EUTECH INSTRUMENTS ALPHA CON 1000 CONDUCTIVITY... User manual

68X216802
Instruction Manual
αlpha-CON1000
Conductivity Controller/Transmitter
MEAS
8.088.08mS
25.025.0
oC
ATC
Conductivity Controller
ααlpha CON1000
CAL
ENTER
ALARM
REL A
REL B
AUTO
MANU
REL A
REL B
ESC

68X216802
2
Preface
This manual serves to explain the use of the αlpha-CON1000 series Conductivity
controller/transmitter. The manual functions in two ways:firstly as a step by step guide to help the
user operate the instrument. Secondly, it serves as a handy reference guide. This instruction
manual is written to cover as many anticipated applications of the αlpha-CON1000 Conductivity
controller/transmitter. If you have doubts in the use of the instrument, please do not hesitate to
contact the nearest Eutech Instruments’ Authorised Distributor.
The information presented in this manual is subject to change without notice as improvements are
made, and does not represent a commitment on part of Eutech Instruments Pte Ltd.
Eutech Instruments cannot accept any responsibility for damage or malfunction of the unit due to
improper use of the instrument.
Copyright 1998 Eutech Instruments Pte Ltd. Version 1.0. All rights reserved.
EUTECH INSTRUMENTS PTE LTD
Blk 55 Ayer Rajah Crescent, #04-16/24, Singapore 139 949. Tel: (65) 778 6876; Fax: (65) 773 0836
TABLE OF CONTENTS
1 Introduction........................................................................................................3
1.1 Description of Unit.................................................................................3
1.2 Applications...........................................................................................4
2Assembly and Installation...................................................................................5
2.1 Measurement and Control System...........................................................5
2.2 Unit Dimensions.....................................................................................5
3Electrical Connection.........................................................................................6
3.1 Connection Diagram...............................................................................6
3.2 Back Panel .............................................................................................7
4Overview.........................................................................................................8
4.1 Keypad and Display................................................................................8
4.2 Function Groups.....................................................................................9
4.3 Control Concept ...................................................................................10
5MEASUREMENT............................................................................................11
5.1 Display in Measurement mode..............................................................11
5.2 Security Codes......................................................................................11
6Calibration Mode.............................................................................................14
6.1 Conductivity Calibration.......................................................................14
7ADVANCED SET-UP MODE.............................................................................15
7.1 Temperature Coefficient sub-function...................................................15
7.2 Temperature calibration (ATC mode only)............................................16
7.3 Control Relay A/Control Relay B (SP1/SP2) sub-function.....................17

Operating Instructions αlpha-CON1000
68X216802
3
7.4 Controller (Cntr) sub-function...............................................................19
7.5 Measurement Range sub-function.........................................................20
7.6 Current Output (rng) sub-function.........................................................21
7.7 Configuration (ConF) sub-function .......................................................22
7.8 Calibration (CAL) sub-function ............................................................23
8Auto/Manual Mode..........................................................................................24
8.1 Auto mode (mode after switch-on)........................................................24
8.2 Manual mode........................................................................................24
9Technical Specifications...............................................................................25
10 Accessories Assembly Accessories....................................................................26
11 General Information.........................................................................................26
11.1 Warranty ..............................................................................................26
11.2 Packaging.............................................................................................26
11.3 Return of Goods ...................................................................................26
11.4 Guidelines for Returning Unit for Repair...............................................26
12 Appendices...............................................................................................27
12.1 Appendix 1 – Jumper Positions.............................................................27
12.2 Appendix 2 – Measurement Ranges available in the Controller..............28
12.3 Appendix 3 – Conductivity at Related Temperature Coefficients (25oC) 28
12.4 Appendix 4 – Conductivity of Various Aqueous Solutions at 25oC........30
12.5 Appendix 5 - Simple Explanation on the Function of Hysteresis............31
12.6 Appendix 6 – General Instructions Concerning Controller Setting .........32
1 Introduction
1.1 Description of Unit
Thank you for purchasing Eutech’s ¼ DIN alpha-1000 series Conductivity process controllers. This
unit is used for measuring the Conductivity of a solution, either in micro-Siemens or milli-Siemens,
one at a time, and the operational mode is switchable from the menu. You can use this unit to
measure Conductivity with limit control. This controller has many user-friendly and safety features
which include:
•Menu-driven program that simplifies set-up
•Ten rangesof Conductivity measurements-software selectable (Appendix 2).
•Built-in memory backup to ensure that calibration data and other information are not erased
if power supply fails
•Automatic temperature compensation (ATC) with Pt100 or Pt 1000
•Manual temperature compensation with independent setting for calibration and process
temperature
•Temperature coefficient variable between 0.00 to 10.00 % per oC. Separate pure water
compensation curve stored in memory. Reference temperature at 25oC.
•0to 1999second time delay adjustment on all relays – minimise false alarms
•Separately adjustable high and low set point hysteresis (dead bands) prevent chattering of
relays around the set points

