ellemka LMK-500D User manual

LMK 500 Black
User’s Guide version 1.0 - English ( )
©2016-2017 by ELLEMKA s.à r.l. All Rights Reserved
LMK 500 Black
User’s Guide version 1.0 - English ( )
©2016-2017 by ELLEMKA s.à r.l. All Rights Reserved
iv. Safety Notes
A. Warning:
1. Do not use in the event of malfunction, turn off or remove the batteries
immediately if you notice smoke or unusual smell coming from the unit to
avoid burns.
2. Do not use the unit in the presence of inflammable gases, as this could
result in explosion or fire.
3. Do not leave the unit where it will be exposed to extremely high
temperature, failure to observe could result damage, skin burn or fire.
4. Do not use worn-out batteries or different specifications for two
batteries altogether. Otherwise, this may leak, hurt and cause injury.
5. Do not operate and use the unit when it is damp with overflow or water
vapor condensation, failure to observe could result in injury or damage.
6. Do not assemble or disassemble the unit, providing technical support and
maintenance internally could result injury or fault. In the event of
malfunction, only an authorized technician or qualified service personnel
is allowed to repair.
7. Keep the unit out of reach of children. If broken, any small parts could
cause injury, damage, choke or death if swallowed inside the mouth.
Consult a physician immediately if the child swallows any part of the unit.
8. Keep the battery out of reach of children, battery contains a poisonous
chemical. If broken, this could cause death or illness if taken inside the
mouth or body.
9. Keep the strap out of reach of children, playing or placing around the neck
could strangulate or choke the infant or child.
B. Precations:
1. The Pulse Oximeter is applicable to use for examining the user ages
beyond 2 years old.
2. Do not directly expose your eyes to the light-emitting components of the
unit, as this could cause harm to your eyes.
3. While curing disease such as (blood vessel) proliferative lesion using
photodynamic therapy, avoid using the Pulse Oximeter for measurement,
because the light (within the wavelength range) emitted by the probe of the
Pulse Oximeter may influence non-lesion tissues of the human body.
4. Do not use the unit for purpose other than specified in this manual, failure
to follow the guidelines could result in minor incidents
5. Do not spray, pour, spill any liquid on the unit and accessories, as this could
result in damage.
6. Do not use new, worn, different types of batteries. Otherwise, it might leak,
heat or break and cause injury.
7. Ask for medical care and advise if the measured value goes beyond the
normal range.
8. Properly consult relevant medical literature or professionals for details
about clinical limitations and contraindications.
v. Quick Start Guide
Follow these steps for a quick start with the LMK-500D Pulse Oximeter.
1. Unpack and check the packaging contents ( section ii).
2. Attach the lanyard into the notch or hole ( section III.A).
3. Install the batteries ( section III.B).
4. Power on ( section IV.A).
5. Insert and test the fingertip ( section IV.F).
6. Read the measured vital signs: SpO2, PR and PI ( section IV.E).
7. Power off ( section IV.G).
SYMBOLS CONVENTIONS SYMBOLS CONVENTIONS
It indicates that waste
electrical and electronic
equipment is to be
collected separately.
Facebook
Alarm Twitter
Beep Pinterest
Display Google+
Pulse Rate (PR) Authorized Distributor
Oxygen Saturation (SpO2) Manufacturer
i. Notices
©2017 by ELLEMKA s.à r.l.™. All Rights Reserved
No part of the user’s manual may be reproduced, transmitted, transcribed and
stored in any form, by any means, without written authorization from Ellemka.
And Ellemka will not be held liable for any damages resulting from the use of
this product.
Without prior notice, Ellemka reserves the right
to change the specifications of the hardware
and software as described in this user’s manual
at anytime. Please check regularly for the pdf file
format of this manual that can be downloaded
easily from the website.
While every effort has been made to ensure that the information is accurate and
comprehensive, we would appreciate your effort to inform Ellemka if any typo
errors can be found in this manual, see for more contact details ( section IX).
