ESCO Technologies Powdermax PW1 Series User manual

User
Manual
Thank you for purchasing this Esco Powder Weighing Balance
Enclosure. Please read this manual thoroughly to familiarize yourself
with the many unique features and exciting innovations we have
uilt into your new equipment. Esco provides many other resources
at our we site, www.escoglo al.com, to complement this manual
and help you enjoy many years of productive and safe use of your
Esco products.
For Technical Service, contact
North America
Esco Technologies, Inc.
2940 Turnpike Drive, Units 15-16 • Hat oro, PA 19040, USA
Toll-Free USA and Canada 1-877-479-3726
Tel 215-441-9661 • Fax 215-441-9660
us.escoglo al.com • usa@escoglo al.com
Rest of World
Esco Micro Pte. Ltd.
21 Changi South Street 1 • Singapore 486 777
Tel +65 6542 0833 • Fax +65 6542 6920
www.escoglo al.com • mail@escoglo al.com
Esco
PW1 User
Manual
Version A - Released November 2013
Powder Weig ing Balance Enclosure

Copyright nformation
© Copyrig t 2013 Esco Micro Pte Ltd. All rig ts reserved.
T e information contained in t is manual and t e accompanying product is copyrig ted and all rig ts are
reserved by Esco.
Esco reserves t e rig t to make periodic minor design c anges wit out obligation to notify any
person or entity of suc c ange.
Sentinel™ and Powdermax
®
are registered trademarks of Esco.
“Material in this manual is provided for informational purposes only. The contents and the product descri ed in
this manual (including any appendix, addendum, attachment or inclusion), are su ject to change without
notice. Esco makes no representations or warranties as to the accuracy of the information contained in this
manual. In no event shall Esco e held lia le for any damages, direct or consequential, arising out of or related
to the use of this manual.”

i
Powder Weighing Balance Enclosure
TableofContents
INTRODUCTORY PAGES
i Table of Contents
ii Manual Revision History
iii
Warranty Terms and Conditions
v
Introduction
. Products Covered
2. Safety Warning
3. Limitation of Liability
4. European Union Directives on WEEE and RoHS
vii Declaration of Conformity
USER SECTION
1
Chapter 1
–
Product Information
. Quick View
Chapter 2 – Control System
3 2. Sentinel Control System
4 2.2 Menu Options
7 Chapter - Basic Enclosure Operation
7
3.
Starting and Shutting Down the
Enclosure
7 3.2 Working in the Enclosure
7 3.3 Stopwatch and Experiment Timer
7 3.4 Working Ergonomics
9 Chapter 4 – Service and Maintenance
9
4
. Scheduled Maintenance
9 4.2 Maintenance/Service Log
APPENDIX