68X216802
4
•Three control modes: limit, proportional pulse length or proportional pulse frequency
•Large dual display LCD for easy reading with clear multiple annunciators, alarm status and
operational message annunciators
•Two switching contacts as set-point triggering relays and an alarm output relay
•Separate alarm relay alerts you when set points have exceeded the limits and if the
Pt100/Pt1000 wires are broken or disconnected during the ATC function
•Hold function freezes output current (0/4...20mA) and releases control relays
•LED indicators signal control activities to monitor controller status from a distance
•Protection against electromagnetic interference - galvanically isolated 0/4..20mA output
provides safety for data logging and control purposes
1.2 Applications
Use this controller in panel mounted enclosures for applications in water treatment and
demineralization, waste water treatment and neutralization processes.

68X216802
2Assembly and Installation
2.1 Measurement and Control System
A typical measurement system consists of:
•a αlpha-CON1000 processcontroller
•a suitable Conductivity electrode with the appropriate Cell constant and integrated temperature
sensorPt 1000 or Pt 100,
•an immersion, flow or process assembly
•a final control element such as pump or valve and
•a chart recorder.
2.2 Unit Dimensions
P
o
w
e
r
M
a
i
n
s
(
2
2
0
/
1
1
0
V
A
C
)
l
F
l
o
w
A
s
s
e
m
b
l
y
t
o
D
o
s
i
n
g
P
u
m
p
s
P
r
o
c
e
s
s
A
s
s
e
m
b
l
y
w
i
t
h
E
l
e
c
t
r
o
d
e
C
h
a
r
t
R
e
c
o
r
d
e
r
P
t
1
0
0
/
P
t
1
0
0
0
T
e
m
p
e
r
a
t
u
r
e
S
e
n
s
o
r
p
h
a
-
C
O
N
1
0
0
0
C
o
n
t
r
o
l
l
e
r
A
l
a
r
m
/
S
i
r
e
n
S
y
s
t
e
m
max.175 max.45
Mounting Cut-Out
F
l
a
t
G
a
s
k
e
t
(
1
m
m
)
(TobeInserted ByCustomer)
Note:TheTapedCornersHavetoBeOnTop
92
32
56
9
6
92 +0.5
92 +0.5

68X216802
6
The field-tested control panel housing is 96 x 96 mm; with protection class IP 54 (front).
Electrical Connection
3.1 Connection Diagram
-+
PE/S
PEAC:NL
125
4
3 6 7 8 9 10 11 12 13 14 15 17
15 16
*)indicated contactpositionsareforcurrentless conditions
SignalOutputHoldInput
Alarm
RelayB
RelayA
PowerMains
CON
mA
SignalInputConductivity
S
/
V
PE/S
CON
O
O
18 19 20 21 22
S
/
S
/
Pt 1000/100

Operating Instructions αlpha-CON1000
68X216802
7
3.2 Back Panel
The back panel consists of two connectors. The first connector is the 17-way PCB edge connector and
the other is the 5-way connector.
Connection for the 17-way screw terminals (from left to right):
1. AC mains live wire 10. Alarm relay resting position (NO)
2. AC mains neutral wire 11. Alarm relay common
3. AC mains protective earth wire 12. Alarm relay working position (NC)
4. Low set relay resting position (NC) 13. Hold function switch terminal 1
5. Low set relay common 14. Hold function switch terminal 2
6. Low set relay working position (NO) 15. No connection
7. High set relay resting position (NC) 16. 0/4 - 20 mA for -ve connection
8. High set relay common 17. 0/4 - 20 mA for +ve connection
9. High set relay working position (NO)
Connections for the 5-way screw terminals:
18. Pt1000/Pt100 lead 1 terminal
19. Pt1000/Pt100 sense lead terminal
20. Pt1000/Pt100 lead 2 terminal
21. Conductivity lead 1
22. Conductivity lead 2
IMPORTANT: The Alarm relay functions as an “Active Low” device i.e. it switches OFF under
Alarm condition. Therefore the Alarm display device should be connected to the ‘NC’ contacts of
the relay.
Pt100/
RELAY1
LNPE
63mA(F)
FUSE 250VAC
HOLD
ALARMRELAY2
Pt1000
NC
-
+
cell
J2

