Euromag MC 406 User manual

1
MC 406 CONVERTER
USE AND MAINTENANCE MANUAL

2
SUMMARY
1. INTRODUCTION…………………………………………………………………………………………...……3
1.1 MANUFACTURER’S NOTICE………………………………………………………………………………..4
1.2 PACKAGING CHECK……………………………………………………………………………………..… 4
1.3 PRELIMINARY NOTES…………………………………………………………………………………….…4
1.4 PRODUCT IDENTIFICATION………………………………………………………………………………..4
2 PRODUCT DESCRIPTION……………………………………………………………….…………………….5
2.1 GENERAL DESCRIPTION………………………………………………………………….………………..5
2.2 COUPLING……………………………………………………………………………………….…………….5
2.3 MEASUREMENT CHARACTERISTICS………………………………………………………….…………5
2.4 DATA STORAGE……………………………………………………………………………………….……..5
2.5 DISPLAY………………………………………………………………………………….……………………5
2.6 BATTERY……………………………………………………………………………….……………………..6
3 CONVERTER……………………………………………………………………………………………………7
3.1 SYSTEM START…………………………………………………………………………………..…….……7
3.2 USER INTERFACE……………………………………………………………………………..…….………8
3.3 DISPLAY OF THE TOTALIZER OVERFLOW COUNTER…………………………………...…………10
3.4 CELULAR COMMUNICATION………………………………………………………………….…………10
3.5 FIRST ACTIVATION……………………………………………………………………………….………..11
3.6 DISPLAY TEST………………………………………………………………………………….…………..11
3.7 FUNCTIONS…………………………………………………………………………………………………12
4. BATTERY LIFE………………………………………………………………………………………….……16
5. AUTO DIAGNOSTICS………………………………………………………………………………….……17
6. PC INTERFACE SOFTWARE………………………………………………………………………………18
6.1 FIRMWARE UPDATE………………………………………………………………………………………19
6.2 LOG OF DOWNLOAD ATTEMPT…………………………………………………………………………20
7. GSM……………………………………………………………………………………………………………21
7.1 MAIN FEATURES…………………………………………………………………………………………..22
7.2 GSM CONNECTION………………………………………………………………………………………..23
7.3 DEVICE SETTINGS………………………………………………………………………………………...24
7.4 PHONE NUMBER AND EMAIL SETTINGS……………………………………………………………..25
7.5 SCHEDULING…………………………………………………………………………………………….…26
7.6 DATA SAVING…………………………………………………………………………………………….…27
7.7 TEST………………………………………………………………………………………………………….28
8. 4-20mA OUTPUT MODULE…………………………………………………………………………………29
9. BLUETOOTH –MODBUS MODULE………………………………………………………………………30
9.1 BLUETOOTH COMMUNICATION………………………………………………………………………..31
9.2 PC CONNECTION THROUGH SOFTWARE INTERFACE…………………………………………...32
10. CONVERTER INSTALLATION………………………………………………………………………..….39
10.1 GROUNDING RECOMMENDATION……………………………………………………………...…...34
10.2 SEPARATE CONFIGURATION…………………………………………………………………………35
11. DISPOSAL OF THE CONVERTER………………………………………………………………………36
12. REPAIR REQUEST FORM (RMA)… ……………………………………………………………………37

3
1. INTRODUCTION
IMPORTANT WARNING
IT IS VERY IMPORTANT THAT ALL THE STAFF OPERATING THE EQUIPMENT HAVE READ THE
INCLUDED THE INSTRUCTIONS AND INDICATIONS PROVIDED IN THIS MANUAL AND THAT FOLLOWS
THEM BEFORE USING THE EQUIPMENT ITSELF
THE MANUFACTURER DOES NOT ASSUME ANY LIABILITY FOR THE CONSEQUENCES ARISING FROM
IMPROPER USE BY THE OPERATOR
The operator is responsible for the suitability of the device for the specific purpose.
Inadequate installation and use of the devices (systems) voids the warranty.
• The manufacturer will not be liable for any kind of damage resulting from the use of its products, including, but not limited
to those deemed direct, indirect, incidental, punitive and consequential. The installation, connection, commissioning and
maintenance must be performed by staff specifically qualified and authorized for that purpose.
• The installation personnel must ensure that the measuring system is correctly connected in accordance with the wiring
diagram.
For applications that require high working pressures or substances that may be hazardous to the people, the environment,
equipment or anything else if a pipe breakage occurs, Euromag International recommends, that before installing the
sensor the operator takes the appropriate precautions such as precautions such as the correct location, protection or the
mounting of a guard or safety valve. The device contains live electrical components, thus the installation, monitoring and
maintenance must be carried out by qualified and experienced staff fully aware of all the necessary precautions. Before
opening any inner part, please disconnect the power supply. The flow meter consists of metal and plastic parts, all of
which must be in compliance with local rules and requirements concerning waste disposal.
1.1 MANUFACTURER’S NOTICE
•While designing the device stresses and loads possibly caused by earthquakes, traffic, strong winds and fire
damage were not taken into account.
•Do not install the device so that it acts as a focus for pipeline stresses. In the configuration of the device please
take into account any external loads.
•During operation do not exceed the pressure and/or temperature ratings indicated on the nameplate or in this
Operating Manual.