ii
MANUAL REVISION HISTORY
REV. NO. DATE
RELEASED DESCRIPTION OF CHANGE REFERENCE SERIAL # OF
ST
UNIT
FOR IMPLEMENTATION

iii
Powder Weighing Balance Enclosure
Warranty Terms and Conditions
Esco products come with a limited warranty. The warranty period will vary depending on the product
purchased, beginning on the date of shipment from any Esco international warehousing location. To determine
which warranty applies to your product, refer to the appendix below.
Esco's limited warranty covers defects in materials and workmanship. Esco's liability under this limited
warranty shall be, at our option, to repair or replace any defective parts of the equipment, provided if proven
to the satisfaction of Esco that these parts were defective at the time of being sold, and that all defective parts
shall be returned, properly identified with a Return Authorization.
This limited warranty covers parts only, and not transportation/insurance charges.
This limited warranty does not cover:
•Freight or installation (inside delivery handling) damage. If your product was damaged in transit, you
must file a claim directly with the freight carrier.
•Products with missing or defaced serial numbers.
•Products for which Esco has not received payment.
•Problems that result from:
oExternal causes such as accident, abuse, misuse, problems with electrical power, improper
operating environmental conditions.
oServicing not authorized by Esco.
oUsage that is not in accordance with product instructions.
oFailure to follow the product instructions.
oFailure to perform preventive maintenance.
oProblems caused by using accessories, parts, or components not supplied by Esco.
oDamage by fire, floods, or acts of God.
oCustomer modifications to the product
•Consumables such as filters (HEPA, ULPA, carbon, pre-filters) and fluorescent / UV bulbs.
•Esco is not liable for any damage incurred on the objects used on or stored in Esco equipment. If the
objects are highly valuable, user is advised to have in place independent external preventive
measures such as connection to a centralized alarm system.
Factory installed, customer specified equipment or accessories are warranted only to the extent guaranteed by
the original manufacturer. The customer agrees that in relation to these products purchased through Esco, our
limited warranty shall not apply and the original manufacturer's warranty shall be the sole warranty in respect
of these products. The customer shall utilize that warranty for the support of such products and in any event
not look to Esco for such warranty support.
Esco encourages all users to register their equipment online at www.escoglobal.com/warranty or complete the
warranty registration form included with each product.
ALL EXPRESS AND IMPLIED WARRANTIES FOR THE PRODUCT, INCLUDING BUT NOT LIMITED TO ANY IMPLIED
WARRANTIES AND CONDITIONS OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE
LIMITED IN TIME TO THE TERM OF THIS LIMITED WARRANTY. NO WARRANTIES, WHETHER EXPRESS OR
IMPLIED, WILL APPLY AFTER THE LIMITED WARRANTY PERIOD HAS EXPIRED. ESCO DOES NOT ACCEPT LIABILITY
BEYOND THE REMEDIES PROVIDED FOR IN THIS LIMITED WARRANTY OR FOR SPECIAL, INDIRECT,
CONSEQUENTIAL OR INCIDENTAL DAMAGES, INCLUDING, WITHOUT LIMITATION, ANY LIABILITY FOR THIRD-
PARTY CLAIMS AGAINST YOU FOR DAMAGES, FOR PRODUCTS NOT BEING AVAILABLE FOR USE, OR FOR LOST
WORK. ESCO'S LIABILITY WILL BE NO MORE THAN THE AMOUNT YOU PAID FOR THE PRODUCT THAT IS THE
SUBJECT OF A CLAIM. THIS IS THE MAXIMUM AMOUNT FOR WHICH ESCO IS RESPONSIBLE.

iv
These Terms and Conditions shall be governed by and construed in accordance with the laws of Singapore and
shall be subject to the exclusive jurisdiction of the courts of Singapore.
Technical Support, Warranty Service Contacts
USA: -877-479-3726
Singapore: +65 65420833
Global Email Helpdesk: support@escoglobal.com
Visit http://www.escoglobal.com/ to talk to a Live Support Representative
Distributors are encouraged to visit the Distributor Intranet for self-help materials.
Product Appendix, Warranty Listings
Biological Safety Cabinets, Laminar Flow Cabinets,
HEPA-Filtered Cabinets (except Streamline brand)
The warranty periods for BSC may vary by country. Contact
your local distributor for specific warranty details.
Laboratory Fume Hoods 2 years limited.
Ductless Fume Hoods 4 years limited for Ascent Opti, 6 years for Ascent Max.
Cleanroom Equipment year limited.
Laboratory Ovens and Incubators
year limited
.
CO
2
Incubators 2 years limited.
Containment/Pharma Products 2 years limited.
Ultralow Temperature Freezer 5 years limited. 60 months on Compressor.
The warranty period starts two months from the date your equipment is shipped from Esco facility for
international distributors. This allows shipping time so the warranty will go into effect at approximately the
same time the equipment is delivered to the user. The warranty protection extends to any subsequent owner
during the warranty period. Distributors who stock Esco equipment are allowed an additional four months for
delivery and installation, providing the product is registered with Esco. User can register product online at
www.escoglobal.com/warranty or complete the warranty registration form included with each product.
Policy updated on
st
January 20 2 (This limited warranty policy applies to products purchased on or after
st
January 20 2)

