ESD AE-700 Operator's manual

OPERATION MANUAL
INFARED NON-CONTACT
THERMO-CHECKER
[INTRODUCTION]
Thank you for using this non-contact Infrared
Thermo-Checker, AE 700 is a special designed for
monitoring forehead temperature range 32 ℃ to
43 ℃ (89℉to109℉). It has the functions of
ambient temperature compensation and self-
detection. It requires no contact, no earmuffs, no
doubt about cross-infection of the ear canal, and
inaccuracy caused by errors in the measurement
angle of the ear canal.
[WARNING]
To get an accurate reading, the best operating
ambient temperature range is around room
temperature 25 degree ℃+/-3℃.
Keep your forehead dry and clean, no hair is
covering to get the accurate measurements. The
subject does not take a bath or exercise within 30
minutes to ensure accurate measurement.
If you use environmental state conversion and
the temperature changes greatly, please place the
measuring instrument in the environment to be
measured for 20 to 40 minutes. After the sensor
has stably adapted to the ambient temperature, a
more accurate value can be measured. Do not use
a sharp thing to touch the sensor lens.
The data is only for tracking reference, If you
find a temperature error, double check with a
clinical thermometer.
It’s important to study all instruction and
warning before using.
[PARTS DESCRIPTION]
1. Test switch: Turn on the machine. When the
switch is released, it will turn to "H" and the test
display, and the data will be automatically
maintained.
2. Celsius temperature switch: After entering the
calibration mode, the temperature increase button
increases by 0.1 ℃ each time
3. Fahrenheit temperature switch: After entering
the adjustment mode, the temperature increase
button will decrease by 0.1 ℃ each time.
4. Indicator light switch: Press it once, the
indicator light is on, then press the indicator light
off.
5. Beep switch: Press the buzzer once to turn it
on, then press the buzzer off again.
6. Back light on/off button: Pressing the button
turns the backlight on, pressing it again turns it
off.
7. LCD Display.
8. Battery door: When replace battery door,
please using the finger indents to pull open the
battery door.
[OPERATION]
1. Environmental description:
please measure under room temperature around
22℃~28℃, recommend not to use in the sun or
cooling room. and make sure the measurement
area is clean and dry. Better not to measure in 10
minutes after sports, shower or heavy sweat to get
accurate measurement. Before use the product, it
is recommended that the measurement person take
his body temperature to measure his own body
temperature, and make a reference comparison
with the amount of temperature measured by this
product.
2. Measurement
Aiming toward the forehead you want to measure
as near as 1-2cm(not to touch), Press the
measurement button and the temperature
measurement value is immediately displayed on
the LCD screen. Before releasing the button, the
number will no longer change, which is the
highest measured temperature. Release the button
to save the temperature value. After automatic
shutdown and restart, the maximum value will be
deleted.
3. Auto Power off: The device will be power off
in 5 second inactivity, this design is for saving
power and lasting battery life.
4. Replacing battrey: One 9V battery is supplied
with the device, when you see a battery icon
appears on the left top of the screen, you have to
replace the battery.
Open the battery cover, take off old battery and
replace with a new one, then close.
※Notes:
1. A person's body temperature changes at
different times of the day, and is also affected by
other external conditions, such as age, gender, and
skin color.
Age Normal body
temp℃Normal body
temp℉
0-2 36.4-38.0 97.5-100.4
3-10 36.1-37.8 97.0-100.0
11-65 35.9-37.6 96.6-99.7
>65 35.8-37.5 96.4-99.5
2. This machine provides a calibration mode (after
removing the battery for 10 seconds, press and
hold the backlight button No. 6 for 3 seconds at
the same time, and then install the battery. Do not
release these two buttons during the battery
installation. After the battery is released, the
screen displays 00.0 after releasing the two
buttons. At this time, the temperature can be
increased or decreased by pressing the button No.
2 or No. 3. After adjusting the temperature, unplug
the battery and install the battery again to enter the
normal measurement mode. The increased or
decreased temperature indicates that the measured
value will increase or decrease accordingly. For
example, the measured temperature is 36.5
degrees. After entering the adjustment mode and
reduced by 0.5 degrees, the normal measurement
mode is restored, and the measured temperature is
36 degrees.) There is an error between the
standard measurement value (preferably measured
with a mercury thermometer), and you can enter
the adjustment mode to make adjustments.
Adjust the temperature range to ± 3℃.
[TROUBLE SHOOTING]
This device is designed to get an accurate
measurement for self-diagnoses on Forehead
temperature within 32℃ to 43℃(89℉to109℉).If
you intend to measure other objects other than
human forehead, it won’t be recommend as though
you may be able to see a reading under 32℃(89℉)
or above 43 ℃ (109℉), the reading won ’ t be
referred. “Lo”or “Hi” error message will appear
once the measurement exceeds the temperature
range.
The information “LO” or “HI” appears in the
following situations for reference:
Reason for the
message "LO" or
"HI"
Recommended
The temperature
value is affected by
hair and sweat
Make sure that there is no
obstruction during
temperature measurement
Temperature is
affected by air flow
change
Make sure that the air
remains stable during
temperature measurements
The measuring
distance is too far Note the measuring
distance (1-3) cm
Walk into the room
from low or high
temperature
outdoors.
Wait for 30 minutes, after
the subject's temperature
adapts to the measurement
environment, then measure
[SOECIFCATION]
Measure range: 32℃~43℃(89 to109℉)
Resolution:0.1℃
Accuracy:
32-34.9℃/
89.6-94.8℉±0.3℃
(0.5℉)
35-42℃/
95-107.6℉±0.2℃
(0.4℉)
42.1-43.0℃/
107.7-109.4℉±0.3℃
(0.5℉)
Response time:500 ms
Measure temp:16-35℃
Supplied:9V battery and operation
manual
Optional accessory: Carrying pouch
Device size:165×72×41mm
Device weight:155g
[WARRANTY]
The product is warranted to be free from defects in
material and workmanship for a period of one year
from the date purchase. This warranty covers
normal operation and dose not cover battery,
misuse, abuse, alteration, tampering, neglect,
improper maintenance, or damage resulting from
leaking batteries.
Above picture and content just for your
reference. Please be subject to the actual
products if anything different or updated.
Please pardon for not informing in advance.

