Eickemeyer OXYVET IV User manual

USER MANUAL
OXYVET IV
OXYGEN CONCENTRATOR
Item no. 213180
Version 1, 04 / 2020
TELEPHONE +49 7461 96 580 0
www.eickemeyer.com

2
USER MANUAL OXYVET IV OXYGEN CONCENTRATOR
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com

3
USER MANUAL OXYVET IV OXYGEN CONCENTRATOR
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
CONTENT
1. Important Safeguards.............................................................................................................. 4
2. Introduction .............................................................................................................................. 5
3. How your Oxyvet IV Oxygen Concentrator works ................................................................ 5
4. Important Parts of your Concentrator.................................................................................... 5
Front View (Figure 1).............................................................................................................................6
Back View (Figure 2) .............................................................................................................................7
5. Setting up your Oxygen Concentrator ................................................................................... 7
6. Before Operating your Oxygen Concentrator........................................................................ 8
7. Operating your OxyVet IV Oxygen Concentrator .................................................................. 9
8. Reserve Oxygen System ........................................................................................................10
9. Caring and periodic Maintenance for your Unit.................................................................. 10
Oxygen Humidifi er (reusable Bottles) ......................................................................................... 10
Cannula, Mask and Tubing................................................................................................................ 10
Air Filter and Oxygen Outlet Connector...................................................................................... 10
Exterior Cabinet................................................................................................................................... 11
Periodic Maintenance for your Unit .............................................................................................. 11
10. Minor Troubleshooting ..........................................................................................................12
11. Specifi cations..........................................................................................................................13
12. Symbol ....................................................................................................................................14
13. Electromagnetic Immunity....................................................................................................15

4
USER MANUAL OXYVET IV OXYGEN CONCENTRATOR
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
Warning!
Read instruction guide before operating this equipment.
Danger!
No smoking.
Indications for Use
The OxyVet IV Oxygen Concentrator is intended for use as an oxygen concentrator to provide supplemental low fl ow oxygen
therapy in the home, nursing homes, patient care facilities, etc.
Warning!
Under certain circumstances, oxygen therapy can be hazardous.
Seeking medical advice before using an oxygen concentrator is advisable.
1. IMPORTANT SAFEGUARDS
Read this entire guide before using your OxyVet IV Oxygen Concentrator.
Important information are highlighted by these terms throughout this guide:
Danger!
Urgent safety information for hazards that will cause serious injury or death.
Warning!
Important safety information for hazards that might cause serious injury.
Caution!
Information for preventing damage to the product.
Note!
Information which you should pay special attention.
Important safeguards are indicated throughout this guide; pay special attention to all safety Information.
Danger!
No smoking.
Save these instructions.

5
USER MANUAL OXYVET IV OXYGEN CONCENTRATOR
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
2. INTRODUCTION
This instruction guide will acquaint you with the OxyVet IV Oxygen Concentrator. Make sure that you read and understand this
guide before operating your unit. Important safeguards are indicated throughout this guide; pay special attention to all safety
information. Contact EICKEMEYER® if you have any questions.
Warning!
For your safety, the oxygen concentrator must be used according to the prescription determined by your physician.
Danger!
Oxygen causes rapid burning. Do not smoke while your oxygen concentrator is operating, or when you are near a
person utilizing oxygen therapy. Do not use within 5 feet (1.6m) of hot, sparking objects or naked sources of fl ame.
3. HOW YOUR OXYVET IV OXYGEN CONCENTRATOR WORKS
Oxygen concentrators are the most reliable, effi cient, and convenient source of supplemental oxygen available today, the
oxygen concentrator is electrically operated. The unit separates oxygen from room air which allows high-purity supplemental
oxygen to be delivered to you through the oxygen outlet, although the concentrator fi lters the oxygen in a room, it will not
affect the normal amount of oxygen in your room.
4. IMPORTANT PARTS OF YOUR CONCENTRATOR
Please take time to familiarize yourself with your OxyVet IV Oxygen Concentrator before operating.
Before using this product, make sure you have the following items:
• 1 User manual
• 1 Oxygen concentrator set
• 1 Adapter – for connect Humidifi er
• 1 Oxygen Outlet Connector