Please be sure to read thoroughly for your safety and be guided in using the unit.
ii.Packaging Contents
Please check and confirm the packaging
contains the following items:
■Pulse Oximeter LMK-500D
■Lanyard
■2 pcs. batteries (AAA)
■User’s manual (this booklet)
iii. Symbol & Conventions
SYMBOLS CONVENTIONS SYMBOLS CONVENTIONS
Refer to a specific page/
figure/section Perfsuion Index (PI)
User's manual SN Serial Number
European Conformity (CE)
is a mandatory
conformity marking for
certain products sold
within the European
Economic Area (EEA).
The product system
conforms to essential
requirement of the
Medical Oximeter
Directive 93/42/EEC.
Not for continous
monitoring
FDA Food and Drug
Administration Brightness
BF-type application part
(refer to IEC 60601-
1:1995)
Settings
% of Oxygen Saturation Restore
beat per minute / Pulse
Rate Postal Address
IP22
IP Code, International
Protection Marking
IP22 is protected against
insertion of fingers and
will not be damaged or
become unsafe during
a specified test in which
it is exposed to dripping
water.
Mailing Address
Battery Power Indication Telephone
Plethysmogram Fax Number
Attention, Warning,
Precaution and Caution Email
Note Website
Within this user’s
manual, Illustrations are for
reference only and may differ
slightly from the appearance
of the actual product.
Do not accept, if any of
the specified items above
are not completely enclosed
inside the packaging box or if
any sign of damage is visible,
refer to Customer Service
and Support section for
specific details( section IX).
PAGE | 01
PAGE | 02

LMK 500 Black
User’s Guide version 1.0 - English ( )
©2016-2017 by ELLEMKA s.à r.l. All Rights Reserved
LMK 500 Black
User’s Guide version 1.0 - English ( )
©2016-2017 by ELLEMKA s.à r.l. All Rights Reserved
C. Know the Unit
LEGEND:
1- lanyard hole or
notch
2- battery
compartment
3- power button
4- OLED screen
5- probe of the unit
Figure 1: Know the Unit
III. Installation
A. Lanyard Installation
1. Locate the lanyard notch or hole within the edges of the unit.
2. Find and check the provided lanyard to use inside the packaging box,
otherwise look for a lanyard
with threads enough to fit
into the hole.
3. Insert the lanyard into the
hole. Take the smallest end
of the lanyard using the
index finger and thumb,
push it through one end of
the hole and out of the outer
side.
4. Pull the lanyard through it,
enough to safeguard the unit.
Attach the lanyard to safeguard the unit from falling or any damage.
B. Battery Installation
1. Open the battery compartment of the unit. Check and clear the battery
compartment from any unwanted objects that may cause any harm to the
user or may cause to explode or damage the unit.
2. Find and check 2 pieces of AAA batteries provided to use inside the
packaging box, otherwise look for the specified batteries to use.
3. Properly insert the batteries
into the battery compartment.
Check the correct polarity
and properly install each
battery as indicated inside
the battery compartment as
(+) positive and (-) negative.
Otherwise, it might leak,
heat and cause injury.
4. Close the battery
compartment with the cover
before using the unit.
For short time use of the unit, please remove the batteries.
I. Introduction
In Pulse Oximetry, Peripheral Oxygen Saturation (SpO2) refers the indirect
measurement of the specific oxygen content in the blood. More specifically, it is
expressed as a percentage of haemoglobin containing oxygen (oxygen-saturated
haemoglobin) to the total amount of haemoglobin in the blood
(oxygen-saturated/unsaturated haemoglobin).
The haemoglobin is an iron-containing protein bound to red blood cells that
transports oxygen from the lungs to the rest of the body
SpO2is a very important parameter to know the health status of an individual.
The normal SpO2values for a healthy individual are between 95 -100 percent.
