator to
overheat, causing the thermal reset to shut down the Aspirator.
• DO NOT drop or insert any object into any opening or hole.
• Keep the cord away from heated surfaces.

3
SPECIFICATIONS
Operating Vacuum Range: 0-25 inHg (0-635 mmHg)
Dimensions:
Length: 13.25 in (33.7 cm)
Width: 8.00 in (20.3 cm)
Height: 9.75 in (24.8 cm)
Weight: 11.4 lbs (5.2 kg)
Shipping Weight: 13.7 lbs (6.2 kg)
Electrical: 108 -132 VAC, 60 Hz, 5 Amps
Operating Temperature Range: 50°Fto104°F(10°Cto40°C)
Storage Temperature Range: -4°Fto104°F(-20°Cto40°C)
Max 95% Noncondensing
Filter: Inline, Hydrophobic
Safety Systems: Automatic Resetting Thermal Limit
Switch
Noise Level: 69db @1 meter
Equipment Classification:
Class Il - with respect to protection from electrical shock
Type BF - degree of protection against electrical shock
IPX0 - degree of protection against ingress of liquids
Mode of Operation - continuous
Specifications are subject to change without prior notice.

4
COMPONENT DESCRIPTION
CAUTION
If for any reason any label becomes illegible or lost, contact
Precision Medical, Inc.
Vacuum
Gauge
Suction Feet
Control
Knob
On/Off
Switch
Patient
Tubing
Connection
Standard 800cc
Suction Canister
with Float Shutoff
Inline
Hydrophobic
Filter
Power Cord
Power Cord
Keeper

5
OPERATING INSTRUCTIONS
CAUTION
Inspect the Easy Vac Aspirator for visual damage, DO NOT USE
if damaged.
1. Place Easy Vac Aspirator near electrical receptacle on a table,
floor or other horizontal, smooth, hard surface.
Allow a minimum of 4 in. (10 cm) clearance on all sides to ensure
adequate air flow for cooling of Aspirator.
2. Press down on Aspirator until Suction Feet are secure.
3. BesuretheOn/OffSwitchisinthe“OFF”position.
4. Be sure the Inline Hydrophobic Filter is properly connected to the
vacuum port on the Suction Canister lid.
• TheInlineHydrophobicFilterpreventsliquidfromentering
the vacuum pump.
CAUTION
DO NOT run the Easy Vac Aspirator without Inline Hydrophobic
Filter. This will void the warranty.
5. Plug the Aspirator into receptacle.
6. TurntheOn/OffSwitchtothe“ON”position.
7. Block the patient connection on the Suction Canister and adjust
the Vacuum Control Knob to the desired vacuum.
To INCREASE vacuum - Turn knob CLOCKWISE
To DECREASE vacuum - Turn knob COUNTERCLOCKWISE
• VacuumshouldbesetasprescribedbyProvider.

6
8. Connect one end of the Patient Suction Tubing to the port on the
Canister marked Patient Tubing.
9. Connect the other end of the Patient Suction Tubing to the
suction catheter and begin procedure as instructed by Provider.
10. When procedure is completed, press the On/Off Switch to the
“OFF”position.
• DONOToverfillthecanister
11. Empty the Suction Canister as follows:
• Removethecanisterfromitsbracketandemptyasdescribed
by your Healthcare Professional or Healthcare Facility
protocol.
CAUTION
Canister is for single patient use only.
12. Replace Suction Canister and reconnect all suctioning tubing.

7
CLEANING / MAINTENANCE
WARNING
Turn the Easy Vac Aspirator off and disconnect from any
external power source before cleaning.
Inline Hydrophobic Filter:
This filter cannot be cleaned. If it becomes contaminated or clogged,
it must be replaced.
To replace Inline Hydrophobic Filter:
1. Disconnect filter assembly from Suction Canister and remove
Suction Canister from the Aspirator.
2. Remove filter assembly from the Aspirator.
3. Dispose of filter.
4. Install new filter (Part #1690) and reconnect tubing and elbow.
Tubing is single patient use only. DO NOT clean.
Cleaning for Exterior Components:
1. Disconnect the Aspirator from power source and other
connections before cleaning.
2. Use a cloth or sponge dampened with soap and water.
3. Wipe dry with a clean cloth or paper towel.
4. Store the Aspirator in a clean area free from grease, oil, and other
sources of contamination.
RETURNS
Returned products require a Returned Goods Authorization (RGA)
number, contact Precision Medical, Inc. All returns must be packaged
in sealed containers to prevent damage. Precision Medical, Inc. will not
beresponsibleforgoodsdamagedintransit.RefertoPrecisionMedical,
Inc. Return Policy available on the Internet; www.precisionmedical.com.
Manuals available on our Website; www.precisionmedical.com

