Fairchild T5220 Manual

FAIRCHILD T5220 SERIES
ELECTRO-PNEUMATIC TRANSDUCER
Installation, Operation and Maintenance Instructions
Figure 1. Model T5220 Outline Dimensions
Figure 2. Mounting Kit 15268. (Sold Separately)
INSTALLATION
The Model T5220 can be mounted directly onto a flat surface The Model T5220 can be Panel Mounted using the
using the two 10-32 tapped mounting holes in the base of the mounting bracket supplied in Mounting Kit 15268. For
housing. For more information, see Figure 1. “T5220 Outline more information, see Figure 2. above.
Dimensions” above.
1

Installation (Continued) NOTE:
The Model T5420 can be mounted on a 2" pipe using the For outdoor use, mount the T5220 in an upright
Optional Mounting Kit 14596. For more information, see position. This will help prevent rain water from
Figure 3. below. entering into the unit. For more information, see
Figure 3. “Mounting Kit 14596” below.
Figure 3. Optional Mounting Kit 14596. (Sold Separately)
2

INSTALLATION
The Model TXPD5220 can be mounted directly onto a flat 4. “TXPD5220 Outline Dimensions” below.
Mounting Kit
surface using the four 1/4-20 x 7/16 tapped mounting holes
14140 is available for Panel or 2" Pipe Mounting. For
in the base of the housing . For more information, see Figure
more information, see Figure 5 below.
Figure 4. TXPD5220 Outline Dimensions.
Figure 5. Mounting Kit 14140. (Sold Separately)
3

Pneumatic Connections
Clean all pipelines to remove dirt and scale before instal-
lation.
Apply a minimum amount of pipe compound to the male
threads of the fitting only. Do Not use teflon tape as a
sealant. Start with the third thread back and work away
from the end of the fitting to avoid the possibility of
contaminating the transducer. Install the transducer in the
air line.
NOTES: Instrument quality air, per ISA Standards
D7.3-1981, is required. Use a filter to re-
move dirt and liquid in the air line ahead of
the transducer for correct performance. If
an air line lubricator is used, it MUST be
located downstream, beyond the trans-
ducer.
The user is responsible for insuring that the
environment in which the unit will be
installed, and the operating gas, are com-
patible with the materials in the trans-
WARNING:
To prevent possible ignition of hazardous
atmosphere, DO NOT REMOVE COVER from
the TXPD5220 Transducer while current is
Electric Connection
For the T5220 Transducer, make connections to theTermi-
nal Block through a 1/2 NPT Conduit Connector (not
supplied) and a plastic sleeve as shown below in Figure 7.
“Electrical Connections”.
For the TXPD5220Transducer, make connections from the
Sealed Connector Wires to an externalTerminal Block. The
TXPD5220 Transducer has an additional ground wire that is
used. For more information, see Figure 7. “Electrical
Connections” below.
NOTE: The T5220 & TXPD5220 should be wired in
accordance with the wiring diagram inside of
cover.
Wiring in Hazardous Areas
Wiring in hazardous areas should be performed in accor-
dance with the Table 1. and any local codes that apply.
Table 1. Hazardous Location Wiring Practices.
Country Agency Code
U.S.
Canada
FM
CSA
ANSI/ISA RP 12.6 & ANSI/NFPA70
CEC Part 1
Intrinsically Safe Connections
Refer to the latest revision of the indicated drawing.
on.
To maintain Explosion-Proof capability, DO
NOT DAMAGE mating surfaces between
cover and base. NEVER ADD GASKETS.
Table 2. Intrinsically Safe Installation Drawings.
Underwriting Group Drawing Number
FM (Factory Mutual)
CSA (Canadian Standards)
EC-14007
EC-16097
Figure 6. Electrical Connections.
4

CALIBRATIONS / ADJUSTMENTS
Equipment Required for Calibration:
• Pneumatic Supply capable of delivering up to 150 psig.
• Current Supply capable of delivering up to 50 mA.
• Voltage Supply capable of delivering up to 10 VDC.
• Pressure Gage capable of a digital readout up to 50 psig
with an accuracy of .1%.
• Digital Volt Meter capable of a readout up to 30 mA or
10 VDC with an accuracy of .02%.
The following adjustments are provided:
Full Range Operation
Forward/Reverse Mode
Calibration - Zero and Span
FULL RANGE OPERATION
Forward Acting Mode Adjustment
NOTE: Air Supply must be on before adjustments are
made.
1. Connect the input signal to the transducer as shown
in Figure 6. “Electrical Connections.”
• Forward Acting Calibration-Zero
2. Apply the minimum input signal and adjust the Zero
screw for minimum output pressure. Turn screw
counterclockwise to increase pressure and clock-
wise to decrease pressure.
• Forward Acting Calibration-Span
3. Apply the maximum input signal and adjust the Span
screw on the potentiometer for maximum output
pressure. Turn screw clockwise to increase pres-
sure and counterclockwise to decrease pressure.
4. Repeat steps 2-3 until the desired output range is
obtained. For more information, see Figure 7. “T5220/
TXPD5220Calibration Configuration.”
Reverse Acting Mode Adjustment
NOTE: Intrinsically safe units cannot be set for re-
verse acting mode in field.
5. Connect the input signal to the transducer as shown
in Figure 6. “Electrical Connections.”
• Reverse Acting Calibration-Zero
6. Apply the minimum input signal and adjust the Zero
screw for maximum output pressure. Turn screw
counterclockwise to increase pressure and clock-
wise to decrease pressure.
• Reverse Acting Calibration-Span
7. Apply the maximum input signal and adjust the Span
screw on the potentiometer for minimum output pres-
sure. Turn screw counterclockwise to increase pres-
sure and clockwise to decrease pressure.
8. Repeat steps 6-7 until the desired output range is
obtained. For more information, see Figure 8. “T5200/
TXPD5200 Calibration Configuration.”
Figure 7. T5220 / TXPD5220 Calibration Configuration.
5

