MRC CD-4307SD User manual

SD card real time data logger
TDS ( Total dissolved solids ), Salt
CONDUCTIVITY METER
Your purchase of this
CONDUCTIVITY METER
with SD CARD
DATA LOGGER marks a
step forward for you
into the field of
precision measurement.
Although this meter a
complex and delicate
instrument, its durable
structure will allow
many years of use if
proper operating
techniques are
developed. Please read
the following
instructions carefully
and always keep this
manual within easy
reach.
OPERATION MANUAL
Model : CD-4307SD
3, Hagavish st. Israel 58817 Tel: 972 35595252, Fax: 972 3 5594529 [email protected]
MRC. 2.19

TABLE OF CONTENTS
1. FEATURES................................................................
.
1
2. SPECIFICATIONS......................................................
.
2
2-1 General Specifications..........................................
.
2
2-2 Electrical Specifications........................................
.
4
3. FRONT PANEL DESCRIPTION.....................................6
4. MODE SELECTION.....................................................8
5. CONDUCTIVITY/TDS MEASURING and
CALIBRATION PROCEDURE.......................................
.
8
5-1 Conductivity measurement...................................
.
9
5-2 TDS ( PPM ) measurement...................................
.
10
5-3 Calibration...........................................................11
6. SALT MEASURING and calibration PROCEDURE...........13
6-1 Salt measurement................................................13
6-2 Calibration...........................................................14
7. OTHER FUNCTION....................................................
.
14
7-1 Data Hold............................................................14
7-2 Record ( Max./ Min. reading )...............................14
7-3 LCD Backlight ON/OFF.........................................
.
15
8. DATALOGGER...........................................................
.
15
8-1 Preparation before execute datalo
gg
er function..................15
8-2 Auto Datalo
gg
er ( Set samplin
g
time 1 second )............≧16
8-3 Manual Datalo
gg
er ( Set samplin
g
time = 0 second ).........
.
17
8-4 Check time information.....................................................17
8-5 Check samplin
g
time information.......................................18
8-6 SD Card Data structure.....................................................18
9. Saving data from the SD card to the computer............20
10. ADVANCED SETTING...............................................22
11. POWER SUPPLY from DC ADAPTER..........................
.
28
12. BATTERY REPLACEMENT.........................................
.
28
13. SYSTEM RESET.......................................................
.
29
14. RS232 PC SERIAL INTERFACE..................................29
15. OPTIONAL ACCESSORIES........................................
.
31
16. PATENT..................................................................
.
32

1. FEATURES
* One meter for multi purpose operation :
Conductivity, TDS ( Total dissolved solids ), Salt measurement.
*Conductivity : 200 uS/2 mS/20 mS/200 mS.
* Salt : 0 to 12 % salt ( % weight ).
* Conductivity measurement can select Temp. Coefficient
of measurement solution.
*ATC for the conductivity measurement.
* Separate probe, easy for operation of different
measurement environment.
* Real time SD memory card Datalogger, built-in Clock
and Calendar, sampling time can set from 1 sec to 8
hour 59 min. 59 sec.
* Manual datalogger is available, during execute the
manual datalogger function, it can set the different
location no. ( position 1 to position 99 ).
* Innovation and easy operation, computer is not need to
setup extra software, after execute datalogger, just take
away the SD card from the meter and plug in the SD card
into the computer, it can down load the all the measured
value with the time information ( year/month/date/
hour/minute/second ) to the Excel directly, thenuser can
make the further data or graphic analysis by themselves.
* SD card capacity : 1 GB to 16 GB.
* LCD with green light backlight, easy reading.
* It can default auto power off or manual power off.
* Data hold, record max. and min. reading.
* Microcomputer circuit, high accuracy.
* Power by UM3/AA ( 1.5 V ) x 6 batteries or DC 9V adapter.
* RS232/USB PC COMPUTER interface.
* Wide applications: water conditioning, aquariums,
beverage, fish hatcheries, food processing, photography,
laboratory, paper industry, plating industry, quality
control, school & college, water conditioning.
1