6
USER MANUAL OXYVET IV OXYGEN CONCENTRATOR
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
Front View (Figure 1)
1
7
2
8 / 9 10
4
6
5
3
11
12
Front View (Figure 1)
1. Power Switch
I = ON
O = OFF
2. Circuit breaker – (selected units) resets the unit after electrical overload shutdown
3. Humidifi er Holder
4. Oxygen outlet – oxygen is dispersed through this port
5. Flow meter
6. Flow meter knob
7. Monitor – indicate the condition of oxygen purity status and hour meter
8. Green Power light – illuminates when your concentrator is operating
9. Normal Oxygen (green) light (the LED display normal oxygen) / optional
10.Low Oxygen (yellow) light (the LED display low oxygen) / optional
11.Red Service Required light – when illuminated contact EICKEMEYER®
12.Castor

7
USER MANUAL OXYVET IV OXYGEN CONCENTRATOR
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
Back View (Figure 2)
13
14
15
16
Rear View (Figure 2)
13. Air fi lter – prevents dirt, dust, and lint from entering your unit
14. Product specifi cation label
15. Exhaust
16. Power cord and/ or IEC power connector
5. SETTING UP YOUR OXYGEN CONCENTRATOR
1. Position your unit near an electrical outlet in the room where you spend most of your time.
Danger!
Keep the oxygen concentrator at least 5 feet (1.6 m) from hot, sparking objects or naked sources of fl ame.
Note!
Do not connect to an electrical outlet controlled by a wall switch. NO other appliances should be plugged into the
wall outlet.
2. Position your unit at least 6 inches (16 cm) from walls, draperies, or any other objects that might prevent the proper fl ow
of air in and out of your oxygen concentrator. The Oxygen concentrator should be located so as to avoid pollutants or
fumes.

8
USER MANUAL OXYVET IV OXYGEN CONCENTRATOR
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
6. BEFORE OPERATING YOUR OXYGEN CONCENTRATOR
1. Before operating your unit always check to be sure the air fi lter (located on the back of your unit) is existed. Proper replace
of this fi lter is discussed in the Caring for Your Concentrator section on page 10.
2. Attach the appropriate oxygen accessories to the oxygen outlet.
Oxygen Tubing Connection:
a. Attach the oxygen tubing directly to the connector (Figure 3).
Figure 3
Oxygen Tubing Connection with Humidifi cation:
If your physician has prescribed an oxygen humidifi er as part of therapy, follow these steps:
a. Fill the humidifi er bottle with distilled water. Do not overfi ll.
b. Attach the oxygen tubing directly to the humidifi er bottle outlet fi tting (Figure 4).
Figure 4
Your physician has prescribed either a nasal cannula, catheter, or face mask, in most cases, they are already attached to the
oxygen tubing. If not, follow the manufacture’s instructions for attachment.
Remove the power cord completely from the line cord strap. Make sure the power switch is in the “Off” position, and insert the
plug into the wall outlet. The unit is double insulated to guard against electric shock.
Warning!
Improper use of the power cord and plugs can cause a burn, fi re or other electric shock hazards. Do not use the unit
if the power cord is damaged.

9
USER MANUAL OXYVET IV OXYGEN CONCENTRATOR
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
7. OPERATING YOUR OXYVET IV OXYGEN CONCENTRATOR
Danger!
Oxygen causes rapid burning. Do not smoke while your oxygen concentrator is operating, or when you are near a
person utilizing oxygen therapy. Keep the oxygen concentrator at least 5 feet (1.6 m) from hot, sparking objects or
naked sources of fl ame.
1. Press the power switch to the “On” position, when the unit is turned on, the power light will illuminate, and the patient
alert system – “Service Required” light and audible signal — will briefl y alarm.
Note!
• If the “Service Required” light illuminates and the audible signal alarms but the unit is not operating, there
is no power to the unit, refer to the Minor Troubleshooting chart not page 12 and contact EICKEMEYER® if
necessary.
• If an audible low-frequency vibration sound is detected, the unit is not operating properly. Refer to the Minor
Troubleshooting chart on page 12, and contact EICKEMEYER® if necessary.
2. Check the fl ow meter to make sure that the fl ow meter ball is centered on the line next to the prescribed number of your
fl ow rate.
Caution!
It is very important to follow your oxygen prescription. Do not increase or decrease the fl ow of oxygen-consult
your physician.
Note!
• EICKEMEYER® may have preset the fl ow meter so that it can not be adjusted.
• If the fl ow meter knob is turned clockwise, the fl ow decreases (and eventually will shut off the oxygen fl ow).
If the knob is turned counterclockwise, the fl ow increases.
The OxyVet IV Oxygen Concentrator monitors the oxygen produced by your unit. When the units is turned “ON”, all four light
(Power, Low Oxygen, Normal Oxygen, and Service Required) on the front panel will illuminate briefl y. After a few seconds, only
the Power and Normal Oxygen lights will remain lit.
The lights on the front panel are defi ned as follows:
• Green Normal Oxygen light-acceptable oxygen level (the LED will display “Normal Oxygen”)
• Yellow Low Oxygen light-below an acceptable oxygen level (the LED will display “Low Oxygen”)
If the oxygen purity falls below the acceptable level, the green Normal Oxygen light will shut off and the yellow Low Oxygen
light will illuminate. Switch to your reserve oxygen system. Refer to the Troubleshooting section in this guide on page 12 and
contact EICKEMEYER®.
As an added safety feature should the oxygen purity continue to drop, an intermittent audible signal will sound. Contact
EICKEMEYER® immediately. Do not attempt any other maintenance.