v
Powder Weighing Balance Enclosure
Introduction
1. Products Covered
Esco PowderMax Powder Weighing Balance Enclosure
Electrical Rating 0.9 meters ( feet)
220-240 V AC, 50Hz, 1Φ
PW
-
3A
110-1 0 V AC, 60Hz, 1Φ
PW
-
3A
2
220-240 V AC, 60Hz, 1Φ
PW
-
3A
3
100 V AC, 50/60 HZ, 1Φ
PW
-
3A7
2. Safety Warning
•Anyone working with, on or around this equipment should read this manual. Failure to read,
understand and follow the instructions given in this documentation may result in damage to the unit,
injury to operating personnel, and / or poor equipment performance.
•Any internal adjustment, modification or maintenance to this equipment must be undertaken by
qualified service personnel.
•The use of any hazardous materials in this equipment must be monitored by an industrial hygienist,
safety officer or some other suitably qualified individual.
•Before you process, you should thoroughly understand the installation procedures and take note of
the environmental / electrical requirements.
•In this manual, important safety related points will be marked with the symbol.
•If the equipment is used in a manner not specified by this manual, the protection provided by this
equipment may be impaired.
3. Limitation of Liability
The disposal and / or emission of substances used in connection with this equipment may be governed by
various local regulations. Familiarization and compliance with any such regulations are the sole responsibility
of the users. Esco’s liability is limited with respect to user compliance with such regulations.
4. European nion Directive on WEEE and RoHS
The European Union has issued two directives:
• Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE)
This product is required to comply with the European Union’s Waste Electrical &
Electronic Equipment (WEEE) Directive 2002/96/EC. It is marked with the following
symbol:
Esco sells products through distributors throughout Europe. Contact your local Esco
distributor for recycling/disposal.
• Directive 2002/95/EC on Restriction on the use of Hazardous Substances (RoHS)
With respect to the directive on RoHS, please note that this hood falls under category 8 (medical devices) and
category 9 (monitoring and control instruments) and is therefore exempted from requirement to comply with
the provisions of this directive.

vi

vii
Powder Weighing Balance Enclosure
Declaration of Conformation
In accordance to EN ISO/IEC 7050- :20 0
We, Esco Micro Pte. Ltd.
of 2 Changi South Street
Singapore, 486777
Tel: +65 6542 0833
Fax: +65 6542 6920
declare on our sole responsibility that the product:
Category : Powder Weighing Balance Enclosure
Brand : PowderMax
Model : PW1- A1
in accordance with the following directives:
2006/95/EEC : The Low Voltage Directive and its amending directives
92/ 1/EEC : The Electromagnetic Compatibility Directive and its amending
directives
has been designed to comply with the requirement of the following Harmonized Standard:
Low Voltage : EN 61010-1:2010
EMC : EN 61 26-1:2006 Class B
More information may be obtained from Esco’s authorized distributors located within the European Union. A
list of these parties and their contact information is available on request from Esco
.
_______________________________
XQ Lin
Group CEO, ESCO
This Declaration of Conformity is only applicable for 230V AC 50Hz units

viii

1
Powder Weighing Balance Enclosures
Chapter 1 - Product Information
1.1Quick View
1. Sash Window
2. Baffle
3. ArmOpening
4. Sentinel ontrol System
5. Service Opening
6. Mobile art (Optional)
6
1
2
3
4
5

2

3
Powder Weighing Balance Enclosures
Chapter 2–Sentinel Control System
2.1 Sentinel Control System
1. Fan Button
oTurns on and turn off the fan.
2. Lamp Button
oTurns on and turn off the fluorescent lamps.
3. Socket Button
oNot used.
4. UV/AUX Button
oNot used.
5. Up (▲) and Down (▼) Arrow Button
oMove the menu options upwards and downwards.
oIncrease and decrease corresponding value inside one of the menu options.
oAccessing the stopwatch and experiment timer function.
6. Set or Mute Button
ohoose the menu or sub-menu currently displayed on the L D screen.
oProceed to the next step or sequence inside one of the menu options.
7. Menu Button
oTo enter and exit from the menu options.
oTo go back to the previous level of the menu options.
oTo access maintenance mode from error condition.
Menu
Button
Set
Button
Up
Button
Down
Button
L D
Display
Fan
Button
Lamp
Button
Socket
Button
UV/AUX
Button
When you are entering menu options, the alarm will sound to indicate that the microprocessor is
not monitoring the operation of the
enclosure
. No further warnings will be given.