In general, the normal values of oxygen saturation shall not be lower than 94
percent. If the measured value of oxygen saturation is lower than 94 percent, an
insufficient supply of oxygen is considered. Many chronic pulmonary or
respiratory related diseases cause to lower the percentage of oxygen saturation
in the human blood. It is important for people with breathing and heart problems
to monitor their SpO2values in a timely manner.
The Pulse Rate (PR) is the number of pulse beats per minute. Normally, the
pulse rate is consistent with the heart rate. In general, the pulse rate of every
people is 60 to 90 beats per minute.
The Perfusion Index (PI) usually reflects the limb perfusion status of an
examined patient, and shows the detection precision of the instrument as well;
that is, examination can still be performed even in the low or weak perfusion
condition. The PI of a normal human body is 3% or greater.
Pulse Oximetry is particularly convenient for noninvasive continuous method of
measurement for blood oxygen saturation, pulse rate and strength; and
perfusion index.
II. Overview
Pulse Oximeter is a small clip-like device usually placed on a patient’s
fingertip. It is based in full digital technology, non-invasive technique to monitor
the oxygen level in your blood. It functions by shining light through your finger.
The sensors detect how much oxygen is in your blood based on the way the light
passes through your finger. It also measures the pulse rate and strength.
A finger pulse oximeter is simple, easy, portable and convenient to use; cheap
and commonly used. Because of its simplicity of use and the ability to provide
immediate readings that are of critical importance in emergency medicine. It is
a very useful monitoring device for patients with respiratory or cardiac problems
to monitor their oxygen saturation levels in timely manner. It also useful for
mountain climbers and athletes whose oxygen levels may decrease at high
altitudes or any physical activity or exercise.
It now used routinely in critical care, monitoring rooms, hospitals (including
operation rooms of the departments of internal medicine and surgery, the
department of anesthesiology, the department of pediatrics, and intensive care
rooms), oxygen bars, social medical care institutions, sports & health and
typically found in ambulances. It is not advisable to use for a continuous
monitoring to any patient. And it is appropriate to use before and after, but not
during the actual physical activities to undertake.
A. Principle of Pulse Oximeter
The principle of a Pulse Oximetry is based on the red and infrared light
absorption characteristics of oxygenated and deoxygenated hemoglobin.
Oxygenated hemoglobin absorbs more infrared light and allows more red
light to pass through. Deoxygenated (or reduced) hemoglobin absorbs more
red light and allows more infrared light to pass through.
Modern Pulse Oximeter consist of a peripheral probe together with a
microprocessor unit displaying a waveform, the oxygen saturation and the
pulse rate. The probe is placed on the digit, earlobe or nose. Within the probe
are two LEDs, one in the visible red spectrum (660nm) and the other in the
infrared spectrum (940nm).
The beams of light pass through the tissues to the photodetector. During
passage through the tissues some light is absorbed by blood and soft tissues
depending on the concentration of hemoglobin. The amount of light
absorption at each frequency depends upon the degree of oxygenation of
hemoglobin within the tissues.
Detection of oxygen saturation of hemoglobin by spectrophotometry is based
on Beer-Lambert law, which relates the concentration of solute to the
intensity of light transmitted through a solution.
B. Features
■User-friendly
■Convenient, handy and low weight
■Portable with elegant design and based on advanced
digital technology
■Precise measurement of SpO2, PR
and PI
■Perfusion Index (PI) display
■One button operation
■Colored OLED screen
■Plethysmogram
■Operation settings
■Brightness settings
■Alarm and beeping sounds
■Bar graph indicates pulse rate
strength
■Demo mode
■Restore default settings
■Battery power indication
■Auto power off within 10 seconds
■Four directions display and six
display modes
Figure 2: Lanyard Installation
Figure 3: Battery Installation
PAGE | 04
PAGE | 03

LMK 500 Black
User’s Guide version 1.0 - English ( )
©2016-2017 by ELLEMKA s.à r.l. All Rights Reserved
LMK 500 Black
User’s Guide version 1.0 - English ( )
©2016-2017 by ELLEMKA s.à r.l. All Rights Reserved
IV. Operation Guide
A. Power on
Press the power on button for about one second, release when the screen
shows the Default Main Display ( Figure 4).