10
USER MANUAL OXYVET IV OXYGEN CONCENTRATOR
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
8. RESERVE OXYGEN SYSTEM
As a precaution, EICKEMEYER® may supply you with a reserve oxygen system. If your unit loses electrical power or fails to
operate correctly, the patient Alert System will sound to signal you to switch to your reserve oxygen system (if provided) and
contact EICKEMEYER®.
9. CARING AND PERIODIC MAINTENANCE FOR YOUR UNIT
Note!
Use no lubricants, oils or grease.
Warning!
Before attempting any cleaning procedures, turn the unit “off.”
Oxygen Humidifi er (reusable Bottles)
If your physician has prescribed humidifi cation, clean your humidifi er bottle daily. Follow the instructions supplied by the
manufacturer. If no cleaning instructions were supplied, follow these steps:
1. Wash the humidifi er in a solution of hot water and dishwashing detergent.
2. Soak the humidifi er in a solution of one part white vinegar to three parts hot water for 30 minutes. This solution acts as
a germicidal agent.
3. Rinse thoroughly with warm tap water and refi ll with distilled water for use. Do not overfi ll.
Cannula, Mask and Tubing
Clean and replace the cannula, mask, and tubing as instructed by EICKEMEYER®.
Air Filter and Oxygen Outlet Connector
The connector should not be cleaned at least once a week. To clean, follow these steps:
1. Remove the oxygen outlet connector (if used).
2. Wash in a solution of warm water and dishwashing detergent.
3. Rinse thoroughly with warm tap water and towel dry. The fi lter should be completely dry before reinstalling.
Air fi lter need to be replaced after device operates each 3,000 hours. Air fi lter is disposable after used. Not replace the air fi lter
regularly will affect the performance of concentration.
Caution!
To prevent product damage, do not attempt to operate the unit without the air fi lter or while the fi lter is still damp.

11
USER MANUAL OXYVET IV OXYGEN CONCENTRATOR
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
Exterior Cabinet
Warning!
To avoid electric shock, do not remove the concentrator cabinet. The cabinet should only be removed by a qualifi ed
EICKEMEYER® technician. Do not apply liquid directly to the cabinet or utilize any petroleum-based solvents or
cleaning agents.
Clean the concentrator exterior cabinet by using a damp cloth or sponge with a mild house-hold cleaner and wipe it dry.
Periodic Maintenance for your Unit
1. Clean the air fi lter at least once a week. The steps of cleaning see the Air Filter and Oxygen Outlet Connector.
2. Change the felt of the Inlet Pre-Filter at least once half a year. Follow those steps:
Open the Important Safeguard pull out the Inlet Pre-Filter. Pull out the old felts from the Inlet Pre-Filter (two sides). Then
insert the new felts to the Inlet Pre-Filter each side, insert the Inlet Pre-Filter and close the Important Safeguard.
3. Change the Bacterial Filter at least once a year. You can not change this by yourself, you must contact EICKEMEYER®
change this for you.
Note!
• The use of certain humidifi ers and accessories may impair the oxygen concentrator’s performance.
• If the Service Required light illuminates and the audible signal alarms but the unit is not operating, there is no
power to the unit. Refer to the Minor Troubleshooting chart on page 12 and contact EICKEMEYER®.