68X216802
4Overview
4.1 Keypad and Display
4.1.1 Keypad
•Perform rapid calibration
•Allows entry to Set up mode
•Select individual functions within the function group of Set up mode
•Store input data in the Set up mode
•Start calibration in the calibration mode
•Select various function groups in the Set up mode.
•Set parameters and numerical values in sub functions of Set up mode
If pressed continuously, the setting speed increases
•Control the relays in the MANUAL function
•Return to the Measurement mode when both keys are pressed together
•Switch between AUTO and MANUAL relay operation using a code
•Display limit set-point values for the switch contacts in AUTO relay operation mode
•Switch between RELAY A and RELAY B in MANUAL relay operation mode
4.1.2 Display
The LCD display features two numerical displays that show status messages and measured values for
easy, quick reference. The display provides short-text information for setting parameters and
configuration.
•HOLD: Relay position and current output are frozen
•SETUP: Set-up mode of function groups
•MEAS: Measurement mode
•CAL: Calibration mode of Conductivity
•READY: Comes on after a successful calibration
•
ATC: Comes on in the ATC mode. Disappears in the
Manual temperature Compensation mode. “ATC”
flashes if the temperature probe is faulty in its ATC
mode
•
Range No.: Indicates the measurement range
selected
•Display for RELAY A/B. Green LED indicates measured value within limit while
RED LED indicates measured value outside limit.
•Alarm display if limit value overshoot or the ATC connection is broken.
-8.8.8.8-8.8.8.8
1.8.8.81.8.8.8ATC
HOLD SETUP MEAS CAL
READY
µµ
S
mS
o
C
88

Operating Instructions αlpha-CON1000
68X216802
9
4.2 Function Groups
The main function and sub-function groups are organised in a matrix format for configuration and
selection of parameters. The main function groups are:
1) Temperature Coefficient settings (tCtC)
2) Temperature Measurement / compensation settings (SSEEtt ooCC)
3) Control relay 1 configuration (SP1SP1)
4) Control relay 2 configuration (SP2SP2)
5) Control type (CntrCntr)
6) Current output (rngrng)
7) Configuration (ConFConF)
8) Calibration (CALCAL dOdO)
The set-up parameters can be viewed or changed by entering a security code. See Section 5.2 for
security code information.
4.2.1 How to view operating parameters without access to change them:
a) Press the ENTER key. The display will prompt the user to enter a
security code (S.Cd). Leave the security code at “000” (do not enter a
security code).
b) Press ENTER key again. This allows you only to view (not
change) any sub-functions’ settings.
c) Press the ∆∆ or ∇∇ keys to scroll through the sub-functions.
d) Press the ENTER key at a particular sub-function to view in detail.
e) Press the ENTER key to return to the sub-function menu.
f) Press the ∆∆ and ∇∇ keys simultaneously (as an Escape key) at any time to return to the
Measurementmode.
Note: To simplify operations, the controller will not display parameters that are not relevant to a
particular sub-function. For example: If the user set the controller for Limit control, it will not display
pulse length/frequency settings.
000000
S.CdS.Cd
tctc
HOLD SETUP
SEt
SEt
oo
CC
HOLD SETUP
SPSP
11
HOLD SETUP
SP2SP2
HOLD SETUP
ConFConF
HOLD SETUP
rngrng
HOLD SETUP
CntrCntr
HOLD SETUP
CALCAL
ConCon
HOLD SETUP

68X216802
10
4.3 Control Concept
The main function and sub-function groups are organised in a matrix format as shown below. These
functions can be accessed via the front keypad for configuration and selection of parameters.
The controller offers two levels of password protection: (1) for direct access to calibration function and
(2) for setting or editing specific controller parameters or functions in the SETUP mode to suit individual
requirements.
Note: The passwords are not user-defined and have been set by factory. It is very important to keep
these passwords strictly confidential to avoid unauthorised tampering of the system at all times.
Note: If the user reads parameters only, the controller automatically reverts to Measurement mode if
none of the keysis pressed for 30 seconds.