8
DISPOSAL INSTRUCTIONS
Dispose of Bio-waste products as described by your Healthcare
Professional or Healthcare Facility protocol.
Please Recycle
WARNING
If Aspirator is used for treatment of infectious disease, consult
your Physician or Healthcare Professional for recommended
procedures for proper disposal.
REPLACEMENT PARTS
Description Part #
Patient Suction Tubing 1955
Filter, Hydrophobic 1690
Filter Assembly (with tube & elbow) 502012
Elbow Canister Outlet 1616
800cc Canister, (Case-10 ea.) 502519-10

9
TROUBLESHOOTING
If the Easy Vac Aspirator fails to function, consult the Troubleshooting
Guide below. If the problem cannot be corrected, consult your Provider.
Problem Probable Cause Remedy
Aspirator will not
operate
1. On/Off switch in
“OFF”position
2. Power Cord not
connected properly
3. Thermal limit
exceeded
1. Push On/Off switch
to“ON” position
2. Check plug at outlet
3. Shut Aspirator off,
wait 15 minutes; be
sure vents are clear
NO vacuum
on gauge / NO
vacuum at Patient
Connection Tubing
1. Hydrophobic Filter
not connected
properly
2. Suction Canister
not connected or
assembled properly
3. Defective Suction
Canister
4. Defective Regulator
1. Confirm and tighten
connections
2. Confirm and tighten
3. Replace Suction
Canister
4. Contact your
Provider
NO vacuum on
gauge / vacuum
at Patient
Connection Tubing
1. Defective vacuum
gauge 1. Contact your
Provider
Vacuum on gauge
/ NO vacuum at
Patient Suction
Tubing
1. Clogged
Hydrophobic filter
2. Canister overfilled
1. Replace
Hydrophobic filter
2. Empty canister
Unable to adjust
vacuum
1. Defective regulator 1. Contact your
Provider

LIMITED WARRANTY
AND
LIMITATION OF LIABILITY
Precision Medical, Inc. warrants that EASY VAC Aspirator
(the Product) will be free of defects in workmanship and/or material for
the following period:
Five (5) years from shipment
Should any failure to conform to this warranty appear within the applicable
period, Precision Medical, Inc. shall, upon written notification thereof and
substantiation that the goods have been stored, installed, maintained and
operatedinaccordancewithPrecisionMedical,Inc.’sinstructionsand
standard industry practice, and that no modifications, substitutions, or
alterations have been made to the goods, correct such defect by suitable
repair or replacement at its own expense.
ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES.
The representatives of Precision Medical, Inc. or any retailers are not
authorized to make oral warranties about the merchandise described in
this contract, and any such statements shall not be relied upon and are
not part of the contract for sale. Thus, this writing is a final, complete and
exclusive statement of the terms of that contract.
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY WARRANTY
OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR
OTHER WARRANTY OF QUALITY, WHETHER EXPRESS OR IMPLIED.
Precision Medical, Inc. shall not under any circumstances be liable for
special, incidental or consequential damages including but not limited to
lost profits, lost sales, or injury to person or property. Correction of non-
conformities as provided above shall constitute fulfillment of all liabilities
of Precision Medical, Inc. whether based on contract, negligence, strict
tort or otherwise. Precision Medical, Inc. reserves the right to discontinue
manufacture of any product or change product materials, designs, or
specifications without notice.
PrecisionMedical,Inc.reservestherighttocorrectclericalortypographical
errors without penalty.

Precision Medical, Inc.
300 Held Drive
Northampton, PA 18067 USA
Toll Free Tel: 800-272-7285
Toll Free Fax: 800-353-1240
Tel:(+001) 610-262-6090
Fax:(+001) 610-262-6080
1907 Rev10 08/30/12 (?M) Printed in USA
Table of contents
Popular Medical Equipment manuals by other brands

Texinov
Texinov FLUXMEDICARE System user manual

Medizintechnik
Medizintechnik ATMOS S 201 Thorax operating instructions

Cincinnati Sub-Zero Products, LLC
Cincinnati Sub-Zero Products, LLC BLANKETROL III 233 Operation manual

Keeler
Keeler Spectra Iris Instructions for use

ADC
ADC Adscope Sprague Use, care & maintenance

NSK
NSK Varios 750 Operation manual

Teleflex
Teleflex WECK EFx Shield quick start guide

Max-Ability
Max-Ability PRESSALIT R8961000 user guide

NSK
NSK Primado2 Operation manual

VitalGo
VitalGo Total-Lift Bed 425T Assembly, Operating and Maintenance Instruction

natus
natus ICS Impulse quick guide

InTouch Health
InTouch Health InTouch Mini user guide