4
1.2 PACKAGING CHECK
At the time of purchase and / or receipt of the product, the buyer is strongly invited to check the quality of the packaging
which must be intact, free of obvious signs of dents and completely and correctly closed.
When opening the packaging, also check that the product in regards of quality and verify all components / accessories as
reported in the package.
1.3 PRELIMINARY NOTES
The main parts that make up the electromagnetic flow meter are:
• The sensor - is installed in the pipes with the help of flanges or threaded connections.
• The converter - can be installed on the sensor (in the compact version) or in its vicinity (in the remote version) connected
by two cables.
Electromagnetic flow meters have many important advantages over their mechanical counterparts, which include among
others: exceptional long-term stability, maximum process reliability, zero maintenance. As a result, these sensors can
provide accurate and reliable measurements for many years.
See the following paragraphs for more detailed indications on correct use and installation.
1.4 PRODUCT IDENTIFICATION
An adhesive identification plate is attached to each MC406 Converter produced by Euromag International
on which the following are reported.
•Model: sensor and converter model
•CE conformity mark
•S/N: serial number that identifies the converter and the sensor
•Y: year of construction
•Size: nominal diameter, standard and nominal pressure flanges
•Q3: nominal flow rate and ratio R (Qnom / Qmin)
•Power Supply: supply voltage and / or battery
•MAP: instrument nominal pressure
•Press. loss class: max pressure drop class in the sensor
•Env. class: environmental class
•EMC class: Electromagnetic Compatibility class
•T: totalizer that cannot be reset
P: resettable partial totalizer
THE IDENTIFICATION PLATE MUST NOT BE REMOVED, DAMAGED OR IMPAIRED. IT MUST ALSO BE KEPT
CLEAN FROM DIRT AND MISCELLANEOUS ADHESIONS AS THE DATA CONTAINED IN IT REPRESENTS THE
ONLY SAFE AND UNIQUE WAY TO RECOGNIZE THE TYPE OF CONVERTER IN YOUR POSSESSION WHICH IS
INFORMATION NEEDED TO FILL IN THE REPAIR REQUEST FORM ATTACHED IN THIS MANUAL.

5
2 PRODUCT DESCRIPTION
2.1 GENERAL DESCRIPTION
MC406 is an electronic converter that, paired with an Euromag electromagnetic sensor, is able to offer
high precision and stability in compact dimensions. The measured flow rate is displayed on the LCD screen together with
one of the 4 totalizers available; other information and some basic settings are directly accessible using the 4-button
interface.
The two independent positive and negative pulse outputs allow connection with any external meter (max 30 V ac or dc /
100 mA). The required power supply is between 12 and 24 V DC and a battery pack can be connected to attain
uninterrupted service even in the event of a power failure. A GSM module can also be installed to allow remote
transmission of measurements *
2.2 COUPLING
The maximum diameter of the sensor that can be used with this converter is 600 mm; backup battery life is affected by the
size of the sensor.
The flow meter can be ordered both in a compact and separate version, with a maximum length of 30mt cable.
2.3 MEASUREMENT CHARACTERISTICS
Flow speed range: from 0.015 m/s to 10 m/s
Liquid conductivity> = 20uS / cm
Sampling: 3.125Hz nominal powered unit (depends on the diameter); battery mode (low power): from 1/5Hz at 1/60Hz
(default 1/15Hz)
2.4 DATA STORAGE
All parameters, totalizers and the register are stored in a non-volatile memory.
2.5 DISPLAY
The LCD can display an 8 and a 6-digit number plus different information icons, allowing the user to view different
information and set many parameters. The details of what can be shown:
•Instant flow
•Positive totalizer (T+)
•Negative totalizer (T-)
•Partial Positive (P+)
•Partial negative (N-)
•Date and time
•Converter temperature
•Code and value of the corresponding parameter
* the data transmitted remotely are not considered legally relevant and have the sole purpose of providing one
remote viewing. The integrity and correctness of the data transmitted are not covered by the MID certification.