4
2.2 Menu Options
Please refer to the following diagram for complete reference to all menu options available.
MENU
SETTINGS
ALIBRATION
ADMIN
SET MODE
SET TIME
WARM UP
POST PURGE
EXP TIMER
VELO ITY UNIT
TEMP UNIT
SET ONSTANT
ZERO SENSOR
ALIB SENSOR
NEW ADMIN PIN
NEW FAN PIN
A/F MONITOR
RESET B/H/M
NORMAL
RESET DEFAULT
MAINTENAN E
HH:MM
XX MIN
HH:MM:SS
M/S | FPM
EL IUS | FAHRENHEIT
XX MIN
XXXX
XXXX
ENABLE | DISABLE
2.2.1 Settings
Users may use the settings menu function to customize the operation of the enclosure to meet specific
application requirements.
2.2.1.1 Set Clock (Time)
Users can set the time by increasing/decreasing the hour and minute values. The correct time will be
maintained even after the unit is turned off.
2.2.1.2 Warm up Time
There will be a period of warm-up, before the fan is fully functioning. This is to ensure that the sensors, the
blower, and the control system are stabilized, as well as purging the work zone of contaminants. The default
setting is 3 minutes and the user can set it between 3 to 15 minutes.
MENU SETTINGS WARM UP TIME XX MINUTES

5
Powder Weighing Balance Enclosures
2.2.1.3 PostPurge Time
After the user switches off the enclosureblower, there will be a post-purge period, to ensure that all
contaminants are purged from the work zone. The default setting is zero minute (disable) and user can set it
between 0 to 15 minutes. However, it is recommended to purge the enclosure for around 3 minutes after the
work is complete.
MENU SETTINGS POST PURGE XX MINUTES
2.2.1.4 Experiment Timer
Experiment timer is a countdown timer that can be used for critical experiment. Experiment timer can be set
between “00:00:00” and “17:59:59”.
MENU SETTINGS EXP TIMER HH:MM:SS
2.2.1.5Air Velocity Unit Selection
Using this option, the user can select the unit in which air velocity is measured and displayed. The user can
choose between metric (m/s) and imperial (fpm).
MENU SETTINGS VELOCITY UNIT
FPM
M/S
2.2.1.6Temperature Unit Selection
Using this option, the user can select the unit in which temperature is measured and displayed. The user can
choose between elsius and Fahrenheit.
MENU SETTINGS TEMP UNIT
FAHRENHEIT
CELSIUS
2.2.2 Calibration
The purpose of calibration is to ensure the accuracy of the airflow display and alarm (if present). This involves
measuring airflow with reference instrumentation and establishing reference between airflow sensor(s) on the
enclosureto the standard reference. alibration should only be carried out by trained personnel. This section
presents a brief overview of the calibration menu function. For more information, refer to test report.
SET CONSTANT
MENU CALIBRATION ZERO SENSOR
CALIB SENSOR
2.2.2.1 Set Constant
Every sensor manufactured by Esco has a specific sensorconstant which is used for temperaturecompensation
performed by the temperature sensor.
2.2.2.2 Zero Sensor
This option is to let the controller record a specific sensor output voltage and correspond it to 0cmh or 0cfm.
2.2.2.3Calib Sensor
This option allows proper calibration and operation of the airflow sensor alarm. There are two points that have
to be calibrated, namely airflow fail point and airflow nominal point
2.2.3 Admin Settings
The admin menu allows you to change both FAN and ADMIN PIN. The reset blowermeter functions are usually
used after you change the bloweras they can easily give you the indication on when to do maintenance. While
the reset default function will return the options in the settings menu to their factory settings.