Operating Instructions αlpha-CON1000
68X216802
11
5MEASUREMENT
5.1 Display in Measurement mode
When the controller is initially powered on, it automatically enters into the Measurement mode after the
large dual LCD displays all segments briefly.
The upper display shows the measured Conductivity value, while the lower display shows the
temperature value. Annunciators at the right side of the display indicate whether the controller is set for
µS or mS measurement.Similarly annunciators or icons at the top or left side of the display shows the
current status of controller, e.g. “HOLD”, “SETUP”, “MEAS”,“CAL”,“READY”, etc.
5.1.1 Check electrode performance
To read current electrode slope and offset values without changing them:
1) Press the CAL key followed by the ENTER key without adjusting the security code (leave code at
“000”). The upper display shows the condition of the probe. The lower display reading shows the
temperature reading.
Note: If security code is changed to a value other than “000”, pressing the ENTER key will return
to the Measurement mode, without displaying electrode information.
2) Press the ENTER key a second time to return to Measurement mode.
5.1.2 Checking set points
To read current set point values without changing them:
Press the RELAY Selection (Rel A /Rel B) key. The upper display shows the set point for Relay A;
the lower display shows “SP1”.
After two seconds the upper display shows the set-point value for Relay B; the lower display shows
“SP2”.
After an additional two seconds, the controller returns to the Measurement mode.
5.2 Security Codes
This controller has two levels of security protection with separate security codes. The first level allows
entry into the Calibration mode: security code = 11; the second allows entry into the SETUP mode:
security code = 22.
The security codes protects the controller from unauthorised tampering of its current setting. The
parameters cannot be changed unless the security code is entered.
5.2.1 How to enter and change parameters in Calibration mode
1) Press the CAL key. The upper display shows “000” and the
lower display shows “C.Cd” to prompt the user to enter the Calibration
security code.
2) Press the ∆∆ or ∇∇ keys to scroll upper display to Calibration
security code “11”.
3) Press the ENTER key. The display shows “CAL CON”.
4) Press ENTER key again to begin calibration. Refer to Section 6 for full details on calibration.
5) Press the ∆∆and ∇∇keys simultaneously (escape) to return to the Measurement mode.
11
11
C.CdC.Cd

68X216802
12
NOTE: To view (not change) the SETUP parameters, push the ENTER key when the security code
reads “000”.

Operating Instructions αlpha-CON1000
68X216802
13
5.2.1.1 Clearing the Calibration security code from the display
The calibration security code automatically resets from “11” to “000” after you return to Measurement
mode, so you do not need to clear the security code from the display.
5.2.2 How to enter and change parameters in Advanced Setup mode
1) Press the ENTER key once. The upper display shows “000” and the lower display shows “S.Cd” to
prompt you to enter the Advanced Setup security code.
2) Press the ∆∆ or ∇∇ keys to scroll the display to Setup security code “22”. NOTE: Pressing the
ENTER key at a value other than “22” causes the controller to revert to the Measurement mode.
3) Press the ENTER key.
4) The upper display reads “tc”.
5) You are now in the Advanced Setup mode. See Section 7 for complete instructions. To return to
Measurementmode,press the ∆∆and ∇∇ keys simultaneously (escape).
NOTE: If you want to view (not change) set up parameters, push the ENTER key when the security
code reads “000”.
5.2.2.1 Clearing the Advanced Setup security code from the display
After you have entered the security code and returned to the Measurement mode, the security code “22”
still appears on the display whenever you press the ENTER key. To conceal the security code, you must
manually reset the code. To clear the Advanced Setup security code from the display:
1) Press the ENTER key in the Measurement mode.
2) Set to any security code (not 11 or 22) and complete by pressing ENTER.
NOTE: When you enter the Calibration mode with code “11” or Advanced Setup mode with security
code “22”, the unit automatically enters into the HOLD mode until you return back to Measurement
mode. The HOLD annunciator is displayed at the upper left of the display. While on HOLD, the
current output is frozen and set point relays are deactivated.