6
2.6 BATTERY
Lithium batteries are primary energy sources with high energy content, and are designed to satisfy the highest possible
safety standards. However, they can pose a potential hazard if are subject to bad electrical or mechanical use. In many
cases, this is associated with a production of excessive heat, in which the increased internal pressure could cause the cell
to break.
The following precautions should therefore be observed when handling and using lithium batteries fundamental:
•Do not short circuit, charge, overcharge or connect to incorrect polarity
•Do not expose the battery to temperatures higher than those specified or incinerate it
•Do not crush, puncture or open cells or disassemble battery packs
•Do not solder or braze on the battery body or battery packs
•Do not expose the contents to water
Lithium batteries are regulated under the United Nations Model Regulations on Transport of Dangerous goods (United
Nations model rules on the transport of dangerous goods), UN ST/SGAC document. 10-1, 12th edition updated, 2001.
The UN document n. 3091 class 9 covers lithium batteries packed with equipment or within it. The UN document no. 3090
class 9 covers the transport of stand-alone batteries.
The following basic precautions should therefore be followed when transporting lithium batteries:
•Transport only in special packages with special shipping labels and documents
•Use caution when handling, transporting and packing to avoid shorting the batteries.
•The gross mass of the package must be limited according to the type of transport. In general, a gross mass of
less than 5 kg is acceptable for all forms of transport.
•The batteries comply with the requirements stated in "UN Manual of tests and criteria, Part III, subsection 38.3" for
transport by air and in accordance with ADR regulations for transport by truck / ship
REMOVE THE BATTERY FROM THE TRANSMITTER BEFORE SENDING THE
FLOWMETER TO EUROMAG INTERNATIONAL IN THE EVENT OF MAINTENANCE
OR INTERVENTION IN WARRANTY

7
3 CONVERTER
3.1 SYSTEM START
Once the physical installation of the instrument is complete, the electronic converter can be switched on by connecting the
12/24V power supply and/or by connecting the battery pack to the connector. Different information is displayed to the user
and a basic configuration is required
BOOTLOADER VERSION AND CHECKSUM
The bootloader is a separate software required to download new firmware to the
converter.
FIRMWARE VERSION AND CHECKSUM
This refers to the main firmware
BATTERY REPLACEMENT NOTIFICATION
If the unit is powered by a new battery pack, it is necessary to select «YES» by
pressing P4 to reset the battery energy meter; if the unit has just been disconnected
and turned back on with the same battery pack, the correct selection is "NO" (P1).
DATE AND TIME SETTING
Check and if necessary, adjust the date and time
The flashing value is the one that can be adjusted using P3 / P4.
P2 selects the next value in the sequence: year> month> day> hour> minute.
After selecting the last value, a further press of P2 confirms the value and go to the
next screen
P1 confirms the current settings and exits
MAIN DISPLAY SCREEN
The converter is now ready.
The positive totalizer appears on the first line and the instantaneous flow rate on the
second
The current technical unit is shown on the right side.

8
3.2 USER INTERFACE
Starting from the main display screen (flow rate and total positive total), the 4 buttons can be used to perform various
operations. In detail:
Scroll through the display options of the first and second lines
P3: scrolls the instantaneous flow rate, T+ in maximum resolution, T- in maximum resolution, the time and the board
temperature
P4: scrolls the totalizers (reduced resolution) and the date in this order: T +> T-> P +> P-> date

9
P2: scrolls the number of active alarms (first row) / total alarms (second row) In a normal working condition no alarms will
be signaled.
If one (or more) alarm condition is registered, it will be scrolled
P1: entering the password to access the function selection (the preset value is 000000)
To increase / decrease the current value of the digit use P4 / P3; go to the next digit with P2.
When the last digit is entered, a further press of P2 will unlock access to the function selection if the entered password is
correct, otherwise the figure will return to the first one
Function selection code: each value corresponds to a menu or parameter set function (see FUNCTIONS); for selecting
the function code use P4 / P3; to enter the selected code use P2
Pressing P1 returns to the main screen.
Within a specific function, the functionality of the buttons may be different (see FUNCTIONS); in this example, P4/P3
increases and decreases the value of the parameter shown on the first line, P2 confirms the value and returns when the
function is selected, P1 stops the operation and returns to the main screen