6
2.2.3.1 New ADMIN PIN (default 0009)
ADMIN PIN restricts access to some of the more delicate menu functions, namely admin and field calibration,
which should only be accessed by a qualified personnel. User must enter four digits ADMIN PIN before accessing
these menus.
ADMIN PIN can also be used to switch to maintenance mode from error condition.Setting the PIN to 0000 will
disable this feature.
MENU ADMIN NEW ADMIN PIN XXXX
2.2.3.2 New FAN PIN (default 0000 DISABLED)
FAN PIN restricts access to fan control and some parts of the menu, settings and set mode. User must enter
four-digit PIN before switching fan on or off. It is recommended that the Fan PIN be issued only to personnel
authorized to use the enclosure.
Setting the PIN to 0000 will disable this feature.
MENU ADMIN NEW FAN PIN XXXX
2.2.3.3 Airflow Monitor
Whenever the air velocity falls below the fail point, the air fail alarm will be triggered. This option is used to
enable/disable alarm. The alarm is enabled by default.
MENU ADMIN A/F MONITOR
DISABLED
ENABLED
2.2.3.4 Reset B/H/M
This option is used to reset the blower hour meter. The blower hour meter indicates how long the blower has
been in operation. This value can also provide some help in setting up maintenance schedule.
2.2.3.5 Reset Default
User can reset the default setting by choosing this option. The features being reset are warm-up period (3
minutes), post-purge period (0 minute), UV timer (60 minute), measurement unit (Metric), Airflow monitor
(enabled), ADMIN PIN (0009), and FAN PIN (0000).
Note that the calibration settings cannot be reset as it may cause the enclosureto operate in an unsafe
manner. The hour meters cannot be reset either.
2.2.4 Setting Mode
PW1 has two working mode; normal mode and maintenance mode.
MENU SET MODE
NORMAL
MAINTENANCE
2.2.4.1 Normal Mode
The normal mode is activated by factory default. In this mode, all alarms and interlocks are enabled.
2.2.4.2 Maintenance Mode
Maintenance mode should only be accessed by qualified personnel during maintenance. In this mode, all
alarms are disabled and all interlocks are defeated.

7
Powder Weighing Balance Enclosures
Chapter 3 – Basic Enclosure Operation
3.1 Starting and Shutting Down the Enclosure
3.1.1 Turning on the Enclosure
1. Turn on the fan by pressing the FAN button. Input the Fan PIN if asked (if PIN ≠ 0000). This will start
the warm up procedure (default: 3 minutes). All buttons are diabled during warm up period.
2. The enclosureis ready for work.
3.1.2 Turning off the Enclosure
1. Turn off the fan by pressing the FAN button. Input the Fan PIN if asked (if PIN ≠ 0000). This will start
the post purge procedure (default: 0 minute). All buttons are disabled during post purge period.
3.2 Working in the Enclosure
•Surface-decontaminate the work area before and after using the enclosure.
•Allow the enclosureto purge any contaminant by allowing the blower to operate at least 3 minutes
before and after using the enclosure.
•Minimize room activity (personnel movements, closing and opening of doors, etc.) since these
external airflow disturbances may adversely affect the enclosure’sinternal airflow, thereby possibly
impairing the containment capabilities of the enclosure.
•While working in the enclosure, move your hands slowly and in a controlled manner.
•PW1 uses alarms to indicate that the condition inside the enclosure is not safe for the operator, so
check the L D display to understand the cause of these alarms.
3.3 Stopwatch and Experiment Timer
•The stopwatch function can be started by pressing the UP button while the sash is closed. Pressing UP
button again while the stopwatch function will stop and resume the timer. Pressing DOWN button will
leave the stopwatch function and reset the timer. The timer in the stopwatch function is counting up
and shown using the HH:MM:SS format.
•The experiment timer can be started by pressing the DOWN button while the sash is closed. Pressing
DOWN button while the experiment timer function is working will stop and resume the timer.
Pressing the UP button will leave the experiment timer function and reset the timer. The timer in the
experiment timer function is counting down and shown using the HH:MM:SS format. Operator can
use the SETTINGS | EXP TIMER menu (refer to section 2.2.1.4) to set the experiment timer.
3.3 Working Ergonomics
On most occasions, you would most likely be operating the enclosurein sitting rather than standing posture.
There are some obvious advantages of the sitting posture:
•The physiological energy cost and fatigue involved in sitting are relatively less
•Sitting posture provides the body with a stable support
However, sitting position has some drawbacks too:
•The working area available is fairly limited
•There is a potential risk of being constrained in the same posture for a long time
•Sitting posture is one of the most stressful postures for one’s back