Power key button is also known as one button operations. As it can be used
to change the mode of display, to select items from the user menu, to switch
display user interfaces ( Figure 8).
Before using, check the power button of the unit if it is functioning and
batteries have enough power voltage to supply.
B. Display User Interfaces
V2.01.D/E - indicates the software
version of the unit.
(*) - denotes for the selected item from
the menu. It only moves from the top to
bottom of the selected menu.
Hold or press the power button to shift,
select, view and setup the display user
interfaces then release the button.
Figure 4:
Default Main Display
Figure 5:
Alarm Setup Menu
Figure 6:
Sound Setup Menu
C. Demo Mode
Before using for
actual measurement,
set the Demo mode
to “off”.
Figure 7:
Main Display
Demo mode “on”
If the Demo mode of the unit is set to“on” the Main Display shows the
expected parameters: default or preset values of SpO2, PR and PI, bar graph
indicates pulse strength and waveforms are for demonstration purposes only.
The main screen shows a label Demo if it is set to “on”.
D. Display Mode
Press any sides of the One Button Operations key to shift, view and select
mode of display to use, as shown below.
Figure 8: Display Mode
E. Reading the Display Data
Figure 9: Reading the Display Data
LEGEND:
1- Power-on button/ one button operations key
2- Battery power indication
3- Pulse Rate strength bar graph
4- Plethysmogram
5- % of Oxygen Saturation symbol
6- Measured value of Oxygen Saturation (SpO2)
7- beat per minute Pulse Rate symbol
8- Measured value of Pulse Rate(PR)
9- Perfusion Index symbol
10 - Measured value of Perfusion Index (PI)
11 - Standby mode, weak or no signal is detected
F. Measuring the SpO2, PR and PI
1. Open up the Pulse Oximeter’s clip with the one hand.
2. Insert the fingertip completely into the probe of the Pulse Oximeter’s where
there is a rubber pad.
3. Keep the fingertip surface in upward position, then release the Pulse
Oximeter’s clip.
4. When the readings are stable, read the measured data of Oxygen
Saturation (SpO2), Pulse Rate (bpmPR), Perfusion Index (PI), Pulse Rate
strength bar graph, Plethysmogram and also the status of the battery
power supply.
5. After measurement, open
up the Pulse Oximeter’s
clip and take off the
fingertip from the probe
of the unit, then release it.
6. After using, remove the
batteries from the battery
compartment and store
properly.
Before and after measurement, clean the fingertip to be measured and the
rubber pad by using a piece of dry soft cloth dipped with 75% medical alcohol for
disinfection purposes.
G. Power off
If the Pulse Oximeter is in standby mode, no signal is detected or not
operated within 10 seconds, it will power off automatically.
V. Settings
A. Alarm Setup Menu - allows to change the settings of alarm, beep, demo
and restore to either on or off.
Figure 10: Measuring the SpO2, PR and PI
PAGE | 05
PAGE | 06

LMK 500 Black
User’s Guide version 1.0 - English ( )
©2016-2017 by ELLEMKA s.à r.l. All Rights Reserved
LMK 500 Black
User’s Guide version 1.0 - English ( )
©2016-2017 by ELLEMKA s.à r.l. All Rights Reserved
Figure 11: Alarm (Alm)
When Alm is set to “on”, the Pulse
Oximeter provides an alarm sound if
the measured PR value or SpO2value
goes beyond the upper or below the
lower limit values as set by default;
otherwise, it will not provide alarm
sounds.
Figure 12: Beep
When Beep is set to “on”, the Pulse
Oximeter provides a ticking sound
along with the pulse beats when use
to measure the Pulse Rate, otherwise
it will not provide beeping sounds.