2. SPECIFICATIONS
2-1 General Specifications
Circuit Custom one-chip of microprocessor LSI
circuit.
Display LCD size : 52 mm x 38 mm
LCD with green backlight ( ON/OFF ).
Measurement * Conductivity ( uS, mS )
Function * TDS ( Total Dissolved Solids, PPM )
* Salt ( % Weight )
* Temperature ( , )℃℉
Temperature Automatic from 0 to 60 (32 - 140 ),℃℉
Compensation with temperature compensation factor
variable between 0 to 5.0% per C.
Conductivity Carbon rod electrode for long life.
Probe
Datalogger Auto 1 sec to 8 hour 59 min. 59 sec.
Sampling Time
@ Sampling time can set to 1 second,
Setting range
but memory data may loss.
Manual Push the data logger button
once will save data one time.
@ Set the sampling time to
0 second.
@ Manual mode, can also select the
1 to 99 position ( Location ) no.
Memory Card SD memory card. 1 GB to 16 GB.
Advanced * SD memory card Format
setting * Set clock time ( Year/Month/Date,
Hour/Minute/ Second )
*Set sampling time
*Auto power OFF management
*Set beep Sound ON/OFF
*Decimal point of SD card setting
*Set temperature unit to or℃℉
*Set CD temperature compensation factor
Data Hold Freeze the display reading.
Memory Recall Maximum & Minimum value.
2

Sampling Time Approx. 1 second.
of Display
Data Output RS 232/USB PC computer interface.
* Connect the optional RS232 cable
UPCB-02 will get the RS232 plug.
* Connect the optional USB cable
USB-01 will get the USB plug.
Operating Meter 0 to 50 .℃
Temperature Probe 0 to 60 .℃
Operating Less than 85% R.H.
Humidity
Power Supply
*
A
Alkaline or heavy duty DC 1.5 V battery
( UM3, AA ) x 6 PCs, or equivalent.
*
A
DC 9V adapter input. ( AC/DC power
adapter is optional ).
Power Current Normal operation ( w/o SD card save
data and LCD Backlight is OFF) :
Approx. DC 14 mA.
When SD card save the data and LCD
Backlight is OFF) :
Approx. DC 37 mA.
*
A
If LCD backlight on, the power
consumption will increase approx.
12 mA.
Weight 489 g/1.08 LB.
Dimension Meter 177 x 68 x 45 mm
(7.0 x 2.7x 1.9 inch)
Probe Round,
22 mm Dia. x 120 mm length.
Accessories * Instruction manual......................1 PC
Included * Conductivity/TDS/Salt probe,
CDPB-03.................................... 1 PC
* Hard carrying case ( CA-06 ).......
.
1 PC
3

Optional * 1.413 mS Conductivity Standard
Accessories Solution...............................CD-14
SD memory card ( 1 GB )
SD memory card ( 2 GB )
AC to DC 9V adapter.
USB cable, USB-01.
RS232 cable, UPCB-02.
Data Acquisition software,SW-U801-WIN.
2-2 Electrical Specifications (23±5 )℃
Conductivity ( uS, mS )
Range Measurement Resolutio
n
Accuracy
200 uS 0 to 200.0 uS 0.1 uS
2 mS 0.2 to 2.000 mS 0.001 mS ± (2% F.S.+1d)
20 mS 2 to 20.00 mS 0.01 mS * F.S. -
200 mS 20 to 200.0 mS 0.1 mS full scale
* Temperature Compensation :
Automatic from 0 to 60 ( 32 - 140 ), with temperature℃℉
compensation factor variable between 0 to 5.0% per C.
* The accuracy is specified under measurement value 100 mS.≦
* mS - milli Simens * @ 23± 5℃
TDS ( Total Dissolved Solids )
Range Measurement Resolutio
n
Accuracy
200 PPM 0 to 132 PPM 0.1 PPM
2,000 PPM 132 to 1,320 PPM 1 PPM ± (2% F.S.+1d)
20,000 PPM 1,320 to 13,200 PPM 10 PPM * F.S. -
200,000 PPM 13,200 to 132,000 PPM 100 PPM full scale
* Temperature Compensation :
Automatic from 0 to 60 ( 32 - 140 ), with temperature℃℉
compensation factor variable between 0 to 5.0% per .℃
* The accuracy is specified under measurement value 66,000 PPM.≦
* PPM - parts per million * @ 23± 5℃
4