12
USER MANUAL OXYVET IV OXYGEN CONCENTRATOR
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
10. MINOR TROUBLESHOOTING
Symptom Possible Cause Remedy
Unit does not operate. Power light is off
when the power switch is “on”. Audible
alarm is pulsing and Service Required
light is fl ashing.
Power cord not properly inserted into
wall outlet.
Check power connection at the wall
outlet. On 230 volt units, check the back
of the unit.
No power at wall outlet. Check your home circuit breaker and
reset if necessary. Use a different wall
outlet if the situation occurs again.
Oxygen concentrator circuit breaker
activated (selected units).
Press the concentrator circuit breaker
button (if equipped) located below the
power switch. Use a different wall
outlet if the situation occurs again.
Unit operates; the Power light is on
when the Power switch is “On”. Red
Service Required light is illuminated.
Audible alarm may be sounding.
Air fi lter is blocked. Check the air fi lter. If the fi lter is dirty,
wash it following the cleaning
instructions on page 10.
Exhaust is blocked. Check the exhaust area; make sure
there is nothing restricting the unit
exhaust.
Blocked or defective cannula, catheter,
face mask or oxygen tubing.
Detach cannula, catheter, or face mask.
If proper fl ow is restored, clean or
replace if necessary. Disconnect the
oxygen tubing at the oxygen outlet. If
proper fl ow is restored, check oxygen
tubing for obstructions or kinks.
Replace if necessary.
Blocked or defective humidifi er bottle. Detach the humidifi er from the oxygen
outlet. If proper fl ow is obtained, clean
or replace humidifi er.
Flow meter set too low. Set fl ow meter to prescribed fl ow rate.
If the above remedies do not work,
contact EICKEMEYER®.
Unit operates the power light is on
when power switch is “on“, audible
low-frequency vibration sound is
detected.
Turn your unit “off”. Switch to your
reserve oxygen system, and contact
EICKEMEYER® immediately.
If any other problems occur with your
oxygen concentrator.
Turn your unit “Off”. Switch to your
reserve oxygen system, and contact
EICKEMEYER® immediately.
Both the green Normal Oxygen and the
yellow Low Oxygen lights are either on
or off.
O.C.I malfunction Contact EICKEMEYER®.
Yellow Low Oxygen light is on or the
yellow low Oxygen light is on and the
intermittent audible Signal is sounding.
Flow meter is not properly yet. Ensure the fl ow meter is properly set to
the prescribed.
Air fi lter is blocked. Check the air fi lter. If the fi lter is dirty,
wash it following the cleaning
instructions on page 10.
Exhaust is blocked. Check the exhaust area: make sure
there is nothing restricting the unit
exhaust.

13
USER MANUAL OXYVET IV OXYGEN CONCENTRATOR
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
Symptom Possible Cause Remedy
Red Service Required light is on and an
intermittent audible signal is sounding.
Flow meter is not properly yet. Ensure the fl ow meter is properly set to
the prescribed.
Air fi lter is blocked. Check the air fi lter. If the fi lter is dirty,
wash it following the cleaning
instructions on page 10.
Exhaust is blocked. Check the exhaust area: make sure
there is nothing restricting the unit
exhaust. If the above remedies do not
work, contact EICKEMEYER®.
11. SPECIFICATIONS
No. Item Rating
1. Delivery Rate 0.5 to 5 LPM
2. Outlet Pressure 8.5–13 psi (68.5–90 kpa)
3. Electrical Rating 230 V, 50 Hz, 1.4A (Max.)
4. Oxygen Percentage 0.5 ~ 5 LPM 93% ± 3%
5. Average oxygen concentration At 2 l/min: 93% ± 3%
6. Operating Environment Range Temperature 10°C to 40 °C
Humidity 15% to 95%
Atmospheric pressure 50 ~ 106 kpa (7.3 ~ 15.4 psi)
7. Weight 15.3kg
8. Sound Level (ISO 8359:1996 from front) ≤55dbA
9. Dimensions (in mm) L 342 x W 368 x H 572
10. Operating System Time Cycle / Pressure Swing
11. Storage Conditions Temperature -40 °C to 70 °C
Humidity 10% to 100%
Atmospheric pressure 50 ~ 106 kpa (7.3 ~ 15.4 psi)
12. Equipment Class and Type The type of protect against electric shock: Class II
The degree of protect against electric shock: TYPE B Type B Applied part
13. Pressure Relief Valve 276 kPa ± 34.5 kPa
14. Warranty Period limit 1 Year