Figure 13: Demo
When Demo is set to “on”, the Main
Display shows the possible readings.
Be sure to set the Demo to “off” to
get the actual and accurate values of
SpO2, PR and PR; otherwise the
readings to be displayed are the
default values as set for
demonstration purposes.
Figure 14: Restore
When Restore is set to “on”, the Pulse
Oximeter will restore the default
factory settings. If the Demo Mode is
set to “on”, the “ok” option of Restore
is not visible in the menu; otherwise
set the Demo Mode to “off” if you will
intend to restore the default settings.
Figure 15: Brigthness
Select Brightness to change the
display screen
brightness, 1 is the lowest and 5 is the
highest value.
Figure 16: Exit
Select Exit to go back to the main
display screen.
B Sound Setup Menu
Please make sure to set the Alm to “on” in Alarm Setup menu if you intend to
set up alarm sounds for the upper and lower limit values of SpO2& PR. For
technical parameters to use, check the specific details ( section VI).
Figure 17: SpO2Alm Hi
When SpO2Alm Hi is set to “100”,
the Pulse Oximeter provides alarm
sounds if the measured SpO2value
reaches the upper limit set value of
“100” otherwise, it will not provide
alarm sounds.
Figure 18: SpO2Alm Lo
When SpO2Alm Lo is set to “94”, the
Pulse Oximeter provides alarm sound
if the measured SpO2value reaches
the lower limit set value of “94”
otherwise, it will not provide alarm
sounds.
Figure 19: PR Alm Hi
When PR Alm Hi is set to “130”, the
Pulse Oximeter provides alarm sound
if the measured PR value reaches
the upper limit set value of “130”
otherwise, it will not provide alert or
alarm sounds.
Figure 20: PR Alm Lo
When PR Alm Lo is set to “50”, the
Pulse Oximeter provides alarm sound
if the measured PR value reaches the
lower limit set value of “50”
otherwise, it will not provide alert or
alarm sounds.
Figure 21: +/-
In “+” mode, it will increase the value
of the corresponding selected menu
item by clicking the functional key
otherwise, in “-” mode it will decrease
the value.
VI. Technical Specifications
TECHNICAL SPECIFICATIONS
PRODUCT NAME Pulse Oximeter
PRODUCT MODEL LMK-500D
SOFTWARE VERSION Version 2.01 D/E
AVAILABLE COLOR Black
CERTIFICATIONS FDA, CE
DISPLAY Screen Type OLED (color)
Directions 4
Modes 6
DIMENSIONS(mm) Width 62.2 mm
Depth 37.0 mm
Height 33.1 mm
POWER SUPPLY AND CURRENT Internal Power Type 2 x AAA Alkaline
Battery
Working Current 35 mA
PEAK WAVELENGTH RANGE
OF THE LIGHT EMITTED FROM
PROBE
Red Light 660nm ± 3
Infrared Light 905nm ± 5
MAXIMUM OPTICAL OUTPUT
POWER OF THE PROBE
Infrared
light(905nm) 1.2mW
PAGE | 07
PAGE | 08

LMK 500 Black
User’s Guide version 1.0 - English ( )
©2016-2017 by ELLEMKA s.à r.l. All Rights Reserved
LMK 500 Black
User’s Guide version 1.0 - English ( )
©2016-2017 by ELLEMKA s.à r.l. All Rights Reserved
TECHNICAL SPECIFICATIONS
NORMAL WORKING CONDITION Working
Temperature
5ºC to 40ºC (41ºF
to 104ºF)
Storage
Temperature -10ºC to -50ºC
Relative Humidity
15% to 80%,
non-condensing in
operation
10% to 93%,
non-condensing in
storage
Atmospheric Pres-
sure 50kPa to 106kPa
Rated Voltage DC 3.0V
MEASUREMENT
PERFORMANCE IN LOW
PERFUSION CONDITION
0.3%
DEFAULT VALUES FOR PR, SPO2
Oxygen Saturation Upper Limit 100
Lower Limit 94
Pulse Rate Upper Limit 130
Lower Limit 50
Alert Condition
When the alarm (Alm) is set to “on” and
the actual measured value goes beyond
or below the range of preset alert
parameters, the Pulse Oximeter gives an
alert sound.