10
3.3 DISPLAY OF THE TOTALIZER OVERFLOW COUNTER
If the totalizers reach the saturation value of the display, they perform the rollover and continue to count; pressing P2 and
P3 together while one of the four totalizers is displayed will show its overflow counter
3.4 CELULAR COMMUNICATION*
Some basic information on cellular activity is shown directly on the MC406 display; Here are the details:
* available only if optional cellular hardware is installed

11
3.5 FIRST ACTIVATION
Depending on the instrument configuration, the converter can be shipped with the battery pack disconnected or already
connected.
If disconnected: the unit is automatically operational once connected (see page 11)
If already connected: the unit is hibernated and must be activated.
To activate a hibernated unit, the P2 and P3 keys must be kept pressed for 5 seconds
3.6 DISPLAY TEST
The integrity of the LCD display can be checked by pressing P1 and P4 together from the current measurement screen;
all segments of the display will be activated and deleted in sequence. In addition, the version and revision are displayed
bootloader revision (VV.RR b) / firmware version and revision (VV.RR f) and the corresponding checksum. To the
completed, the previous measurement screen will be restored.

12
3.7 FUNCTIONS
Below is a list of all the functions accessible from the user interface.
read only
editable
FUNC
DESCRIPTION
ACTION P1
ACTION P2
ACTION P3
ACTION P4
MIN VALUE
MAX VALUE
NOTES
01
Instant flowrate
unit
Back to
main
screen
Confirm +
return to
function
selection
Decrease
value
Increase
value
1
6
1=m;2=m3;3=L;4=ML;
5
=ft3;6=gal;7=BBL.
Setting 1 (meters) the
time base is
automatically set to 1
second
02
Time base unit
Back to
main
screen
Confirm +
return to
function
selection
Decrease
value
Increase
value
1
4
1=s;2=m;3=h;4=d
03
Counter unit
Back to
main
screen
Confirm +
return to
function
selection
Decrease
value
Increase
value
1
4
1=m3;2=L;3=ML;4=gal
;
5=BBL;6=Aft;7=Ain
04
Zero calibration
Back to
main
screen
Cancel
operation +
return to
function
selection
No action
Start
calibration
-
-
The display shows the
time left to complete
the operation. At the
end of the operation,
the main screen is
displayed.
05
Ka factor
Back to
main
screen
Return to
function
selection
Back to
main screen
Back to
main screen
-10.0
10.0
06
Sensor diameter
Back to
main
screen
Return to
function
selection
Back to
main screen
Back to
main screen
1
4000
The parameter is
electrically blocked;
read only
07
Pulse volume
Back to
main
screen
Return to
function
selection
Back to
main screen
Back to
main screen
1
10000
The parameter is
electrically blocked;
read only
08
Pulse technical
unit
Back to
main
screen
Return to
function
selection
Back to
main screen
Back to
main screen
1
4
1=mL;2=L;3=m3;4=gal
09
Pulse duration
Back to
main
screen
Confirm +
return to
function
selection
Decrease
value
Increase
value
5
500
Expressed in ms
(seconds / 1000)
10
Measurement
frequency
Back to
main
screen
Confirm +
return to
function
selection
Decrease
value
Increase
value
5
60
Interval between two
measurements [s]
11
Date and time
Back to the
previous
field; if in
years
setting
field, back
to main
screen
forward to
the next
field; if in
minutes
setting field,
back to main
screen
Decrease
value
Increase
value
2015/01/01
00:00
2099/31/12
23:59
12
Damping
Back to
main
screen
Confirm +
return to
function
selection
Decrease
value
Increase
value
2
100
Number of samples
averaged
13
Bypass
Back to
main
screen
Confirm +
return to
function
selection
Decrease
value
Increase
value
2
95
Bypass filter threshold,
% of full scale