14
USER MANUAL OXYVET IV OXYGEN CONCENTRATOR
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
12. SYMBOL
General warning, caution, risk of danger
Please read the instructions carefully before operating the product.
Type B applied part
Class II equipment
Manufacture’s serial number
Date of manufacture
Manufacture
Waste Electrical and Electronic Equipment Directive 2002 / 96 / EC (WEEE Directive) after over the valid
term within 3 years, the machine should be treated on local law or regulation requirement. To avoid hurt
users and pollute environment.
The equipment bears CE mark indicating its conformity with the provision of Council Directive 93 / 42 / EEC
concerning medical devices.
FRAGILE
Contents of the transport package are fragile therefore it shall be handled with care.
THIS WAY UP
Indicates correct upright position of the transport package.
KEEP AWAY FROM RAIN
Transport package shall be kept away from rain.
Non-ionizing electromagnetic radiation

15
USER MANUAL OXYVET IV OXYGEN CONCENTRATOR
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
13. ELECTROMAGNETIC IMMUNITY
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The OxyVet IV Oxygen Concentrator is intended for use in the electromagnetic environment specifi ed below. The customer
or the user of the OxyVet IV Oxygen Concentrator should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment-guidance
Conducted RF
IEC 61000-4-6
Conducted RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 V
3 V/m
Portable and mobile RF communications
equipment should be use no closer to any
part of the OxyVet IV Oxygen Concentrator
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance
d = √P
( )
3.5
V1
d = √P
( )
3.5
E1
= 80 MHz to 800 MHz
d = √P
( )
7
E1
= 800 MHz to 2.5 GHz
Where p is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).b
Field strengths from fi xed RF transmitters,
as determined by an electromagnetic site
survey, a should be less than the
compliance level in each frequency range.b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic is affected by absorption and refl ection from
structures, objects and people.
a Field strengths from fi xed transmitters, such as base stations for radio (cellular / cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fi xed RF transmitters, an electromagnetic site survey should be
considered. If the measured fi eld strength in the location in which the OxyVet IV Oxygen Concentrator is used exceeds
the applicable RF compliance level above, the OxyVet IV Oxygen Concentrator should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating
the OxyVet IV Oxygen Concentrator.
b Over the frequency range 150 kHz to 80 MHz, fi eld strengths should be less than 3 V/m.

GERMANY
EICKEMEYER KG
Eltastraße 8
78532 Tuttlingen
T +49 7461 96 580 0
F +49 7461 96 580 90
info@eickemeyer.de
www.eickemeyer.de
SWITZERLAND
EICKEMEYER AG
Sandgrube 29
9050 Appenzell
T +41 71 788 23 13
F +41 71 788 23 14
info@eickemeyer.ch
www.eickemeyer.ch
UNITED KINGDOM
EICKEMEYER Ltd.
3 Windmill Business Village
Brooklands Close
Sunbury-on- Thames
Surrey, TW16 7DY
T +44 20 8891 2007
info@eickemeyer.co.uk
www.eickemeyer.co.uk
POLAND
EICKEMEYER Sp. z o.o.
Al. Jana Pawła II 27
00-867 Warszawa
T +48 22 185 55 76
F +48 22 185 59 40
info@eickemeyer.pl
www.eickemeyer.pl
DENMARK
EICKEMEYER ApS
Solbakken 26, Hammelev
6500 Vojens
T +45 7020 5019
info@eickemeyer.dk
www.eickemeyer.dk
NETHERLANDS
EICKEMEYER B.V.
Bellweg 44
4104 BJ Culemborg
T +31 345 58 9400
info@eickemeyer.nl
www.eickemeyer.nl
ITALY
EICKEMEYER S.R.L.
Via G. Verdi 8
65015 Montesilvano (PE)
T +39 0859 35 4078
F +39 0859 35 9471
info@eickemeyer.it
www.eickemeyer.it
CANADA
EICKEMEYER Inc.
617 Douro Street, Suite #205
Stratford, Ont. Canada
N5A 0B5
T +1 519 273 5558
F +1 519 271 7114
info@eickemeyervet.ca
www.eickemeyer.com
This manual suits for next models
1
Table of contents
Other Eickemeyer Medical Equipment manuals

Eickemeyer
Eickemeyer LAVC-2000-D User manual

Eickemeyer
Eickemeyer HAICO TELGTE II User manual

Eickemeyer
Eickemeyer SYRINGE PUMP User manual

Eickemeyer
Eickemeyer INFUSOVET CONNECT User manual

Eickemeyer
Eickemeyer LIFEVET PT User manual

Eickemeyer
Eickemeyer CryoIQ DERM plus User manual

Eickemeyer
Eickemeyer HIRAY 30 User manual

Eickemeyer
Eickemeyer HIRAY 5 User manual

Eickemeyer
Eickemeyer LifeVet CP User manual

Eickemeyer
Eickemeyer KOWA SL-17 User manual