TECHNICAL PARAMETERS
Display Range Oxygen Saturation
(SpO2)35% to 100%
Pulse Rate (PR) 25 bpm to 250 bpm
Resolution Oxygen Saturation
(SpO2)1%
Pulse Rate (PR) 1 bpm
Measurement Precision Oxygen Saturation
(SpO2)
± 2% (70% to 100%)
No requirement (≤
69%)
Pulse Rate (PR) 25 bpm to 250 bpm
Alert Range Oxygen Saturation
(SpO2)
Upper & Lower
Limit: 50% to 100%
Pulse Rate (PR)
Upper & Lower
Limit: 50% to 100%
Oxygen Saturation
Alert Error Oxygen Saturation
(SpO2)
± 1% of the preset
value
Pulse Rate
The greater of ±10%
of the preset value
and ±bpm
Perfusion Index (PI) Weak Minimum 0.2%
Normal >= 3%
CAPACITY TO RESIST AMBIENT
LIGHT INTERFERENCE
Compare the measured values of oxygen
saturation in the case of indoor natural
light and existing illumination sources
with the measured values of oxygen
saturation in the darkroom condition. The
deviation is smaller than ± 1%.
SAFETY TYPE
Anti-electric-shock type Internal power supply device
Anti-electric-shock degree BF-type application part
Running mode Continuous working
Waterproof grade IP22
VII. Product Handling
Do not use any strong dissolving agent such as acetone. Do not rub the body
of the unit using any metallic and non-metallic pointed objects. Do not clean the
inner and outer surface of the unit using any washing liquid.
A. Cleaning
1. Power off the unit and remove the batteries before cleaning.
2. Ensure that the unit is dirt and dust free before using.
3. Check the internal surface of the unit if it is free from any unwanted
objects.
4. Clean the outer surface of the unit (including the screen and rubber pad)
using a piece of dry soft cloth with 75% medical alcohol.
B. Maintenance
1. When the battery power indication flashes, it indicates running out of
power supply. Turn off the unit and replace the AAA batteries with new
ones (same type) at the same time.
2. Properly install, remove and store the batteries before and after normal
use.
3. If the Pulse Oximeter is not used or operated, remove the batteries inside
the battery compartment.
4. Properly dispose discarded unit, batteries or accessory in accordance with
the environment protection ordinances and regulations.
5. Periodically check the unit for any damage may it be in outer and internal
surface.
6. Avoid using the Pulse Oximeter in an environment with inflammable gases
or in an environment where the temperature or humidity is excessively high
or low.
7. Check the accuracy of the SpO2, PR and PI readings by using an appropriate
calibration apparatus.
C. Storage and Transportation
Temperature
10ºC ~ +50ºC
Keep away
from moisture
Place Upward
Atmospheric Pressure
50kPa ~ 106kPa
Relative Humidity
10% ~ 93%
Non Hook
VIII. Troubleshooting
A. Inaccurate Readings
■Inaccurate measurement may be caused by autoclaving, ethylene,
sterilizing oxide, liquid immersing in the sensors, high-frequency electro
surgical interference.
■Avoid unnecessary body movements to be able to get a stable and an
accurate SpO2, PR and PI values.
■The patient may suffer from hypertension, severe vascular atrophy, severe
anemia or low oxygen.
■The patient is in sudden cardiac arrest or shock state.
■The fingertip nail to be measured is using a fake nail or has a nail polish.
■For longer measurements, periodically change the contact position
between the probe of the Pulse Oximeter and the fingertip.
■Check the integrity of the skin and the blood circulation condition of the
finger.