Temperature
Function Measuring Range Resolutio
n
Accuracy
℃0 to 60℃℃ 0.1 ℃±0.8℃
℉32 to 140℉℉ 0.1 ℉±1.5℉
* @ 23± 5℃
Salt
Measurement 0 to 12 % salt ( % weight ).
Range
Resolution 0.01 % salt.
Accuracy 0.5 % salt value
* F.S. : full scale.
5

3. FRONT PANEL DESCRIPTION
6

3-1 Display
3-2 Power Button ( Backlight Button )
3-3 Hold Button ( ESC Button )
3-4 REC Button ( Enter Button )
3-5 Mode Button ( Button )▲
3-6 Range Button ( Button )▼
3-7 Time Button
3-8 Logger Button ( SET Button, Sampling check )
3-9 Stand
3-10 Battery Compartment/Cover
3-11 Battery Cover Screw
3-12 Tripod Fix Nut
3-13 Probe Socket
3-14 DC 9V Power Adapter Input Socket
3-15 Reset Button
3-16 RS-232 Output Terminal
3-17 SD card socket
7

4. MODE SELECTION
1)Turn on the meter by pressing the " Power Button "
( 3-2, Fig. 1 ) momentarily.
*Pressing the " Power Button " ( 3-2, Fig. 1 )
continuously and > 2 seconds again will turn off the
meter.
2)The meter can select 3 kind Mode as :
a. Conductivity, TDS measurement
b.TDS measurement
c. Salt measurement
Pressing the " Mode Button " ( 3-5, Fig. 1 ) once, the
Display will show the following text in sequence :
Cd Conductivity measurement
tdS 'TDS measurement
SALt Salt measurement
Until the Display show the desired mode the meter will
execute this Mode with default.
5. CONDUCTIVITY/TDS MEASURING
and CALIBRATION PROCEDURE
The meter default function are following :
*The display unit is set to conductivity ( uS, mS ).
*The temperature unit is set to .℃
*Temp. compensation factor is set to 2.0% per C.
*Auto range.
*Auto power off.
*The sampling time of data logger function is
2 seconds.
8

Fig. 2
5-1 Conductivity measurement
1)Prepare the Conductivity Probe ( CDPB-03, standard
accessory, included ), install the " Probe Plug " ( 5-1,
Fig. 2 ) into the " Probe Socket " ( 3-13, Fig. 1 ).
2)Power on the meter by pressing " Power Button "
( 3-2, Fig. 1 ) once.
Select the Meter's measurement Mode to " Cd "
( Conductivity measurement ), refer to chapter 4, page 8.
3)Hold the " Probe Handle " ( 5-2, Fig. 2 ) by hand
and let the " Sensing head " ( 5-3, Fig. 2 )
immersed wholly into the measured solution. Shake
the probe to let the probe's internal air bubble drift
out from the sensing head.
Display will show the conductivity mS ( uS ) values.
at the same time the left bottom display will show the
Temp. value of the measured solution.
9

Manual range operation
The meter is default to be used for the auto range
mode. Push the " Range Button " ( 3-6, Fig. 1 ) once in
sequence will change the range from 200 uS, 2 mS,
20 mS, 200 mS and auto range.
Change the Temp. unit to ℉
If intend to change the Temp. unit from to , please℃℉
refer to chapter 10-7 page 26.
Change the Temp. Coefficient Factor
The default Temp. compensation factor value of the
measurement solution is to 2.0% per . If intend to℃
change it, please refer to chapter 10-8, page 27.
Zero adjustment
If the probe not immerse the measurement solution and
display not show zero value, pressing the " Button "▲
( 3-5, Fig. 1 ) continuously at least 10 seconds will let
display show zero. The zero function only valid for the
200 uS range and the not zero value is < 2.0 uS.
5-2 TDS ( PPM ) measurement
The measuring procedures are same as above
5-1 Conductivity ( uS, mS ) measurement,
except to change the display unit from uS, mS to PPM.
The detail procedures please refer chapter 4, page 8.
10