Popular Thermometer manuals by other brands

York Survey Supply 32058 operating instructions

York Survey Supply

York Survey Supply 32058 operating instructions

Withings Thermo Quick installation guide

Withings

Withings Thermo Quick installation guide

Milwaukee C12 LTGH Original instructions

Milwaukee

Milwaukee C12 LTGH Original instructions

Emos E8860 manual

Emos

Emos E8860 manual

Emos E0129 manual

Emos

Emos E0129 manual

Physio Logic DiGiPro user guide

Physio Logic

Physio Logic DiGiPro user guide

OWIM 279818 Operation and safety notes

OWIM

OWIM 279818 Operation and safety notes

Levenhuk Wezzer PLUS LP30 user manual

Levenhuk

Levenhuk Wezzer PLUS LP30 user manual

Sea-Bird Electronics SBE 38 user manual

Sea-Bird Electronics

Sea-Bird Electronics SBE 38 user manual

Fluke 574 user manual

Fluke

Fluke 574 user manual

inventum TMC609 instruction manual

inventum

inventum TMC609 instruction manual

Hanna Instruments Pronto Plus HI99551 Series user manual

Hanna Instruments

Hanna Instruments Pronto Plus HI99551 Series user manual

Williams-Sonoma MAV219 user manual

Williams-Sonoma

Williams-Sonoma MAV219 user manual

ident-a-kid Temp Chek Plus Assembly instructions

ident-a-kid

ident-a-kid Temp Chek Plus Assembly instructions

Omron i-Temp manual

Omron

Omron i-Temp manual

Ambient Weather WH31P user manual

Ambient Weather

Ambient Weather WH31P user manual

Ginger Hill Thermee instruction manual

Ginger Hill

Ginger Hill Thermee instruction manual

Omega OS540 instruction sheet

Omega

Omega OS540 instruction sheet

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.