The functional tester shall not be used to evaluate the accuracy of the probe
of the Pulse Oximeter.
The Pulse Oximeter has a specific calibration curve, and is precise for the
combination of the mainframe and probe of the Pulse Oximeter. If the functional
tester can measure the portion of overall errors of the mainframe-probe system
of the Pulse Oximeter attributable to the mainframe of the Pulse Oximeter, the
functional tester can test the accuracy of a Pulse Oximeter that duplicates the
calibration curve.
PAGE | 09
PAGE | 10

LMK 500 Black
User’s Guide version 1.0 - English ( )
©2016-2017 by ELLEMKA s.à r.l. All Rights Reserved
LMK 500 Black
User’s Guide version 1.0 - English ( )
©2016-2017 by ELLEMKA s.à r.l. All Rights Reserved
ELLEMKA s.à r.l.
BP 56 L-6905 Niederanven Luxembourg/EU
+352 77 9191
ellemka.com
USER’S GUIDE
Pulse Oximeter
LMK-500 D
Oxygen Saturation Pulse Rate Perfusion Index
English-Version 1.0
B. Possible Problems, Causes and Solutions
POSSIBLE
PROBLEM POSSIBLE CAUSE POSSIBLE SOLUTION
The Pulse
Oximeter
fails to
display data
about the
SpO2, PR
and PI.
1. The finger is not placed
at the correct
position inside the probe
of the Pulse Oximeter.
2. The patient does not
concentrate on the
examination.
1. Place the finger at the
correct position.
2. Try many times of
measurement. If you are
sure that the unit is
normally functioning,
advise the patient to seek
for a medical care in time.
The
displayed
value of the
SpO2and PR
fluctuates.
1. The finger is not properly
inserted inside the probe
of the Pulse Oximeter.
2. The finger of the patient’s
body is moving.
1. Completely or properly
insert the finger again.
2. Remind the patient to stay
calm and to not move
the body or finger during
actual measurement.
The Pulse
Oximeter
cannot be
powered on.
1. The batteries are not
properly installed or no
batteries are installed at
all.
2. The batteries are run out
of power supply.
3. The power on button or
the Pulse Oximeter is
damaged.
1. Open the battery
compartment and check if
batteries are properly
installed.
2. Replace the batteries.
3. Refer Customer Support
section for more
details ( section IX).
The screen
is suddenly
off.
1. If the Pulse Oximeter
does not detect a signal
within 10 seconds, it will
automatically power off.
2. The batteries are not
properly installed or no
batteries are installed
at all.
3. The Pulse Oximeter or
the screen is damage.
1. It is normal
2. Replace the batteries
3. Refer to Customer Service
Support section for more
details ( section IX).
IX. Customer Service and Support
The user has no user-serviceable internal components. Try to resolve any
maintenance issues with the product by referring to the Troubleshooting section
( section VIII). If you are unable to resolve the problem, please check the
contact details provided below.
A. Instructions for Amazon customers:
For customers who purchased their Pulse Oximeter medical device in our
Amazon based Online Shop, please contact our after-sales service provider,
namely Amazon Customer Support for any questions or problems you may
have with the device. Amazon Customer Services will provide support and
instructions on how to proceed.
Please follow the instructions provided below:
1. Go to http://www.amazon.co.uk.
2. Log into your Amazon customer account.
3. Scroll down to the “Let us help you” section on the right hand side of the
Amazon UK homepage.
4. Select the subject of interest.
5. Follow the instructions.
B. Instructions for Ellemka Website customers:
For customers who purchased their Pulse Oximeter medical device from
our website, please contact us directly or visit our website and search
relevant solutions from our Support Page.
Please follow the instructions provided below:
1. Go to http://www.ellemka.com
2. Find “Support Page”
3. Choose the specific category on what support are you searching for.
4. Select among the displayed results.
5. Then download the necessary files you will be needing.
6. Otherwise, contact Ellemka Customer Services directly.
PAGE | 11
Table of contents