5-3 Calibration
1)Prepare the standard conductivity solution ( optional )
For example :
2 mS range calibration solution :
1.413 mS Conductivity Standard Solution, CD-14
200 uS range calibration solution :
80 uS Conductivity Standard Solution
20 mS range calibration solution :
12.88 mS Conductivity Standard Solution
or other Conductivity Standard Solution
2)Install the " Probe Plug " ( 5-1, Fig. 2 ) into the
" CD Socket " ( 3-13, Fig. 1 ).
3)Power on the meter by pressing " Power Button "
( 3-2, Fig. 1 ) once.
4)Hold the " Probe Handle " ( 5-2, Fig. 2 ) by hand
and let the " Sensing head " ( 5-2, Fig. 2 )
immersed wholly into the measured solution. Shake
the probe to let the probe's internal air bubble drift
out from the sensing head.
Display will show the conductivity mS ( uS ) values.
5)Use the two fingers to press the " REC Button " ( 3-4,
Fig 1 ) and " HOLD Button " ( 3-3, Fig. 1 ) at the same
time. the display will show the following screen as
example, release the both fingers.
Cd
CAL
11

6)Press the " Enter Button " ( 3-4, Fig. 1 ), the measuring
value will present on both upper and lower Display.
Use " Button " ( 3-5, Fig. 1 ), " Button " ( 3-6, Fig. 1 )▲▼
to adjust the up display value exact same as the standard
conductivity value. Press the " Enter Button " ( 3-4. Fig. 1 )
will save the calibration data and finish the calibration
procedures.
1.318 1.413 Standard value
1.318 1.318
*
If only intend to make the one point calibration,
just execute the 2 mS range ( 1.413 mS Cal. )
is enough.
*
Multi-points calibration procedures should
execute the 2 mS range ( 1.413 mS Cal. )
calibration at first, then make other ranges
(20 uS range, 20 mS range or 200 mS range )
calibration procedures following if necessary.
12

6. SALT MEASURING and
CALIBRATION
6-1 Salt measurement
1)Prepare the Probe ( CDPB-03, same probe as the
conductivity measurement, refere to Fig. 2, page 9 ),
install the " Probe Plug " ( 5-1, Fig. 2 ) into the "
Probe Socket " ( 3-13, Fig. 1 ).
2)Power on the meter by pressing " Power Button "
( 3-2, Fig. 1 ) once.
Select the Meter's measurement Mode to " SALt ", refer
to Chapter 4, page 8.
3)Hold the " Probe Handle " ( 5-2, Fig. 2 ) by hand
and let the " Sensing head " ( 5-3, Fig. 2 )
immersed wholly into the measured solution. Shake
the probe to let the probe's internal air bubble drift
out from the sensing head.
Display will show the Salt values ( % weight ).
6-2 Calibratio
n
If the conductivity range already make the calibration
completely then the Salt measurement is not necessary to
make the calibration again.
13

7. OTHER FUNCTION
7-1 Data Hold
During the measurement, press the " Hold Button " ( 3-3,
Fig. 1 ) once will hold the measured value & the LCD will
display a " HOLD " symbol.
Press the " Hold Button " once again will release the data
hold function.
7-2 Data Record ( Max., Min. reading )
1)The data record function records the maximum and
minimum readings. Press the " REC Button " ( 3-4, Fig.
1 ) once to start the Data Record function and there
will be a " REC. " symbol on the display.
2)With the " REC. " symbol on the display :
a)Press the " REC Button " ( 3-4, Fig. 1 ) once, the
" REC. MAX. " symbol along with the maximum value
will appear on the display.
If intend to delete the maximum value, just press
the " Hold Button " ( 3-3, Fig. 1 ) once, then the
display will show the " REC. " symbol only & execute
the memory function continuously.
b)Press the " REC Button " ( 3-4, Fig. 1 ) again, the
" REC. MIN. " symbol along with the minimum value
will appear on the display.
If intend to delete the minimum value, just press
the " Hold Button " ( 3-3, Fig. 1 ) once, then
the display will show the " REC. " symbol only &
execute the memory function continuously.
c)To exit the memory record function, just press the
" REC " button for 2 seconds at least. The display will
revert to the current reading.
14