13
FUNC
DESCRIPTION
ACTION P1
ACTION P2
ACTION P3
ACTION P4
MIN VALUE
MAX VALUE
NOTES
14
Peak cut
Back to
main
screen
Confirm +
return to
function
selection
Decrease
value
Increase
value
1
90
Pulse suppression
filter threshold,% of full
scale
15
Cut off
Back to
main
screen
Confirm +
return to
function
selection
Back to
main screen
Back to
main screen
0.00
10.00
[m/s] - the parameter
is electrically locked
16
Flowrate display
digits
Back to
main
screen
Confirm +
return to
function
selection
Decrease
value
Increase
value
0
4
Number of decimal
digits of the instant
flow shown on the
display
17
Totalizers
display digits
Back to
main
screen
Confirm +
return to
function
selection
Decrease
value
Increase
value
0
4
Number of decimal
places of the totalizers
18
Empty pipe
threshold
Back to
main
screen
Confirm +
return to
function
selection
Decrease
value
Increase
value
200
5000
19
Full scale
Back to
main
screen
Confirm +
return to
function
selection
Decrease
value
Increase
value
1.0
10.0
Full scale expressed
in m/s (not connected
to the diameter value)
20
Partial positive
totalizer reset
Back to
main
screen
Return to
function
selection
No action
Partial
positive tot.
reset (Must
be hold
pressed for
5 seconds)
No action
No action
The display shows the
positive partial
totalizer
21
Partial negative
totalizer reset
Back to
main
screen
Return to
function
selection
No action
Partial
negative tot.
reset (Must
be hold
pressed for
5 seconds
No action
No action
The display shows the
negative partial
totalizer
22
Firmware version
Back to
main
screen
Return to
function
selection
No action
No action
-
-
Show the current
version. Firmware
revision
23
Firmware
checksum
Back to
main
screen
Return to
function
selection
No action
No action
-
-
Shows the installed
firmware CRC32
24
Empty pipe
detection mode
Back to
main
screen
Confirm +
return to
function
selection
Decrease
value
Increase
value
0
3
0: Disabled;
1: Enabled on the
fourth electrode;
2: Enabled on the
measuring electrodes;
3: Enabled on both
25
Password setup
Back to
main
screen
forward to
the next
digit, confirm
and return
Decrease
value
Increase
value
000000
999999
26
Bootloader
version
Back to
main
screen
Return to
function
selection
No action
No action
-
-
Shows the current
version. Revision of
the bootloader
27
Bootloader
checksum
Back to
main
screen
Return to
function
selection
No action
No action
-
-
Show the current
bootloader CRC32
28
Line frequency
Back to
main
screen
Confirm +
return to
function
selection
Decrease
value
Increase
value
0
1
Set the frequency of
the power supply 0: 50
Hz 1: 60 Hz

14
FUNC
DESCRIPTION
ACTION P1
ACTION P2
ACTION P3
ACTION P4
MIN VALUE
MAX VALUE
NOTES
29
Load user
parameters
Back to
main
screen
Return to
function
selection
Decrease
value
(Long press)
load user
parameters
-
-
Replaces the
parameters of the
instrument with those
saved as
user copy
30
Save user
parameters
Back to
main
screen
Return to
function
selection
Decrease
value
(Long press)
save user
parameters
-
-
Save the instrument
parameters as a user
copy
31
Load factory
parameters
Back to
main
screen
Return to
function
selection
Decrease
value
(Long press)
load factory
parameters
-
-
It replaces the
parameters of the
instrument with the
factory ones
32
Damping in low
power mode
Back to
main
screen
Confirm +
return to
function
selection
Decrease
value
Increase
value
1
100
Number of samples to
be averaged over
33
Auto power off
Back to
main
screen
Return to
function
selection
Decrease
value
Increase
value
0
7
Shutdown time
0=20s;1=1 min;2=3
34
Sensor offset
Back to
main
screen
Return to
function
selection
No action
No action
-99999999
99999999
Shows the offset of
the paired sensor
35
Log of firmware
download
attempts
Back to
main
screen
Return to
function
selection
Show
previous
record
Show next
record
Oldest registry
(0)
Last registry
available
Show log of firmware
download attempts
(see section)
36
Service code
Non
accessible
Non
accessible
Non
accessible
Non
accessible
Non accessible
Non
accessible
function not
accessible; an attempt
to access
will show "denied" on
the screen
37
Negative pulse
output
configuration
Back to
main
screen
Return to
function
selection
Decrease
value
Increase
value
0
3
0: Negative flow;
1: Positive Flow;
2: Net Positive flow;
3: Error report
38
Selection of the
measurement to
map on 4-20mA
output
Back to
main
screen
Return to
function
selection
Decrease
value
Increase
value
0
4
0: Flow;
1: Full scale percent;
2: Pressure;
3: Temperature T1
4: Temperature T2
39
Measurement
value to be
mapped on 4mA
output
Back to
main
screen
Return to
function
selection
Decrease
value
Increase
value
0
1000
Measurement value to
be mapped on 4mA
output (enabled only
with 4-20mA card)
40
Measurement
value to be
mapped on
20mA output
Back to
main
screen
Return to
function
selection
Decrease
value
Increase
value
0
1000
Measurement value to
be mapped on 20mA
output (enabled only
with 4-20mA card)
41
Flow send
Back to
main
screen
Return to
function
selection
Decrease
value
Increase
value
0
1
0: signals error in case
of reverse flow
42
Flow simulation
Back to
main
screen
Return to
function
selection
Decrease
value
Increase
value
0
1
0: flow simulation
disabled
1: flow simulation
enabled (enabled only
with 4-20mA card)
43
Flow simulation
value
Back to
main
screen
Return to
function
selection
Decrease
value
Increase
value
0
1000
Value of the flow to be
simulated. (enabled
only with 4-20mA
card)