MAINTENANCE
To clean the Orifice, use the following procedure: To clear probblems caused by Magnet malfunction,
1. Turn off the air supply to the transducer. If the use the following procedure:
transducer is a common supply unit, remove the Pipe 1. Shut off the valve that is supplying air to transducer. It
Plug from the In Port located on the rectangular base. is not necessary to remove the Transducer from air
If the transducer is a separate supply unit, remove the line.
Air Line Fitting from the In Port. 2. Remove three Screws (12) holding Magnet (11) to
2. Remove the Orifice Assembly (9) from the unit by Transducer (8).
threading a 1/4-20" Screw into the orifice housing and
pulling it out through the inlet port. For more informa- 3. Remove Magnet (11) and clean gap of any foreign
tion, see Figure 8. “Exploded Drawing” below. material with masking tape. For more information, see
View B, below.
3. Clean with alcohol and dry with compressed air.
NOTE: Parts must be completely dry before reas- 4. Place Magnet (11) over Coil (10) and tighten with three
Screws (12).
sembling.
To clean the Inner Valve Assembly, use the following 5. Using a non-matalic rod, Raise up on the arm to make
procedure: sure Coil (10) is not sticking in Magnet (11).
1. Shut off the valve that is supplying air to transducer. It NOTE: If Coil is sticking, loosen Screws and repo-
is not necessary to remove the Transducer from air sition Magnet. Tighten Screws and repeat
line. step 5.
2. Remove Screws (7) holding the Retainer Cap (6) and To replace PC Board, use the following procedure:
Plate (5) to Booster (1). 1. Disconnect wires from Terminal Block on PC Board (14).
3. Pull out Inner Valve Assembly (4) and wash the seat
carefully. 2. Unsolder Red (+) and Black (-) Leads on PC Board (14).
3. Remove four Screws (13) holding PC Board (14) to
4. Wipe off any particles that may be attached to rubber Transducer (8). For more information, see Figure 8.
Seat Ring Assembly (2). For more information, see “Exploded Drawing” below.
Detail A, below.
NOTE: Avoid such solvents as acetone, carbon tetra- 4. Replace PC Board (14) with the appropriate PC Board
chloride and trichlorethylene. from Table 3. on page 7.
Figure 8. Exploded Drawing.
6

Maintenance (continued) TROUBLE-SHOOTING
Table 3. T5220 Series Components.
Item Qty. Part No. Description
1
2
3
4
5
6
7
8
9
10
11
12
13
14
14
14
14
14
14
14
14
1
1
1
1
1
1
2
1
1
4
1
1
1
1
1
1
1
1
3
1
1
NS
NS
NS
NS
NS
NS
NS
NS
NS
NS
NS
NS
NS
13744-2
15638-2
13745-2
15483-2
14639-2
15640-2
13746-2
15639-2
1
2
1
2
1
2
1
2
Transducer
Orifice
Coil
Magnet
Screw
Screw
Booster
Seat Ring Assembly
Screen
Inner ValveAssembly
Retainer Plate
Retainer Cap
Screw
1-5 mA PC Board
1-5 mA PC Board
4-20,10-50 mA PC Board
4-20,10-50 mA PC Board
1-5 VDC PC Board
1-5 VDC PC Board
1-9 VDC PC Board
1-9 VDC PC Board
1
2
For T5220, TFI5220, TFN5220, TFNI5220, TXPD5220,
& TXPDI5220 Units Only.
For TCI5220 Only.
Table 4. Trouble-Shooting.
Problem Solution (check)
No Output Supply Pressure
Clogged Orifice
Power Supply
Leakage Pneumatic Connections
Low or Improper
Span Adjust
Zero and SpanAdjust
Supply Pressure Low
Output Leakage
Erratic Operation DC Signal
Loose Wires or Connections
Liquid in Air Supply
Dirt in Magnet Gap
Unit Does Not
Operate
Replace PC Board
NOTE: Booster Repair Kits are available for the T5220.
Ratio Kit Number
1:1 15288-1
1:2 15288-2
1:3 15288-3
1:4 15656-4
NS: (Not Serviceable) In the Part No. column, NS indi- 1:5 15656-5
cates that the part is not available for field replacement. 1:6 15656-6
7

IS-500T5220
Litho in USA
Rev. M 06/07
This manual suits for next models
2
Table of contents
Other Fairchild Transducer manuals

Fairchild
Fairchild TXI7800 Manual

Fairchild
Fairchild T6000 User manual

Fairchild
Fairchild T7800 Series General instructions

Fairchild
Fairchild T79 Series Operating and maintenance manual

Fairchild
Fairchild T6000 Guide

Fairchild
Fairchild T7800 Series Manual

Fairchild
Fairchild T6100 Manual

Fairchild
Fairchild T6000 User manual

Fairchild
Fairchild T7900 User manual

Fairchild
Fairchild T5200 Series Manual