68X216802
6Calibration Mode
You can reach the Calibration mode directly from the Measurement mode by pressing the CAL key and
entering the Calibration security code. You can also reach the Calibration mode from the Advanced
Setup mode.
6.1 Conductivity Calibration
Calibration is always carried out in the specific range selected. The Conductivity Controller allows a
one-point calibration.
1) Enter Calibration mode. While in the Measurement mode,
push the CAL key and scroll to Calibration code “11”. Push the
ENTER key again. The upper display reads “CAL” and the lower
display shows “Con”.
2) Press the ENTER key. The controller displays the status of
the Cell resulting from the previous calibration.Press the ENTER
key once again to carry out calibration.
3) Immerse the Conductivity cell in a suitable standard
solution, whose value is within the measurement range selected
in the controller. Agitate the Cell in the solution to remove any
trapped air-bubbles.
4) Once the reading has stabilized, use the ∆∆ or ∇∇ keys to
adjust the measured value to that of the standard solution. Press
the ENTER key to accept the value and the controller displays
the revised condition of the probe. Note: if the percentage
difference between the measured value and the standard
solution is greater than 40%, the controller will not accept
the calibration. Change the Cell.
5) Press the ENTER key. If you entered the calibration mode
using the CAL key, the controller will return to the Measurement
mode. If you entered the calibration mode from the Advanced
Set-up mode, the controller will return to the sub-function menu.
Note: If there is a calibration error, the controller displays “ERR”. If this happens, push both the ∆∆
and ∇∇keys (escape) to restart the calibration beginning from step 1.
Note: When calibrating with manual temperature compensation, the controller automatically changes
from the preset process temperature to the calibration temperature. After leaving the Calibration mode,
the controller switches back to process temperature (for setting the calibration temperature and the
process temperature, see section 7.4.3).
CALCAL
ConCon
HOLD SETUP
100100.0.0
CCEELL
HOLD CAL
HOLD CAL
14131413
25.025.0
µµ
S
oC
ATC
100100.0.0
CELCEL
HOLD CAL
%
%
6
6
6
6

68X216802
7ADVANCED SET-UP MODE
7.1 Temperature Coefficient sub-function
This sub-function allows you to select the correct temperature coefficient for optimum operations. For
applications in the pure water or ultr-pure water industries, simply select the “Pur” temperature
coefficient option. For all other applications, select “Lin” temperature coefficient. The controller allows
further input of temperature coefficient values, independently for the process and for the calibrating
solutions. Default is at 2.10 %.
7.1.1 Selecting Pure-water or Linear Temperature Coefficient
1) Enter Advanced set-up mode. Push the ENTER key and
scroll to Advanced Set-up security code “22”. Push the ENTER
key again. Controller displays “Lin” in the upper display and “tC”
in the lower display.
2) Press the ∆∆ or ∇∇ keys to select between “Pur” and “Lin”
temperature coefficients. If “Pur” tC is be selected, press the
ENTER key and controller goes back to “tC”. If “Lin” tC is selcted,
press the ENTER key and proceed to step 3.
3) The upper display shows “2.10%”, while the lower display
show “P.tC”.This option allows you to input the Process
Temperature coefficient, from 0 to 10%. Default is 2.10%. Use the
∆∆ or ∇∇keys to enter the required temperature coefficient value.
4) Press the ENTER key. Next, enter the calibration solution
temperature coefficient value, from 0 to 10%. Default is 2.10%.
5) Press the ∆∆ or ∇∇ keys to input the desired calibration
solutiontemperaturecoefficient.
6) Press the ENTER key to accept the value.
7) Continue with additional Advance setup procedures or return
to Measurement mode by pressing the ∆∆ and ∇∇ keys (escape)
simultaneously.
HOLD SETUPHOLD SETUP HOLD SETUP HOLD SETUP
HOLD SETUP
tctc
2.102.10
P.tCP.tC
2.102.10
C.tCC.tC
LinLin
tCtC %
ATC
%
ATC
PurPur
tCtC
tctc
HOLD SETUP
LinLin
tCtC
HOLD SETUP
HOLD SETUP
2.102.10 %%
P.tCP.tC ATCATC
HOLD SETUP
2.102.10 %%
C.tCC.tC ATCATC