8
Therefore you should pay careful attention to the following guidelines in order to achieve comfortable and
healthy working conditions:
•Always ensure that your legs have enough legroom.
•Keep your lower back comfortably supported by your chair. Adjust the chair or use a pillow behind
your back whenever necessary.
•You should place your feet flat on the floor or on a footrest. Don’t dangle your feet and compress
your thighs.
•You should keep varying your sitting position throughout the day at regular intervals so that you are
never in the same posture for too long.
•Observe the following precautions with respect to your eyes:
oGive your eyes frequent breaks. Periodically look away from the work area and focus at a
distant point.
oKeep your glasses clean.
•Arrange the items/apparatus frequently used in your work in such a way that you can minimize the
physical strain involved in handling them.
•Exercise regularly
The enclosure’snoise emission has been tested and found to be in compliance with EN 12469, ISO 4871 and
NSF/ANSI 49 which is important to ensure health and comfort for the operator.
Ergonomics accessories available with Esco include:
•Armrest padding
•Lab chair
•Footrest
Please contact your local distributor or Esco for more information.

9
Powder Weighing Balance Enclosures
Chapter 4–Ser ice and Maintenance
4.1 Scheduled Maintenance
Proper and timely maintenance is crucial for trouble free functioning of any device and your Esco enclosureis
no exception to this rule. We strongly recommend that you follow the maintenance schedule suggested
hereunder in order to obtain optimal performance from your Esco enclosure.
No. Description of Task to Perform Maintenance to be carried out every
Week Month Quarter 1 Year
1
lean the interior work surface
,
walls
and sash
√
2 Inspect alarm and measure basic airflow √
3 Inspect the controls of the enclosure √
4 lean the exterior surfaces of the enclosure √
5 heck for any blockages at the hood baffles √
6 Inspect the enclosure’s face velocity √
7 Inspect the enclosure for any physical abnormalities or malfunction √
8 hange the fluorescent lamps √
9 Re-certification √
C eaning the enc osure
•lean the work surface and walls with appropriate disinfectant agent and soap water afterward
•lean the sash window using an appropriate disinfectant agent and glass cleaner afterward
•Use a damp cloth to clean the exterior surface of the enclosure, particularly on the front and top in
order to remove dust that accumulated there
•Use clean water to finish the cleaning and wash away any residue of disinfectant agent, soapy water
and glass cleaner.
Check the enc osure’s functiona ity
•heck the enclosure’s mechanical functionality; ex: sash window – lubricate if necessary
•heck the enclosure’s electrical functionality; ex: fluorescent lamp – replace if necessary
•heck the enclosure’s for any defect, repair immediately
Re-certification
All enclosuremust be re-certified annually by a certified engineer. See test report for recertification procedure.
4.2 Maintenance/Service Log
It is good practice (and in some cases regulatory requirement) to maintain a log of all maintenance work
carried out on your enclosure.

10
This manual suits for next models
4
Table of contents
Popular Accessories manuals by other brands

Samsung
Samsung SIA-0100Q user manual

Observator Instruments
Observator Instruments ANALITE NEP-5000-LINK manual

Hornbach
Hornbach 10252573 Installation and user manual

American Standard
American Standard Ariana 6021 installation instructions

Fluke
Fluke 3562 Getting started guide

Takex
Takex PB-30TK instruction manual

Screen Technics
Screen Technics ELECTRICINEMA installation instructions

Chacon
Chacon 84164 quick start guide

Decagon Devices
Decagon Devices GS3 manual

Microsonic
Microsonic crm plus 25/DIU/TC/E operating manual

Tri-Tronics
Tri-Tronics Mini Label Eye installation manual

Pepperl+Fuchs
Pepperl+Fuchs VIM8 Series manual