CONFIGURATION AND ADJUSTMENT
The EE660 as ordered is ready for use immediately and requires no configuration by the user. If required, the optional USB
configuration adapter and the E+E Product Configuration Software (EE-PCS) can be used for changing the factory setup as
well as for adjusting of the air velocity measurement.
Note: The EE660 must not have any additional power supply when using the USB configuration adapter HA011066.
PC
HA011066
For product data sheets EE-PCS please see www.epluse.com.
The E+E Product Configuration Software (EE-PCS) is free and can be downloaded from www.epluse.com/configurator.
TECHNICAL DATA
(Modification rights reserved)
ACCESSORIES
• USB configuration adapter HA011066
• Product configuration software EE-PCS (free download: www.epluse.com/EE660)
• Power supply adapter V03 (see data sheet Accessories)
Measurand
Working range 1) 0...1 m/s (0...200 ft/min)
0...1.5 m/s (0...300 ft/min)
0...2 m/s (0...400 ft/min)
Accuracy at 20 °C 2) (68 °F), 0.15...1 m/s (30...200 ft/min) ± (0.04 m/s (7.9 ft/min) + 2 % of mv)
45 % RH, 1013 hPa 0.15...1.5 m/s (30...300 ft/min) ± (0.05 m/s (9.8 ft/min) + 2 % of mv)
0.15...2 m/s (30...400 ft/min) ± (0.06 m/s (11.8 ft/min) + 2 % of mv)
Response time τ90
3) typ. 4 sec or typ. 1 sec (at constant temperature)
Output
Analogue 0 - 10 V and 4 - 20 mA
0...1 m/s / 0...1.5 m/s / 0...2 m/s 1) -1 mA < IL< 1 mA RL< 450 Ω(linear, 3-wires)
Digital interface RS485 with max. 32 devices on one bus
Protocol Modbus RTU or BACnet MS/TP
General
Power supply (Class III) 24 V AC/DC ± 20 %
Current consumption (max.) AC supply -
no display
DC supply -
no display
AC supply -
with display
DC supply -
with display
Analogue ouput 74 mA rms 41 mA 180 mA rms 85 mA
Digital output 120 mA rms 50 mA - -
Angular dependence < 3% of the measured value at | ∆α | < 10°
Electrical connection screw terminals max. 1.5 mm2(AWG 16)
Cable gland M16x1.5
Electromagnetic compatibility EN61326-1 EN61326-2-3
Industrial Environment
Housing material Polycarbonate, UL94V-0 (with Display UL94HB) approved
Protection class Enclosure IP65 / NEMA4, remote probe IP20
Temperature range working temperature probe -25 ... +50 °C (-13...122 °F)
working temperature electronic -10 ... +50 °C (14...122 °F)
storage temperature -30 ... +60 °C (-22...140 °F)
Working range humidity 5...95 % RH (non-condensing)
1) Selectable by jumper, only for analogue output
2) The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation).
The accuracy was calculated in accordance with EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement).
3) Selectable by jumper (analogue) and slide switch (digital)