68X216802
7.2 Temperature calibration (ATC mode only)
1) Select “ATC on” as described above in Section 7.4.1.
2) Press the ENTER key. The upper display indicates the
current temperature offset. The current measured temperature is
shown in the lower display.
3) Compare the current measured temperature on the controller
display to a thermometer known to be accurate. Note down the
correcttemperaturevalue.
4) Press the ∆∆ or ∇∇ keys to scroll the lower display to match
the correct value. The upper display will now show the offset value.
You can offset temperature up to ±5 oC.
5) Press the ENTER key to confirm your selection.
6) Continue with additional Advanced Set-up procedures, or
return to the Measurement mode by pressing the ∆∆ and ∇∇ keys
(escape) simultaneously.
7.2.1 Setting manual temperature compensation
Note: This parameter is blanked out when the controller is set for
ATC operation.
For manual temperature compensation, you can set two different
temperatures: process and calibration. This allows calibration at a
temperature other than your process temperature. Example: setting
a calibration temperature of 25 oC lets you calibrate using standard
buffer solutions at 25 oC, even if your process temperature is a
differenttemperature.
1) Select “ATC oFF” as described above in section 7.4.1
2) Press the ENTER key. The upper display shows the current
process temperature and the lower display shows “P.oC” to indicate
process temperature.
3) Press the ∆∆ or ∇∇ keys to adjust the process temperature
value. You can adjust the value from –9.9 to 125 oC.
4) After you set the process temperature value, press the
ENTER key. The upper display shows the current calibration
temperature and the lower display shows “C.oC” to indicate
calibrationtemperature.
5) Press the ∆∆ or ∇∇keys to adjust the calibration temperature value. You can adjust this value from
–9.9 to 125 oC.
6) Press the ENTER key to confirm your selection.
7) Continue with additional Advanced Set-up procedures, or return to Measurement mode by
pressing the ∆∆ and ∇∇keys (escape) simultaneously.
o
C
0.00.0
25.025.0 ATCATC
25.025.0
CC.C.C
oFFoFF
AtcAtc
25.025.0
P.CP.C
HOLD SETUP
HOLD SETUP
HOLD SETUP
onon
AtcAtc
HOLD SETUP
HOLD SETUP

Operating Instructions αlpha-CON1000
68X216802
17
7.3 Control Relay A/Control Relay B (SP1/SP2) sub-function
The SP1 option sets the operating parameters for Relay A; and SP2 for relay B. Since these groups
have the same set-up parameters, they are described together.
7.3.1 Entering the Set point 1 (Set point 2) sub-function
1) Enter Advanced Set-up mode. Push the ENTER key and
scroll to Advanced Set-up security code “22”. Push the ENTER
key again.
2) Press the ∆∆ or ∇∇ keys to scroll until the upper display
shows SP1 (SP2).
7.3.2 Selecting the set point values
This lets you choose the value that will cause your controller to
activate (Default: SP1 = 6.20mg/l; SP2 = 4.00mg/l).
1) Follow directions in 7.5.1 to enter Control Relay mode. If
you are in this mode, skip to step 2.
2) Press the ENTER key. The upper display shows the
current set point value and the lower display shows SP1 (SP2).
3) Press the ∆∆ or ∇∇ keys to select your value for Set point 1
(Set point 2). Your controller will activate at the value you select.
4) Press the ENTER key to confirm your selection.
5) Proceed to 7.5.3, or return to Measurement mode by
pressing the ∆∆ and ∇∇keys simultaneously (escape).
7.3.3 Choosing High or Low set points
Select a low set point to activate controller when the dissolved oxygen value undershoots the low set
point; select a high set point to activate controller when the value overshoots the high set point. Using
both SP1 and SP2, you can select lo/lo, lo/hi, hi/lo, or hi/hi set points (Default: SP1 = Lo; SP2 = Lo).
1) Follow directions in 7.5.2to enter Control Relay mode.
2) Press the ENTER key. Scroll with the ∆∆ or ∇∇ keys until the upper display shows Lo or Hi (for low
or high set point) and the lower display shows SP1 (SP2).
3) Press the ∆∆ or ∇∇keys to select low (lo) or high (hi) set point for SP1 (SP2).
4) Press the ENTER key to confirm your selection.
5) Proceed to 7.5.4, or return Measurement mode by pressing the ∆∆ and ∇∇ keys simultaneously
(escape).
7.3.4 Selecting a hysteresis (dead band) value (0.1 to 1.0 mg/l or 1 to 10%)
Hysteresis prevents rapid contact switching if your value is fluctuating near the set point. It does this by
overshooting the set point value to a specified hysteresis value (default is 0.2 mg/l or 2.5%).You can set
the hysteresis value from 0.1 to 10mg/l, 0.1 to 10 ppm or 1 to 10%.
HOLD SETUPHOLD SETUP HOLD SETUP HOLD SETUP HOLD SETUP HOLD SETUP
SP 1SP 1
LoLo
SP 1SP 1 2020
HYSHYS 00
On.dOn.d 00
Of.dOf.d
100100
SP 1SP 1
SPSP 11
LoLo
SPSP 11
100100
µµSS
SPSP 11
HOLD SETUP
HOLD SETUP
HOLD SETUP