15
FUNC
DESCRIPTION
ACTION P1
ACTION P2
ACTION P3
ACTION P4
MIN VALUE
MAX VALUE
NOTES
44
Bluetooth access
interval
Back to
main
screen
Return to
function
selection
Decrease
value
Increase
value
0
9
0: Always off;
1: On every 30 sec;
2: On every 60 sec;
3: On every 2 min;
4: On every 5 min;
5: On every 10 min;
6: On every 15 min;
7: On every 30 min;
8: On every 60 min;
9: Always on
(Enabled only with
Bluetooth-Modbus
card)
45
Bluetooth access
duration
Back to
main
screen
Return to
function
selection
Decrease
value
Increase
value
3
30
Duration in seconds of
the Bluetooth access
(enabled only with
Bluetooth-Modbus
card installed)
46
MODBUS
access interval
Back to
main
screen
Return to
function
selection
Decrease
value
Increase
value
0
12
0: Always off;
1: On every 5 sec;
2: On every 10 sec;
3: On every 15 sec;
4: On every 30 sec;
5: On every 1 min;
6: On every 2 min;
7: On every 5 min;
8: On every 10 min;
9: On every 15 min;
10: On every 30 min;
11: On every 60 min;
12: Always on;
(enabled only with
Bluetooth-Modbus
card installed)
47
MODBUS
access duration
Back to
main
screen
Return to
function
selection
Decrease
value
Increase
value
1
30
Duration in seconds of
the Modbus access
(enabled only with
Bluetooth-Modbus
card installed)
48
MODBUS baud
rate
Back to
main
screen
Return to
function
selection
Decrease
value
Increase
value
0
6
Baudrate Modbus
0: 2400;
1: 4800;
2: 9600;
3: 19200;
4: 38400;
5: 57600;
6: 115200 (default)
(enabled only with
Bluetooth-Modbus
card installed)
49
Number of
MODBUS bits
Back to
main
screen
Return to
function
selection
Decrease
value
Increase
value
0
1
Number of modbus
bits
0: 8 data bits
1: 7 data bits
50
MODBUS parity
Back to
main
screen
Return to
function
selectio0n
Decrease
value
Increase
value
0
2
MODBUS parity 0:
None
1: Odd

16
FUNC
DESCRIPTION
ACTION P1
ACTION P2
ACTION P3
ACTION P4
MIN VALUE
MAX VALUE
NOTES
51
MODBUS stop
bit
Back to
main
screen
Return to
function
selection
Decrease
value
Increase
value
0
1
MODBUS stopbit
0: 1 bit stop
1: 2 bit stop
(enabled only with
Bluetooth-Modbus
card installed)
52
MODBUS
address
Back to
main
screen
Return to
function
selection
Decrease
value
Increase
value
0
255
MODBUS address
(enabled only with
Bluetooth-Modbus
card installed)
53
MODBUS byte-
order
Back to
main
screen
Return to
function
selection
Decrease
value
Increase
value
1
5
MODBUS byte order
0: little endian
1: big endian
2: little endian byte
swap
4. BATTERY LIFE
The converter is supported by an optional lithium battery pack
and can operate for over 10 years in battery mode (depending
on the configuration and working conditions). The residual life
comes estimated by calculating the use of energy for the
duration of the tool, so when the battery is replaced it is
necessary reset the battery usage counter. The ignition the unit
automatically takes the user to a screen selection (see SYSTEM
START). Frequently waking up the unit or entering very long
shutdown times (feature 33) can shorten the expected battery
life.
Expected battery life (years) according to sensor diameter and sampling rate