7-3 LCD Backlight ON/OFF
After power ON, the " LCD Backlight " will light
automatically. During the measurement, press the
" Backlight Button " ( 3-2, Fig. 1 ) once will turn OFF the
" LCD Backlight ".
Press the " Backlight Button " once again will turn ON the
" LCD Backlight " again.
8. DATALOGGER
8-1 Preparation before execute datalogger function
a. Insert the SD card
Prepare a " SD memory card " ( 1 GB to 16 GB, optional ),
insert the SD card into the " SD card socket " ( 3-12, Fig. 1 ).
The front panel of the SD card should face against the
down case.
b. SD card Format
If SD card just the first time use into the meter, it
recommend to make the " SD card Format " at first. ,
please refer chapter 10-1, page 23.
c. Time setting
If the meter is used at first time, it should to adjust the
clock time exactly, please refer chapter 10-2, page 23.
d. Decimal format setting
The numerical data structure of SD card is
default used the " . " as the decimal, for
example "20.6" "1000.53" . But in certain
countries ( Europe ...) is used the " , " as the
decimal point, for example " 20, 6 "
"1000,53". Under such situation, it should
change the Decimal character at first, details
of setting the Decimal point, refer to Chapter
10-6, page 26.
15

8-2 Auto Datalogger ( Set sampling time 1 second )≧
a. Start the datalogger
Press the " REC Button ( 3-4, Fig. 1 ) once , the LCD will
show the text " REC ", then press the " Logger Button "
( 3-8, Fig. 1 ), the bottom text " DATALOGGER " will
flashing, at the same time the measuring data along the
time information will be saved into the memory circuit.
Remark :
*
How to set the sampling time, refer to Chapter 10-3
page 24.
*
How to set the beeper sound is enable, refer to
Chapter 10-5, page 25.
b. Pause the datalogger
During execute the Datalogger function , if press the
" Logger Button " ( 3-8, Fig. 1 ) once will pause the
Datalogger function ( stop to save the measuring data
into the memory circuit temporally ). In the same time
the the text of " DATALOGGER " will be no flashing.
Remark :
If press the " Logger Button " ( 3-8, Fig. 1 ) once again
will execute the Datalogger again, the bottom text of "
DATALOOGER " will flashing .
c. Finish the Datalogger
During pause the Datalogger, press the " REC Button "
( 3-4, Fig. 1) continuously at least two seconds, the " REC "
indication will be disappeared and finish the Datalogger.
16

8-3 Manual Datalogger ( Set sampling time = 0
second )
a. Set sampling time is to 0 second
Press the " REC Button ( 3-4, Fig. 1 ) once , the LCD will
show the text " REC ", then press the " Logger Button " (
3-8, Fig. 1 ) once, the bottom text " DATALOGGER " will
flashing once and Beeper will sound once, at the same
time the measuring data along the time information will be
saved into the memory circuit. The lower Display will show
the Position ( Location ) no. and saved into the SD card too.
Remark :
During execute the Manual Datalogger, press the " ▲
Button " ( 3-5, Fig, 1 ) the lower no. ( position no. ) will
flashing. It can use the " Button " ( 3-5, Fig. 1) or "▲
Button " ( 3-6, Fig. 1 ) to set the measuring Location▼
no. ( 1 to 99, for example room 1 to room 99 ) to
identify the measurement location , the lower Display will
show P x ( x = 1 to 99 ).
b. Finish the Datalogger
Press the " REC Button " ( 3-4, Fig. 1) continuously at
least two seconds, the " REC " indication will be
disappeared and finish the Datalogger.
8-4 To check the time information
During the normal measurement screen ( not execute
the Datalogger ),
1)If press " Time Button " ( 3-7, Fig. 1 ) once , the lower
LCD display will present the time information of
Hour/Minute/Second ( h.m.s ) in the lower Display.
17

2)If press " Time Button " ( 3-7, Fig. 1 ) once again , the
lower LCD display will present the time information of
Year/Month/Date ( yy.mm.dd ) in the lower Display.
3)If press " Time Button " ( 3-7, Fig. 1 ) once again ,
the LCD will return to normal screen.
8-5 Check sampling time information
During the normal measurement screen ( not execute the
Datalogger ), If press " Sampling Button " ( 3-8, Fig. 1 )
once , the lower LCD display will present the Sampling
time information in second unit.
8-6 SD Card Data structure
1)When the first time, the SD card is used into the meter,
the SD card will generate a route :
CDA01
2)If the first time to execute the Datalogger,
under the route CDA01\, will generate a new
file name CDA01001.XLS.
After exist the Datalogger, then execute again,
the data will save to the CDA01001.XLS until
Data column reach to 30,000 columns, then
will generate a new file, for example CDA01002.XLS
3)Under the folder CDA01\, if the total files more
than 99 files, will generate anew route, such as
CDA02\ ........
18
Table of contents