68X216802
18
Example: You have set your set point 1 (Lo) at 6.00 µS and your hysteresis limit value is at 0.50µS. If
your measured value undershoots the low set point of 6.00 µS, the controller’s relay activates, which in
turn activates an external device such as a pump or valve. The actions of the external device will cause
the value to rise above 6.00µS. When the value has increased to 6.50 µS, the relay,and hence the
pump will switch off. 1) Follow directions in 7.5.3to enter Control Relay mode.
2) Press the ENTER key. Scroll with the ∆∆ or ∇∇ keys until the upper
display shows the hysteresis (dead band) value and the lower display
shows“HYS”.
3) Press the ∆∆ or ∇∇keys to enter your hysteresis value for Set point
1 (Set point 2). Your controller will activate at the value you select. Note: All settings for SP1 and
SP2 are completely independent of each other.
4) Press the ENTER key to confirm your selection.
5) Proceed to 7.5.5, or return to Measurement mode by pressing the ∆∆ and ∇∇ keys simultaneously
(escape).
NOTE: Please refer to Appendix 3 for a graphical representation of the Hysteresis.
7.3.5 Setting an on-delay time lag
You can set as time delay for each relay, which stops the relay from switching on the moment the set
point is exceeded. This controller lets you set a 0 to 1999secondstime delay before the relay activates.
1) Follow directions in 7.5.4to enter Control Relay mode.
2) Press the ENTER key. Scroll with the ∆∆ and ∇∇ keys until the
upper display shows “0” time and the lower display shows “On.d”.
3) Press the ∆∆ or ∇∇ keys to enter on-delay time for Set point 1
(Set point 2). The controller will delay activation for the number of
seconds (0 to 1999) you select.
4) Press the ENTER key to confirm your selection.
5) Proceed to 7.5.6, or return to Measurement mode by pressing the ∆∆ and ∇∇ keys simultaneously
(escape).
7.3.6 Setting an off-delay time lag
You can set as time delay for each relay, which stops the relay from switching off the moment the value
reached the set point and hysteresis. This controller lets you set a 0 to 1999 secondstime delay before
your relay deactivates. 1) Follow directions in 7.5.5to enter Control Relay mode.
2) Press the ENTER key. Scroll with the ∆∆ or ∇∇ keys until the
upper display shows “0” time and the lower display shows “OF.d”.
3) Press the ∆∆ or ∇∇ keys to enter on-delay time for Set point 1
(Set point 2). Your controller will delay activation for the number of
seconds (0 to 1999) you select.
4) Press the ENTER key to confirm your selection.
5) Continue with Advanced Set-up mode procedures, or return to Measurement mode by pressing
the ∆∆ and∇∇keys simultaneously (escape).
2020
µµ
S
S
HYSHYS
HOLD SETUP
00
On.dOn.d
HOLD SETUP
00
OF.dOF.d
HOLD SETUP