17
5. AUTO DIAGNOSTICS
During the measurement process, the MC406M automatically detects if an error condition occurs and possibly stops the
measurement by reporting an alarm indication on the display. The following table describes the various conditions of
alarm and converter behavior.
ANOMALY DETECTED
POSSIBLE CAUSE
CONVERTER BEHAVIOR
ICONS DISPLAYED AND
CORRESPONDING ERROR
CODE
Excitation failure
Damaged sensor cables;
Sensor disconnected;
Damaged sensor coils
The measurement is forced to 0
and the error condition is displayed
and recorded.
Empty tube detected on the fourth
electrode
The sensor is not completely full of
liquid;
an air intake causes turbulence;
sensor disconnected;
damaged sensor cables;
liquid conductivity too low.
The measurement is forced to 0
and the error condition is displayed
and recorded.
To save battery, the excitation of
the coils is stopped until the error
condition is overcome.
Empty pipe detected / invalid
measurement
on the measuring electrodes
The sensor is at least half empty;
of the air drawn in due to
turbulence; sensor disconnected;
damaged sensor cables; liquid
conductivity too low; an electrical
leak on the process influences the
measurement; a high source of
electromagnetic interference
influence the measure.
The measurement is forced to 0
and the error condition is displayed
and recorded.
To save battery, the excitation of
the coils is stopped until the error
condition is overcome.
The temperature exceeds the
nominal operating range of the
converter
The ambient temperature is above
the working range; the converter is
installed with direct sunlight
exposure.
The error condition is displayed.
Wet electronic board
The cap or a cable gland has not
been closed properly causing
water or moisture condensation to
enter the converter housing
The error condition is displayed.
Supply voltage out of limit
The battery is damaged or
discharged; an incorrect power
source is connected to the
converter.
The error condition is displayed.
Pulse overlap
Pulse settings are not compatible
with the measured process.
The error condition is displayed.
Communication error with 4-20mA
card
Current loop interrupted.
Problems on connection cable
between 4-20mA board and
measurement board.
4-20mA board malfunction
The error condition is displayed.
Incorrect Bluetooth parameters
Warning
Bluetooth power-on interval too
short, battery life will be reduced
considerably
The firmware checksum differs
from the expected value
The memory of the internal
program has changed.
The error condition is displayed
and the program execution
stopped.
The checksum of the parameters
differs from the expected value.
Parameter data is corrupt.
The error condition is displayed
and the program execution
stopped.
A watchdog reset has been
performed
A disturbance has resulted in
unmanaged behavior in the
regular execution of the software.
The error condition is displayed
and the program execution
stopped.

18
6. PC INTERFACE SOFTWARE
The supplied PC communication software is based on a proprietary protocol; its use is possible with the interface
Euromag IrComm (USB to infrared adapter) and allows you to:
•View live measurements in real time
•View all the parameters
•Modify the unprotected parameters
•Download the internal datalogger
•Update firmware
MAIN DASHBOARD - real live values and basic information
GRAPHIC - Recording of live measurements / Visualization

19
DATA LOG - Internal process log download and CSV file export
CONFIGURATION - Parameter setting & configuration

20
6.1 FIRMWARE UPDATE
To update the converter with a newer firmware version, select the firmware .hex file and press «Download new firmware
to converter». The progress of the firmware download is displayed while the instrument displays the sequence of internal
operations.
The previously installed firmware is saved in backup on a dedicated memory area, therefore in case of error during
unloading, the converter will automatically restore normal functionality. During the operation the red LED turns on lights up
when the FLASH memory is cleared or written.
At the end of the update process, the unit will restart automatically.
0
Firmware check installed
6
Checking the downloaded firmware
1
Awaiting download request
7
Firmware update installed
2
Download request received
8
Backup of the new firmware installed
3
Clear download memory
9
Update OK
4
Firmware download
10
Firmware reset installed (only during startup)
5
Checking the downloaded firmware
11
Firmware reset installed (after error during update)
Table of contents
Other Euromag Media Converter manuals