Troubleshooting
4 5 6
Configure Software
3
1.Download Data: ElitechLog soware will automacally
access the logger and download the recorded data
to local computer if it finds the logger is connected.
If not, manually click “Download Data” to download
the data.
2.Filter Data: Click “Filter Data” under Graph tab to
select and view your desired me range of the
data.
3.Export data: Click “Export Data” to save Excel/PDF
format files to local computer.
4.Configure opons: Set logger me, log interval,
start delay, high/low limit, date / me format, email
etc. (Check User Manual for default parameters)
Note: New configuraon will inialize previous
recorded data. Please make sure to back up all
necessary data before you apply new
configuraons.
Refer to “Help” for more advanced funcons. More
product informaon are available on the company
website www.elitechlog.com.
If... Please...
only a few data was logged. check whether the baery is installed; or check if it
was installed correctly.
the logger does not log aer
start up
the logger cannot stop logging
by pressing the buon .
check whether start delay is enabled in the soware
configuraon.
check parameter sengs to see if buon customizaon
is enabled (default configuraon is disabled.)
Technical Specifications
Recording Opons Mul-Use Cerficaons EN12830, CE, RoHS
Validaon Cerficate Hardcopy
Soware ElitechLog Win or Mac (latest version)
Report Generaon Automac PDF report
Password Protecon Oponal on request
Connecon Interface USB 2.0, A-Type
Alarm Configuraon Oponal, up to 5 points
Reprogrammable With free Elitech Win or MAC soware
Demensions 80mmx33mmx14mm(LxWxH)
Weight 35g
Temperature Range
Temperature Accuracy
Temperature Resoluon
Data Storage Capacity 32,000 readings
Shelf Life/Baery Six months
1
/CR2032 buon cell
Recording Interval 10s~24 hour adjustable
Startup Mode Buon
Stop Mode Buon, soware or stop when full
Protecon Class IP67 1. Depending on opmal storage condions
(±15
°C
to +23
°C
/45% to 75% rH)
-30
°C
to 70
°C
±0.5(-20
°C
/+40
°C
);±1.0(other range)
0.1
°C