Operating Instructions αlpha-CON1000
68X216802
19
7.4 Controller (Cntr) sub-function
You can set the controller’s parameters in this sub-function.
7.4.1 Entering the Controller sub-function
1) Enter Advanced Set-up mode. Push the ENTER key and scroll to Advanced set-up security code
“22”. Push the ENTER key again.
2) Press the ∆∆ or ∇∇keys to scroll until the upper display shows “Cntr”.
7.4.2 Choosing the controller type (limit or monitor)
This mode lets you choose your controller type: limit control or control off.
-Use control off to operate controller as a monitor only or to prevent relays from switching.
1) Follow directions in 7.6.1 to enter Controller mode.
2) Press the ENTER key. The upper display shows the current controller type and the lower display
shows“tyP”.
3) Press the ∆∆ or ∇∇keys to select your controller type.
-L.Ct = limit value pickup (on/off control).
-oFF = controller off.
4) Press the ENTER key to confirm your selection.
5) Proceed to 7.6.3, or return to Measurement mode by pressing the ∆∆ and ∇∇ keys simultaneously
(escape).
CntrCntr
HOLD SETUP L.CtL.Ct
tyPtyP
HOLD SETUP
HOLD SETUP
dEEndEEn
rELrEL
HOLD SETUP
oFFoFF
tyPtyP

68X216802
7.4.3 Choosing break/make contact relay type
Note: If the controller type “oFF” is set, the parameters listed in 7.5.3, 7.5.4, 7.5.5 and 7.5.6are blanked
out.
This mode lets you determine the relay-state under Non-Alarm condition – dEEN (de-energised) or EN
(energised). 1) Follow directions in 7.6.2to enter Controller mode.
2) Press the ENTER key. Scroll until the lower display shows “rEL”
and the upper display shows the current selection (de-energised =
dEEN or energised = EN).
3) Press the ∆∆ or ∇∇ keys to choose de-energised or energised
relay state.
4) Press the ENTER key to confirm your selection.
5) Continue with Advanced Set-up mode procedures, or return to
Measurement mode by pressing the ∆∆ and ∇∇ keys simultaneously
(escape).
7.5 Measurement Range sub-function
In this sub-function, the appropriate range is selected with the appropriate cell constant.
7.5.1 Selecting the Measuring Range sub-function
1) Enter Advanced Set-up mode. Push the ENTER key and scroll to
Advanced Set-up security code “22”. Push the ENTER key again.
2) Press the ∆∆ or ∇∇ keys to scroll until the upper display shows
“rng”.
3) Press the ENTER key. The upper display shows the maximum
measurement range, while the lower display shows the Cell constant. In
the lower left-corner of the LCD, the number corresponding to the
respective range is displayed.
4) Press the ∆∆ or ∇∇keys to select the correct range. (Please refer to
Appendix 2 for the full list of measurement ranges).
5) Proceed to 7.8 to set current output or return to Measurement mode by pressing the ∆∆ and
∇∇ keys simultaneously (escape).
dEEndEEn
rELrEL
HOLD SETUP
EnEn
rELrEL
HOLD SETUP
rngrng
HOLD SETUP
200200..00
0.10.1
HOLD SETUP
4
µ
S
Other manuals for ALPHA CON 1000 CONDUCTIVITY...
1
This manual suits for next models
1
Table of contents
Other EUTECH INSTRUMENTS Controllers manuals

EUTECH INSTRUMENTS
EUTECH INSTRUMENTS ALPHA DO 1000 DISSOLVED OXYGEN... User manual

EUTECH INSTRUMENTS
EUTECH INSTRUMENTS ALPHA CON 100 CONDUCTIVITY... User manual

EUTECH INSTRUMENTS
EUTECH INSTRUMENTS WD-19505-40 User manual

EUTECH INSTRUMENTS
EUTECH INSTRUMENTS ALPHA CON 200 CONDUCTIVITY... User manual

EUTECH INSTRUMENTS
EUTECH INSTRUMENTS ALPHA CON 100 - REV 1.1 User manual

EUTECH INSTRUMENTS
EUTECH INSTRUMENTS Alpha-CON560 User manual

EUTECH INSTRUMENTS
EUTECH INSTRUMENTS ALPHA CON 1000 CONDUCTIVITY... User manual

EUTECH INSTRUMENTS
EUTECH INSTRUMENTS ALPHA CON 200 CONDUCTIVITY... User manual